Clinical evaluation of the efficacy of a new bone cement-injectable cannulated pedicle screw in the treatment of spondylolysis-type lumbar spondylolisthesis with osteoporosis: a retrospective study

Abstract

Purpose

To investigate the clinical efficacy and safety of a bone cement-injectable cannulated pedicle screw (CICPS) in the treatment of spondylolysis-type lumbar spondylolisthesis with osteoporosis.

Methods

A retrospective study was conducted on 37 patients (Dual-energy X-ray bone density detection showed different degrees of osteoporosis) with spondylolysis-type lumbar spondylolisthesis who underwent lumbar spondylolisthesis reduction and fusion using a new type of injectable bone cement screw from May 2011 to March 2015. Postoperative clinical efficacy was evaluated by the Visual Analogue Scale (VAS) scores and the Oswestry Disability Index (ODI). Imaging indexes were used to evaluate the stability of internal fixation of the devices 1, 3, 6, and 12 months after surgery and annually thereafter. The safety of the CICPS was assessed by the prevalence of intraoperative and postoperative complications.

Results

A total of 124 CICPS were implanted intraoperatively. Bone cement leakage occurred in 3 screws (2.42%), and no clinical discomfort was found in any patients. All 37 patients were followed up with an average follow-up time of 26.6 ± 13.4 months (12–58 months). In the evaluation of the clinical effects of the operation, the average postoperative VAS score of the patients decreased from 4.30 ± 1.58 before surgery to 0.30 ± 0.70 after surgery (P < 0.001), and the ODI decreased from 47.27% ± 16.97% before surgery to 3.36% ± 5.70% after surgery (P < 0.001). No screw was loose, broken or pulled out.

Conclusion

CICPS is safe and effective in the treatment of spondylolysis-type lumbar spondylolisthesis complicated by osteoporosis.

Peer Review reports

Background

Lumbar spondylolisthesis is a common disease that causes chronic lumbar and leg pain in middle-aged and older people. The Wiltse Classification System classified lumbar spondylolisthesis into 5 types, including dysplasia, spondylolysis, degeneration, trauma and pathology, among which spondylolysis has the highest incidence [1, 2]. The main surgical treatment for spondylolysis-type lumbar spondylolisthesis is pedicle screw system. However, the postoperative screw loosening rate is extremely high in lumbar spondylolysis-type slippage patients with osteoporosis, due to the reduced bone mineral density, bone microarchitecture, and thinning of cortical bone. As Galbusera et al. reported that spondylolisthesis patients with osteoporosis had a screw loosening rate of 60% [3].To overcome screw loosening, cannulated pedicle screw (CPS) augmented by poly-methyl-methacrylate (PMMA) has become the most used method in spinal diseases with osteoporosis [4]. In our previous study, we designed a new type of pedicle screw named bone cement-injectable cannulated pedicle screw (CICPS) as shown in Fig. 1. Biomechanical test results proved CICPS had greater torque than the OMEGA cannulated pedicle screw and conventional pedicle screw, and finite element analysis showed none excessive stress at the screw-cement–bone interface in the CICPS group [5]. Afterwards, we have applied our CICPS in relevant clinical applications [6], all of which achieved satisfactory outcomes, and obtained national patent.

Fig. 1
figure 1

Gross observation of CICPS. A Top view of CICPS. B Side view of CICPS. C Three side holes distributed longitudinally from small to large

In this study, we retrospectively reviewed the preoperative and postoperative clinical manifestations and imaging parameters of 37 consecutive patients with spondylolysis-type lumbar spondylolisthesis complicated by osteoporosis using CICPS. The incidence of complications and the clinical efficacy of our CICPS were summarized.

Materials and methods

General data

From May 17, 2011 to March 12, 2015, clinical data of 37 patients with spondylolysis-type lumbar spondylolisthesis and osteoporosis were reviewed in this study. Inclusion criteria were as follows: 1. Confirmed diagnoses of spondylolysis-type lumbar spondylolisthesis by positive lumbar X-ray and plain lumbar Computed Tomography (CT) results with related clinical symptoms; 2. Confirmed diagnoses of osteoporosis by preoperative dual-energy X-ray bone density absorptiometry, T-score < -2.5 SD [7]; and 3. CICPS were used with no surgical contraindications. Exclusion criteria were as follows: 1. Allergy to the implant; 2. Presence of other spine diseases; and 3. Infections, blood system-related diseases, or other surgical contraindications. General information of patients was listed in Table 1. The study was approved by the Southwest Hospital ethics committee (KY201927), and carried out in accordance with ethical guidelines of Army Medical University. All patients included in this study gave their informed consents.

Table 1 Baseline demographic and clinical characteristics of 37 patients

CICPS design

CICPS (produced by Kanghui Medical Devices, Jiangsu, China) used in this study had diameters of 5.0 mm to 6.0 mm, lengths of 45 to 50 mm, nail heads adopting unidirectional and universal designs, unidirectional heads that can move with 360° rotation, and a longitudinal hollow center. The bone cement outflow diameter was 2.2 mm and was closed distally. Three side holes were designed in the front 2/5 of the screw and were distributed longitudinally from small to large. The side holes near the nail head were round with a diameter of 2 mm. The opposite side was separated by a thread with an oval side hole, a long diameter of 3 mm and a short diameter of 2 mm. The u-shaped side hole with a penetrating tip was 4 mm long and 2 mm wide (Fig. 1).

Surgical method

Transforminal lumbar interbody fusion was routinely performed. All the cases underwent vertebrae fusion, and whether single or double vertebrae were used depending on the condition of patients, all the cages used were purchased from Johnson & Johnson (USA). The surgical procedure used a conventional posterior midline approach with a nailing method similar to that of ordinary pedicle screws and a nailing angle slightly larger than for normal pedicle screws (Fig. 2A). To ensure that the screw hole side was away from the walls, the screw was placed 80% to 90% into the vertebra from the front wall, far away from the paries posterior of the vertebral body, to avoid leakage of bone cement into the spinal canal. Before nailing, a probe was used to ensure the integrity of the pin track was not damaged. Under c-arm fluoroscopy monitoring (Fig. 2B), bone cement in "toothpaste" period was injected through a customized injection system (Fig. 2C). Before bone cement injection, a negative pressure aspirator (Fig. 2C at the red arrow) was used to ensure the side hole of screw is not blocked by bone chips, otherwise, negative pressure should be repeatedly applied or the screws should be rotated to ensure an unobstructed side hole.

Fig. 2
figure 2

Surgical procedure of CICPS. A CICPS nailing angle slightly larger than normal pedicle screws, red line represents CICPS nailing angle and green line represents normal screws nailing angle. B Intraoperative bone cement dispersion under c-arm fluoroscopy monitoring. C Customized injection system of CICPS, red arrow indicates negative pressure aspirator

Considering the optimal strengthening effect need 1–2 ml of bone cement [6], and 1.5 ml of bone cement will remain in the supporting injection system, a total of 3 ml bone cement was injected during the operation. Decompression was performed on the spondylolisthesis, and the crushed autologous bone was filled into cages of appropriate size for bone fusion. After the bone cement was solidified, the connecting rod was bended and the sliding vertebral body was lifted for resetting (Figs. 3 and 4 for typical cases).

Fig. 3
figure 3

A 59 years old female, BMD -2.7SD, treated with CICPS fixation by PMMA augmentation. A Preoperative radiographs showed L4 vertebral body was of level II forward slip, spondylolysis type. B and C No bone cement leakage was observed one week after surgery. D CT results indicated successful fusion six months after surgery that continuous bone callus passed in the intervertebral space of fusion. E and F 56 months after the operation, CT showed good intervertebral fusion, no loosening or pulling out of screw was observed

Fig. 4
figure 4

A 54 years old female, BMD -2.8SD, treated with CICPS fixation by PMMA augmentation. A Preoperative radiographs showed L5 vertebral body was of level IV forward slip, spondylolysis type. B X-ray examination showed good vertebral reduction and no leakage of bone cement one week after operation. C CT results showed good fusion 36 months after surgery. D 60 months after surgery, no obvious double ring sign, no loosening, no fracture or pulling out was observed

In order to save medical costs, common pedicle screw internal fixation was given because the surgeon felt that the bone of some patients was fine during the operation. All patients were given anti-osteoporosis therapy, including calcium/vitamin D supplementation, aluminum phosphate or zoledronic acid salt according to the degree of osteoporosis.

Evaluation method

The efficacy of CICPS was assessed via comparison between preoperative and postoperative VAS and ODI scores at 1, 3, 6, and 12 months and each year thereafter. As the compliance of patients varied, some patients may not follow-up accurately according to the suggested time points, but all the final evaluation were performed more than one-year post operation. The stability of the CICPS was related to the intervertebral height and the Taillard index at the last follow-up. The intervertebral height between the posterior margin of the spondylolisthesis and the upper endplate of the lower vertebra was denoted by H-y1, and the vertical distance between the anterior edge of the lower vertebra and the lower endplate of the spondylolisthesis was denoted by H-y2. The average value of the two was denoted by H-y0 (Fig. 5A). For the Taillard index, we measured the horizontal distance L- × 2 of the upper vertebral body relative to the lower vertebral body slippage and the width L- × 1 of the upper endplate of the lower vertebral body (Fig. 5B). The ratio of the former to the latter is denoted by L- × 0.

Fig. 5
figure 5

CICPS efficacy evaluation. A Height measurement of intervertebral space. B Spondylolisthesis degree measurement. C CICPSs displacement degree measurement

The criterion for successful fusion was the presence of a continuous trabecular passage through the fusion zone on the lumbar vertebra clearance [8]. The stability of the CICPS was evaluated by the changes in the distance (x) of the screw tip from the anterior edge of the vertebral body and the distance (y) (Fig. 5C) of the lumbar lateral screw tip from the endplate of the vertebral body after surgery and at the last follow-up. The criterion for CICPS loosening was as follows: [1] lateral CICPS displacement greater than 1 mm [9, 10] and [2] the appearance of a double ring sign around the CICPS [11]. Intraoperative bone cement leakage was classified as described by Yeom [12].

All the imaging data were measured by three experienced spinal surgeons on the INFINITT PACS System (INFINITT Healthcare Co., Ltd, Hangzhou, China), and the mean value was taken. All measurement data were recorded as the mean ± standard deviation. As the statistical indicators follow normal distribution, a paired T test was used to analyze the differences between key statistical indicators (VAS, ODI, H-y0, L- × 0, X, and Y), and data processing was completed with the IBM SPSS 25 version.

Results

Basic description of the surgery

All patients had lumbosacral pain of varying degrees, accompanied or not by pain and numbness in one or both lower limbs. Positive and lateral dynamic X-ray, CT and Magnetic Resonance Imaging (MRI) examination showed lumbar spondylolisthesis of different degrees of severity, accompanied by different degrees of severity of intervertebral disc herniation and spinal canal stenosis. All patients underwent conservative treatment for more than half a year without obvious improvement of symptoms. As shown in Table 1, the mean age of the patients was 60.11 ± 7.83 years old, the dual-energy X-ray bone density was -3.16 ± 0.59, the operative time was 223 ± 51 min, the intraoperative blood loss was 427 ± 277 ml, and the mean postoperative hospital stay was 6.2 ± 2.8 days. All the 37 patients were followed up, with an average follow-up time of 26.6 ± 13.4 months. A total of 124 CICPS and 36 pedicle screws were implanted in 37 patients, and 3 of the CICPSs were stopped due to bone cement leakage.

Effectiveness assessment of CICPS

The 37 patients had varying degrees of relief or even disappearance of symptoms after surgical treatment, and none of them had postoperative deterioration dysfunction or increase in pain relative to the preoperative state. During the follow-up, the VAS score of the patients was significantly decreased from 4.30 ± 1.58 before the surgery to 0.30 ± 0.70 (P < 0.001) at the last follow-up (Fig. 6A). The ODI index decreased from 47.27% ± 16.97% before surgery to 3.36% ± 5.70% at the last follow-up (P < 0.001) (Fig. 6B). Suggesting the clinical benefit of CICPS usage was definite.

Fig. 6
figure 6

Effectiveness assessment of CICPS. A VAS score before operation and at the last follow-up; B Preoperative ODI score and the last follow-up ODI score

Stability assessment of CICPS

The preoperative Taillard index L- × 0 of the patients decreased from 0.31 ± 0.02 to 0.07 ± 0.02 (P < 0.001) after surgery and to 0.07 ± 0.01 at the last follow-up (P < 0.001 compared with preoperative (Fig. 7A), indicating the intraoperative lifting and reduction effect of CICPS was efficient. No vertebral re-slip was observed at the last follow-up, and subsequently, the orthopaedic effect was stable.

Fig. 7
figure 7

Stability assessment of CICPS. A Taillard index changes before surgery, one week after surgery and during the last follow-up; B Changes of intervertebral space height before surgery, one week after surgery and at the last follow-up

The height of intervertebral space H-y0 increased from 7.78 ± 0.56 mm preoperatively to 12.33 ± 0.36 mm postoperatively (P < 0.001) and to 11.66 ± 0.35 mm at the last follow-up (P < 0.001) (Fig. 7B). Besides, positive and lateral radiographs of the lumbar spine showed continuous trabecular bone passing in all fusion segments, indicating successful fusion.

The postoperative X value of the patient was 4.92 ± 0.43 mm and at the last follow-up was 4.97 ± 0.42 mm (P = 0.563, Fig. 8A). Postoperative Y value was 7.44 ± 0.48 mm and at the last follow-up was 7.48 ± 0.48 mm (P = 0.657, Fig. 8B). The absolute difference between the postoperative measured values of X and Y and the final follow-up values of X and Y in all patients was less than 1 mm (Fig. 8C). No obvious double ring sign was observed around CICPS on the lumbar spine frontal radiography, indicating that the CICPS were not loosening or pulling out.

Fig. 8
figure 8

Stability assessment of CICPS. A The horizontal distance of CICPS from the apex to the anterior margin of the vertebra (X) after surgery and at final follow-up; B The vertical distance of CICPS from the apex to the superior endplate(Y) after surgery and at final follow-up; C Represents the absolute difference of x value and y value between the postoperative and the last follow-up

Complications

Intraoperative vertebral leakage occurred in 3 screws (2.42%), all of which were Yeom type S screws [12]. No symptomatic bone cement pulmonary embolism, deep vein thrombosis occurred after the operation. One patient suffered from a surgical incision infection, which was improved after antibiotic treatment.

Discussion

Spondylolysis-type lumbar spondylolisthesis occurs when facet joints lose their ability to prevent the upper vertebral bodies from sliding forward. Which leads to the increase of the burden of supraspinal ligaments and interspinous ligaments, and the acceleration of spinal degeneration. Finally, the function of the stable structure between vertebral bodies will lost, and vertebral slip can progress over the III degrees. Up to now, pedicle screw system is still the best surgical method to reconstruct the displaced vertebral body and perform bone fusion for patients with lumbar spondylolisthesis. However, the stability of pedicle screw would be reduced in osteoporosis vertebral bodies, as the thickness of trabeculae and bone cortex decrease and makes screws easy to loosen and pull out. Studies reported that pedicle screw stability is often inadequate for patients with bone mineral density (BMD) below 80 mg/cm3 [13], and the screw loosening rate in the osteoporotic vertebral body is up to 60% [14].

To improve the biomechanical strength of pedicle screw in osteoporotic vertebral body, studies have been conducted on adjusting the length, diameter, thread design and reinforcement material of pedicle screws [15,16,17,18,19,20]. Due to the limited size of pedicles and the presence of osteoporosis, most of the pedicle screws failed to achieve the desired effect. Recently, some studies reported that inflatable pedicle screws [21, 22] and bone cement screws [23, 24]achieved good internal fixation results in patients with osteoporosis associated spinal diseases. However, the efficacy of above-mentioned strategies was mostly limited to patients with mild osteoporosis (Meyerding classification of spondylolysis, level I and II), as cases of vertebral spondylolisthesis above level III were explicitly exclude [22, 23]. Hence, we focused on spondylolysis-type lumbar spondylolisthesis, which is more specific in this study.

Among the various strengthening measures for pedicle screws, PMMA was considered as the best choice owning to its high fixation strength and rapidly coagulation property [4, 25, 26], which meets the demands of intraoperative use of pedicle screws for lifting and reduction of the spondylolisthesis. For the design of bone cement-injectable screws, it is suggested that there would be higher bone cement leakage risk if the proximal side hole of the screw was closer to the nail head [27]. Hence, to improve the fixation strength and reduce the risk of bone cement leakage, the CICPS in this study were designed with three side holes in the front 2/5 of the screw which distributed longitudinally from small to large. Apart from 10 level I and 20 level II patients, we also proved the feasibility of CICPS in 6 level III cases and 1 level IV case (Table 1). No screw fracture was found in any of the follow-up cases, and the bone cement leakage rate was 2.42%, lower than that of similar types of injectable bone cement screws (5%-62.3%) [28], which greatly reduced the possibility of nerve injury.

Fatal pulmonary embolism caused by intraoperative using bone cement screws has been another problem. Insa Jansen [29]and Martin-Fernandez [30] analyzed 1330 and 1780 bone cement screws respectively, and found that most cases of pulmonary embolism occurred in thoracic or thoracolumbar internal fixation and that there was no pulmonary embolism or suspected cases of pulmonary embolism when simply using bone cement screws in lumbar vertebra. In this study, to reduce the risk of pulmonary embolism, only a small amount of bone cement (1–2 ml) was injected into each screw to achieve the fixation strength. As the results proved, the application of bone cement screws in the lumbosacral region alone would not cause pulmonary embolism, which further confirmed the safety of CICPSs in the treatment of lumbar spondylolisthesis (spondylolysis-type) with osteoporosis.

Because of the absence of revision cases in this study, the operability of removing revision screws after surgery is not yet known. After comparing general pedicle screws enhanced by PMMA with hollow side hole cement screws, CHAO et al. found no significant difference in pulling out the screws [31]. M. Martin-Fernandez also declared success in pulling out 180 bone cement reinforced injectable cement screws in various clinical applications [32].

The fusion rate is an important index of internal fixation [33]. In the imaging follow-up of the present study, postoperative L- × 0 was 0.07 ± 0.02 mm, close to final follow-up L- × 0 0.07 ± 0.01 mm (P > 0.05). The absolute difference between X and Y was less than 1 mm at the last follow-up, and there was no obvious double ring sign around CICPSs in the lumbar frontal radiography. All the above results indicated that no re-slip of the reduced vertebral body was observed, and no loosening or pulling out of CICPSs was observed.

Although postoperative loss of intervertebral height occurred, the difference was not statistically significant (Fig. 7B). We believe that there are two main reasons for postoperative intervertebral height loss. Firstly, patients are commonly complicated with osteoporosis, and the resistance of bone to pressure is relatively weakened. Studies demonstrated that osteoporosis is an important risk factor for the sinking of the interbody fusion apparatus [34,35,36]. Secondly, due to the incomplete reduction of the spondylolisthesis during the operation, the biomechanics of the spine changed and the local pressure increased, which also led to the loss of the postoperative intervertebral height. Although no sinking of the cage has been observed due to the relatively small number of cases in this study, the surgeon should still be vigilant for possible sinking of the cage.

This study has several limitations. Firstly, the sample size is small and the follow-up time is not long enough that longer follow-up duration with more cases is needed. Secondly, due to ethical limitations, here only clinical evaluation was performed. A controlled randomized study could be suggested for further study. Finally, standard operating procedure for patients of different degrees was not established, which will be our future research focus.

Conclusion

In this study, the clinical efficacy and safety of CICPS in the treatment of spondylolysis-type lumbar spondylolisthesis with osteoporosis was investigated, within an average follow-up time of 26.6 ± 13.4 months. Only 3 out of 124 CICPS (2.42%) had bone cement leakage, and no clinical discomfort was found in any patients. Both the VAS score and the ODI decreased significantly preoperative to postoperative. Indicating, CICPSs are safe and effective in the treatment of spondylolisthesis with osteoporosis.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

CICPS:

Cement-injectable cannulated pedicle screw

VAS:

Visual Analogue Scale

ODI:

Oswestry Disability Index

CPS:

Cannulated pedicle screw

PMMA:

Poly-methyl-methacrylate

CT:

Computed Tomography

MRI:

Magnetic Resonance Imaging

BMD:

Bone mineral density

References

  1. Farrokhi MR, Rahmanian A, Masoudi MS. Posterolateral versus posterior interbody fusion in isthmic spondylolisthesis. J Neurotrauma. 2012;29(8):1567–73.

    Article  PubMed  Google Scholar 

  2. Tsirikos AI, Sud A, McGurk SM. Radiographic and functional outcome of posterolateral lumbosacral fusion for low grade isthmic spondylolisthesis in children and adolescents. Bone Joint J. 2016;98-B(1):88–96.

    Article  CAS  PubMed  Google Scholar 

  3. Galbusera F, Volkheimer D, Reitmaier S, Berger-Roscher N, Kienle A, Wilke HJ. Pedicle screw loosening: a clinically relevant complication? Eur Spine J. 2015;24(5):1005–16.

    Article  PubMed  Google Scholar 

  4. Frankel BM, Jones T, Wang CJSJ. Segmental polymethylmethacrylate-augmented pedicle screw fixation in patients with bone softening caused by osteoporosis and metastatic tumor involvement: a clinical evaluation. Neurosurgery. 2007;61(3):531–7; discussion 537-8.2006;6(5):52S–3S.

  5. Liu Y, Xu J, Sun D, Luo F, Zhang Z, Dai F. Biomechanical and finite element analyses of bone cement-Injectable cannulated pedicle screw fixation in osteoporotic bone. J Biomed Mater Res B Appl Biomater. 2016;104(5):960–7.

    Article  CAS  PubMed  Google Scholar 

  6. Dai F, Liu Y, Zhang F, Sun D, Luo F, Zhang Z, et al. Surgical treatment of the osteoporotic spine with bone cement-injectable cannulated pedicle screw fixation: technical description and preliminary application in 43 patients. Clinics (Sao Paulo). 2015;70(2):114–9.

    Article  Google Scholar 

  7. Hough S. The diagnosis of osteoporosis. J Endocrinol Metab Diabetes S Afr. 2012;17(2):67–8.

    Google Scholar 

  8. Lee C, Dorcil J, Radomisli TE. Nonunion of the spine: a review. Clin Orthop Relat Res. 2004;419(419):71–5.

    Article  Google Scholar 

  9. Wu JC, Huang WC, Tsai HW, Ko CC, Wu CL, Tu TH, et al. Pedicle screw loosening in dynamic stabilization: incidence, risk, and outcome in 126 patients. Neurosurg Focus. 2011;31(4):E9.

    Article  PubMed  Google Scholar 

  10. Ko CC, Tsai HW, Huang WC, Wu JC, Chen YC, Shih YH, et al. Screw loosening in the Dynesys stabilization system: radiographic evidence and effect on outcomes. Neurosurg Focus. 2010;28(6):E10.

    Article  PubMed  Google Scholar 

  11. Dakhil-Jerew F, Jadeja H, Cohen A, Shepperd JJ. Inter-observer reliability of detecting Dynesys pedicle screw using plain X-rays: a study on 50 post-operative patients. Eur Spine J. 2009;18(10):1486–93.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Yeom JS, Kim WJ, Choy WS, Lee CK, Chang BS, Kang JW. Leakage of cement in percutaneous transpedicular vertebroplasty for painful osteoporotic compression fractures. J Bone Joint Surg Br. 2003;85(1):83–9.

    Article  CAS  PubMed  Google Scholar 

  13. Weiser L, Huber G, Sellenschloh K, Viezens L, Puschel K, Morlock MM, et al. Insufficient stability of pedicle screws in osteoporotic vertebrae: biomechanical correlation of bone mineral density and pedicle screw fixation strength. Eur Spine J. 2017;26(11):2891–7.

    Article  PubMed  Google Scholar 

  14. El Saman A, Meier S, et al. Reduced loosening rate and loss of correction following posterior stabilization with or without PMMA augmentation of pedicle screws in vertebral fractures in the elderly. Eur J Trauma Emerg Surg. 2013;39(5):455–60.

    Article  PubMed  Google Scholar 

  15. Polly DW Jr, Orchowski JR, Ellenbogen RG. Revision pedicle screws. Bigger, longer shims–what is best? Spine (Phila PA 1976). 1998;23(12):1374–9.

    Article  Google Scholar 

  16. Hirano T, Hasegawa K, Washio T, Hara T, Takahashi H. Fracture risk during pedicle screw insertion in osteoporotic spine. J Spinal Disord. 1998;11(6):493–7.

    Article  CAS  PubMed  Google Scholar 

  17. Zindrick MR, Wiltse LL, Widell EH, Thomas JC, Holland WR, Field BT, et al. A biomechanical study of intrapeduncular screw fixation in the lumbosacral spine. Clin Orthop Relat Res. 1986;203(203):99–112.

    Google Scholar 

  18. Hasegawa T, Inufusa A, Imai Y, Mikawa Y, Lim TH, An HS. Hydroxyapatite-coating of pedicle screws improves resistance against pull-out force in the osteoporotic canine lumbar spine model: a pilot study. Spine J. 2005;5(3):239–43.

    Article  PubMed  Google Scholar 

  19. Asnis SE, Ernberg JJ, Bostrom MP, Wright TM, Harrington RM, Tencer A, et al. Cancellous bone screw thread design and holding power. J Orthop Trauma. 1996;10(7):462–9.

    Article  CAS  PubMed  Google Scholar 

  20. Krenn MH, Piotrowski WP, Penzkofer R, Augat P. Influence of thread design on pedicle screw fixation. Laboratory investigation. J Neurosurg Spine. 2008;9(1):90–5.

    Article  PubMed  Google Scholar 

  21. Cook SD, Barbera J, Rubi M, Salkeld SL, Whitecloud TS. Lumbosacral fixation using expandable pedicle screws. an alternative in reoperation and osteoporosis. Spine J. 2001;1(2):109–14.

    Article  CAS  PubMed  Google Scholar 

  22. Gazzeri R, Roperto R, Fiore C. Surgical treatment of degenerative and traumatic spinal diseases with expandable screws in patients with osteoporosis: 2-year follow-up clinical study. J Neurosurg Spine. 2016;25(5):610–9.

    Article  PubMed  Google Scholar 

  23. Lubansu A, Rynkowski M, Abeloos L, Appelboom G, Dewitte O. Minimally Invasive Spinal Arthrodesis in Osteoporotic Population Using a Cannulated and Fenestrated Augmented Screw: Technical Description and Clinical Experience. Minim Invasive Surg. 2012;2012:507826.

    PubMed  PubMed Central  Google Scholar 

  24. Liu YY, Xiao J, Yin X, Liu MY, Zhao JH, Liu P, et al. Clinical efficacy of bone cement-injectable cannulated pedicle screw short segment fixation for lumbar spondylolisthesis with Osteoporosise. Sci Rep. 2020;10(1):3929.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  25. Elder BD, Lo SF, Holmes C, Goodwin CR, Kosztowski TA, Lina IA, et al. The biomechanics of pedicle screw augmentation with cement. Spine J. 2015;15(6):1432–45.

    Article  PubMed  Google Scholar 

  26. Hamasaki T, Tanaka N, Kim J, Okada M, Ochi M, Hutton WC. Pedicle screw augmentation with polyethylene tape: a biomechanical study in the osteoporotic thoracolumbar spine. J Spinal Disord Tech. 2010;23(2):127–32.

    Article  PubMed  Google Scholar 

  27. Chen LH, Tai CL, Lai PL, Lee DM, Tsai TT, Fu TS, et al. Pullout strength for cannulated pedicle screws with bone cement augmentation in severely osteoporotic bone: influences of radial hole and pilot hole tapping. Clin Biomech (Bristol, Avon). 2009;24(8):613–8.

    Article  Google Scholar 

  28. Chang MC, Kao HC, Ying SH, Liu CL. Polymethylmethacrylate augmentation of cannulated pedicle screws for fixation in osteoporotic spines and comparison of its clinical results and biomechanical characteristics with the needle injection method. J Spinal Disord Tech. 2013;26(6):305–15.

    Article  PubMed  Google Scholar 

  29. Janssen I, Ryang YM, Gempt J, Bette S, Gerhardt J, Kirschke JS, Meyer B. Risk of cement leakage and pulmonary embolism by bone cement-augmented pedicle screw fixation of the thoracolumbar spine. Spine J. 2017;17(6):837–44.

    Article  PubMed  Google Scholar 

  30. Martín-Fernández M, López-Herradón A, Piñera AR, Tomé-Bermejo F, Duart JM, Vlad MD, et al. Potential risks of using cement-augmented screws for spinal fusion in patients with low bone quality. Spine J. 2017;17(8):1192–99.

  31. Chao KH, Lai YS, Chen WC, Chang CM, Cheng CKJME, Physics. Biomechanical analysis of different types of pedicle screw augmentation: a cadaveric and synthetic bone sample study of instrumented vertebral specimens. Med Eng Phys. 2013;35(10):1506–12.

  32. Martín-Fernández M, López-Herradón A, Piñera AR, Tomé-Bermejo F, Duart JM, Vlad MD, et al. Potential risks of using cement-augmented screws for spinal fusion in patients with low bone quality. Spine J. 2017:S1529943017301869.

  33. Harrington JF Jr, Park MC. Single level arthrodesis as treatment for midcervical fracture subluxation: a cohort study. J Spinal Disord Tech. 2007;20(1):42–8.

    Article  PubMed  Google Scholar 

  34. Cheng C, Wang K, Zhang C, Wu H, Jian F. Clinical results and complications associated with oblique lumbar interbody fusion technique. Ann Transl Med. 2021;9(1):16.

    Article  PubMed  PubMed Central  Google Scholar 

  35. Jones BC, Okano I, Salzmann SN, Reisener MMJ, Chiapparelli E, Shue MJ, et al. Endplate volumetric bone mineral density is a predictor for cage subsidence following lateral lumbar interbody fusion: a risk factor analysis - ScienceDirect. Spine J. 2021;21(10):1729–37.

    Article  PubMed  Google Scholar 

  36. He W, He D, Sun Y, Xing Y, Wen J, Wang W, et al. Standalone oblique lateral interbody fusion vs. combined with percutaneous pedicle screw in spondylolisthesis. BMC Musculoskelet Disord. 2020;21(1):184.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

None.

Funding

This work was supported by grants from “New clinical technology of grade III of Army Medical University” (No. 2019–3-1–34).

Author information

Authors and Affiliations

Authors

Contributions

Fei Dai and Ting-ting Zheng designed/performed most of the investigation, data analysis, and revised the draft. Lei Song and Jun Xiao performed the operations, imaging analyze, and wrote the manuscript; Rui Zhou and Cong-can Li contributed to acquisition and interpretation of the data and analyses. All of the authors have read and approved the manuscript.

Corresponding authors

Correspondence to Ting-ting Zheng or Fei Dai.

Ethics declarations

Ethics approval and consent to participate

The study was approved by the Southwest Hospital ethics committee (KY201927). All patients included in this study gave their informed consents.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Song, L., Xiao, J., Zhou, R. et al. Clinical evaluation of the efficacy of a new bone cement-injectable cannulated pedicle screw in the treatment of spondylolysis-type lumbar spondylolisthesis with osteoporosis: a retrospective study. BMC Musculoskelet Disord 23, 951 (2022). https://doi.org/10.1186/s12891-022-05904-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-022-05904-7

Keywords