Biomechanical analysis of two medial buttress plate fixation methods to treat Pauwels type III femoral neck fractures

Abstract

Background

Femoral neck fractures in young people are usually Pauwels Type III fractures. The common treatment method are multiple parallel cannulated screws or dynamic hip screw sliding compression fixation. Due to the huge shear stress, the rate of complications such as femoral head necrosis and nonunion is still high after treatment. The aim of our study was to compare the stabilities of two fixation methods in fixating pauwels type III femoral neck fractures.

Methods

All biomimetic fracture samples are fixed with three cannulated screws combined with a medial buttress plate. There were two fixation groups for the buttress plate and proximal fracture fragment: Group A, long screw (40 mm); Group B, short screw (6 mm). Samples were subjected to electrical strain measurement under a load of 500 N, axial stiffness was measured, and then the samples were axially loaded until failure. More than 5 mm of displacement or synthetic bone fracture was considered as construct failure.

Results

There were no significant differences in failure load (P = 0.669), stiffness (P = 0.842), or strain distribution (P > 0.05) between the two groups.

Conclusions

Unicortical short screws can provide the same stability as long screws for Pauwels Type III Femoral Neck Fractures.

Peer Review reports

Introduction

Femoral neck fractures are fairly common in clinical work, but they are rare in young patients (20–55 years old), who account for only about 3–5% of all cases [1, 2]. Femoral neck fractures in young people are typically high-energy vertical shear injuries, and most are Pauwels type III fractures [3]. Pauwels’ 1935 publication was the first biomechanical classification of femoral neck fractures [4]. Although there has been some debate about the exact angles that define the categories, the underlying theoretical principle is that the more vertical the fracture line, the greater the shear force it bears, and the greater the risk of complications such as nonunion; Pauwels’ III fracture (Pauwels angle > 50°) is the most vertical [5]. Due to the large vertical shear force, the stable fixation of Pauwels III femoral neck fracture is very difficult, and the prognosis is unacceptable [6]. Arthroplasty is usually quickly ruled out for younger patients since implant normally don’t endure more than 20 years, and the surgery is associated with multiple complications including infections and aseptic loosening [7, 8]. The best fixation strategy for this fracture remains controversial [9]. The currently available internal fixation methods for femoral neck fracture include cannulated screws, dynamic hip screw (DHS), cephalomedullary nails, and proximal femoral locking plates [10,11,12]. However, the postoperative failure rate of these fractures is very high, ranging from 20 to 80% [13,14,15]. Shear force is dominant in femoral neck vertical fractures, so the internal fixator must be able to resist these shearing forces during bone healing [16]. The DHS was developed to address this issue, but their use is associated with an increased risk of osteonecrosis [17, 18]. Cannulated screws remain the most promising and commonly used devices because of their minimal invasiveness, easy handling, and ability to induce dynamic compression [19].

To improve fixation stability, Mir et al. [20] proposed the use of a medial buttress plate to treat femoral neck fractures in young adults based on previous studies. A medial buttress plate can clamp the fracture apex, neutralize shearing forces, and transfer them into compressive forces into the plane of cannulated screws or other typical construct [21]. Kunapuli and colleagues [22] first studied the strength of augmented versus nonaugmented medial buttress plates for stabilizing vertical shear femoral neck fractures and found that augmentation increased the load to failure by 183% and improved the construct’s stiffness by 35%. Li et al. [16] used finite element analysis to compare outcomes for the combination of medial buttress plate with cannulated screws to those of cannulated screws alone. The medial buttress plate combination can provide greater stability, which indicates better healing of femoral neck fractures. A recent clinical study by Ye et al. [23] reviewed the results of 28 vertical femoral neck fractures treated with cannulated screws augmented with buttress plates and found that augmentation of cannulated screw fixation resulted in an 89% union rate, which was significantly higher than the use of cannulated screws alone. As a result, conventional sliding compression fixation combined with a medial buttress plate has become the latest treatment for young patients with Pauwels III femoral neck fractures. Despite good results with this approach, there is no uniform criterion for the fixation of proximal fracture block. On one hand, a unicortical long screw may be more stable, but it will interfere with cannulated screw placement and bring great inconvenience to the operator; the single unicortical screw has little effect on cannulated screws, but it is not clear whether it offers sufficient stability is enough.

In this study, strain at different positions was measured to study the mechanical performance of two fixation methods, in addition to recording the vertical load and displacement. The bonding strain gauge (SG) was first reported in 1938, followed a few years later by the first application to bone biomechanics [24]. SGs have been widely used in bone biomechanics since; Because of their accuracy and high-frequency response, they are still the gold standard for bone strain measurement [25]. The working principle of SG is to read the increase or decrease in resistance of metal materials when they are elongated/contract by external tensile/compressive force [26].

This study tested the hypothesis that compared with a unicortical long screw, a unicortical short screw can provide the same stability. We compared the biomechanical performance of two fixation methods of cannulated screws combined with medial buttress plates on synthetic bone and assessed the biomechanical properties of these fixation methods by strain electrical measurement to provide a theoretical reference for clinical operation method selection.

Materials and methods

Specimen preparation

A total of 14 synthetic bone models of the right femur (Model 2200, Synbone, Zizers, Switzerland) were used for this biomechanical analyses, which allowed us to increase the sample size compared to a cadaveric study. The artificial femurs are covered with a thin layer of cortex and filled with low-density polyurethane foam corresponding to cortical and cancellous bone. The mechanical properties and cortical bone thickness are representative of cadaveric bone from younger patients [3]. The models were 455 mm in length, with a 135° neck shaft angle, 15° anteversion, 48 mm head diameter, and a minimum femoral shaft diameter of 28 mm. A protractor was used to draw the fracture line under the head of the femur (60° to the line of the horizontal). The specimen was osteotomized with a band saw, and the cross-section was perpendicular to the neck axis to simulate a Pauwels type III fracture. All samples were initially implanted with cannulated screws under fluoroscopic guidance before osteotomy to facilitate anatomical reduction. The medial buttress plate was implanted after osteotomy, because the femoral neck will be shortened. If the medial buttress plate is pre-drilled and initially implanted before osteotomy, then the broken end of the fracture will not be completely attached after reimplantation of the medial buttress plate, and anatomical reduction will not be achieved.

Fourteen synthetic bone models were divided into two groups and subjected to different fixation methods (Fig. 1). All samples were fixed with three 7.0-mm cannulated screws with a triangular construction. The lowermost screw was positioned in the femoral calcar region, with the starting point above the lesser trochanter. The remaining two were located above, near the anterior or posterior cortex of the femoral neck. All screws are located 5 mm from the subchondral bone of the femoral head. A 3.5 mm thick locking plate with four holes was bent accordingly to fit the lower edge of the neck of the femur. At the distal end of the fracture line, two bicortical screws were used to fix the plate to the femur at the distal end of the fracture line. The two groups were treated as follows: Group A, the proximal end of the fracture line was fixed to the femoral head with a long unicortical screw (length 40 mm); Group B, the proximal fragment was fixed with a short unicortical screw (length 6 mm).

Fig. 1
figure 1

Pauwels III fractures were simulated via osteotomy by creating a fracture oriented 60° (*) from the horizonta. The two groups all used three cannulated screws combined with medial buttress plate fixation, but the fixation of the proximal fracture block was different: A) long screws with a length of 40 mm, B) Short screw with a length of 6 mm

Synthetic femoral models were cut from the middle diaphyseal region to a final size of 200.0 mm. The distal end was firmly fixed with polymethylmethacrylate, and each specimen was oriented in 15° of adduction in the coronal plane and aligned vertically in the sagittal plane to imitate the standing posture of one leg. Adduction of 15° simulates the physiological load of the proximal femur during standing on one leg of the gait and has been used in previous biomechanical studies [27]. The bone surface was cross-smoothed with sandpaper along the SG axis at 45° [28]. The surface was cleaned with ethanol followed by acetone and allowed to dry naturally [29]. SGs (BX120-3AA, TESTING INSTRUMENT FACTORY HUANGYAN ZHEJIANG, Zhengjiang, China) were attached in the preset position of the sample with cyanoacrylate adhesive (T-1, Beihua Chemical Works, Beijing, China). The lead of the SG was welded to the adjacent terminal and connected to a strainmeter by a wire cable. The wiring method is quarter bridge, and the temperature compensation is external. The SG position distribution is shown in Fig. 2.

Fig. 2
figure 2

The location of the strain gauge

Testing

We used an MTS model 55,100 material testing machine (Material Testing Systems, MTS Systems Corp., Eden Prairie, MN, USA) with a 1000.0 KN capacity for vertical load testing. The load is transmitted through the load sensor, and the displacement is transmitted through the tie rod displacement transducer (KTR-12, MIRAN, Guangzhou, China). The displacement transducer (range 15 mm, resolution 0.01 mm) measures the relative displacement between fracture fragments under load; we chose this approach over vertical displacement because it represents the overall displacement of the model rather than the relative displacement of the fracture fragments (Fig. 3). Before starting the measurement, the specimen was preloaded to 100 N to provide close contact with the biomechanical testing machine and reduce error due to the elastic creep effect [30]. Three biomechanical experiments were carried out. 1) Strain measurement: the machine provides a vertical compression load with a peak value of 500 N at a rate of 2 mm/min and records the strain value of each measurement point at 500 N. 2) Stiffness test: a vertical compression load is provided at a device movement rate of 2 mm/min until 500 N, and use the slope of the load-displacement curve in the linear elastic region to calculate the axial stiffness. 3) Failed load test: the machine provides a vertical compression load at a rate of 2 mm/min until specimen failure. The peak load over this duration represents the structural collapse of the fixation construct and is referred to as failure load. The construct failure modes were recorded. Failure was defined as 5-mm linear or synthetic bone fracture as previously reported [10, 31, 32]. The justification for this threshold was given by Alho et al. [33], who found that a change in fracture position by 5 mm at 3-month follow-up was strongly associated with local complications and the need for revision surgery.

Fig. 3
figure 3

Synthetic femoral models were cut from the middle diaphyseal region with a final size of 200.0 mm. The distal end was firmly fixed with polymethylmethacrylate to each specimen was oriented in 15° (*) of of adduction in the coronal plane. The displacement transducer (#) measures the relative displacement between the fracture fragments under load

Statistical analysis

All statistical calculations were carried out using SPSS (SPSS Version 26; SPSS Inc., Chicago, IL, USA). Descriptive analysis of the data was performed after confirming the normality of the data (Kolmogorov–Smirnov test). Unpaired Student’s t-test.

was used to analyze the data from the two structural groups to and identify differences in the maximum load and stiffness. Post-hoc pair-wise comparisons were made with Holm-Sidak tests. In all cases, differences were considered significant at P < 0.05.

Results

We measured the strain distributions, displacements, and loads for two types of internal fixation. The values are listed in Table 1 and Table 2.

Table 1 Stress distribution of each group under 500 N load (Gpa)
Table 2 Biomechanical properties of the three fixation techniques

Strain measurement

We referred to the literature to calculate the stress value of each measuring point with Hooke’s law [19]: σ = E ε (where σ is stress, ε is strain, and E is the elastic modulus). The calculated elastic modulus was 1076. In terms of strain distribution, there was no significant difference between the two groups (P > 0.05). The stress at the joint between the medial buttress plate and screw was greater than the stress near the cannulated screws, indicating that that the plate can reduce stress. The maximum stress and strain occurred at measuring points 13, 14, 15, and 16; the addition of the medial buttress plate improved shear force resistance for fracture stability and healing. Less stress was measured near the hollow lag screw hole (5, 6, 7, and 8). The minimum stress and strain were at measuring points 3, 9 and 10 (Table 1).

Failure load and stiffness

Among the 14 specimens, catastrophic failure occurred in 2, and the other 12 had displaced proximal fractures > 5 mm. The average maximum loads of Group A and B were 1968.09 N (SD = 227.29 N), 1911.86 N (SD = 215.55 N) respectively. There was no statistically significant difference in the average maximum load among the two groups (P = 0.669). The average stiffnesses of groups A and B were 390.70 N/mm (SD = 91.2 N/mm), 390.70 N/mm (SD = 104.99 N/mm) respectively (P = 0.842).

Discussion

Femoral neck fractures in young adults are usually the result of high-energy trauma [34], and they are associated with high rates of postoperative osteonecrosis, nonunion, malunion, and revision surgery [13, 35, 36]. Anatomical reduction and stable internal fixation are necessary for satisfactory treatment of femoral neck fractures [37]. Some biomechanical studies reported that constructs augmented with medial buttress fixation had significantly higher stiffness and load to failure [22, 23, 38, 39]. However, the most suitable medial buttress plate fixation method is controversial. The present study evaluated and compared different kinds of internal fixation for Pauwels type III femoral neck fractures. Fourteen fracture models were generated to analyze two fixation styles under identical loading conditions. The maximum loads and strain distributions were compared to investigate biomechanical differences [38].

In terms of the fixation screws for the proximal bone block, there were no significant differences in the mechanical results when we compared the unicortical long screw and the short screw. That is, a unicortical short screw provided the same stability as a unicortical long screw.

The strain measurement findings revealed that the length of the screw at the proximal end of the steel plate did not obviously influence the strain distribution, which was not significantly different among the two groups. The medial buttress plate can provide an additional path to transfer force between the fractured fragments, which dissipates stress on cannulated screws and the 12 and 13 measuring points. Correspondingly, the stress increased and strain concentration occurred at the joint of the medial buttress plate and screws (13, 14, 15, and 16 measuring points; Table 1). Catastrophic failure occurred in 2 samples, and the location of the fracture corresponded to the stress concentration location (Fig. 4). Similarly, Zeng et al. performed finite element analysis and showed that strain concentration occurred at the joint between the medial buttress plate and screws [38]. However, strain is the major determinant of crack initiation and development before bone fracture [40], suggesting that stress concentrated at the junction of the plate and screw may cause periprosthetic fractures.

Fig. 4
figure 4

The sample fractured during the loading process. The fracture site is near the 15 and 16 measuring points, which represents the place where the stress is concentrated

In this present study, we also found no difference in the maximum loads and stiffnesses of Groups A and B. In other words, a proximal fracture can be fixed with a short screw to achieve the same stability as with a long screw. The reason may be that the cortical bone has typical mechanical load-bearing characteristics; screw penetration through the cortical bone is the key, whereas screw length does not dramatically affect stability [41].

Moreover, the fixation of proximal femoral neck fractured fragments with unicortical short screw has many advantages compared with unicortical long screw. First, the unicortical short screw will not interfere with cannulated screw placement, making the operation more convenient for the surgeon. Secondly, the unicortical short screw will not enter the hip joint cavity, prevents the screw from being too long to damage the articular surface. Finally, the unicortical short screw minimizes the damage to the blood supply of the femoral head.

There are some limitations of the current study. First, we employed synthetic bones rather than cadaveric bones, so the results do not directly translate to the anatomy of the femoral trabeculae and the forces they can withstand. The artificial setting also does not truly reproduce the manner in which this fracture develops. Synthetic bones also eliminate patient variables and ensured that the biomechanical properties were comparable between groups. Second, femoral neck fractures were idealized by making smooth saw cuts perpendicular to the neck axis. This may not accurately simulate the jagged features of the interface between bone fragments in clinical situations. Specifically, the lower interface roughness may have underestimated stiffness and strength values. Thirdly, we lengthened the screw channel at the proximal end of the buttress plate by 2 mm, but it remains unclear how much enlargement is optimal.

Conclusions

Our study found no statistically significant difference in axial load models between unicortical short locking screws (or combined screw channel enlargement) and unicortical long locking screws. The application of a unicortical short screw fixes the proximal end of the medial buttress plate and expands the screw path to the proximal end, which might be a viable surgical plan.

Availability of data and materials

All data generated or analyzed during this study are included in this article.

References

  1. Hedlund R, Lindgren U, Ahlbom A. Age- and sex-specific incidence of femoral neck and trochanteric fractures. An analysis based on 20,538 fractures in Stockholm County, Sweden, 1972-1981. Clin Orthop Relat Res. 1987;(222):132–9.

  2. Zetterberg C, Elmerson S, Andersson GB. Epidemiology of hip fractures in Göteborg, Sweden, 1940-1983. Clin Orthop Relat Res. 1984;(191):43–52.

  3. Kemker B, Magone K, Owen J, Atkinson P, Martin S, Atkinson T. A sliding hip screw augmented with 2 screws is biomechanically similar to an inverted triad of cannulated screws in repair of a Pauwels type-III fracture. Injury. 2017;48(8):1743–8.

    Article  CAS  Google Scholar 

  4. Bartonícek J. Pauwels' classification of femoral neck fractures: correct interpretation of the original. J Orthop Trauma. 2001;15(5):358–60.

    Article  Google Scholar 

  5. Liporace F, Gaines R, Collinge C, Haidukewych GJ. Results of internal fixation of Pauwels type-3 vertical femoral neck fractures. J Bone Joint Surg Am. 2008;90(8):1654–9.

    Article  Google Scholar 

  6. Collinge CA, Mir H, Reddix R. Fracture morphology of high shear angle "vertical" femoral neck fractures in young adult patients. J Orthop Trauma. 2014;28(5):270–5.

    Article  Google Scholar 

  7. Bhandari M, Devereaux PJ, Swiontkowski MF, Tornetta P 3rd, Obremskey W, Koval KJ, et al. Internal fixation compared with arthroplasty for displaced fractures of the femoral neck. A meta-analysis. J Bone Joint Surg Am. 2003;85(9):1673–81.

    Article  Google Scholar 

  8. Alolabi B, Bajammal S, Shirali J, Karanicolas PJ, Gafni A, Bhandari M. Treatment of displaced femoral neck fractures in the elderly: a cost-benefit analysis. J Orthop Trauma. 2009;23(6):442–6.

    Article  Google Scholar 

  9. Luttrell K, Beltran M, Collinge CA. Preoperative decision making in the treatment of high-angle "vertical" femoral neck fractures in young adult patients. An expert opinion survey of the Orthopaedic trauma Association's (OTA) membership. J Orthop Trauma. 2014;28(9):e221–5.

    Article  Google Scholar 

  10. Aminian A, Gao F, Fedoriw WW, Zhang LQ, Kalainov DM, Merk BR. Vertically oriented femoral neck fractures: mechanical analysis of four fixation techniques. J Orthop Trauma. 2007;21(8):544–8.

    Article  Google Scholar 

  11. Li J, Zhao Z, Yin P, Zhang L, Tang P. Comparison of three different internal fixation implants in treatment of femoral neck fracture-a finite element analysis. J Orthop Surg Res. 2019;14(1):76.

    Article  Google Scholar 

  12. Burke NG, Kennedy J, Cousins G, Fitzpatrick D, Mullett H. Locking plate fixation with and without inferomedial screws for proximal humeral fractures: a biomechanical study. J Orthop Surg (Hong Kong). 2014;22(2):190–4.

    Article  Google Scholar 

  13. Damany DS, Parker MJ, Chojnowski A. Complications after intracapsular hip fractures in young adults. A meta-analysis of 18 published studies involving 564 fractures. Injury. 2005;36(1):131–41.

    CAS  PubMed  Google Scholar 

  14. Pauyo T, Drager J, Albers A, Harvey EJ. Management of femoral neck fractures in the young patient: a critical analysis review. World J Orthop. 2014;5(3):204–17.

    Article  Google Scholar 

  15. Kofoed H. Femoral neck fractures in young adults. Injury. 1982;14(2):146–50.

    Article  CAS  Google Scholar 

  16. Li J, Yin P, Zhang L, Chen H, Tang P. Medial anatomical buttress plate in treating displaced femoral neck fracture a finite element analysis. Injury. 2019;50(11):1895–900.

    Article  Google Scholar 

  17. Nauth A, Creek AT, Zellar A, Lawendy A-R, Dowrick A, Gupta A, et al. Fracture fixation in the operative management of hip fractures (FAITH): an international, multicentre, randomised controlled trial. Lancet. 2017;389(10078):1519–27.

    Article  Google Scholar 

  18. Zhang YL, Chen S, Ai ZS, Gao YS, Mei J, Zhang CQ. Osteonecrosis of the femoral head, nonunion and potential risk factors in Pauwels grade-3 femoral neck fractures: a retrospective cohort study. Medicine. 2016;95(24):e3706.

    Article  Google Scholar 

  19. Ben S, Zhao J, Rabczuk T. Does Hooke's law work in helical nanosprings? Phys Chem Chem Phys. 2015;17(32):20990–7.

    Article  CAS  Google Scholar 

  20. Mir H, Collinge C. Application of a medial buttress plate may prevent many treatment failures seen after fixation of vertical femoral neck fractures in young adults. Med Hypotheses. 2015;84(5):429–33.

    Article  Google Scholar 

  21. Stacey SC, Renninger CH, Hak D, Mauffrey C. Tips and tricks for ORIF of displaced femoral neck fractures in the young adult patient. Eur J Orthop Surg Traumatol. 2016;26(4):355–63.

    Article  Google Scholar 

  22. Kunapuli SC, Schramski MJ, Lee AS, Popovich JM Jr, Cholewicki J, Reeves NP, et al. Biomechanical analysis of augmented plate fixation for the treatment of vertical shear femoral neck fractures. J Orthop Trauma. 2015;29(3):144–50.

    Article  Google Scholar 

  23. Ye Y, Chen K, Tian K, Li W, Mauffrey C, Hak DJ. Medial buttress plate augmentation of cannulated screw fixation in vertically unstable femoral neck fractures: surgical technique and preliminary results. Injury. 2017;48(10):2189–93.

    Article  Google Scholar 

  24. Gurdjian ES, Lissner HR. Mechanism of head injury as studied by the cathode ray oscilloscope preliminary report. J Neurosurg. 1944;1(6):393–9.

    Article  Google Scholar 

  25. Grassi L, Isaksson H. Extracting accurate strain measurements in bone mechanics: a critical review of current methods. J Mech Behav Biomed Mater. 2015;50:43–54.

    Article  Google Scholar 

  26. Roberts VL. Strain-gage techniques in biomechanics. Exp Mech. 1966;6(3):19A–22A.

    Article  Google Scholar 

  27. Kuzyk PR, Zdero R, Shah S, Olsen M, Waddell JP, Schemitsch EH. Femoral head lag screw position for cephalomedullary nails: a biomechanical analysis. J Orthop Trauma. 2012;26(7):414–21.

    Article  Google Scholar 

  28. Aamodt A, Lund-Larsen J, Eine J, Andersen E, Benum P, Husby OS. Changes in proximal femoral strain after insertion of uncemented standard and customised femoral stems. An experimental study in human femora. J Bone Joint Surg Br. 2001;83(6):921–9.

    Article  CAS  Google Scholar 

  29. Basso T, Klaksvik J, Syversen U, Foss OA. A biomechanical comparison of composite femurs and cadaver femurs used in experiments on operated hip fractures. J Biomech. 2014;47(16):3898–902.

    Article  Google Scholar 

  30. Kim YH, Kim JS, Cho SH. Strain distribution in the proximal human femur. An in vitro comparison in the intact femur and after insertion of reference and experimental femoral stems. J Bone Joint Surg Br. 2001;83(2):295–301.

    Article  CAS  Google Scholar 

  31. Husby T, Alho A, Hoiseth A, Fonstelien E. Strength of femoral neck fracture fixation. Comparison of six techniques in cadavers. Acta Orthop Scand. 1987;58(6):634–7.

    Article  CAS  Google Scholar 

  32. Baitner AC, Maurer SG, Hickey DG, Jazrawi LM, Kummer FJ, Jamal J, et al. Vertical shear fractures of the femoral neck. A biomechanical study. Clin Orthop Relat Res. 1999;367:300–5.

    Article  Google Scholar 

  33. Filipov O, Stoffel K, Gueorguiev B, Sommer C. Femoral neck fracture osteosynthesis by the biplane double-supported screw fixation method (BDSF) reduces the risk of fixation failure: clinical outcomes in 207 patients. Arch Orthop Trauma Surg. 2017;137(6):779–88.

    Article  Google Scholar 

  34. Giordano V, Paes RP, Alves DD, Amaral AB, Belangero WD, Giordano M, et al. Stability of L-shaped and inverted triangle fixation assemblies in treating Pauwels type II femoral neck fracture: a comparative mechanical study. Eur J Orthop Surg Traumatol. 2018;28(7):1359–67.

    Article  Google Scholar 

  35. Slobogean GP, Sprague SA, Scott T, Bhandari M. Complications following young femoral neck fractures. Injury. 2015;46(3):484–91.

    Article  CAS  Google Scholar 

  36. Asnis SE, Wanek-Sgaglione L. Intracapsular fractures of the femoral neck. Results of cannulated screw fixation. J Bone Joint Surg Am. 1994;76(12):1793–803.

    Article  CAS  Google Scholar 

  37. Panteli M, Rodham P, Giannoudis PV. Biomechanical rationale for implant choices in femoral neck fracture fixation in the non-elderly. Injury. 2015;46(3):445–52.

    Article  Google Scholar 

  38. Zeng W, Liu Y, Hou X. Biomechanical evaluation of internal fixation implants for femoral neck fractures: a comparative finite element analysis. Comput Methods Prog Biomed. 2020;196:105714.

    Article  Google Scholar 

  39. Giordano V, Alves DD, Paes RP, Amaral AB, Giordano M, Belangero W, et al. The role of the medial plate for Pauwels type III femoral neck fracture: a comparative mechanical study using two fixations with cannulated screws. J Exp Orthop. 2019;6(1):18.

    Article  Google Scholar 

  40. Wolfram U, Schwiedrzik J. Post-yield and failure properties of cortical bone. Bonekey Rep. 2016;5:829.

    Article  Google Scholar 

  41. Augat P, Schorlemmer S. The role of cortical bone and its microstructure in bone strength. Age Ageing. 2006;35(Suppl 2):ii27–31.

    Article  Google Scholar 

Download references

Acknowledgements

None.

Funding

There was no funding of this study.

Author information

Authors and Affiliations

Authors

Contributions

P CD, L JC, and L ZW conceived the idea; L JC and L ZW performed research, D J conducted the analyses; all authors contributed to the writing and revisions. The author(s) read and approved the final manuscript.

Corresponding author

Correspondence to Chengdong Piao.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

None.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Liu, J., Li, Z., Ding, J. et al. Biomechanical analysis of two medial buttress plate fixation methods to treat Pauwels type III femoral neck fractures. BMC Musculoskelet Disord 23, 49 (2022). https://doi.org/10.1186/s12891-022-05014-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-022-05014-4

Keywords