Skip to main content

Biomechanical study between percutaneous vertebroplasty combined with cement pedicle plasty improves vertebral biomechanical stability: A finite element analysis

Abstract

Objective

To investigate the biomechanical effects of percutaneous vertebroplasty combined with cement pedicle plasty (PVCPP) on the unstable osteoporotic vertebral fractures (OVFs) through finite element (FE) analysis. The study compares the biomechanical stability of finite element models between percutaneous vertebroplasty (PVP) and percutaneous vertebroplasty combined with cement pedicle plasty.

Methods

Two patients with unstable OVFs underwent computed tomography (CT) examination at the thoracolumbar vertebral body levels, respectively. The CT images were reconstructed into three-dimensional finite element models to simulate stress conditions across six dimensions and to evaluate the vertebral von Mises stress before and after bone cement reinforcement.

Results

The study found that stress distribution differed between groups mainly at the pedicle base. In the surgical vertebral bodies, the maximum stress in the PVP group decreased during flexion and left bending, while it increased in other states. In the PVCPP group, all maximum stresses decreased. In the inferior vertebral bodies, the maximum stress in the PVP group generally increased, while it decreased in the PVCPP group. In the superior vertebral bodies, postoperatively, the maximum stress in the PVP group generally increased, while it almost remained unchanged in the PVCPP group. PVP group had higher cement stress and displacement.

Conclusion

PVCPP is an effective treatment method for patients with unstable OVFs. It can quickly relieve pain and enhance the stability of the three columns, thereby reducing the risk of some complications.

Peer Review reports

Introduction

Osteoporotic vertebral fractures (OVFs) are a common complication of osteoporosis and are becoming increasingly prevalent due to population aging and nearly 90% of spinal fractures occur in the thoracolumbar region [1]. These may include unstable vertebral fractures with damage to the anterior, middle, and posterior columns and pedicles [2,3,4]. These patients are at higher risk for daily trauma, which, if left untreated, can lead to severe vertebral fractures that can lead to ongoing pain and even neurological symptoms.

Percutaneous vertebroplasty (PVP) is a minimally invasive surgery performed under local anesthesia, known for its quick recovery, significant therapeutic effects, and high acceptance rate. It is increasingly considered a standard surgical method for treating osteoporotic vertebral fractures [5]. The bone cement-strengthened area of PVP is mainly concentrated in the front 2/3 of the vertebral body, and the unfilled bone cement area in the back 1/3 of the vertebral body forms an area with relatively weak stiffness and strength [6]; however, it plays a vital role in stability and preventing vertebral body collapse and kyphosis. Unstable OVFs are usually accompanied by fractures of the posterior 1/3 of the vertebral body (including fractures of the central column, pedicle base, etc.). Some scholars believe that postoperative complications such as bone cement mass displacement, vertebral body collapse, and kyphosis may occur after PVP treatment for patients with such fractures, leading to the risk of recurrence of pain and even surgical failure [7, 8]. Therefore, traditional PVP treatment, such as vertebral fractures, has been controversial. As early as 2004, Eyheremendy et al. [9] reported using bone cement to fill the vertebral pedicles while performing percutaneous vertebroplasty to treat patients with osteoporotic vertebral fractures. Moreover, percutaneous vertebroplasty combined with cemented pedicle plasty has been used in the treatment of Kummell disease [10]. Compared with single PVP, the former shows better advantages in preventing the displacement of bone cement and postoperative follow-up [11]. In a preliminary clinical retrospective study, it has been previously reported in the literature that percutaneous vertebroplasty combined with cement pedicle plasty (PVCPP) was used to treat osteoporotic thoracolumbar burst fractures and achieved good clinical results [12]. However, at present, the scope of application of this surgical method is relatively limited, and adequate biomechanical verification still needs to be investigated.

The finite element (FE) model was considered to be helpful in biomechanical studies [13], and it can avoid the use of human specimens and minimize the variations due to inherent differences among individual parameters [14, 15]. To further understand the impact of PVCPP on vertebral bodies and bone cement, we reviewed the data of two patients undergoing PVP and PVCPP treatments for a single segment of unstable OVF, respectively. We constructed a vertebral biomechanical finite element model to preliminarily observe and discuss the biomechanical characteristics of both surgical methods, providing a theoretical basis for the rational application of the PVCPP technique.

Materials and methods

Patients with unstable OVFs

In this research, two female volunteers with unstable osteoporotic vertebral fractures (64 years old/68 years old, BMD: -2.6/-2.7) underwent CT scans at the levels of T10-T12 and T11-L1, respectively, both before and after undergoing treatment with PVP or PVCPP. These scans had a slice thickness of 0.625 mm. Subsequently, the CT images were reconstructed to create 3D FE models. On T1-weighted MRI, the affected vertebral body exhibited a low signal, while showing a high signal on T2-weighted MRI, ruling out other pathological fractures. All CT images were archived in the digital imaging and communications in medicine (DICOM) format. This study obtained an exemption from ethical approval from the Ethics Committee of the Affiliated Hospital of Traditional Chinese Medicine of Southwest Medical University due to the use of anonymized data. Informed consent from the patients was obtained.

The construction of 3D models and fracture models

Two 3D models were developed based on the preoperative CT scans of T10-T12 and T12-L2 vertebral bodies, respectively. The CT images of 2 patients in DICOM format were imported into Mimics to generate the 3D model of the T10-T11-T12 and T12-L1-L2 vertebral bodies, respectively, including the cortical (1 mm thick) and cancellous bone. The vertebral bodies and intervertebral disc models use SOLID187 ten-node tetrahedral mesh element type. The following steps from the segmentation menu were performed: threshold segmentation was used to separate the bone and soft tissue, and editing mask tools were used to edit the image shape, select the desired area, fill the image area appearing in the gap, and split out the required contour layer by layer. Finally, 3D models were reconstructed through the edit mask option. The reconstructed 3D models were saved in STL format.

STL models were imported into the Geomagic automatic reverse engineering software for processes such as noise reduction, feature removal, and the application of structural patches and fitting surfaces. This resulted in the generation of the surface for 3D models of vertebral bodies, which were subsequently saved in STP format. Among them, Solidworks was used to run commands such as stretch, move, copy, and combine (delete) to create the intervertebral disc. A sketch was drawn on the upper and lower surfaces of the intervertebral disc to divide the intervertebral disc into the upper and lower endplates, and then divided the nucleus pulposus and annulus fibrosus. Optimization of these models was carried out using Geomagic Studio software before importing them into SolidWorks. There, the vertebral bodies were assembled, aligned, and manipulated using the surface command to create the endplates, cartilage, and intervertebral discs (comprising both the nucleus pulposus and annulus fibrosus). We used a linear spring element that only bears tension and not compression to simulate the solid ligaments, and the linear stiffness equation was as follows: F = kΔL. All specific anatomical data and corresponding material parameters for the ligaments are detailed in Table 1. Where, F is the force exerted by the ligament, k is the stiffness constant of the ligament, and ΔL is the deformation (displacement) of the ligament. The comprehensive 3D model was achieved by assembling these components. To simulate the vertebral fracture line, a previously established simulation method was employed, utilizing the surface command to incise the vertebral body, thus creating a 0.5-mm fracture line [16].

Table 1 Material properties of finite element analysis models

Post-surgery augment models and Finite element analysis models

STP models were imported into SolidWorks for further processing. Extract and establish postoperative bone cement models from postoperative CT scans of vertebral bodies, each with a volume of 5 ml. In the T11 and L1 vertebral models, the bone cement was centrally positioned using the assembly feature. Subsequently, through the application of the software's Boolean operations, unnecessary bone material was excised, integrating the bone cement model seamlessly into the vertebral body, as illustrated in Fig. 1. This process yielded a 3D model of the T11 and L1 vertebral bodies post-bone cement augmentation. Table 1 presents the material properties utilized in recent studies concerning OVFs. FE Analysis was conducted using ANSYS software, wherein the 3D models—including those of the cortical and cancellous bones, bone cement, endplates, cartilage, and intervertebral discs (nucleus pulposus and annulus fibrosus)—were imported for comprehensive analysis. The contact between small joints is defined as facet-to-facet contact. The interfaces between vertebral bodies and endplates, endplates and intervertebral discs, and vertebral bodies and bone cement are all set to be fully bonded (Fig. 2).

Fig. 1
figure 1

T10-T12, T12-L2 3D models and fracture models

Fig. 2
figure 2

Finite element analysis models. A Finite element model of PVCPP group and PVP group. B Sagittal view of finite element models (showing cross-section of cortex and cancellous substance)

Force analysis of finite element model

To constrain the boundary conditions of the thoracolumbar spine models after processing, all nodes and degrees of freedom in all directions on the lower surface of T12 and L1 vertebral bodies are fully constrained to make them fixed throughout the simulation process. Vertical downward compressive loads of 500N are applied respectively on the upper surfaces of T10 and T11 vertebral bodies to simulate the weight of the upper body, while 10 N.m torque is added on the upper surface of the superior vertebrae to achieve changes in the direction of model motion, simulating flexion, extension, lateral bending, and left/right axial rotation of the vertebrae (Fig. 3) [17, 25]. According to the three-column concept of the spine, loads and moments are applied to the upper endplate and articular surface of the T12 vertical body, with 85% of the loads and moments applied to the anterior-middle columns and 15% to the posterior columns [3, 26]. This study aims to assess the overall biomechanical changes of vertebral bodies after surgery using patient-specific CT scans. The equation of all von Mises stress calculation was as follows: a= (((a1-a2)^2 + (a2-a3)^2 + (a3-a1)^2)/2)^0.5. Where, a is the maximum von Mises stress, and a1, a2, and a3 are the principal stresses in the three principal stress directions of the material. All vertebral bodies, bone cement stress results, and stress cloud diagrams obtained after the analysis can be exported to the computer. We calculated the overall stress of the surgical vertebral bodies, the adjacent vertebral bodies, and the bone cement. Due to the anatomical differences between the two patient groups, the final results are expressed as the percentage change in maximum stress before and after surgery to assess bone cement augmentation's effectiveness.

Fig. 3
figure 3

Boundary loading conditions for the finite element models. A Finite element model of PVP group. B Finite element model of PVCPP group

Result

Mesh convergence verification of finite element models

By varying the mesh size of the T12-L2 segment to create four different mesh resolutions (with mesh size differences not exceeding 25%), the maximum stress in the T12-L2 flexed position was compared. The mesh element sizes were 3.3mm, 2.5mm, 1.9mm, and 1.4mm, respectively. The results showed that for mesh sizes of 2.5mm and 1.4mm, the maximum stress values in T12-L2 were similar (the maximum difference did not exceed approximately 5%). However, further reducing the element size from 2.5mm to 1.9mm did not significantly change the peak stress, but greatly extended the solution time (Table 2). Similarly, such calculations were also performed in the motion direction of all finite element models. Therefore, a mesh density with an element size of 2.5mm was considered more reasonable.

Table 2 Mesh convergence verification of the finite element model

Calculation results of the maximum von Mises stress of the surgical vertebral bodies for PVP and PVCPP

The maximum von Mises stress cloud diagrams of the two groups of surgical vertebral bodies are shown in Fig. 4. The calculation results and statistical histograms after loading the same load are shown in Table 3 and Fig. 5A. The results indicated that under the conditions of forward flexion, extension, lateral bending, and rotation, the main difference in stress distribution between the two groups of patients' surgical vertebral bodies lies at the pedicle base in the PVCPP group. The postoperative maximum stress in the PVP group decreased during forward flexion and left lateral bending, while the stress increased in the remaining states. The postoperative maximum von Mises stress variation in the PVP group was -2.24%, + 27.29%, -7.47%, + 2.41%, + 16.60%, and + 10.47%. In contrast, the postoperative maximum von Mises stress in the PVCPP group decreased, with a maximum stress variation of -26.08%, -6.25%, -29.19%, -30.55%, -29.40%, and -28.8%.

Fig. 4
figure 4

Stress cloud diagram under different working conditions of each group of models: A PVP; B PVCPP; C Bone cement

Table 3 Calculation results of vertebral bodies von Mises stress (MPa) before and after PVP and PVCPP surgery (results were rounded to one decimal place)
Fig. 5
figure 5

Statistical graphs of results under different experiments: A-C Statistical histogram of surgical, superior, and inferior vertebral stress experimental data for PVP and PVCPP (Percentage change). D Statistical graphs of results bone cement stress and displacement

Calculation results of the maximum von Mises stress of the adjacent vertebral bodies for PVP and PVCPP

The maximum von Mises stress cloud diagrams of the two groups of adjacent vertebral bodies are shown in Fig. 4. The calculation results and statistical histograms after loading the same load are shown in Table 3 and Fig. 5B-C. Based on the calculations for the superior vertebral bodies, under the conditions of forward flexion, extension, lateral bending, and rotation, the postoperative stress distributions of the two finite element model groups showed little difference, and the maximum von Mises stress was reduced in both. For the superior vertebral body in the PVP group, the maximum stress variation was -27.95%, -10.87%, -7.29%, -13.71%, -11.93%, and -13.12%. In contrast, the PVCPP group showed a much smaller maximum stress variation in the superior vertebral body, with values of -0.24%, -0.18%, -0.40%, -0.22%, -0.20%, and -0.20%, respectively. Similarly, based on the calculations for the inferior vertebral bodies, the postoperative maximum von Mises stress generally increased in the PVP group, with stress changes of + 10.04%, -4.50%, + 12.82%, + 1.75%, + 0.97%, and + 2.62%. In the PVCPP group, the postoperative stress uniformly decreased, with changes of -1.62%, -1.26%, -1.61%, -1.58%, -1.64%, and -1.56%.

Calculated results of the experimental data of bone cement

Stress on the implant is one of the most important indices for evaluating its stability. A smaller maximum von Mises stress on the bone cement indicates a lower possibility of loosening and displacement of bone cement. The relative displacement of bone cement also reflects its stability. The distribution of the maximum von Mises stress for the two groups of bone cement is shown in Fig. 4C. The statistical histogram is presented in Fig. 5D. The calculation results showed that the bone cement in the PVP group exhibited higher maximum von Mises stress in flexion, left bending, right bending, and right rotation, with stress values of 7.3482 MPa, 9.1908 MPa, 7.9208 MPa, and 8.6053 MPa, respectively. This indicated that cement loosening or displacement risk in these four conditions was much higher than in the PVCPP group.

Furthermore, the stress distribution in the bone cement of the PVP group was uneven, suggesting that the local average stress on the bone cement was higher than that in the PVCPP group, increasing the probability of local complications. However, the PVCPP group exhibited higher maximum von Mises stress in extension and left rotation, with values of 8.2542 MPa and 8.1251 MPa, respectively. Overall, the bone cement in the PVP group is subject to higher average stress than that in the PVCPP group, indicating that the PVCPP group has more excellent stability.

The relative displacement of bone cement also reflects its stability. The smaller the relative displacement, the better the stability of the bone cement. In this experiment, the displacement of the bone cement in the vertebral bodies relative to its initial position was calculated. The calculation results and statistical histograms of the relative displacement of different groups of bone cement are shown in Fig. 5D. Under the same load. The PVP group exhibited poorer bone cement stability regarding extension, lateral bending, and rotation. Its cement displacement was the highest, with extension showing 0.16191 mm and 0.10397 mm, left bending showing 0.15062 mm and 0.10859 mm, right bending showing 0.14665 mm and 0.11955 mm, left rotation showing 0.13213 mm and 0.10631 mm, and right rotation showing 0.13011 mm and 0.11452 mm. In terms of flexion, the measurement results of the PVP and PVCPP groups were not significantly different. The evaluation results indicate that the PVP group has the lowest stability of bone cement, with the highest relative displacement.

Discussion

In 1984, Galibert and others first applied PVP to the treatment of vertebral hemangiomas, achieving good clinical outcomes [27]. Over decades of development, this technique has become the gold standard for treating osteoporotic vertebral fractures clinically, effectively increasing vertebral bodies strength and stiffness, and quickly alleviating pain and restoring daily life [28]. However, a limitation exists where PVP mainly targets the anterior two-thirds of the vertebral body, leaving the posterior third unfilled with cement. Being a zone of relative weakness, this unfilled region is critical in ensuring vertebral stability and preventing deformities [6]. In PVP procedures, the "blind spot" in the posterior third of the vertebral body is often overlooked. Therefore, there is a controversy over its use in treating unstable OVFs. In 2002, Eyheremendy et al. [9] applied pedicle strengthening surgery to treat pedicle osteolytic lesions for the first time and achieved good clinical results. Later, in 2004, Van der Schaaf et al. [29] combined bilateral PVP with bilateral pedicle cement injection for treating Kummell's disease, effectively connecting the vertebral columns into a single structure with cement, yielding good results. They believed this method could effectively avoid the displacement of bone cement and prevent the posterior wall of the vertebral body from moving into the spinal canal later, causing spinal neurological symptoms. In the meantime, they speculate that bone cement pedicle plasty is the critical factor.

Percutaneous vertebroplasty combined with cement pedicle plasty was developed based on PVP. Bone cement is slowly injected into the first two-thirds of the vertebral body. Then, slowly retract the bone cement pusher and inject an appropriate amount of bone cement simultaneously. While the catheter is slowly withdrawn, sufficient bone cement is left in the pedicle. The three columns are gradually filled with bone cement through careful operation during the operation. It is worth noting that percutaneous vertebroplasty combined with bone cement pedicle plasty has been used to treat Kummell disease and pedicle osteolytic metastatic tumors and has achieved good clinical results [10, 30]. Theoretically, the advantages of this approach include providing mass effect, preventing vertebral bodies from collapsing, and supporting the three columns and pedicles to enhance structural stability. Moreover, this surgical method allows the bone cement of the vertebral body to be fixed in the vertebral body through the bone cement in the pedicle so that the bone cement is completely connected to the normal bone tissue, which can effectively prevent the bone cement from shifting. Such patients have poor vertebral body stability and severe osteoporosis. We perform routine surgery on these patients, and the postoperative prognosis is good. However, this surgical method's long-term clinical follow-up efficacy and biomechanical stability analysis are currently needed.

The FE method has proven to be effective in simulating spine motion thanks to its predictive solid ability [31, 32]. It offers the advantage of low cost for numerical experiments and allows for adjusting various parameters, including material properties [33]. Furthermore, the model can provide results that are difficult to obtain through invasive procedures [34]. By conducting FE analysis, it is possible to accurately simulate the anatomical dynamics of the human spine and observe biomechanical changes that closely resemble natural anatomical conditions [34, 35]. Early studies indicated that PVP altered the biomechanical strength of fractured vertebral bodies but could increase the risk of fractures in adjacent vertebral bodies [36]. Most literature reports on lumbar 3D finite element models are based on bony structures without considering the spine's posterior elements. Therefore, there may be inaccuracies in assessing the impact on the stress of vertebral bodies after traditional PVP surgery, which has limitations. This study is the first to explore finite element biomechanics after PVCPP treatment for unstable OVFs, using real pre- and postoperative CT data for 3D biomechanical finite element modeling and simulating the characteristics of osteoporotic vertebral material based on the physical properties of different tissues. This approach brings the 3D finite element model closer to the biomechanical properties of the vertebral bodies, more accurately reflects postoperative spinal biomechanical parameters, and detects stress changes in each vertebra more precisely after PVCPP treatment [37].

Our study found that under conditions such as flexion, extension, lateral bending, and rotation, the main difference in stress distribution between the patients’ surgical vertebral bodies lies at the pedicle base in the PVCPP group. Post-PVP, the maximum von Mises stress decreased during flexion and left bending, while it increased in other states. In contrast, post-PVCPP, the maximum von Mises stress decreased in all conditions. In the same six dimensions, the stress distribution on the superior vertebral bodies between the two finite element model groups showed little difference, with a decrease in maximum von Mises stress. Similarly, calculations for the inferior vertebral bodies indicated a general increase in maximum von Mises stress post-PVP, while in the PVCPP group, stress in the inferior vertebral body decreased. Therefore, the PVCPP group experienced more uniform stress distribution on the adjacent vertebral body without increasing the risk of fractures. Since vertebral fractures are closely related to the distribution and magnitude of stress, this technique can reduce the risk of fractures in adjacent vertebral bodies [38]. Moreover, bone cement's stress and relative displacement are essential to its stability. The smaller the maximum von Mises stress on the bone cement, the less likely it is to loosen and displace. In this experiment, the average stress on the bone cement in the PVP group was higher than in the PVCPP group, indicating more substantial stability in the PVCPP group. Moreover, the relative displacement of bone cement was highest in the PVP group. Therefore, the vertebral body's overall stability is more substantial after bone cement pedicle augmentation. The observed results are primarily due to the advantages of PVCPP: Firstly, bilateral pedicle approach puncture allows for a more uniform distribution of bone cement, as previously reported [39]. Additionally, beyond the anterior two-thirds of the vertebral body filled in traditional vertebroplasty, bone cement also fills the weaker posterior third's “blind spot” and the interior of the pedicle, leading to more uniform stress distribution on the vertebral bodies. The "bone cement screw" effect effectively integrates the fracture area, preventing displacement of fracture fragments and cement, and further effectively prevents uneven stress distribution caused by kyphotic deformity leading to sagittal imbalance.

There are some limitations for this research. Firstly, The models in this study were all based on patient-specific CT data modeling. The finite element model established before and after surgery was a pathological model. The model was not compared with the normal model to verify its effectiveness. Secondly, the research did not use specific parameters for the intervertebral disc proposed by Niemeyer et al. [40], 2012 for precise modeling and the callus behaves as a poroelastic medium, and studying stress, as well as fluid entry into the callus, is necessary to reduce the efficiency of the ossification process [41]. Thirdly, this study focuses on the overall biomechanics of the skeleton, involving preliminary macroscopic analysis, without considering the influence of bone density and skeletal heterogeneity [42]. Fourthly, because the clinical manifestations and imaging of the two patients did not involve disc herniation, intervertebral disc protrusion related to pain was not considered [43]. Fifthly, static loading does not fully simulate the dynamic kinematics of human activities throughout the day [44]. Lastly, this study is based on individual-specific CT data, and the same results found in this paper cannot be guaranteed to be applicable to the shape of any patient's skeleton [45, 46].

In summary, PVCPP is a minimally invasive, rapidly recovering, and effective surgical method that distributes the load evenly across the surgical and adjacent vertebral bodies and reduces stress. It is concluded that this technique significantly reduces the risk of vertebral fractures and prevents cement displacement, providing a reference for the clinical adoption of PVCPP in treating unstable OVFs. Although similar to vertebroplasty, the surgical risk increases due to cement injection near the spinal cord and nerve structures. Therefore, high-quality fluoroscopic guidance and high-viscosity bone cement are vital to this technique.

Conclusion

Based on the results of finite element analysis experiments and previous clinical effects, percutaneous vertebroplasty combined with bone cement pedicle plasty is an effective treatment method for patients with this type of unstable OVFs. This surgical approach offers rapid pain relief and enhances the stability of the three columns, thereby reducing the chances of vertebral bodies refracture and preventing the intrusion of posterior wall fracture fragments into the spinal canal, which could damage the spinal nerves.

Availability of data and materials

All data generated or analyzed during this study are included in this published article.

Abbreviations

ALL:

Anterior longitudinal ligament

PLL:

Posterior longitudinal ligament

ISL:

Interspinous ligament

SSL:

Supraspinal ligament

LF:

Ligamentum flavum

CL:

Capsular ligament

LI:

Ligamenta interspinalia

Pre:

Pre-augmented

Post:

Post-augmented

PVP:

Percutaneous vertebroplasty

PVCPP:

Percutaneous vertebroplasty combined with cement pedicle plasty

SugVB:

Surgical vertebral body

SupVB:

Superior vertebral body

InfVB:

Inferior vertebral body

OVFs:

Osteoporotic vertebral fractures

FE:

Finite element

DICOM:

Digital imaging and communications in medicine

T3D2:

Truss-3D-2node

3D:

Three-dimension

References

  1. Cahueque M, Cobar A, Zuñiga C, et al. Management of burst fractures in the thoracolumbar spine. J Orthop. 2016;13(4):278–81. https://doi.org/10.1016/j.jor.2016.06.007.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Panjabi MM, Oxland TR, Kifune M, et al. Validity of the three-column theory of thoracolumbar fractures. Spine. 1995;20(10):1122–7. https://doi.org/10.1097/00007632-199505150-00003.

    Article  CAS  PubMed  Google Scholar 

  3. Denis F. The three column spine and its significance in the classification of acute thoracolumbar spinal injuries. Spine. 1983;8(8):817–31. https://doi.org/10.1097/00007632-198311000-00003.

    Article  CAS  PubMed  Google Scholar 

  4. Spiegl U, Anemüller C, Jarvers J-S, et al. Hybrid stabilization of unstable osteoporotic thoracolumbar vertebral body fractures: clinical and radiological outcome after a mean of 4 years. Eur Spine J. 2019;28(5):1130–7. https://doi.org/10.1007/s00586-019-05957-8.

    Article  PubMed  Google Scholar 

  5. Beall DP, Phillips TR. Vertebral augmentation: an overview. Skeletal Radiol. 2023;52(10):1911–20. https://doi.org/10.1007/s00256-022-04092-8.

    Article  PubMed  Google Scholar 

  6. Distefano D, Scarone P, Isalberti M, et al. The “armed concrete” approach: stent-screw-assisted internal fixation (SAIF) reconstructs and internally fixates the most severe osteoporotic vertebral fractures. J Neurointerv Surg. 2021;13(1):63–8. https://doi.org/10.1136/neurintsurg-2020-016597.

    Article  PubMed  Google Scholar 

  7. Lee HM, Park SY, Lee SH, et al. Comparative analysis of clinical outcomes in patients with osteoporotic vertebral compression fractures (OVCFs): conservative treatment versus balloon kyphoplasty. Spine J. 2012;12(11):998–1005. https://doi.org/10.1016/j.spinee.2012.08.024.

    Article  PubMed  Google Scholar 

  8. Yu W, Guan WM, Hayashi D, et al. Vertebral fracture severity assessment on anteroposterior radiographs with a new semi-quantitative technique. Osteoporos Int. 2024;35(5):831–9. https://doi.org/10.1007/s00198-024-07024-5.

    Article  CAS  PubMed  Google Scholar 

  9. Eyheremendy E, De Luca SE, Sanabria E. Percutaneous pediculoplasty in osteoporotic compression fractures. J Vasc Interv Radiol. 2004. https://doi.org/10.1097/01.rvi.0000136969.96466.99.

    Article  PubMed  Google Scholar 

  10. Wang B, Wang Y, Zhao Q, et al. Pediculoplasty combined with vertebroplasty for the treatment of Kummell’s disease without neurological impairment: robot-assisted and fluoroscopy-guided. Am J Transl Res. 2020;12(12):8019–29.

    CAS  PubMed  PubMed Central  Google Scholar 

  11. Liu T, Gu G, Zhan C, et al. Comparison of percutaneous vertebroplasty and percutaneous vertebroplasty combined with pediculoplasty for Kümmell’s disease: a retrospective observational study. Journal of Orthopaedic Surgery and Research. 2023;18(1):0–0. https://doi.org/10.1186/s13018-023-03957-5.

    Article  Google Scholar 

  12. Xiao C, Wang H, Lei Y, et al. Percutaneous kyphoplasty combined with pediculoplasty for the surgical treatment of osteoporotic thoracolumbar burst fractures. J Orthop Surg Res. 2024;19(1):87. https://doi.org/10.1186/s13018-024-04562-w.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Zahaf S, Kebdani S. Study and analysis of mechanical behavior between rigid and dynamic fixation systems analyzed by the finite element method. Journal of Biomimetics, Biomaterials and Biomedical Engineering. 2017;33:12–31. https://doi.org/10.4028/www.scientific.net/jbbbe.33.12.

    Article  CAS  Google Scholar 

  14. Winking M, Stahl JP, Oertel MF, et al. Treatment of pain from osteoporotic vertebral collapse by percutaneous PMMA vertebroplasty. Acta neurochirurgica. 2004;146(5):469–76. https://doi.org/10.1007/s00701-004-0259-7.

    Article  CAS  PubMed  Google Scholar 

  15. Kosmopoulos V, Keller TS. Damage-based finite-element vertebroplasty simulations. Eur Spine J. 2004;13(7):617–25. https://doi.org/10.1007/s00586-003-0651-7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  16. Liang D, Ye LQ, Jiang XB, et al. Biomechanical effects of cement distribution in the fractured area on osteoporotic vertebral compression fractures: a three-dimensional finite element analysis. J Surg Res. 2015;195(1):246–56. https://doi.org/10.1016/j.jss.2014.12.053.

    Article  CAS  PubMed  Google Scholar 

  17. Zhang X, Chen T, Meng F, et al. A finite element analysis on different bone cement forms and injection volumes injected into lumbar vertebral body in percutaneous kyphoplasty. BMC Musculoskelet Disord. 2022;23(1):621. https://doi.org/10.1186/s12891-022-05522-3.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  18. Zhao WT, Qin DP, Zhang XG, et al. Biomechanical effects of different vertebral heights after augmentation of osteoporotic vertebral compression fracture: a three-dimensional finite element analysis. J Orthop Surg Res. 2018;13(1):32. https://doi.org/10.1186/s13018-018-0733-1.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Bereczki F, Turbucz M, Kiss R, et al. Stability evaluation of different oblique lumbar interbody fusion constructs in normal and osteoporotic condition - a finite element based study. Front Bioeng Biotechnol. 2021;9: 749914. https://doi.org/10.3389/fbioe.2021.749914.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Salvatore G, Berton A, Giambini H, et al. Biomechanical effects of metastasis in the osteoporotic lumbar spine: a finite element analysis. BMC Musculoskelet Disord. 2018;19(1):38. https://doi.org/10.1186/s12891-018-1953-6.

    Article  PubMed  PubMed Central  Google Scholar 

  21. Wang QD, Guo LX. Prediction of complications and fusion outcomes of fused lumbar spine with or without fixation system under whole-body vibration. Med Biol Eng Comput. 2021;59(6):1223–33. https://doi.org/10.1007/s11517-021-02375-1.

    Article  PubMed  Google Scholar 

  22. Yang P, Zhang Y, Ding HW, et al. Pedicle screw fixation with kyphoplasty decreases the fracture risk of the treated and adjacent non-treated vertebral bodies: a finite element analysis. J Huazhong Univ Sci Technolog Med Sci. 2016;36(6):887–94. https://doi.org/10.1007/s11596-016-1680-x.

    Article  PubMed  Google Scholar 

  23. Zhang Q, Chon T, Zhang Y, et al. Finite element analysis of the lumbar spine in adolescent idiopathic scoliosis subjected to different loads. Comput Biol Med. 2021;136: 104745. https://doi.org/10.1016/j.compbiomed.2021.104745.

    Article  CAS  PubMed  Google Scholar 

  24. Huang S, Zhou C, Zhang X, et al. Biomechanical analysis of sandwich vertebrae in osteoporotic patients: finite element analysis. Front Endocrinol (Lausanne). 2023;14: 1259095. https://doi.org/10.3389/fendo.2023.1259095.

    Article  PubMed  Google Scholar 

  25. Zuo XH, Chen YB, Xie P, et al. Finite element analysis of wedge and biconcave deformity in four different height restoration after augmentation of osteoporotic vertebral compression fractures. J Orthop Surg Res. 2021;16(1):138. https://doi.org/10.1186/s13018-021-02225-8.

    Article  PubMed  PubMed Central  Google Scholar 

  26. Chung SK, Kim YE, Wang KC. Biomechanical effect of constraint in lumbar total disc replacement: a study with finite element analysis. Spine (Phila Pa 1976). 2009;34(12):1281–6. https://doi.org/10.1097/BRS.0b013e3181a4ec2d.

    Article  PubMed  Google Scholar 

  27. Galibert P, Deramond H, Rosat P, et al. [Preliminary note on the treatment of vertebral angioma by percutaneous acrylic vertebroplasty]. Neurochirurgie. 1987;33(2):166–8.

    CAS  PubMed  Google Scholar 

  28. Li K, Ji C, Luo D, et al. Role of percutaneous vertebroplasty with high-viscosity cement in the treatment of severe osteoporotic vertebral compression fractures. Sci Rep. 2021;11(1):4602. https://doi.org/10.1038/s41598-021-84314-6.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  29. van der Schaaf I, Fransen H. Percutaneous vertebroplasty as treatment for Kummell’s disease. JBR-BTR. 2009;92(2):83–5.

    PubMed  Google Scholar 

  30. Zetao Wu, Zhengyin L, Xiaodong X, et al. Clinical application of CT-guided percutaneous pedicle plasty in the treatment of pedicle osteolytic metastases. Journal of Practical Radiology. 2007;23(7):950–2.

    Google Scholar 

  31. Xu L, Qiu Y, Chen Z, et al. A re-evaluation of the effects of dual growing rods on apical vertebral rotation in patients with early-onset scoliosis and a minimum of two lengthening procedures: a CT-based study. J Neurosurg Pediatr. 2018;22(3):306–12. https://doi.org/10.3171/2018.3.PEDS1832.

    Article  PubMed  Google Scholar 

  32. Wu W, Chen C, Ning J, et al. A novel anterior transpedicular screw artificial vertebral body system for lower cervical spine fixation: a finite element study. J Biomech Eng. 2017;139(6):null. https://doi.org/10.1115/1.4036393.

    Article  Google Scholar 

  33. Keddar B, Aour B, Zahaf S, et al. Static study and finite element analysis of a new method of fixation of a medial humerus fracture by an intramedullary nailing system analyzed by the ANSYS workbench 16.2 calculus code. Nano Biomedicine and Engineering. 2019;11(3):0–0. https://doi.org/10.5101/nbe.v11i3.p272-289.

    Article  CAS  Google Scholar 

  34. Levadnyi I, Awrejcewicz J, Zhang Y, et al. Finite element analysis of impact for helmeted and non-helmeted head. J Med Biol Eng. 2018;38(4):587–95. https://doi.org/10.1007/s40846-017-0324-3.

    Article  PubMed  Google Scholar 

  35. Mestar A, Zahaf S, Zina N, et al. Development and validation of a numerical model for the mechanical behavior of knee prosthesis analyzed by the finite elements method. Journal of Biomimetics, Biomaterials and Biomedical Engineering. 2018;37:12–42. https://doi.org/10.4028/www.scientific.net/jbbbe.37.12.

    Article  CAS  Google Scholar 

  36. Borensztein M, Camino Willhuber GO, Posadas Martinez ML, et al. Analysis of risk factors for new vertebral fracture after percutaneous vertebroplasty. Global Spine J. 2018;8(5):446–52. https://doi.org/10.1177/2192568217732988.

    Article  PubMed  Google Scholar 

  37. Keddar I, Aour B, Zahaf S. Comparative study of the fractured humerus fixation by intramedullary nailing and compression plate. J Fail Anal Prev. 2022;22(5):1905–23. https://doi.org/10.1007/s11668-022-01459-w.

    Article  Google Scholar 

  38. Zhou C, Meng X, Huang S, et al. Biomechanical study of different bone cement distribution on osteoporotic vertebral compression fracture-a finite element analysis. Heliyon. 2024;10(5): e26726. https://doi.org/10.1016/j.heliyon.2024.e26726.

    Article  PubMed  PubMed Central  Google Scholar 

  39. Haibier A, Yusufu A, Lin H, et al. Effect of different cement distribution in bilateral and unilateral Percutaneous vertebro plasty on the clinical efficacy of vertebral compression fractures. BMC Musculoskelet Disord. 2023;24(1):908. https://doi.org/10.1186/s12891-023-06997-4.

    Article  PubMed  PubMed Central  Google Scholar 

  40. Cappetti N, Naddeo A, Naddeo F, et al. Finite elements/Taguchi method based procedure for the identification of the geometrical parameters significantly affecting the biomechanical behavior of a lumbar disc. Comput Methods Biomech Biomed Engin. 2016;19(12):1278–85. https://doi.org/10.1080/10255842.2015.1128529.

    Article  CAS  PubMed  Google Scholar 

  41. Travascio F, Buller LT, Milne E, et al. Mechanical performance and implications on bone healing of different screw configurations for plate fixation of diaphyseal tibia fractures: a computational study. Eur J Orthop Surg Traumatol. 2021;31(1):121–30. https://doi.org/10.1007/s00590-020-02749-5.

    Article  PubMed  Google Scholar 

  42. Zahaf S, Habib H, Mansouri B, et al. The effect of the eccentric loading on the components of the spine. Glob J Res Eng. 2017.

  43. Amirouche F, Solitro GF, Siemionow K, et al. Role of posterior elements in the disc bulging of a degenerated cervical spine. Int J Spine Surg. 2015;9: 13. https://doi.org/10.14444/2013.

    Article  PubMed  PubMed Central  Google Scholar 

  44. Travascio F, Eltoukhy M, Cami S, et al. Altered mechano-chemical environment in hip articular cartilage: effect of obesity. Biomech Model Mechanobiol. 2014;13(5):945–59. https://doi.org/10.1007/s10237-013-0545-5.

    Article  PubMed  Google Scholar 

  45. Garay RS, Solitro GF, Lam KC, et al. Characterization of regional variation of bone mineral density in the geriatric human cervical spine by quantitative computed tomography. PLoS One. 2022;17(7): e0271187. https://doi.org/10.1371/journal.pone.0271187.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  46. Solitro GF, Mainnemare F, Amirouche F, et al. A novel technique with reduced computed tomography exposure to predict vertebral compression fracture: a finite element study based on rat vertebrae. Med Biol Eng Comput. 2019;57(4):795–805. https://doi.org/10.1007/s11517-018-1918-9.

    Article  PubMed  Google Scholar 

Download references

Funding

This study has received the funding by the Nanjing Drum Tower Hospital’s talent introduction and scientific research start-up project (RC2023-040), and the fundings for Clinical Trials from the Affiliated Drum Tower Hospital, Medical School of Nanjing University (2024-LCYJ-PY-53).

Author information

Authors and Affiliations

Authors

Contributions

Sen Li and Hongtao Li contributed to the study design. Hongtao Li contributed to the drafting of the manuscript. Changming Xiao and Hongyu Pan conducted the literature search, quality assessment, data collection, and analysis. Yang Lei and Haozhong Wang solved the cases of doubt. All authors have read and approved the final manuscript.

Corresponding author

Correspondence to Sen Li.

Ethics declarations

Ethics approval and consent to participate

This study obtained an exemption from ethical approval from the Ethics Committee of the Affiliated Hospital of Traditional Chinese Medicine of Southwest Medical University due to the use of anonymized data, in accordance with the Declaration of Helsinki. The authors obtained informed consent from the patients.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Li, H., Xiao, C., Pan, H. et al. Biomechanical study between percutaneous vertebroplasty combined with cement pedicle plasty improves vertebral biomechanical stability: A finite element analysis. BMC Musculoskelet Disord 25, 597 (2024). https://doi.org/10.1186/s12891-024-07689-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-024-07689-3

Keywords