The first 100 patients treated with a new anatomical pre-contoured locking plate for clavicular midshaft fractures

Abstract

Background

Pre-contoured locking plates were recently introduced in the management of clavicular midshaft fractures. These plates may offer advantages such as no necessity for intraoperative bending and reduced plate irritation. The purpose of this study was to review the clinical and radiographical outcome of the first 100 patients treated with a new anatomical pre-contoured locking plate.

Methods

In a retrospective single-center study, 100 consecutive patients (16 female, 84 male) with a median age of 40 years (range 15–82) who underwent surgery for clavicular midshaft fractures with a VariAx locking plate (Stryker Corporation Kalmazoo, MI, USA) between March 2012 and January 2016 were included. Postoperative follow-up was performed until union was clinically and radiographically achieved. Fracture type, surgical time, intraoperative need for contouring the plate, further surgery such as revision or hardware removal and complications were recorded.

Results

One-hundred patients with a dislocated midshaft clavicular fracture with a mean follow-up of 21.9 months (standard deviation 13.2) were included. Ninety-three patients reported normal shoulder function at latest follow-up. Median surgical time was 75.5 min (range, 35–179). In three patients, intraoperative bending of the plate was necessary. In two patients, plates designed for the other side were implanted. Five patients needed revision surgery: One patient with wound healing problems, one patient with a re-fracture after early (13 months) hardware removal and minor trauma, one patient with postoperative shoulder stiffness and two patients with failed osteosynthesis because of surgical implantation fault. One asymptomatic nonunion without further treatment was observed. In 30 patients, the plate was removed after a mean of 17.5 months (SD 4.2) because of subjective plate discomfort.

Conclusions

With this new pre-contoured locking plate, good to excellent intraoperative fit to the anatomical shape of the clavicle can be achieved. The implant seems to be reliable regarding handling and complications. Clinical and radiological results are comparable to results reported in the literature. Hardware removal rate is comparable to other studies with a pre-contoured plate and lower compared to non-pre-contoured.

Peer Review reports

Background

Clavicle fractures are common with an average of 2.6% of all fractures and most often occur in the middle third of the clavicle [1]. Even though many of these fractures can be treated conservatively, recent evidence suggests that dislocated fracture types show an increased complication rate including a higher non- or malunion rate and therefore operative treatment of displaced and/or comminuted clavicle midshaft fractures has become more popular in the last decades [2,3,4,5,6]. Due to the distinct anatomy of the clavicle, plate fixation in displaced multifragmentary midshaft fractures can be technically challenging. In the past, reconstruction plates were often used due to their contourability and their lower profile. From a biomechanical point of view, the drawback was a diminished implant rigidity, which could lead to hardware failures such as breaking or bending [3, 7]. Furthermore, implant removal rate because of irritating plates is still high up to 38% [8]. Recently, the anatomic pre-shaped clavicle plates were introduced, to address the above-mentioned issues. They offer a higher biomechanical stability [9] without the need for intraoperative bending [10] and thus facilitate fixation. Moreover, these plates may reduce plate irritation due to the low profile and beveled edges and thus have shown to lead to a decreased hardware removal rate [11, 12]. Postoperative range of motion (ROM) and subjective scores does not seem to be affected by the choice of a contoured or non-contoured plate [12].

In this study, the VariAx Clavicle Locking Plate system (Stryker Corporation Kalmazoo, MI, USA), an anatomically pre-shaped, variable angled locking plate system with various plates designed for the most common clavicle fractures was used. It was designed using a CT Scan Database of 160 clavicles of different size, gender, and ethnicity. Made of titanium alloy and treated with Type II anodization, these plates are designed to carry the loads that are required them while remaining low profile. Kienast et al. [13] compared the plate-to-bone fitting of manually bent reconstruction plates with the precontoured VariAx Clavicle locking plate developed using by an anatomical 3D computed tomography-bone database. These data showed a higher degree of plate fit for the osteosynthesis plate developed by using the anatomic 3D database.

However, it remains unclear if the improved plate-to-bone fit has an impact on the clinical and radiographical outcome. The purpose of this study was to evaluate the clinical and radiological outcomes including the rate of plate prominence, rate of hardware removal and rate of complications as well as intraoperative plate-to-bone fit, postoperative ROM and subjective scores in the first 100 patients that have received an open reduction and internal fixation using a pre-shaped locking plate.

Methods

This study was approved by the local ethical committee (EKNZ BASEC 2016–00454). Informed consent was obtained from all the patients before surgery. In a retrospective single-center study, 100 patients (16 female, 84 male) with a median age of 40 years (range 15–82) who underwent surgery for clavicular midshaft fractures with a VariAx locking plate between March 2012 and January 2016 were included. The indications for surgery included (1) fractures with greater than 20 mm of clavicle shortening, (2) substantially displaced fracture fragments lacking cortical apposition (3) impending skin compromise and (4) multifragmentary fractures. Exclusion criteria were pathologic fractures. Surgery was performed in most patients (74%) within 3 days after initial trauma depending on operating room capacity. In 25 patients, surgery was performed after 1 or 2 weeks in case of secondary dislocation in the radiological control with initial try of a conservative management. In one case, time to surgery was about 16 months. In this case, a conservative treatment of a clavicle fracture ended up in a symptomatic non- and malunion, so that a surgery with cancellous bone grafting from iliac crest and stabilization trough VariAx plate was indicated.

Surgical technique

All patients obtained a single-shot antibiotic therapy with a cephalosporin 30 min before surgery. After general anesthesia, patients were positioned in a beach-chair position. The skin incision was either centered or vertical over the fracture with dissection of the platysma. The branches of supraclavicular nerve were identified and protected when possible. The soft tissue was stripped of the subcutaneous surface of the cranial clavicle surface in an epiperiosteal plane. The fracture ends were exposed and debrided of hematoma and interposed soft tissue. Simple or, where possible also multifragmentary fractures were reduced by using clamps and fixed with one or two 2.7 mm lag screws before fixing the fracture with a plate. Where reduction was not possible, fractures were reduced indirectly by bridging the fracture with a plate and then reducing the fragments between the most lateral and most medial fragment and fixing them using an osteosuture (Fiberwire 2.0, Arthrex, Naples, FL, USA). The surgeon chose the plate type intraoperatively depending on subjective best fit. Every surgeon had to describe the anatomical fit of the plate in the surgical report. Good anatomical fit was defined as easy positioning of the plate to the bone, small plate-to-bone distance and no need for bending to achieve this. In all patients, a superior plate was implanted - either standard midshaft (n = 54), bridging midshaft (n = 38) or plates with lateral extension in fractures with large comminution zone (n = 8). The plate was fixed preliminary with clamps. Depending on bone quality, 3.5 mm polyaxial locking screws or non-locking screws were used with at least three bi-cortically placed screws on either side of the fracture. After wound irrigation, the wound was closed in layers using an intraarticular suture with an absorbable suture for skin closure.

Postoperative care and clinical follow-up

All patients were immobilized with a sling for comfort. Active and passive movement of the affected shoulder with a range of motion (ROM) of maximum 90 degrees in flexion and abduction without weight bearing for 6 weeks was recommended. After 6 weeks, free ROM and free weight bearing as tolerated was allowed. Standard thromboembolic prophylaxis was administered only during hospital stay with low molecular weight heparin. Antalgic therapy was implemented by NSAID and paracetamol.

The first postoperative clinical and radiographical follow-up was performed after 6 weeks. Further follow-up has been performed in symptomatic patients 3 or 6 months postoperatively.

In patients complaining of discomfort by hardware, plate removal was performed at the earliest 18 months postoperative except in one case. Here, the plate was removed after 13 months because of patients urge despite our dissuasion. In all cases, where clinical follow-up ended up after 3 months, the patient was called at a minimum of 12 months postoperative and a standardized questionnaire was filled. In this questionnaire, we asked patients about their subjective shoulder function, the ROM compared to the non-affected shoulder, plate discomfort and ability for work. In these cases, the date of questionnaire completion was recorded as the date of the last follow-up.

Outcome measures

The following data were collected for all the patients: Demographic information (age, sex, time to last follow-up, age at surgery, affected side, fracture location, the median American Society of Anesthesiologists (ASA) classification, fracture type and classification according to Robinson [14], surgical time, intraoperative need for contouring the plate, further surgery such as revision or hardware removal, complications such as plate discomfort, and subjective shoulder function (Table 1).

Table 1 Details of the patients: Demographics

Statistics

All statistics were analyzed using GraphPad Prism 7.02 (GraphPad, La Jolla, CA, USA). Fisher’s exact test was performed to calculate Odds ratio (OR). A p-value < 0.05 was considered to be statistically significant. All results are presented in medians (range) or mean (standard deviation).

Results

One-hundred patients with dislocated midshaft clavicular fractures with a mean follow-up of 21.9 months (SD 13.2) were included. Eighty-five patients completed at least a 1-year follow-up. Ten patients had a minimal follow-up of 3 or 6 months. Five patients were lost to follow-up: Two patients with multiple trauma died because of complications unrelated to the clavicle fracture within 8 weeks postoperative, one patient returned to prison after surgery, one patient returned to his native country, and in one case, the patient could not be reached.

Clinical results

From 95 reviewed patients, 93 patients (98%) reported a normal subjective shoulder function at last follow-up. Two patients (2%) complained about slight residual pain during activity.

Ninety-two of these 95 patients showed a symmetrical ROM while 3 patients had a reduced ROM in the affected shoulder compared to the non-affected shoulder: one patient with a shoulder stiffness which was treated with arthroscopic capsulotomy after 5 and 9 months. In the second arthroscopy, Propionibacterium acnes was found and treated with antibiotics and a hardware removal after 1 year showed an implant-associated infection as well. At last follow-up (23 months after clavicle fracture), the patient showed a symmetrical ROM. The other patient with reduced ROM had a non-revised plate dislocation, and the last one had primary osteoarthritis of the shoulder.

Thirty-six patients (38%) complained of plate discomfort. In 5 patients this information was missing because there was no follow-up. In 30 patients (32%), the plate was removed after a median of 17.5 months (range, 12–31). The main reason was subjective plate discomfort.

With the numbers available, female patients had a non-significant 2.063-fold higher risk for plate removal than male (OR 2.063 (CI 95%: 0.740–6.261, p = 0.236).

Radiological results

Standardized radiographs of the clavicle (a.-p. - and tangential view) were performed on the day of the injury, 1 or 2 days postoperatively and in line with the first follow-up after 6 weeks. In the case of no consolidation after 6 weeks and persisting pain, another radiograph has been done at 3 months and/or 1 year postoperatively. A total of 77 patients showed consolidation after a median of 3.5 months (Fig. 1), in one patient an asymptomatic nonunion could be found and in the other 22 patients the information about radiographic consolidation was missing because of no further radiological follow-up due to clinical pain-freedom at 6 weeks or 3 months follow-up (17 patients) or loss of follow-up (the above described 5 patients).

Fig. 1
figure 1

a X-ray in two planes of a 25 years old male patient with multifragmentary midshaft clavicular fracture, b postoperatively, c 5 months after surgery with consolidation of the fracture and d 6 weeks follow-up after hardware removal done 12 months after initial surgery shows solid bone union

Intraoperative findings/results

The median surgical time was 75.5 min (range, 35–179). The anatomical fit of the VariAx plate was reviewed in the surgical rapport and was subjectively described as very good with easy positioning, low plate-to-bone distances and no need of intraoperative bending in most rapports. In only three patients (3%) intraoperative bending of the plate was necessary; in two cases (2%) because of anatomical reasons only and in one case (1%) the plate designed for the other side showed better anatomical fit.

Complications

Five patients (5%) needed revision surgery (Table 2). One patient needed soft tissue revision because of wound healing problem and a non-infected hematoma. One patient needed re-osteosynthesis because of a re-fracture after hardware removal (13 months postoperative) and a minor trauma (Fig. 2). In three patients, failures occurred due to surgical errors where the medial screws were not placed bi-cortically, with subsequent medial plate loosening. One error occurred in experienced and two in less experienced surgeons. Two of these patients needed revision surgery with no further problems in follow-up after revision surgery. One patient was treated for postoperative shoulder stiffness with arthroscopic capsulotomy after 5 and 9 months. One patient developed an asymptomatic nonunion.

Table 2 Results
Fig. 2
figure 2

a X-ray in two planes of a 32 years old female patient with multifragmentary midshaft clavicular fracture, b 5 months after surgery with consolidation of the fracture and c 5 days after hardware removal done 14 months after initial surgery showing refracture after minor trauma and d 6 weeks after revision surgery with a VariAx locking plate

Discussion

Using plates seems to be favorable in the treatment of displaced clavicular midshaft fractures regarding the decreased risk of nonunion by providing higher stability. Regarding implant choice and plate positioning, every system seems to have its own advantages and disadvantages. Superior plate positioning, i.e., provides higher stability regarding the bending forces. Non-pre-contoured plates, i.e. pelvic reconstruction plates, are designed to bend easily at the cost of stability with reported implant failures due to plate weakness. Plate bending of non-contoured plates to achieve an appropriate intraoperative fit of the plate to the variety of human clavicle anatomy is challenging. For experienced surgeons fit accuracy after bending a low profile 3.5 mm pelvic reconstruction plate seems to be better compared to a decent pre-contoured plate (Arthrex Inc., Naples FL, USA) [15]. But the rate of implant related complications like skin irritation or implant prominence still remains high. Pre-contoured plates, therefore, provide biomechanically more stability [9] and may reduce plate irritation and with that lead to less secondary surgery for hardware removal [12, 16, 17]. The VariAx locking plate has been designed 3-dimensionally by using CT-Scan-Database of 160 clavicles, what may improve the anatomical fit to most of the clavicles and so avoid the need for intraoperative bending and reduce the plate irritation. In a comparison study of 5 pre-contoured clavicle plates in 20 embalmed human clavicles, the VariAx plate showed the best anatomical fit of all investigated plates [18]. However, clinical results for this pre-shaped plate are missing.

In our study, we reviewed the first 100 patients treated with pre-shaped VariAx locking plate at our institution and analyzed the clinical and radiological outcomes including the rate of plate prominence, rate of hardware removal and rate of complications as well as intraoperative plate-to-bone fit, postoperative ROM and subjective scores.

Regarding the plate irritation, we still had a high incidence of plate discomfort (36 patients) with a removal rate of 31%. The mean reason was subjective plate discomfort. Our cohort is young and accident mostly happened during sport activity. So, we assume that most patients calculate the risk for complications in a re-operation e.g. plate-removal seems low regarding the benefit of having no more plate irritation especially while sport activity. Furthermore, we suppose, that in some cases the cosmetic aspect assumes a role. This could explain the higher removal rate in female patients which is according to a recent study where female patients are 4 times more likely to undergo hardware removal [16, 19]. Compared to other studies with pre-contoured superior clavicle plates we have a comparable rate of plate discomfort, while the removal rate is still higher (Table 3). So, we obviously complied more often to patients request for plate removal. I.e. VanBeek et al. [12] compared in a retrospective study non-contoured to pre-contoured superior clavicle plate fixation in 52 patients. Patients complained of prominent hardware in 64% in non-contoured plates and 32% in pre-contoured plates with hardware removal in 21 and 10% of the cases respectively. Additionally, the superior positioning of the plate is another risk factor for symptomatic hardware [20].

Table 3 Comparison of reported hardware removal rates for midshaft clavicular fractures treated with noncontoured or precontoured plates

Regarding intraoperative handling of the VariAx plate most surgeons reported good anatomical fit and even less experienced surgeons performed surgery in secure way. Bending of the plate was necessary in only 3 cases to achieve perfect fit to the specific patient’s anatomy. Although bending of the plate is usually not necessary, it can be performed in a secure way when the surgeon avoids sharp bends, reverse bends or bending the device at a screw hole. We had no plate breakage, but a failure of osteosynthesis with loosening of the medial screws and further plate dislocation due to surgeon’s technical fault. Our complication rate is low; only five patients needed revision surgery of which only one patient had a deep, implant-associated infection. No neurovascular complication was observed in any of the 100 patients. In a comparison study of conventional to pre-contoured plates, 2 of 15 patients treated with a conventional plate had to be revised because of plate breakage [17]. Reconstruction plate furthermore showed a failure rate of 12.6% and revision rate of 6.3% in 111 patients [8]. Böstman et al. [21] evaluated the pitfalls of plate fixation of midclavicular fractures in 1997. They showed a total complication rate of 23% in 103 patients with major complication like deep infection, plate breakage, nonunion and refracture after plate removal. The reoperation rate was about 14% including hardware removal.

In conclusion, we can say that the VariAx locking plate is a secure surgery tool for clavicle midshaft fractures with good anatomical fit and a low complication rate. Regarding the costs, in our hospital setting the VariAx plate is one third more expensive than a regular LCP reconstruction plate with 4 locking and 2 non locking screws in each plate.

The clinical outcome was very good with normal subjective shoulder function, symmetrical ROM and return to previous level of acivity in 93 of 95 patients at last follow-up. These results are comparable to other scores in the literature where Constant scores between 86.5 and 96.1 and DASH scores between 4.3 and 5.2 are reported [3, 6].

Value of radiological outcome using the VariAx plate in our first 100 patients is diminished because of lower follow-up rate. From 78 radiographs at last follow-up, 77 showed consolidations and only one patient showed an asymptomatic nonunion without further need for surgical treatment. In 17 patients radiological follow-up ended after 6 weeks because they were absolutely pain free, had a ROM of at least 80% compared to the unaffected shoulder and no further follow-up was wished by the patients. Five patients were lost to follow-up. In a prospective, multicenter clinical trial, 2 patients out of 67 patients who were treated with plate fixation for clavicular midshaft fractures showed nonunion [6]. It is reliable to our results.

This study should be interpreted in light of its potential limitations. Due to the fact, that it is a retrospective study, the data were collected from patient records and the clinical follow-up is not completed for 1 year postoperative in every patient so that some data had to been collected by a phone call. For this reason, subjective shoulder scores, i.e. DASH or Constant score, were not determined. The clinical follow-up was incomplete, and 5 patients were lost to follow-up. Additionally, the radiological follow-up was not performed for every patient. Many patients had a free ROM and no pain at the fracture site during the 6 weeks follow-up examination was found, so no further radiological follow-up was performed, and fracture healing was clinically assumed. Furthermore, it is an uncontrolled study with no control group of our patients treated by VariAx plate fixation for clavicle fractures. Therefore, we relied on comparison with existing literature. All VariAx plates we implanted are superior plates. Plate discomfort risk could decrease by using anterior VariAx plates21, whereas fracture rigidity and stiffness was reduced with anterior plating [9].

Conclusions

With this new pre-shaped locking plate, good to excellent intraoperative fit to the anatomical shape of the clavicle can be achieved. The implant seems to be reliable regarding handling and complications. Clinical and radiological results were very good, and a low complication rate was reported. The hardware removal rate was lower than those reported for non-contoured plates, but higher to other studies using pre-contoured plates although plate discomfort rate is comparable. The decrease of hardware-related plate removal might favor the use of pre-contoured clavicle plates.

Abbreviations

a.-p.:

antero-posterior

ASA:

American Society of Anesthesiologists

CI:

Confidential

CT:

Computed tomography

DASH:

Disability of the arm, shoulder and hand

NSAID:

Non-steroidal anti-inflammatory drugs

OR:

Odds ratio

ROM:

Range of motion

SD:

Standard deviation interval

References

  1. Postacchini F, Gumina S, De Santis P, Albo F. Epidemiology of clavicle fractures. J shoulder Elb Surg. 2002;11(5):452–6.

    Article  Google Scholar 

  2. Mckee RC, Whelan DB, Frcs C, Schemitsch EH, Mckee MD. Operative versus nonoperative care of displaced of randomized clinical trials. J Bone Joint Surg Am. 2012;94–A(8):675–84.

    Article  Google Scholar 

  3. Virtanen KJ, Remes V, Pajarinen J, Savolainen V, Björkenheim J-M, Paavola M. Sling compared with plate osteosynthesis for treatment of displaced midshaft clavicular fractures: a randomized clinical trial. J Bone Joint Surg Am. 2012;94(17):1546–53.

    Article  Google Scholar 

  4. Robinson CM, Goudie EB, Murray IR, Jenkins PJ, Ahktar MA, Read EO, Foster CJ, Clark K, Brooksbank AJ, Arthur A, Crowther MA, Packham I, Chesser TJ. Open reduction and plate fixation versus nonoperative treatment for displaced midshaft clavicular fractures: a multicenter, randomized, controlled trial. J Bone Joint Surg Am. 2013;95(17):1576–84.

    Article  CAS  Google Scholar 

  5. Woltz S, Stegeman SA, Krijnen P, van Dijkman BA, van Thiel TPH, Schep NWL, de Rijcke PAR, Frölke JPM, Schipper IB. Plate fixation compared with nonoperative treatment for displaced Midshaft clavicular fractures: a multicenter randomized controlled trial. J Bone Joint Surg Am. 2017;99(2):106–12.

    Article  Google Scholar 

  6. Canadian Orthopaedic Trauma Society. Nonoperative treatment compared with plate fixation of displaced Midshaft clavicular fractures. A Multicenter, Randomized Clinical Trial. J Bone Joint Surg Am. 2007;89(1):1–10.

    Article  Google Scholar 

  7. Chen Y-F, Wei H-F, Zhang C, Zeng B-F, Zhang C-Q, Xue J-F, Xie X-T, Lu Y. Retrospective comparison of titanium elastic nail (TEN) and reconstruction plate repair of displaced midshaft clavicular fractures. J Shoulder Elb Surg. 2012;21(4):495–501.

    Article  Google Scholar 

  8. Woltz S, Duijff JW, Hoogendoorn JM, Rhemrev SJ, Breederveld RS, Schipper IB, Beeres FJP. Reconstruction plates for midshaft clavicular fractures: a retrospective cohort study. Orthop Traumatol Surg Res. 2016;102(1):25–9.

    Article  CAS  Google Scholar 

  9. Toogood P, Coughlin D, Rodriguez D, Lotz J. A biomechanical comparison of superior and anterior positioning of precontoured plates for midshaft clavicle fractures. Am J Orthop (Belle Mead NJ). 2014;43(10):E226–31.

    Google Scholar 

  10. Huang JI, Toogood P, Chen MR, Wilber JH, Cooperman DR. Clavicular anatomy and the applicability of precontoured plates. J Bone Joint Surg Am. 2007;89(10):2260–5.

    PubMed  Google Scholar 

  11. Ranalletta M, Rossi LA, Bongiovanni SL, Tanoira I, Piuzzi NS, Maignon G. Surgical treatment of displaced midshaft clavicular fractures with precontoured plates. J Shoulder Elb Surg. 2015;24(7):1036–40.

    Article  Google Scholar 

  12. VanBeek C, Boselli KJ, Cadet ER, Ahmad CS, Levine WN. Precontoured plating of clavicle fractures: decreased hardware-related complications? Clin Orthop Relat Res. 2011;469(12):3337–43.

    Article  Google Scholar 

  13. Kienast B, Burgkart RH, Petersik A, Reimers N, Neumann H, Beimel C, Schulz AP. Passform eines neuen auf Basis einer anatomischen Knochendatenbank entwickelten Klavikula-Osteosyntheseplattensystems (VariAx Clavicle) im Vergleich zu individuell von Chirurgen gebogenen Rekonstruktionsplatten. Obere Extrem. 2013;8(4):226–30.

    Article  Google Scholar 

  14. Robinson CM. Fractures of the clavicle in the adult. Epidemiology and classification. J Bone Joint Surg Br. 1998;80(3):476–84.

    Article  CAS  Google Scholar 

  15. Bauer DE, Hingsammer A, Schenk P, Vlachopoulos L, Imam MA, Fürnstahl P, Meyer DC. Are commercially-available precontoured anatomical clavicle plating systems offering the purported superior optimum fitting to the clavicle? A cadaveric analysis and review of literature. Orthop Traumatol Surg Res. 2018;104(6):755–8.

    Article  CAS  Google Scholar 

  16. Rongguang A, Zhen J, Jianhua Z, Jifei S, Xinhua J, Baoqing Y. Surgical treatment of displaced Midshaft clavicle fractures: Precontoured plates versus noncontoured plates. J Hand Surg Am. 2016;41(9):e263–6.

    Article  Google Scholar 

  17. Chandrasenan J, Espag M, Dias R, Clark DI. The use of anatomic pre-contoured plates in the treatment of mid-shaft clavicle fractures. Inj Extra. 2008;39(5):171.

    Article  Google Scholar 

  18. Oppermann J, Ettmann L, Meyer C, Ries C, Stein G, Spies CK, Müller LP, Faymonville C. Analysis of the accuracy of fit of five anatomically precontoured clavicle plate systems. Unfallchirurg. 2018;121(6):457–82.

    Article  Google Scholar 

  19. Naimark M, Dufka FL, Han R, Sing DC, Toogood P, Ma CB, Zhang AL, Feeley BT. Plate fixation of midshaft clavicular fractures: patient-reported outcomes and hardware-related complications. J Shoulder Elb Surg. 2016;25(5):739–46.

    Article  Google Scholar 

  20. Nourian A, Dhaliwal S, Vangala S, Vezeridis PS. Midshaft fractures of the clavicle: a meta-analysis comparing surgical fixation using Anteroinferior plating versus superior plating. J Orthop Trauma. 2017;31(9):461–7.

    Article  Google Scholar 

  21. Böstman O, Manninen M, Pihlajamäki H. Complications of plate fixation in fresh displaced midclavicular fractures. J Trauma. 1997;43(5):778–83.

    Article  Google Scholar 

  22. Kabak S, Halici M, Tuncel M, Avsarogullari L, Karaoglu S. Treatment of midclavicular nonunion: comparison of dynamic compression plating and low-contact dynamic compression plating techniques. J Shoulder Elb Surg. 2004;13(4):396–403.

    Article  Google Scholar 

  23. Lai Y-C, Tarng Y-W, Hsu C-J, Chang W-N, Yang S-W, Renn J-H. Comparison of dynamic and locked compression plates for treating midshaft clavicle fractures. Orthopedics. 2012;35(5):e697–702.

    Article  Google Scholar 

  24. Campochiaro G, Tsatsis C, Gazzotti G, Rebuzzi M, Catani F. Displaced mid-shaft clavicular fractures: surgical treatment with a pre-contoured angular stability plate. Musculoskelet Surg 96 Suppl. 2012;1(S1):S21–6.

    Article  Google Scholar 

  25. Gilde AK, Jones CB, Sietsema DL, Hoffmann MF. Does plate type influence the clinical outcomes and implant removal in midclavicular fractures fixed with 2.7-mm anteroinferior plates? A retrospective cohort study. J Orthop Surg Res. 2014;9:55.

    Article  Google Scholar 

Download references

Acknowledgements

We thank Hanna Zuche for her contribution to the data collection.

This abstract has been previously published in SWISS MEDICAL WEEKLY 147, 28S, 2017.

Funding

Not applicable.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Author information

Authors and Affiliations

Authors

Contributions

TR and RSC wrote the manuscript and interpreted the patient data regarding to the clinical and radiological outcome, both authors contributed equally to this manuscript. RF was involved in the data collection. UR reviewed the text and did the majority of the presented surgeries. NH is head of the department, reviewed the text and is responsible for the data collection at his institution. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Roland S. Camenzind.

Ethics declarations

Ethics approval and consent to participate

Ethical review board approval was given for this study by the local ethical committee (Ethikkommission Nordwest- und Zentralschweiz Business Administration System for Ethics Committees; EKNZ BASEC 2016-00454). Written informed consent was obtained from all the patients before surgery and written informed consent was obtained from legal guardians for those patients under the age of 16.

Consent for publication

Not applicable.

Competing interests

TR, RSC, RF and NH declare that they have no conflict of interest. UR has a medical advisor contract with Stryker Corporation Kalmazoo, MI, USA.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Reisch, T., Camenzind, R.S., Fuhrer, R. et al. The first 100 patients treated with a new anatomical pre-contoured locking plate for clavicular midshaft fractures. BMC Musculoskelet Disord 20, 4 (2019). https://doi.org/10.1186/s12891-018-2396-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-018-2396-9

Keywords