Extending the intermedullary nail will not reduce the potential risk of femoral head varus in PFNA patients biomechanically: a clinical review and corresponding numerical simulation

Abstract

Femoral head varus is an important complication in intertrochanteric fracture patients treated with proximal femoral nail anti-rotation (PFNA) fixation. Theoretically, extending the length of the intramedullary nail could optimize fixation stability by lengthening the force arm. However, whether extending the nail length can optimize patient prognosis is unclear. In this study, a review of imaging data from intertrochanteric fracture patients with PFNA fixation was performed, and the length of the intramedullary nail in the femoral trunk and the distance between the lesser trochanter and the distal locking screw were measured. The femoral neck varus status was judged at the 6-month follow-up. The correlation coefficients between nail length and femoral neck varus angle were computed, and linear regression analysis was used to determine whether a change in nail length was an independent risk factor for femoral neck varus. Moreover, the biomechanical effects of different nail lengths on PFNA fixation stability and local stress distribution have also been verified by numerical mechanical simulations. Clinical review revealed that changes in nail length were not significantly correlated with femoral head varus and were also not an independent risk factor for this complication. In addition, only slight biomechanical changes can be observed in the numerical simulation results. Therefore, commonly used intramedullary nails should be able to meet the needs of PFNA-fixed patients, and additional procedures for longer nail insertion may be unnecessary.

Peer Review reports

Introduction

Femoral intertrochanteric fracture is the most common hip fracture in elderly patients [1, 2]. Although hip arthroplasty is an alternative treatment strategy, internal fixation is still the most commonly used surgical method for these patients [1, 3]. Proximal femoral nail anti-rotation (PFNA) fixation has been widely used to treat intertrochanteric fractures and has achieved a credible clinical prognosis in most patients [4, 5].

The optimal length of the main nail has not been determined during PFNA fixation. Several studies have shown that long nails were developed to reduce the potential risk of stress concentration and resulting complications by extending the load arm of the intermedullary nail [6, 7]. However, several recent studies have shown that the use of longer nails does not obviously reduce the potential risk of complications [8,9,10]. In addition, extending the nail length significantly increases blood loss, surgical time, and even intraoperative radiography [7, 10]. Therefore, the absence of existing guidelines and the treating surgeon’s preference, rather than any objective factors, influence the decision of nail length.

Postoperative femoral head varus was still a commonly observed complication in patients who underwent PFNA fixation [4, 11], which adversely affects patient recovery. As an important [12]intraoperative parameter, the length of the nail significantly affects the instant postoperative local biomechanical environment (i.e., fixation stability and stress distribution on the femoral head) [6]. Given that deterioration of the biomechanical environment could initially trigger femoral head varus, the nail length of PFNA may potentially affect the risk of this complication. However, to our knowledge, this topic has yet to be identified.

In this study, by reviewing the clinical data of PFNA-fixed patients, the clinical effect of nail length on the incidence of femoral head varus was investigated, and the corresponding biomechanical mechanism of clinically observed phenomena was also verified via numerical simulations. Identifying this topic could provide theoretical foundations for the optimization of intraoperative procedures and corresponding clinical prognoses for PFNA-fixed patients.

Materials and methods

Clinical review

Patient collection

The ethics committees of our hospital reviewed and approved the protocol of this study. Informed consent was waived for this retrospective study. We retrospectively reviewed patients who suffered from the intertrochanteric fractures and underwent PFNA fixation from January 2021 to August 2022. The exclusion criteria were as follows: (1) had femur trauma or an operation history; (2) had a pathological fracture caused by primary or metastatic bone tumors, bone tuberculosis, or rheumatic immune diseases; (3) underwent revision surgery within the clinical follow-up period of 6 months for other complications; (4) had conservative treatment; (5) were lost to follow-up or died during the follow-up cycle; and (6) had long-term bed rest (larger than 2 weeks) (Fig. 1) [13, 14]. Patients baseline characteristics, including age, sex, and BMI, were recorded [15, 16]. A well-trained orthopedic surgeon performed all the PFNA operations, and the tip-to-apex distance (TAD) for the anti-rotation blade on the anterior-posterior radiography was < 25 mm, and the blade was located in the inferior half of femoral head for all the enrolled patients [4, 17]. Patients with stable fracture type began weight-bearing on the injured limb (walking on the ground) within 3 days after surgery, and patients with unstable fracture began injured weight-bearing within 2 weeks.

Fig. 1
figure 1

Schematic for patient inclusion and exclusion

Radiographic data collection

All patients underwent anterior-posterior radiography three times, including immediately before and after the operation and nearly 6 months after the screw fixation operation. The femoral neck-trunk angle was measured via anterior-posterior radiography at three different times, and the difference between pre and instant postoperative angles was computed to represent the varus correlation status, and that between instant postoperative and 6 moths’ follow up was computed to represent the femoral head varus progression status [4, 18]. The length of the intramedullary nail in the femoral trunk (below the lesser trochanter plane) and the distance between the lesser trochanter and the distal locking screw were measured separately (Fig. 1). Fracture type was defined in the preoperative anterior-posterior radiography. AO 3.1 A1.1, 1.2, 1.3 and 2.1 types fracture was defined as “stable fracture”, AO 2.2 and 2.3 types fracture was defined as “unstable fracture”. TAD value and the distance between femoral calcar and anti-rotation blade on the central point of femoral neck were measured in the instant postoperative anterior-posterior radiography (D3 and D4, Fig. 2).

Fig. 2
figure 2

The measurement of imaging parameters: The length of the intramedullary nail in the femoral trunk (D1), the distance between the lesser trochanter and the distal locking screw (D2), and the femoral head varus values, TAD in the anterior-posterior radiography (D3), and the distance between femoral calcar and the anti-rotation blade (D4)

Statistical analyses

We conducted the statistical analyses with SPSS software. One week after completing the radiological data measurements, a randomly selected imaging specialist with extensive experience in orthopedic imaging and the aforementioned orthopedic physician re-evaluated the imaging data for 20 patients to assess interrater reliability. Intraclass correlation coefficients (ICCs) and Kappa value were calculated to determine the consistency of continue and binary variables [19, 20]. Excellent interrater reliability was judged when the ICC and Kappa value was > 0.8 [19, 21]. Normality tests for continuous variables were performed, and Pearson and Spearman correlation coefficients between normally distributed continuous variables, nonnormally distributed continuous variables and categorical variables and femoral head varus values were calculated separately [22, 23]. Moreover, we performed linear logistic regression to identify independent risk factors for femoral head necrosis. Univariate analyses of each potential risk factor were performed, and the variables that achieved a significance level of p < 0.1 were entered into multivariate analyses. Variables with P < 0.05 were considered independent risk factors in the multivariate analysis [24, 25].

Numerical surgical simulations and finite element analyses (FEA)

Construction of the intact finite element (FE) model

The proximal femur model was constructed based on the outline of the syn-bone model rather than that of any special patient. The model construction strategy was selected to avoid ethics-related procedures and eliminate the confounding effects caused by individual differences in the outlines of different patients [18, 26]. A thin CT scan was performed on the syn-bone femur model (thickness = 0.55 mm). The range of the proximal femur was defined as the distance from the tip of the femoral head to 30 cm below the lesser trochanter. The outline of the proximal femur model was constructed according to the CT-scanned femur outline in 3D-CAD software [27, 28]. This study consisted of our published studies. The computational efficiency and accuracy of the numerical model constructed by this method were better than those of the traditional reverse model construction strategy [15, 29]. Cortical and cancellous bones were separately constructed, and outlines of these bony structures were constructed separately based on the CT imaging data [30, 31].

Construction of PFNA fixed intertrochanteric fracture models with different nail length

To simulate the fixation stability of different screw fixation strategies in the instability fracture model, the AO 3 − 1 2.3 type intertrochanteric fracture model was constructed [26, 32]. The specific modeling method involved creating the fracture by intersecting three fracture lines. The first fracture line was positioned 10 mm below the greater trochanter, forming a 20° angle with the long axis of the femoral shaft [18, 26]. The second fracture line was set tangent to the upper edge of the lesser trochanter, and the third fracture line connected the intersection of the first and second lines with the vertex of the greater trochanter. The bone within the cut range was removed to complete the reconstruction of the fracture model. The numerical model of a PFNA fixation device was also constructed with 3D-CAD software [30, 32]. When simulating the PFNA fixation operation, the trajectory of the anti-rotation blade was coaxial to the long axis of the femoral neck [11, 18]. By adjusting the nail length and the position of the distal locking screw, the biomechanical significance of changes in clinically observed parameters (i.e., the length of the intramedullary nail in the femoral trunk and the distance between the lesser trochanter and the distal locking screw) has been investigated via numerical models. The detailed model construction strategy is presented in Fig. 3.

Fig. 3
figure 3

Schematic for numerical model construction, and surgical simulations in models with different length of the intramedullary nail in the femoral trunk, and different distances between the lesser trochanter and the distal locking screw

Boundary and loading conditions

The material properties of cortical and cancellous bone and the titanium alloy cannulated screw were separately defined as isotropic material according to the same type of study. The degrees of freedom of the inferior surfaces of the numerical models were completely fixed. Different forces were applied to the femoral head at 10° lateral to the coronal plane and 9° posteriorly to the sagittal plane [3, 33]. Two different loading protocols were applied on the superior surface of the femoral head. A 2100 N load was applied on the femoral head. The stress distributions of the femoral head and anti-rotation blade, the stress distributions of the femoral head and anti-rotation blade, the maximum displacement of the femoral head, and the failure load resulting from fixation failure were computed and recorded to determine the potential risk of femoral head varus (Fig. 4) [30, 34].

Fig. 4
figure 4

Boundary and loading conditions for numerical simulation, computational results in different models

Results

Clinical review and judgement of independent risk factors for femoral head varus in PFNA fixed patients

Excellent intra- and interobserver measurements of imaging-based parameters were judged based on the computed ICC and Kappa values (Table 1). In this study, clinical data from 121 patients were collected. Correlation analysis indicated no significant correlation between nail length parameters (i.e., the nail length in the femoral trunk or the distance between the lesser trochanter and the distal locking screw) and the incidence of femoral head varus, and these parameters were also not found to be independent risk factors for femoral head varus. Besides, larger TAD value and unstable type of fracture were significantly correlated to femoral head varus, but only unstable fracture was an independent risk factor for femoral head varus progression. Meanwhile, the distance between femoral calcar and blade was not significantly correlated to femoral head varus values, and was not an independent risk factor for femoral head varus progression. Moreover, there were no significant difference between D1 and D2, and the femoral head varus correlation status (Tables 2, 3 and 4). Finally, no patients suffered diaphyseal fractures in the absence of additional trauma during the clinical follow-up period. Fixation strength computation.

Table 1 ICC and Kappa values of inter- and intraobserver reliability when measuring imaging based parameters
Table 2 Correlation coefficients between femoral head varus and variates
Table 3 Linear regression analysis of severe femoral head varus
Table 4 Correlation coefficients between instant postoperative femoral head varus correlation and variates

Consistent with the results of previous reviews, a consistent variation in the computed parameters can be observed in the current numerical models. Specifically, in the model simultaneously extending nail length in the femoral trunk and the distance between the lesser trochanter and the distal locking screw, the maximum stress of cancellous bone reduced by more than 5%, differences in other parameters was less than 1% between different models. Moreover, contrary to our expectations, compared to those in the control group, the fixation failure loads in the nail length extension and the distance between the lesser trochanter and the distal locking screw groups were slightly lower (Figs. 4 and 5).

Fig. 5
figure 5

Nephograms for anti-rotation blade stress distributions: Slight difference in stress distribution can be observed in different numerical models

Discussion

Theoretically, extending the intermedullary nail in PFNA-fixed patients can optimize fixation stability by alleviating local stress concentrations [12, 35]. This approach may reduce the risk of biomechanical complications, especially diaphyseal fractures [6, 7]. However, comparable prognoses have also been reported for patients with short or long nails. More significantly, extending the nail length can trigger an increase in blood loss, surgical time, and even intraoperative radiography examination time and inhibit patients’ quick recovery [6, 7]. Moreover, although there were differences in indications between “short and long nail” selection (e.g., long nail was more suitable for patients with subtrochanteric fracture) [6, 7], and all PFNA used in the current patients were traditional “short nail”. However, the effect of nail length changes on fixation stability and corresponding risk of femoral head varus has yet to be identified in patients with “short nail” PFNA fixation.

To verify this topic and provide theoretical and practical guidance for the optimization of PFNA operation, we performed comprehensive research consisting of a clinical review and corresponding numerical simulations. A consistent variation tendency can be observed in this study. That is, extending the nail length will not alleviate the stress concentration on the femoral head or the corresponding risk of femoral head varus. Therefore, based on the current results, commonly used intramedullary nail lengths should be able to meet the needs of PFNA-fixed patients, and additional iatrogenic injury caused by longer nail insertion should be avoided. In addition, although studies have reported that extending the nail length can reduce the risk of postoperative diaphyseal fracture, no patients in the present study experienced this complication. This is mainly due to the relatively high strength of the cortex in the diaphysis and the lower activity in elderly patients with intertrochanteric fractures.

Moreover, although larger TAD value was significantly correlated to femoral head varus, this factor was not an independent risk factor for femoral head varus progression. This result indicate that in patients whose TAD < 25 mm, the influence of TAD value on fixation stability was limited, and this was consisted to published studies [4, 36]. Moreover, studies present that femoral calcar is an important structure to provide the medium support of the femur [1, 37], but the current study indicate that the distance between calcar and anti-rotation blade was not significantly affect the risk of femoral head varus progression. Therefore, we can also deduce that place the blade in the inferior part of femoral neck can construct comparable fixation stability in PFNA operation, but the detail relation between blade position and PFNA fixation stability still should be verified in our future studies.

The following topics from the methodological perspective should be clarified in this study. First, femoral head varus values were recorded to determine patient prognosis. The reoperation rate of PFNA-fixed patients was recorded to determine the complication rate in several studies [9, 10]. Traditionally, only patients who underwent reoperation were considered to suffer complications [10, 38]. However, studies have shown that even in patients who do not undergo reoperation, the progression of femoral head varus can also trigger clinical deterioration, even in patients with fracture union [39, 40]. Therefore, this study evaluated patient prognosis by measuring the femoral head varus angle and can provide a new perspective for optimizing the PFNA operation.

In addition, although mechanical tests can directly reflect fixation stability in models with internal fixation [41, 42], numerical models were selected to determine the biomechanical mechanism of changes in the PFNA strategy in this study. The fresh specimen model-based mechanical test can only reflect the overall fixation stability change, and numerical simulations can directly reflect the stress distribution on the blade-bone interface [24, 43]. More significantly, numerical mechanical simulation can discretize the biomechanical significance of particular variables [15, 16]. In this study, the biomechanical significance of nail length extension and the distance between the lesser trochanter and the distal locking screw were separately investigated, but these factors are not feasible for mechanical tests. Meanwhile, when investigating the effect of a certain variable on the fixation stability, researchers prone to construct numerical models with unstable type of fracture (e.g., A.2.3 fracture type in this study), this was a commonly used model construction strategy on the same type studies [18, 26, 32]. That’s because this model construction strategy can reduce potential incidence of false negative result.

The combination of clinical review and mechanical simulation can optimize the credibility of this study [15, 16]. As a study with negative results (i.e., changes in nail length will not significantly affect the risk of femoral head varus) and if only the clinical review section was involved in this study, researchers may consider that the limited sample size may lead to potential false negatives. Similarly, if the significance of nail length changes has been investigated by numerical simulations alone, researchers may consider that improper model construction strategies or loading conditions can trigger false negative results. In contrast, when negative clinical review results and computational results can be mutually corroborated, the reliability of the study results can be effectively guaranteed. Finally, the syn-bone model was selected to construct the outline of the numerical model. Although this modeling strategy suffers from defects (simplification of cortical and cancellous bony regions in the standard model and the outline of certain anatomical structures) [18, 32], it has been widely used for its inherent advantages, such as the nonrequirement of an ethical review and repeatable results (elimination of confounding effects of patient-specific morphological parameters on computational) [18, 26].

Admittedly, this study has the following limitations. First, only one fracture type was selected to construct the numerical model. Changes in fracture type, especially different grades of medial or lateral wall injury, and the interactions between these parameters and nail length changes should be verified in our future studies [1, 44]. In addition, different trabecular structures, including compressive and tensile trabecular structures, play significant biomechanical roles in maintaining PFNA fixation stability [11, 13]. Due to the lack of a standard model construction strategy, these structures have been simplified to iatrogenic structures in current numerical models. Moreover, limited by the lack of fatigue test related mechanical material properties (e.g., s-n curve), only one cycle of load was applied on the numerical model. As a result, biomechanical significance during the cyclic load can not be identified in this study, and this was an important limitation of the current study. However, as a comprehensive research consisted by clinical review and numerical simulations, an identical tendency can be observed in both two parts of this study. Therefore, we believe that this limitation will not significantly affect the credibility of the current research conclusion.

Meanwhile, the ignorance of interaction between blade trajectory on the coronal and transverse planes was also a main limitation of this study [6, 45]. However, for the lack of instant postoperative CT imaging data, the blade trajectory on the transverse plane can not be directly judged, and which should be identified in our future studies. Finally, although this study revealed that commonly used intramedullary nails should be able to meet the needs of PFNA-fixed patients, according to the basic principle of biomechanics, we believe that the reduction in nail length is not unlimited and that a safe threshold should be identified in our future biomechanical studies.

Data availability

All data generated or analysed during this study are included in this published article and its supplementary information files.

References

  1. Chang SM, Hou ZY, Hu SJ, Du SC. Intertrochanteric femur fracture treatment in Asia: what we know and what the World can learn. Qld Gov Min J. 2020;51:189–205.

    Google Scholar 

  2. Ricci WM. Stability of intertrochanteric femur fractures. J Orthop Trauma. 2023;37:S1–4.

    Article  PubMed  Google Scholar 

  3. Liu W, Zhou D, Liu F, Weaver MJ, Vrahas MS. Mechanical complications of intertrochanteric hip fractures treated with trochanteric femoral nails. J Trauma Acute care Surg. 2013;75:304–10.

    Article  PubMed  Google Scholar 

  4. Nikoloski AN, Osbrough AL, Yates PJ. Should the tip-apex distance (TAD) rule be modified for the proximal femoral nail antirotation (PFNA)? A retrospective study. J Orthop Surg Res. 2013;8:35.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Zhou JQ, Chang SM. Failure of PFNA: helical blade perforation and tip-apex distance. Injury. 2012;43:1227–8.

    Article  PubMed  Google Scholar 

  6. Hong JB, Dan Y, Ouyang L, Liu Y, Xiong LM, Li S, Feng XB, Shao ZW, Yan C, Yang SH, Liu P. Biomechanical study on different lengths of PFNA fixation for unstable intertrochanteric femoral fractures. J Musculoskel Neuronal Interact. 2017;17:299–302.

    Google Scholar 

  7. Horwitz DS, Tawari A, Suk M. Nail length in the management of intertrochanteric fracture of the Femur. J Am Acad Orthop Surg. 2016;24:e50–58.

    Article  PubMed  Google Scholar 

  8. Weiser L, Korecki MA, Sellenschloh K, Fensky F, Püschel K, Morlock MM, Rueger JM, Lehmann W. The role of inter-prosthetic distance, cortical thickness and bone mineral density in the development of inter-prosthetic fractures of the femur: a biomechanical cadaver study. bone Joint J. 2014;96–b:1378–84.

    Article  PubMed  Google Scholar 

  9. Wang F, Zou JL, Shang J. Does matching degree matter for proximal femoral intramedullary nail on reoperation rate in intertrochanteric fractures? J Orthop Surg Res. 2022;17:573.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Luque Pérez R, Checa Betegón P, Galán-Olleros M, Arvinius C, Valle-Cruz J, Marco F. Nailing unstable pertrochanteric fractures: does size matters? Arch Orthop Trauma Surg. 2022;142:145–55.

    Article  PubMed  Google Scholar 

  11. Frei HC, Hotz T, Cadosch D, Rudin M, Käch K. Central head perforation, or cut through, caused by the helical blade of the proximal femoral nail antirotation. J Orthop Trauma. 2012;26:e102–107.

    Article  PubMed  Google Scholar 

  12. Page PRJ, Poole WEC, Shah K, Upadhyay PK. Short or long intramedullary devices for hip fracture? A systematic review of the evidence. J Orthop. 2020;22:377–82.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Napoli N, Jin J, Peters K, Wustrack R, Burch S, Chau A, Cauley J, Ensrud K, Kelly M, Black DM. Are women with thicker cortices in the femoral shaft at higher risk of subtrochanteric/diaphyseal fractures? The study of osteoporotic fractures. J Clin Endocrinol Metab. 2012;97:2414–22.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  14. Jin Z, Xu S, Yang Y, Wei Y, Tian Y, Wang Z, Bai L. Cemented hemiarthroplasty versus proximal femoral nail antirotation in the management of intertrochanteric femoral fractures in the elderly: a case control study. BMC Musculoskelet Disord. 2021;22:846.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Li J, Xie Y, Sun S, Xue C, Xu W, Xu C, Xi Z. Regional differences in bone mineral density biomechanically induce a higher risk of adjacent vertebral fracture after percutaneous vertebroplasty: a case-comparative study. Int J Surg (London England) 2023.

  16. Xi Z, Xie Y, Chen S, Sun S, Zhang X, Yang J, Li J. The cranial vertebral body suffers a higher risk of adjacent vertebral fracture due to the poor biomechanical environment in patients with percutaneous vertebralplasty. Spine Journal: Official J North Am Spine Soc 2023.

  17. Rubio-Avila J, Madden K, Simunovic N, Bhandari M. Tip to apex distance in femoral intertrochanteric fractures: a systematic review. J Orthop Science: Official J Japanese Orthop Association. 2013;18:592–8.

    Article  Google Scholar 

  18. Nie S, Li J, Li M, Hao M, Wang K, Xiong Y, Gan X, Zhang L, Tang P. Finite-element analysis of a Novel Cephalomedullary nail for restricted sliding to reduce risk of Implant failure in unstable intertrochanteric fractures. Orthop Surg. 2022;14:3009–18.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Liu C, Cai HX, Zhang JF, Ma JJ, Lu YJ, Fan SW. Quantitative estimation of the high-intensity zone in the lumbar spine: comparison between the symptomatic and asymptomatic population. Spine Journal: Official J North Am Spine Soc. 2014;14:391–6.

    Article  Google Scholar 

  20. Pompe E, de Jong PA, de Jong WU, Takx RA, Eikendal AL, Willemink MJ, Oudkerk M, Budde RP, Lammers JW. Mohamed Hoesein FA: inter-observer and inter-examination variability of manual vertebral bone attenuation measurements on computed tomography. Eur Radiol. 2016;26:3046–53.

    Article  PubMed  PubMed Central  Google Scholar 

  21. Kelcey B, Shen Z, Spybrook J. Intraclass correlation coefficients for Designing cluster-randomized trials in Sub-saharan Africa Education. Eval Rev. 2016;40:500–25.

    Article  PubMed  Google Scholar 

  22. Akoglu H. User’s guide to correlation coefficients. Turkish J Emerg Med. 2018;18:91–3.

    Article  Google Scholar 

  23. Pripp AH. [Pearson’s or Spearman’s correlation coefficients]. Tidsskrift for den Norske laegeforening: tidsskrift for praktisk medicin, ny raekke 2018, 138.

  24. Li JC, Xie TH, Zhang Z, Song ZT, Song YM, Zeng JC. The Mismatch between Bony endplates and grafted bone increases Screw Loosening Risk for OLIF patients with ALSR fixation biomechanically. Front Bioeng Biotechnol. 2022;10:862951.

    Article  PubMed  PubMed Central  Google Scholar 

  25. Li J, Zhang Z, Xie T, Song Z, Song Y, Zeng J. The preoperative Hounsfield unit value at the position of the future screw insertion is a better predictor of screw loosening than other methods. European radiology 2022.

  26. Li J, Yin P, Zhang L, Chen H, Tang P. Medial anatomical buttress plate in treating displaced femoral neck fracture a finite element analysis. Injury. 2019;50:1895–900.

    Article  PubMed  Google Scholar 

  27. Li J, Xu C, Zhang X, Xi Z, Sun S, Zhang K, Fang X, Xie L, Liu Y, Song Y. Disc measurement and nucleus calibration in a smoothened lumbar model increases the accuracy and efficiency of in-silico study. J Orthop Surg Res. 2021;16:498.

    Article  PubMed  PubMed Central  Google Scholar 

  28. Xu C, Xi Z, Fang Z, Zhang X, Wang N, Li J, Liu Y. Annulus Calibration increases the computational accuracy of the lumbar finite element Model. Global Spine J 2022:21925682221081224.

  29. Li J, Xu C, Zhang X, Xi Z, Liu M, Fang Z, Wang N, Xie L, Song Y. TELD with limited foraminoplasty has potential biomechanical advantages over TELD with large annuloplasty: an in-silico study. BMC Musculoskelet Disord. 2021;22:616.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  30. Liang C, Peng R, Jiang N, Xie G, Wang L, Yu B. Intertrochanteric fracture: Association between the coronal position of the lag screw and stress distribution. Asian J Surg. 2018;41:241–9.

    Article  PubMed  Google Scholar 

  31. Zeng W, Liu Y, Hou X. Biomechanical evaluation of internal fixation implants for femoral neck fractures: a comparative finite element analysis. Comput Methods Programs Biomed. 2020;196:105714.

    Article  PubMed  Google Scholar 

  32. Li J, Han L, Zhang H, Zhao Z, Su X, Zhou J, Li C, Yin P, Hao M, Wang K, et al. Medial sustainable nail versus proximal femoral nail antirotation in treating AO/OTA 31-A2.3 fractures: Finite element analysis and biomechanical evaluation. Injury. 2019;50:648–56.

    Article  PubMed  Google Scholar 

  33. Chen DW, Lin CL, Hu CC, Tsai MF, Lee MS. Biomechanical consideration of total hip arthroplasty following failed fixation of femoral intertrochanteric fractures - a finite element analysis. Med Eng Phys. 2013;35:569–75.

    Article  PubMed  Google Scholar 

  34. Wang J, Ma JX, Lu B, Bai HH, Wang Y, Ma XL. Comparative finite element analysis of three implants fixing stable and unstable subtrochanteric femoral fractures: Proximal Femoral Nail Antirotation (PFNA), Proximal Femoral Locking Plate (PFLP), and Reverse Less Invasive Stabilization System (LISS). Orthopaedics & traumatology, surgery & research: OTSR 2020, 106:95–101.

  35. Galanopoulos IP, Mavrogenis AF, Megaloikonomos PD, Vottis CT, Mitsiokapa E, Koulouvaris P, Mastrokalos DS, Papagelopoulos PJ, Kontogeorgakos VA. Similar function and complications for patients with short versus long hip nailing for unstable pertrochanteric fractures. Sicot-j 2018, 4:23.

  36. Geller JA, Saifi C, Morrison TA, Macaulay W. Tip-apex distance of intramedullary devices as a predictor of cut-out failure in the treatment of peritrochanteric elderly hip fractures. Int Orthop. 2010;34:719–22.

    Article  PubMed  Google Scholar 

  37. Tianye L, Peng Y, Jingli X, QiuShi W, GuangQuan Z, Wei H, Qingwen Z. Finite element analysis of different internal fixation methods for the treatment of Pauwels type III femoral neck fracture. Biomed Pharmacotherapy = Biomedecine Pharmacotherapie. 2019;112:108658.

    Article  Google Scholar 

  38. Cho HM, Choi SM, Park JY, Lee Y, Bae JH. A finite element analysis and cyclic load experiment on an additional transcortical-type hole formed around the proximal femoral nail system’s distal locking screw. BMC Musculoskelet Disord. 2022;23:92.

    Article  PubMed  PubMed Central  Google Scholar 

  39. Wei Z, Chen S, Zhang S. Research progress on anteromedial cortical positive support reduction in treatment of intertrochanteric fracture. Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi = Zhongguo Xiufu Chongjian Waike Zazhi = Chin J Reparative Reconstr Surg. 2019;33:1216–22.

    Google Scholar 

  40. Nagura S. [Coxa Vara and coxa valga]. Seikeigeka Orthop Surg. 1967;18:659–61.

    CAS  Google Scholar 

  41. Renner SM, Natarajan RN, Patwardhan AG, Havey RM, Voronov LI, Guo BY, Andersson GB, An HS. Novel model to analyze the effect of a large compressive follower pre-load on range of motions in a lumbar spine. J Biomech. 2007;40:1326–32.

    Article  PubMed  Google Scholar 

  42. Schilling C, Krüger S, Grupp TM, Duda GN, Blömer W, Rohlmann A. The effect of design parameters of dynamic pedicle screw systems on kinematics and load bearing: an in vitro study. Eur Spine Journal: Official Publication Eur Spine Soc Eur Spinal Deformity Soc Eur Sect Cerv Spine Res Soc. 2011;20:297–307.

    Article  CAS  Google Scholar 

  43. Li JC, Yang ZQ, Xie TH, Song ZT, Song YM, Zeng JC. Deterioration of the fixation segment’s stress distribution and the strength reduction of screw holding position together cause screw loosening in ALSR fixed OLIF patients with poor BMD. Front Bioeng Biotechnol. 2022;10:922848.

    Article  PubMed  PubMed Central  Google Scholar 

  44. Stauff MP, Freedman BA, Kim JH, Hamasaki T, Yoon ST, Hutton WC. The effect of pedicle screw redirection after lateral wall breach–a biomechanical study using human lumbar vertebrae. Spine Journal: Official J North Am Spine Soc. 2014;14:98–103.

    Article  Google Scholar 

  45. Bartoska R, Baca V, Horak Z, Hrubina M, Skala-Rosenbaum J, Marvan J, Kachlik D, Dzupa V. The importance of intramedullary hip nail positioning during implantation for stable pertrochanteric fractures: biomechanical analysis. Surg Radiologic Anatomy: SRA. 2016;38:577–85.

    Article  Google Scholar 

Download references

Acknowledgements

Not Applicable.

Funding

Not Applicable.

Author information

Authors and Affiliations

Authors

Contributions

Conception and design: Xiaofeng Cha, Wenqiang Xu, and Jingchi Li. Acquisition of data: Ju jie Li, and Hong Xu. Analysis and interpretation of imaging data: Xiaofeng Cha, Ju jie Li, and Hong Xu. Statistical analysis: Xiaofeng Cha, Qin Zhou, and Jingchi Li. Manuscript Preparation: Xiaofeng Cha, and Jingchi Li. Manuscript revision and modification: Wenqiang Xu, Jingchi Li.

Corresponding authors

Correspondence to Wenqiang Xu or Jingchi Li.

Ethics declarations

Ethics approval and consent to participate

Approval for this study was obtained from the Ethics committee of the Luzhou Hospital of Traditional Chinese Medicine, Informed consent was waived for this retrospective study.

Consent for publication

Not Applicable.

Conflict of interest

The authors have no conflicts of interest to disclose in relation to this article.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Cha, X., Zhou, Q., Li, J. et al. Extending the intermedullary nail will not reduce the potential risk of femoral head varus in PFNA patients biomechanically: a clinical review and corresponding numerical simulation. BMC Musculoskelet Disord 25, 405 (2024). https://doi.org/10.1186/s12891-024-07334-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-024-07334-z

Keywords