Design and analysis of three-dimensional printing of a porous titanium scaffold

Abstract

Objective

Mechanic strength, pore morphology and size are key factors for the three-dimensional (3D) printing of porous titanium scaffolds, therefore, developing optimal structure for the 3D printed titanium scaffold to fill bone defects in knee joints is instructive and important.

Methods

Structural models of titanium scaffolds with fifteen different pore unit were designed with 3D printing computer software; five different scaffold shapes were designed: imitation diamond-60°, imitation diamond-90°, imitation diamond-120°, regular tetrahedron and regular hexahedron. Each structural shape was evaluated with three pore sizes (400, 600 and 800 μm), and fifteen types of cylindrical models (size: 20 mm; height: 20 mm). Autodesk Inventor software was used to determine the strength and safety of the models by simulating simple strength acting on the knee joints. We analyzed the data and found suitable models for the design of 3D printing of porous titanium scaffolds.

Results

Fifteen different types of pore unit structural models were evaluated under positive pressure and lateral pressure; the compressive strength reduced when the pore size increased. Under torsional pressure, the strengths of the imitation diamond structure were similar when the pore size increased, and the strengths of the regular tetrahedron and regular hexahedron structures reduced when the pore size increased. In each case, the compressive strength of the regular hexahedron structure was highest, that of the regular tetrahedron was second highest, and that of the imitation diamond structure was relatively low. Fifteen types of cylindrical models under a set force were evaluated, and the sequence of comprehensive compressive strength, from strong to weak was: regular hexahedron > regular tetrahedron > imitation diamond-120° > imitation diamond-90° > imitation diamond-60°. The compressive strength of cylinder models was higher when the pore size was smaller.

Conclusion

The pore size and pore morphology were important factors influencing the compressive strength. The strength of each structure reduced when the pore size (400, 600 and 800 μm) increased. The models of regular hexahedron, regular tetrahedron and imitation diamond-120°appeared to meet the conditions of large pore sizes and high compressive strength.

Peer Review reports

Introduction

Bone defects are common in the clinic and are usually associated with diseases, such as infection, osteolysis, original implant loosening or tumor excision. Clinically, bone loss has been addressed with methods such as cement, autogenous bone grafts, and artificial implants [1, 2]. However, autogenous bone grafts are painful and source-limited, and are accompanied with complications, such as donor site morbidity [3]. Bone cement can lead to complications of absorption poisoning, bone absorption poisoning, bone absorption and allergies [4]. In recent years, artificial implant materials, such as calcium phosphate [5], ceramic [6], polymer materials [7] and metal [8], have been developed. Among them, metal is used in the clinic because of its high strength, high load capacity, shape memory, inertness and superelasticity. Common metal scaffolds include tantalum, titanium and titanium-nickel alloy.

Tantalum is a biocompatible material with good tissue compatibility with human tissue, causing almost no side effects. Therefore, this metal is widely used in medical clinics [9, 10]. However, the tantalum scaffold is expensive, and its strong oxidation and high melting point lead to costly and difficult processing. The production cost is high, and the methods are limited, and the tantalum powder shape is irregular; therefore, it is difficult to use 3D printing to generate tantalum scaffolds. Titanium has very high corrosion resistance, low density, and the highest strength / weight ratio of metals in addition to being non-magnetic. Titanium and titanium alloy materials also have good biomechanical properties and biocompatibility. Therefore, titanium is widely used in orthopedic implants [11].

3D printing technology, which is based on computer-aided design data, uses powder chromatography as a layer-by-layer printing method to quickly create objects with a complex 3D structure [12]. Compared with the traditional production technology, such as porous scaffolds, 3D printing technology has greater advantages in the control of the scaffold porosity, pore size, pore volume, spatial arrangement and other surface properties [13]. 3D printing technology has been used in many industries, including the medical industry, such as for orthopedic surgery and bone defects treatment, to make models and help doctors better understand complex anatomy and pathology [14,15,16,17,18]. Porous titanium scaffolds generated with 3D printing have great advantages for the control of porosity, pore size, pore volume, spatial arrangement and other surface features. 3D printing of porous titanium metaphyseal cones has previously been applied to revision TKA [19] and 3D printing of porous titanium is conducive to bone differentiation and new bone formation [20,21,22].

At present, there are some problems in the 3D porous titanium scaffolds. For example, its porous structure will reduce its mechanical properties. Different pore shapes will affect its mechanical properties. And different pore sizes will also affect the mechanical properties and the growth of bone tissue. Reasonable design of porous structure can not only promote the regeneration and growth of bone tissue, but also acquire well mechanical properties. Therefore, the pore shape and pore size are very important for the design of 3D porous titanium scaffolds. Several studies have done this work. One study has designed 3D porous titanium scaffolds with five different pore shapes and the same pore size, and found that under the condition of near porosity and the same pole sectional size, the 3D structure with different unit cells has a great impact on scaffold strength [23]. Another study has designed six types of 3D porous titanium scaffolds, and found the compressive stiffness values covered the range of cortical and trabecular bone [24]. However, these studies did not comprehensively compare the mechanical properties of scaffolds with different pore sizes and the same pore shapes or with different pore shapes and the same pore sizes under the positive, lateral and torsional pressure. This study designed five different pore shapes, and three different pore sizes for each pore shape, and finite element analysis was carried out under the positive, lateral and torsional pressure. The results were converted into force or a safety factor. The design models with large pore sizes and high compressive strength were screened out, which could provide references for the further production of porous titanium scaffolds with the 3D printing technique.

Materials and methods

Scaffold model design

Ti6Al4V has good heat resistance, strength, plasticity, toughness, formability, corrosion resistance and biocompatibility [25, 26]. Ti6Al4V was used as the material used for titanium scaffolds, and its main physical properties include: mass density 4.43 g/cc, yield tensile strength 880 MPa, ultimate tensile strength 950 MPa, modulus of elasticity 113.8 GPa, compressive yield strength 970 MPa, Poisson’s ratio 0.342 ul, shear strength 44 GPa [27,28,29]. Ti6Al4V powder has an average size of 45 μm [30].

Some studies have designed 3D porous titanium scaffolds with different pore shapes of imitation diamond, regular tetrahedron, regular octahedron, three circles type [23] and regular hexahedron [21, 25, 31]. Many studies suggested that large pore size was beneficial to the growth of bone tissue, such as 500 [24] and 700 μm [32]. Therefore, this study chose five common unit structures: imitation diamond-60°, imitation diamond-90°, imitation diamond-120°, regular tetrahedron and regular hexahedron and three pore sizes: 400, 600 and 800 μm.

Five types of unit structures were designed using Autodesk Inventor software (Inventor 2016, Autodesk Inc. San Rafael, California, USA), imitation diamond-60°, imitation diamond-90°, imitation diamond-120°, regular tetrahedron and regular hexahedron. The size of the bracket bar was 400 μm. Each type of unit structure has three pore sizes: 400, 600 and 800 μm, and the aperture was set as shown in Fig. 1. The fifteen kinds of unit structures were aligned to obtain a cylinder with a size of 20 mm and a height of 20 mm (cylinder size and height were finely adjusted for overall integrity).

Fig. 1
figure 1

A Overall top view of the imitation diamond unit structure; R is the size of the aperture. B Regular tetrahedron unit structure has a diameter R in either side. C Size of the aperture R is the inscribed circle diameter on either side of the regular hexahedral unit structure

Finite element analysis

The positive pressure, lateral pressure and torsional pressure were applied to fifteen kinds of unit structures and the corresponding fifteen kinds of cylindrical models with ABAQUS software (ABAQUS 2016, Simulia Inc. Providence, Rhole Island, USA). Finite element analysis was carried out to obtain the mechanical properties and compare the data. These simulations and collation took 3 months approximately.

The configuration of the computer used to run the Autodesk Inventor software and the ABAQUS software included a CPU model: Intel Core i7 6700HQ, quad-core eight threads, 256 GB solid state + 1 TB hard disk, 16G memory, NVIDIA GeForce GTX 1060 graphics card, and Win10 system.

Setting and analysis of unit structure model

When the positive pressure, lateral pressure and torsional pressure were applied to the fifteen kinds of unit structures, and when the minimum safety factor of the model was greater than or equal to 1, the force was the maximum pressure the unit structure could bear (accurate to 0.5 N). When the unit structure was under positive pressure, lateral pressure and torsional pressure, a fixed surface was selected and a force was applied to the offside of the fixed side. The unit structure and force are shown in Fig. 2. The maximum forces of each unit structure in the safe state obtained by software analysis were compared.

Fig. 2
figure 2

Schematic of different unit structures and applied pressure

Safety factor

The safety factor is the ratio of the ultimate stress to the allowable stress. The safety factor is the strength margin, considering factors that accurately calculate the load and stress, the importance of the work of the machine, and the reliability of the material. The value is greater than or equal to 1 (less than 1 indicates permanent deformation) [33, 34]. When the computer simulated the applied force, the point with the minimum value of the safety factor of models (Fig. 3) was determined, which was the point where the model was most likely to be damaged.

Fig. 3
figure 3

Unit structure and cylindrical structure of imitation diamond-60°under lateral pressure (600 N); blue indicates security, and red indicates danger. The reddest point occurred when the safety factor was the smallest

Setting and analysis of cylindrical model

Five types of cylindrical models are shown in Fig. 4. The pressure applied to the fifteen kinds of cylindrical models was set such that the pressure on the knee was approximately twice the weight of the adult (approximately 60 kg). When an adult is standing on one leg, the knee is subjected to a stress of approximately 3 ~ 4 times the person’s weight. The pressure on the knees when climbing stairs is about 3 to 6 times of body weight [35]. When the person kicks the ball, the twisting force is approximately 3000–4000 N. Therefore, when the positive, lateral and torsional pressures were applied to the fifteen kinds of cylindrical models, they were applied with 1 times, 3 times and 5 times the positive pressure and 1 times and 3times the lateral pressure, and the torsional pressure applied was 2000, 3000 and 4000 N. Then, the safety factors were obtained using the ABAQUS software analysis under different values for various forces and compared.

Fig. 4
figure 4

The five types of cylindrical model. A Imitation diamond-60°; B imitation diamond-90°; C imitation diamond-120°; D regular tetrahedron; E hexahedron

Results and analysis

Results and analysis of unit structure model

The maximum positive pressure values that could be sustained on the fifteen kinds of unit structures in the safety state are shown in Fig. 5 and Table 1. The maximum lateral pressure values are shown in Fig. 6 and Table 2, and the maximum torsional pressure values are shown in Fig. 7 and Table 3. Through the analysis of the data in the table, the following conclusions can be drawn:

  1. (1)

    Under positive pressure, the maximum force of the imitation diamond and the regular hexahedron structure decreased with increasing pore size, the maximum force of the regular tetrahedron structure didn’t decrease with increasing pore size, and the maximum forces of the regular tetrahedron and the regular hexahedron structures were much larger than that of the imitation diamond structure, and the compressive strength of the regular hexahedron structure was the highest.

  2. (2)

    Under the lateral pressure, the maximum force of the five kinds of unit structures decreased with increasing pore size, and the maximum forces of the regular tetrahedron and regular hexahedron structures were much larger than that of the imitation diamond structure. The compressive strength of the regular hexahedron structure was the highest.

  3. (3)

    Under the torsional pressure, the maximum force of the imitation diamond-90° structure increased with increasing pore size. The maximum force of the other structure did not change with the pore size. The regular hexahedron structure had the greatest torsional resistance. And the regular tetrahedron structure had the smallest torsional resistance.

  4. (4)

    In the comprehensive analysis of positive pressures, the order of compressive capacity of the five types of unit models, from strong to weak, was regular hexahedron > regular tetrahedron > imitation diamond-120° > imitation diamond-90° > imitation diamond-60°.

  5. (5)

    In the lateral pressure comprehensive analysis, the order of the compressive capacity of the five types of unit models from strong to weak, was regular hexahedron > regular tetrahedron > imitation diamond-120° > imitation diamond-90° > imitation diamond-60°.

  6. (6)

    In the comprehensive analysis of the torsional pressure, the order of the compressive capacity of the five types of unit models, from strong to weak, was regular hexahedron > imitation diamond-120° > imitation diamond-90° > imitation diamond-60° > regular tetrahedron.

Fig. 5
figure 5

When the safety factor is greater than or equal to 1, the unit structure can withstand the maximum positive pressure (accurate to 0.5 N)

Table 1 When the safety factor is greater than or equal to 1, the unit structure can withstand the maximum positive pressure (accurate to 0.5 N)
Fig. 6
figure 6

When the safety factor is greater than or equal to 1, the unit structure can withstand the maximum lateral pressure (accurate to 0.5 N)

Fig. 7
figure 7

When the safety factor is greater than or equal to 1, the unit structure can withstand the maximum torsional pressure (accurate to 0.5 N)

Table 2 When the safety factor is greater than or equal to 1, the unit structure can withstand the maximum lateral pressure (accurate to 0.5 N)

Results and analyses of cylindrical model

The safety factors for the fifteen cylindrical models given the positive, lateral and torsional pressures are shown in Tables 4, 5, 6, 7 and 8. Through the analysis of the data in the tables, the following conclusions could be drawn:

  1. (1)

    Under the positive, lateral and torsional pressures, the safety factors of the imitation diamond, the regular tetrahedron and the regular hexahedron models decreased with increasing pore size.

  2. (2)

    Under positive pressure, considering factors such as the force, pore size and safety factor, the order of the compressive capacity of the five types of cylindrical models, from strong to weak, was regular hexahedron > regular tetrahedron > imitation diamond-90° > imitation diamond-60° > imitation diamond-120°.

  3. (3)

    Under the lateral pressure, considering factors such as the force, pore size and safety factor, the order of the compressive capacity of the five types of cylindrical models, from strong to weak, was regular hexahedron > regular tetrahedron > imitation diamond-120° > imitation diamond-90° > imitation diamond-60°.

  4. (4)

    Under the torsional pressure, considering factors such as the force, pore size and safety factor, the order of the compressive capacity of the five types of cylindrical models, from strong to weak, was regular hexahedron > imitation diamond-120° > imitation diamond-90° > regular tetrahedron > imitation diamond-60°.

  5. (5)

    Based on the above points, the order of the comprehensive compressive capacity of the five types of cylindrical models, from strong to weak, was regular hexahedron > regular tetrahedron > imitation diamond-120° > imitation diamond-90° > imitation diamond-60°. The smaller the pore size in each type of cylindrical model, the greater the compressive strength.

Table 3 When the safety factor is greater than or equal to 1, the unit structure can withstand the maximum torsional pressure (accurate to 0.5 N)
Table 4 Minimum safety factor for the imitation diamond-60° cylindrical model under different pressures
Table 5 Minimum safety factor for the imitation diamond-90° cylindrical model under different pressures
Table 6 Minimum safety factor for the imitation diamond-120° cylindrical model under different pressures
Table 7 Minimum safety factor for the regular tetrahedron cylindrical model under different pressures
Table 8 Minimum safety factor for the regular hexahedron cylindrical model under different pressures

Discussion

Bone defects are common and the important problems to be solved in the clinic. Treatments of bone defects include autologous bone grafts, allogeneic bone grafts, bone cement and the implantation of artificial materials. At present, a variety of artificial materials have been used to treat bone defects, including metals because they have favorable characteristics. Ti6Al4V has good heat resistance, strength, plasticity, toughness, formability, corrosion resistance and biocompatibility. Compared to the space-holder technology [36], oaming method [37], and other methods, 3D printing technology can better control of the porosity, larger pore size [38], pore volume, spatial arrangement and other surface properties of scaffolds. The porous titanium scaffold made by 3D printing technology has good biomechanical properties, biocompatibility and lower elastic modulus [39] compared to titanium and its alloys.

The design of the pore morphology in 3D printed porous titanium scaffolds has a great influence on mechanics, and also affects the cell biologically. The factors that affect the growth of osteoblasts are not only pore size, and other factors such as connectivity, pore shape and porosity [40]. Therefore, the study of pore size and shape is important to the design of 3D printed porous titanium scaffolds. In this study, Autodesk Inventor software was used to design the unit structures and cylindrical models with different structures and different pore sizes. The lower limit of the pore size was 400 μm, which ensured that the pore size of the scaffold was large enough. There were five different types of unit structures, each with different shapes of pores. Any of the structures of the scaffold holes could be connected to any other. By finite element analyses, we found under positive pressure, the regular hexahedron cylindrical model was the strongest, then the regular tetrahedron cylindrical model, the imitation diamond cylindrical model was the worst, consistent with other studies [23, 30, 41]. Compared to other studies, we also carried out analysis under the lateral and torsional pressure, and found results were variably under different pressures, but the comprehensive compressive capacity of regular hexahedron cylindrical model was always the best under each pressure. This study showed the compressive strength of porous structures decreased with increases of pore size, similar to other studies [21, 42]. Most studies just analyzed cylindrical or cube models, the force analysis of the unit structure was also carried out in this study, and the results were consistent with those of the cylindrical models.

The positive pressure for the imitation diamond unit structure could be analyzed with the simple model in Fig. 8.

Fig. 8
figure 8

Positive pressure on the imitation diamond unit structure

The length of the bracket bar was L, the angle between the two bars was θ, the pressure on the bar end was F, the moment of the bracket center point was M, and the following formula could be obtained [23]:

$$\mathrm{M}=\mathrm{FLsin}\frac{\uptheta }{2}$$

From the above formula, when the moment M of the bracket center point was constant, the compressive strength at the point where the bracket was most vulnerable was constant; when θ was constant, L was larger and F was smaller. When L was constant, θ was larger (0° < θ < 180°) and F was smaller.

Therefore, for the imitation diamond structure, when the pore size was consistent, the angle θ was larger and the force F was smaller. When the angle between the two bars was constant, the pore size was larger, L was larger, and the force F was smaller. However, because the pore size of the imitation diamond was the same, θ was larger but L was smaller; therefore, it was not sufficient to use this formula for analysis.

For the regular tetrahedron, the vertex moment M was constant, while at the same time, there were three bars bearing pressure; thus compared to the imitation diamond structure, the regular tetrahedron could withstand greater force.

For the regular hexahedron, the force was in the longitudinal direction of the bar, which was the compressive strength of the titanium bar, and it could therefore withstand the greatest force.

The material of metallic biomaterials can affect the mechanical properties. The digital design and finite element analysis method are also applicable to other metallic biomaterials. Of course, the specific design is contingent on the properties of metallic biomaterials selected, especially the material can be made into finished products by 3D printing. The digital model can be obtained by input the relative physical properties data of the metallic biomaterial, design shapes and pore sizes when modeling, and finite element analysis can be carried out.

There are some shortcomings in this study: (1) Compared with dense titanium,3D printed porous titanium scaffolds have a lower modulus of elasticity, but this study did not analyze the elastic modulus. (2) This study did not measure or analyze the porosity, surface area or other factors. (3) The porous titanium scaffold used to fill knee bone defects is not simply under positive, lateral or torsional pressure, but may also be subject to various directions of the various pressures from various directions; this study failed to analyze this complex condition. (4) Due to the different pore sizes of unit structures contained in the fixed-size cylindrical model, the complete unit structure may not be retained at the edge of the model when intercepting, resulting in some data inconsistent with the theory. For example, in Table 8, when the 600 μm pore size models were subjected to force, the data obtained were not in the middle. (5) This study only used software to design and simulate forces; the obtained data are only a reference for further entity production and testing. Biocompatibility and osteoblast attachment, differentiation and growth on the 3D printed porous titanium scaffold require further studies.

Conclusions

This study evaluated 3D printed porous titanium scaffolds with fifteen different pore structures under positive, lateral and torsional pressures. The order of the comprehensive compressive capacity of the five types of cylindrical models was regular hexahedron > regular tetrahedron > imitation diamond-120° > imitation diamond-90° > imitation diamond-60°, and for each type of cylinder, the smaller the pore size, the greater the compressive strength. The regular hexahedron, regular tetrahedron and imitation diamond-120° models appear to meet the conditions of large pore size and high compressive strength. In addition, the strength of each structure reduced when the pore size (400, 600 and 800 μm) increased. Obtaining the optimal design of the 3D printed porous titanium scaffold can provide a potentially effective clinical solution and ultimately promote the clinical application of titanium scaffolds.

Availability of data and materials

Materials described in the manuscript, including all relevant raw data, can be freely available to any scientist and reader wishing to use them for non-commercial purposes, without breaching participant confidentiality. The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request. All data generated or analysed during this study are included in this published article.

References

  1. Fedorka CJ, Chen AF, Pagnotto MR, Crossett LS, Klatt BA. Revision total knee arthroplasty with porous-coated metaphyseal sleeves provides radiographic ingrowth and stable fixation. Knee Surg Sports Traumatol Arthrosc. 2018;26(5):1500–05.

  2. Villatte G, Erivan R, Salles G, Pereira B, Galvin M, Descamps S, Boisgard S. Acetabular bone defects in THA revision: reconstruction using morsellised virus-inactivated bone allograft and reinforcement ring. Seven-year outcomes in 95 patients. Orthop Traumatol Surg Res. 2017;103(4):543–8.

    Article  CAS  Google Scholar 

  3. Costa Mendes L, Sauvigne T, Guiol J. Morbidity of autologous bone harvesting in implantology: literature review from 1990 to 2015. Rev Stomatol Chir Maxillofac Chir Orale. 2016;117(6):388–402.

    CAS  PubMed  Google Scholar 

  4. Stathopoulos I, Andrianopoulos N, Paschaloglou D, Tsarouchas I. Revision total knee arthroplasty due to bone cement and metal hypersensitivity. Arch Orthop Trauma Surg. 2017;137(2):267–71.

    Article  Google Scholar 

  5. Inzana JA, Olvera D, Fuller SM, Kelly JP, Graeve OA, Schwarz EM, Kates SL, Awad HA. 3D printing of composite calcium phosphate and collagen scaffolds for bone regeneration. Biomaterials. 2014;35(13):4026–34.

    Article  CAS  Google Scholar 

  6. Bose S, Vahabzadeh S, Bandyopadhyay A. Bone tissue engineering using 3D printing. Mater Today. 2013;16(12):496–504.

    Article  CAS  Google Scholar 

  7. Heller M, Bauer HK, Goetze E, Gielisch M, Ozbolat IT, Moncal KK, Rizk E, Seitz H, Gelinsky M, Schrder HC, et al. Materials and scaffolds in medical 3D printing and bioprinting in the context of bone regeneration. Int J Comput Dent. 2016;19(4):301–21.

    PubMed  Google Scholar 

  8. Li X, Wang L, Yu XM, Feng YF, Wang CT, Yang K, Su D. Tantalum coating on porous Ti6Al4V scaffold using chemical vapor deposition and preliminary biological evaluation. Mater Sci Eng C. 2013;33(5):2987–94.

    Article  CAS  Google Scholar 

  9. De Francesco M, Gobbato EA, Noce D, Cavallari F, Fioretti A. Clinical and radiographic evaluation of single tantalum dental implants: a prospective pilot clinical study. Oral Implantol. 2016;9(Suppl 1/2016 to N 4/2016):38–44.

    Article  Google Scholar 

  10. Potter GD 3rd, Abdel MP, Lewallen DG, Hanssen AD. Midterm results of porous tantalum femoral cones in revision total knee arthroplasty. J Bone Joint Surg Am. 2016;98(15):1286–91.

    Article  Google Scholar 

  11. Gao Y, Ou Y, Deng Q, He B, Du X, Li J. Comparison between titanium mesh and autogenous iliac bone graft to restore vertebral height through posterior approach for the treatment of thoracic and lumbar spinal tuberculosis. PLoS One. 2017;12(4):e0175567.

    Article  Google Scholar 

  12. Hong S, Sycks D, Chan HF, Lin S, Lopez GP, Guilak F, Leong KW, Zhao X. 3D printing of highly stretchable and tough hydrogels into complex, cellularized structures. Adv Mater. 2015;27(27):4035–40.

    Article  CAS  Google Scholar 

  13. Roseti L, Parisi V, Petretta M, Cavallo C, Desando G, Bartolotti I, Grigolo B. Scaffolds for bone tissue engineering: state of the art and new perspectives. Mater Sci Eng C Mater Biol Appl. 2017;78:1246–62.

    Article  CAS  Google Scholar 

  14. Tetsworth K, Block S, Glatt V. Putting 3D modelling and 3D printing into practice: virtual surgery and preoperative planning to reconstruct complex post-traumatic skeletal deformities and defects. Sicot-J. 2017;3:16.

    Article  Google Scholar 

  15. Rasperini G, Pilipchuk SP, Flanagan CL, Park CH, Pagni G, Hollister SJ, Giannobile WV. 3D-printed bioresorbable scaffold for periodontal repair. J Dent Res. 2015;94(9 Suppl):153S-157S.

    Article  CAS  Google Scholar 

  16. Shaunak S, Dhinsa BS, Khan WS. The role of 3D modelling and printing in orthopaedic tissue engineering: a review of the current literature. Curr Stem Cell Res Ther. 2017;12(3):225–32.

    Article  CAS  Google Scholar 

  17. Wong KC. 3D-printed patient-specific applications in orthopedics. Orthop Res Rev. 2016;8:57–66.

    PubMed  PubMed Central  Google Scholar 

  18. Belvedere C, Siegler S, Fortunato A, Caravaggi P, Liverani E, Durante S, Ensini A, Konow T, Leardini A. New comprehensive procedure for custom-made total ankle replacements: medical imaging, joint modeling, prosthesis design, and 3D printing. J Orthop Res. 2019;37(3):760–8.

    Article  CAS  Google Scholar 

  19. Faizan A, Bhowmik-Stoker M, Kirk AE, Krebs VE, Harwin SF, Meneghini RM. Development and verification of novel porous titanium metaphyseal cones for revision total knee arthroplasty. J Arthroplasty. 2017;32(6):1946–53.

    Article  Google Scholar 

  20. Zhou X, Zhang D, Wang M, Zhang D, Xu Y. Three-dimensional printed titanium scaffolds enhance osteogenic differentiation and new bone formation by cultured adipose tissue-derived stem cells through the IGF-1R/AKT/mammalian target of rapamycin complex 1 (mTORC1) pathway. Med Sci Monit. 2019;25:8043–54.

    Article  CAS  Google Scholar 

  21. Zheng Y, Han Q, Wang J, Li D, Song Z, Yu J. Promotion of osseointegration between implant and bone interface by titanium alloy porous scaffolds prepared by 3D printing. ACS Biomater Sci Eng. 2020;6(9):5181–90.

    Article  CAS  Google Scholar 

  22. Felgueiras HP, Decambron A, Manassero M, Tulasne L, Evans MD, Viateau V, Migonney V. Bone tissue response induced by bioactive polymer functionalized Ti6Al4V surfaces: in vitro and in vivo study. J Colloid Interface Sci. 2017;491:44–54.

    Article  CAS  Google Scholar 

  23. Wang C, Li X, Luo Y. Design and strength analysis of porous titanium scaffold. J Shanghai Jiaotong Univ (Chin Ed). 2016;50(2):165–8.

    CAS  Google Scholar 

  24. Van Bael S, Chai YC, Truscello S, Moesen M, Kerckhofs G, Van Oosterwyck H, Kruth IP, Schrooten J. The effect of pore geometry on the in vitro biological behavior of human periosteum-derived cells seeded on selective laser-melted Ti6Al4V bone scaffolds. Acta Biomater. 2012;8(7):2824–34.

    Article  Google Scholar 

  25. Bartolomeu F, Sampaio M, Carvalho O, Pinto E, Alves N, Gomes JR, Silva FS, Miranda G. Tribological behavior of Ti6Al4V cellular structures produced by selective laser melting. J Mech Behav Biomed Mater. 2017;69:128–34.

    Article  CAS  Google Scholar 

  26. Lietaert K, Cutolo A, Boey D, Van Hooreweder B. Fatigue life of additively manufactured Ti6Al4V scaffolds under tension-tension, tension-compression and compression-compression fatigue load. Sci Rep. 2018;8(1):4957.

    Article  Google Scholar 

  27. Welsch G, Boyer R, Collings E. Materials properties handbook: titanium alloys. ASM International, Materials Park, OH; 1994.

  28. ASM: Metals handbook vol. 2. ASM International, Materials Park, OH; 1990.

  29. ASM Handbook Committee: Properties and selection: stainless steels, tool materials and special-purpose metals, metals handbook Vol. 3, American Society for Metals, Materials Park, OH, 1980.

  30. Li JP, de Wijn JR, Van Blitterswijk CA, de Groot K. Porous Ti6Al4V scaffold directly fabricating by rapid prototyping: preparation and in vitro experiment. Biomaterials. 2006;27(8):1223–35.

    Article  CAS  Google Scholar 

  31. Bartolomeu F, Costa MM, Alves N, Miranda G, Silva FS. Selective laser melting of Ti6Al4V sub-millimetric cellular structures: prediction of dimensional deviations and mechanical performance. J Mech Behav Biomed Mater. 2021;113:104123.

    Article  CAS  Google Scholar 

  32. Ran Q, Yang W, Hu Y, Shen X, Yu Y, Xiang Y, Cai K. Osteogenesis of 3D printed porous Ti6Al4V implants with different pore sizes. J Mech Behav Biomed Mater. 2018;84:1–11.

    Article  CAS  Google Scholar 

  33. Sibgatullin KE, Sibgatullin ES. Safety factor of anisotropic bars in the space of generalized forces. Mech Compos Mater. 2017;52(6):781–8.

    Article  Google Scholar 

  34. Wu HM, Zhao LX, Yao GY, Zhang YW, Liu GC. Reliability design method based on average safety factor. Mach Des Res. 2009;25(4):14–6.

    Google Scholar 

  35. Costigan PA, Deluzio KJ, Wyss UP. Knee and hip kinetics during normal stair climbing. Gait Posture. 2002;16(1):31–7.

    Article  Google Scholar 

  36. Arifvianto B, Leeflang MA, Zhou J. Diametral compression behavior of biomedical titanium scaffolds with open, interconnected pores prepared with the space holder method. J Mech Behav Biomed Mater. 2017;68:144–54.

    Article  CAS  Google Scholar 

  37. Zhang M, Wang GL, Zhang HF, Hu XD, Shi XY, Li S, Lin W. Repair of segmental long bone defect in a rabbit radius nonunion model: comparison of cylindrical porous titanium and hydroxyapatite scaffolds. Artif Organs. 2014;38(6):493–502.

    Article  CAS  Google Scholar 

  38. Dewidar M, Mohamed HF, Lim JK. A new approach for manufacturing a high porosity Ti-6Al-4V scaffolds for biomedical applications. J Mater Sci Technol. 2008;24(6):931–5.

    CAS  Google Scholar 

  39. El-Hajje A, Kolos EC, Wang JK, Maleksaeedi S, He Z, Wiria FE, Choong C, Ruys AJ. Physical and mechanical characterisation of 3D-printed porous titanium for biomedical applications. J Mater Sci Mater Med. 2014;25(11):2471–80.

    Article  CAS  Google Scholar 

  40. Caparros C, Ortiz-Hernandez M, Molmeneu M, Punset M, Calero JA, Aparicio C, Fernandez-Fairen M, Perez R, Gil FJ. Bioactive macroporous titanium implants highly interconnected. J Mater Sci Mater Med. 2016;27(10):151.

    Article  Google Scholar 

  41. Elsayed H, Rebesan P, Giacomello G, Pasetto M, Gardin C, Ferroni L, Zavan B, Biasetto L. Direct ink writing of porous titanium (Ti6Al4V) lattice structures. Mater Sci Eng C Mater Biol Appl. 2019;103:109794.

    Article  CAS  Google Scholar 

  42. Xu Y, Zhang D, Zhou Y, Wang W, Cao X. Study on topology optimization design, manufacturability, and performance evaluation of Ti-6Al-4V porous structures fabricated by selective laser melting (SLM). Materials. 2017;10(9):1048.

    Article  Google Scholar 

Download references

Acknowledgements

The authors thank School of Biomedical Engineering, Shanghai Jiao Tong University, for finite element analysis.

Funding

This work was supported by the Shanghai Science and Technology Committee Guidance Project (16411971700); Fundamental Research Funds for the Central Universities, Subject Crossing (1501219095); and National Natural Science Foundation of China (81101344/H0604).

Author information

Authors and Affiliations

Authors

Contributions

Chunxi Yang contributed to the conception of the study; Jiajie Yang performed the experiment; Jiajie Yang and Chunxi Yang contributed significantly to analysis and manuscript preparation; Jiajie Yang and Yaqiang Li performed the data analyses and wrote the manuscript; Xiaojian Shi, Meihua Shen, Kaibing Shi and Lingjie Shen helped perform the analysis with constructive discussions. The authors read and approved the final manuscript.

Corresponding author

Correspondence to Chunxi Yang.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not Applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Yang, J., Li, Y., Shi, X. et al. Design and analysis of three-dimensional printing of a porous titanium scaffold. BMC Musculoskelet Disord 22, 654 (2021). https://doi.org/10.1186/s12891-021-04520-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-021-04520-1

Keywords