Metal interference screw fixation combinations show high revision rates in primary hamstring tendon ACL reconstruction
BMC Musculoskeletal Disorders volume 25, Article number: 2 (2024)
Abstract
Background
Different fixation methods in anterior cruciate ligament reconstruction (ACLR) have been associated with different revision rates, specifically in the early postoperative period. However, most previous research has either grouped together different fixation types or evaluated femoral-sided fixation or tibial-sided fixation separately. Therefore, the purpose of this study was to determine ACL revision rates for specific combinations of femoral and tibial fixation methods within 2 years of primary hamstring tendon autograft ACLR based on data from the Swedish National Knee Ligament Registry (SNKLR).
Methods
Patients that underwent primary hamstring tendon autograft ACLR between 2005 and 2018 in the SNKLR were included. The collected data included patient characteristics (age, sex, body mass index [BMI]), activity at time of injury, surgical information (concomitant injuries, time from injury to surgery, fixation types at the femur and tibia), and subsequent revision ACLR. Revision rate within 2 years of the index procedure was chosen, as ACLR fixation is most likely to contribute to ACLR revision within the first 2 years, during graft maturation.
Results
Of the 23,238 included patients undergoing primary hamstring ACLR, 581 (2.5%) underwent revision ACLR within 2 years of the index procedure. Among the combinations used for > 300 patients, the femoral metal interference screw/tibial metal interference screw fixation combination had the highest revision rate followed by metal interference screw/resorbable screw and Endobutton/AO screw fixation combinations, with respective revision rates of 4.0, 3.0, and 3.0%. The lowest revision rate within 2 years of ACLR was found in the Endobutton/metal interference screw with backup Osteosuture fixation combination, used in 433 cases, with a failure rate of 0.9%.
Conclusion
Different early ACL revision rates were found across different combinations of femoral and tibial fixation devices within 2 years of primary hamstring tendon autograft ACLR. Metal interference screw fixation, particularly when performed on both the femoral and tibial sides, most frequently resulted in revision ACLR. These findings may be helpful for surgeons in selecting appropriate fixation devices for hamstring ACLR.
Level of evidence
IV
Introduction
Anterior cruciate ligament reconstruction (ACLR) revision is one of the most thoroughly researched topics in sports medicine, and yet there remain unanswered questions. Several factors contributing to ACL revision risk have been identified, including patient-, technical-, and injury-related factors [1,2,3,4]. Two often discussed technical decisions impacting ACLR revision rates are graft choice and fixation method [3, 5]. While by far the most frequently used graft type in Sweden continues to be hamstring tendon autograft, preferences regarding fixation vary widely among surgeons and healthcare institutions [6]. As a result of consistency in graft choice among Swedish surgeons, short-term outcomes may instead depend upon the method of graft fixation [7,8,9].
Fixation methods in ACLR have been shown to impact revision rate, specifically in the early postoperative period [7, 10]. Most prior literature has either grouped together femoral and tibia-sided fixation types or evaluated femoral-sided fixation or tibial-sided fixation separately [5, 11, 12]. However, previous studies have not investigated the role of specific femoral and tibial fixation combinations on ACLR revision rates within the early postoperative period. Analyzing the interaction between fixation devices on both the femoral and tibial sides is important, as their effect may be complementary or contradictory, which cannot be fully comprehended by evaluating each tibial and femoral fixation type separately.
The purpose of this study was to determine the rate of ACL revision with respect to specific combinations of femoral and tibial fixation devices within 2 years following primary hamstring tendon autograft ACLR based on data from the Swedish National Knee Ligament Registry (SNKLR).
Methods
This registry study was performed in accordance with the Declaration of Helsinki and was approved by the Regional Ethical Board in Stockholm, Sweden (2011/337–31/3), and the Swedish Ethical Review Authority (2022–00913-01).
The data source was the SNKLR, a national registry of cruciate ligament surgical procedures in Sweden. The registry includes both surgeon and patient reported data including patient, injury, and surgical characteristics as well as patient-reported outcome measures. While patient demographic characteristics, injury- and surgery-related factors are reported by the surgeon, questionnaires regarding current knee function are filled out by the patient. The registry has been described in more detail in previous literature [13, 14].
Study population
Patients aged > 13 years undergoing primary ACLR with hamstring tendon autograft between 2005 and 2018 in the SNKLR were included. Patients with prior knee surgery, double-bundle ACLR, concomitant fracture (patella, femur, tibia, or fibula), other ligament injury, or neurovascular injury were excluded from further analysis. Fixation types “other”, “XO-button” as well as fixation types that are only used in ACLR with transtibial technique (Rigidfix, Transfix) were excluded because of the modern understanding of the superiority of the tibial-independent drilling technique at restoring the anatomy of the ACL [15,16,17]. Demographic data including patient age, sex, body mass index (BMI), and activity at time of injury were extracted from the SNKLR. Surgical information on concomitant injuries (meniscus, cartilage), time from injury to surgery, fixation types (Table 1) and subsequent revision ACLR were extracted. Activity performed at the time of injury was divided into six different groups: alpine/skiing, pivoting sport (American football/rugby, basketball, dancing, floorball, gymnastics, handball, ice hockey/bandy, martial arts, racket sports, soccer, volleyball, wrestling), non-pivoting sport (cross-country skiing, cycling, horseback riding, motocross/endure, skateboarding, snowboarding, and surfing/wakeboarding), other physical activity (other recreational sport, exercise, trampoline), traffic related, and other (other outdoor activity and work).
The primary outcome was the ipsilateral revision rate by different femoral and tibial fixation device combinations, reported as “femoral fixation/tibial fixation” with backup fixation if applicable, within 2 years following primary ACLR. Revision rate within 2 years of the index procedure was chosen because ACLR fixation is most likely to contribute to ACLR revision in the first 2 years, during the process of graft maturation [18, 19]. Beyond 2 years, other factors such as tunnel position and patient activity level are likely more meaningful contributors to failure and thus revision.
Statistical analyses
All statistical analyses were performed by using the SAS System for Windows (version 9, SAS Institute, Cary, North Carolina, USA). The count (n) and proportion (%) were used to present categorical variables, while continuous and ordinal data were presented by using the mean and standard deviation (SD), as well as the median with minimum and maximum.
Results
Baseline characteristics
Of the 23,238 patients included in this study, 13,087 (56%) were males (Table 2). The mean age at time of surgery was 27.3 years + 10.4, and the mean BMI was 24.5 + 3.3 kg/m2. The mean time from ACL injury to surgery was 19.1 + 33.7 months.
Fixation types and revision rate
Endobutton and Tightrope were the most common femoral fixation devices (52.5 and 35.2%) (Table 3), while resorbable screw and AO screw were the most frequently used tibial fixation devices (24.3 and 22.9%). Of the included patients, 581 (2.5%) underwent revision surgery within 2 years after the primary ACLR.
Revision by combination of femoral/tibial fixation used in > 20 patients
The most common combination of femoral/tibial fixation leading to revision within 2 years after primary ACLR was Ultrabutton/AO screw combination with a revision rate of 10.5% (Fig. 1). The Endobutton/Suture washer and Retrobutton/Intrafix combinations were also found to have high revision rates (8.8 and 7.5%, respectively).
Revision by combination of femoral/tibial fixation used in > 300 patients
The most common fixation combination leading to revision was metal interference screw/metal interference screw followed by metal interference screw/resorbable screw and Endobutton/AO screw, with revision rates of 4.0, 3.0, and 3.0%, respectively (Fig. 2). The Endobutton/metal interference screw with backup Osteosuture fixation combinations, used in a total of 433 cases, had the lowest revision rate (0.9%). The Endobutton/resorbable screw with backup post fixation combination had the second-lowest revision rate (1.9%) and was used in a total of 533 cases.
Discussion
The main finding of this study was that, among the most widely used fixation combinations, the greatest 2-year ACL revision rate was found when metal interference screws were used for both femoral and tibial fixation.
This study uniquely examines combinations of femoral- and tibial-sided fixation in a large volume of primary hamstring ACLR. While the majority of prior literature compares ACLR revision rates based on either femoral fixation or tibial fixation type in isolation, the evaluation of fixation combinations is an important contribution to our understanding of the relationship between fixation and ACLR failure. For example, aperture (i.e., interference screw) fixation is theorized to lead to a shorter mobile length of the graft, minimizing graft micromotion with potentially improved graft-tunnel healing [7, 8]. Graft fixation with both femoral and tibial interference screws would lead to the shortest mobile graft length, whereas fixation with one screw and one suspensory device would lead to a different mobile graft length. Therefore, the tibial-sided fixation could affect the biomechanical characteristics of the femoral-sided fixation. As another example, if interference screws risk damage to soft tissue grafts, [20, 21] then a construct with two interference screws would theoretically have a higher risk of graft revision than a construct with one screw. Once again, the combination of fixation devices matters more than fixation on one side alone.
Adding to previous literature, this study found that the metal interference screw/metal interference screw fixation combination resulted in a revision surgery most commonly among the more widely used fixation combinations. Previous literature shows varying effects of metal interference screw fixation on ACLR revision rates, for example with increased risk of revision ACLR with femoral metal interference screw fixation [9] and decreased revision risk with tibial metal interference screw fixation compared with other fixation types [22]. Furthermore, some previous literature has demonstrated lower stability rates and higher revision rates followed by ACLR with femoral aperture fixation compared to suspensory fixation [23, 24]. However, revision rates compared between the femoral and tibial metal interference screw combination versus other fixation combinations have not been reported, highlighting the original contribution of this study to the current literature.
The prominent revision rate seen with metal interference screws could be related to the metal threads cutting into the graft, thereby weakening it [20, 21]. This explanation is further supported by the minor revision rate in the current study with metal interference screw fixation combined with Osteosuture backup fixation, in which case the sutures may maintain graft fixation despite destruction of graft fibers by the cutting forces of the screw. In contrast, metal screws on both femoral and tibial sides, as opposed to one side only, may increase the risk of graft injury.
In contrast to previous studies, low revision rates were found in patients treated with femoral suspensory and tibial interference screw fixation combinations [11, 25, 26]. During the late 2000s and early 2010s, the use of anatomic ACLR surged in popularity, with a concurrent increase in the adoption of suspensory and interference screw fixation in tibial-independent anatomic ACLR [6, 27, 28]. Thus, the introduction of this new and complex technique may have involved a higher revision rate due to a learning curve and, subsequently, could have led to the increased revision rates with the fixation combinations frequently used in anatomic ACLR [29, 30]. However, improved outcomes have been observed over the recent years with anatomic ACLR, [29] which may partially explain our findings of low revision rates in patients with femoral suspensory and tibial interference screw fixations combinations. Examples of such fixation methods in the current study are the Endobutton/metal interference screw with backup Osteosuture fixation and the Endobutton/resorbable screw with backup post fixation combinations.
This study had strengths and limitations. Most importantly, the current study included a large patient sample and detailed information on 23,238 patients with ACLR. The SNKLR has previously been described to include data on > 90% of all ACLR performed in Sweden [6]. Consequently, the study population can be considered representative of the overall Swedish ACLR population. One limitation is that this study only included all soft tissue hamstring autograft, limiting the generalizability of the results to other graft choices. Second, this study defined failure as revision surgery, neglecting patients with poor functional outcomes who have clinically failed despite not undergoing revision surgery. The relatively long time from injury to surgery may affect the generalizability of the study results to populations undergoing earlier ACLR. Furthermore, this study did not include information on radiographic alignment, lateral extra articular procedures, postoperative rehabilitation protocols or complications, such as infection, as the data was not available from the registry. Finally, possible variation in concomitant injuries and their treatment were not assessed among the different fixation groups.
Conclusion
Different early ACL revision rates were found across different combinations of femoral and tibial fixation devices within 2 years of primary hamstring tendon autograft ACLR. Metal interference screw fixation, particularly when performed on both the femoral and tibial sides, was found to result most commonly in a revision surgery. These findings may be helpful for surgeons in selecting appropriate fixation devices for hamstring ACLR.
Availability of data and materials
The dataset analyzed during the current study are available from the corresponding author on reasonable request.
References
Wiggins AJ, Grandhi RK, Schneider DK, Stanfield D, Webster KE, Myer GD. Risk of secondary injury in younger athletes after anterior cruciate ligament reconstruction: a systematic review and Meta-analysis. Am J Sports Med. 2016;44(7):1861–76.
Effect of graft choice on the outcome of revision anterior cruciate ligament reconstruction in the multicenter ACL revision study (MARS) cohort. Am J Sports Med. 2014;42(10):2301–10.
Zhao D, Pan JK, Lin FZ, Luo MH, Liang GH, Zeng LF, et al. Risk factors for revision or Rerupture after anterior cruciate ligament reconstruction: a systematic review and Meta-analysis. Am J Sports Med. 2022;51(11):3053–75.
Svantesson E, Hamrin Senorski E, Baldari A, Ayeni OR, Engebretsen L, Franceschi F, et al. Factors associated with additional anterior cruciate ligament reconstruction and register comparison: a systematic review on the Scandinavian knee ligament registers. Br J Sports Med. 2019;53(7):418–25.
Rahardja R, Love H, Clatworthy MG, Monk AP, Young SW. Suspensory versus interference Tibial fixation of hamstring tendon autografts in anterior cruciate ligament reconstruction: results from the new Zealand ACL registry. Am J Sports Med. 2022;50(4):904–11.
XBase The Swedish knee ligament registry. Annual Report 2020. . In.; 2020.
Brand J Jr, Weiler A, Caborn DN, Brown CH Jr, Johnson DL. Graft fixation in cruciate ligament reconstruction. Am J Sports Med. 2000;28(5):761–74.
Ishibashi Y, Rudy TW, Livesay GA, Stone JD, Fu FH, Woo SL. The effect of anterior cruciate ligament graft fixation site at the tibia on knee stability: evaluation using a robotic testing system. Arthroscopy. 1997;13(2):177–82.
Snaebjörnsson T, Hamrin Senorski E, Svantesson E, Westin O, Persson A, Karlsson J, et al. Graft fixation and timing of surgery are predictors of early anterior cruciate ligament revision: a cohort study from the Swedish and Norwegian knee ligament registries based on 18,425 patients. JB JS Open Access. 2019;4(4):e0037.
Kurosaka M, Yoshiya S, Andrish JT. A biomechanical comparison of different surgical techniques of graft fixation in anterior cruciate ligament reconstruction. Am J Sports Med. 1987;15(3):225–9.
Eysturoy NH, Nissen KA, Nielsen T, Lind M. The influence of graft fixation methods on revision rates after primary anterior cruciate ligament reconstruction. Am J Sports Med. 2018;46(3):524–30.
Asif N, Khan MJ, Haris KP, Waliullah S, Sharma A, Firoz D. A prospective randomized study of arthroscopic ACL reconstruction with adjustable- versus fixed-loop device for femoral side fixation. Knee Surg Relat Res. 2021;33(1):42.
Hamrin Senorski E, Svantesson E, Engebretsen L, Lind M, Forssblad M, Karlsson J, et al. 15 years of the Scandinavian knee ligament registries: lessons, limitations and likely prospects. Br J Sports Med. 2019;53(20):1259–60.
Kvist J, Kartus J, Karlsson J, Forssblad M. Results from the Swedish national anterior cruciate ligament register. Arthroscopy. 2014;30(7):803–10.
Rothrauff BB, Jorge A, de Sa D, Kay J, Fu FH, Musahl V. Anatomic ACL reconstruction reduces risk of post-traumatic osteoarthritis: a systematic review with minimum 10-year follow-up. Knee Surg Sports Traumatol Arthrosc. 2020;28(4):1072–84.
Loucas M, Loucas R, D'Ambrosi R, Hantes ME. Clinical and radiological outcomes of anteromedial portal versus Transtibial technique in ACL reconstruction: a systematic review. Orthop J Sports Med. 2021;9(7):23259671211024591.
Desai N, Andernord D, Sundemo D, Alentorn-Geli E, Musahl V, Fu F, et al. Revision surgery in anterior cruciate ligament reconstruction: a cohort study of 17,682 patients from the Swedish National Knee Ligament Register. Knee Surg Sports Traumatol Arthrosc. 2017;25(5):1542–54.
Chu CR, Williams AA. Quantitative MRI UTE-T2* and T2* show progressive and continued graft maturation over 2 years in human patients after anterior cruciate ligament reconstruction. Orthop J Sports Med. 2019;7(8):2325967119863056.
Lutz PM, Achtnich A, Schütte V, Woertler K, Imhoff AB, Willinger L. Anterior cruciate ligament autograft maturation on sequential postoperative MRI is not correlated with clinical outcome and anterior knee stability. Knee Surg Sports Traumatol Arthrosc. 2022;30(10):3258–67.
Zantop T, Weimann A, Schmidtko R, Herbort M, Raschke MJ, Petersen W. Graft laceration and pullout strength of soft-tissue anterior cruciate ligament reconstruction: in vitro study comparing titanium, poly-d,l-lactide, and poly-d,l-lactide-tricalcium phosphate screws. Arthroscopy. 2006;22(11):1204–10.
Sawyer GA, Anderson BC, Paller D, Heard WM, Fadale PD. Effect of interference screw fixation on ACL graft tensile strength. J Knee Surg. 2013;26(3):155–9.
Andernord D, Björnsson H, Petzold M, Eriksson BI, Forssblad M, Karlsson J, et al. Surgical predictors of early revision surgery after anterior cruciate ligament reconstruction: results from the Swedish National Knee Ligament Register on 13,102 patients. Am J Sports Med. 2014;42(7):1574–82.
Prodromos CC, Joyce BT, Shi K, Keller BL. A meta-analysis of stability after anterior cruciate ligament reconstruction as a function of hamstring versus patellar tendon graft and fixation type. Arthroscopy. 2005;21(10):1202.
Speziali A, Delcogliano M, Tei M, Placella G, Bartoli M, Menghi A, et al. Fixation techniques for the anterior cruciate ligament reconstruction: early follow-up. A systematic review of level I and II therapeutic studies. Musculoskelet Surg. 2014;98(3):179–87.
Persson A, Kjellsen AB, Fjeldsgaard K, Engebretsen L, Espehaug B, Fevang JM. Registry data highlight increased revision rates for endobutton/biosure HA in ACL reconstruction with hamstring tendon autograft: a nationwide cohort study from the Norwegian knee ligament registry, 2004-2013. Am J Sports Med. 2015;43(9):2182–8.
Spragg LM, Prentice HA, Morris A, Funahashi TT, Maletis GB, Csintalan RP. Femoral-tibial fixation affects risk of revision and reoperation after anterior cruciate ligament reconstruction using hamstring autograft. Knee Surg Sports Traumatol Arthrosc. 2019;27(11):3518–26.
Persson A, Gifstad T, Lind M, Engebretsen L, Fjeldsgaard K, Drogset JO, et al. Graft fixation influences revision risk after ACL reconstruction with hamstring tendon autografts. Acta Orthop. 2018;89(2):204–10.
Tibor L, Chan PH, Funahashi TT, Wyatt R, Maletis GB, Inacio MC. Surgical technique trends in primary ACL reconstruction from 2007 to 2014. J Bone Joint Surg Am. 2016;98(13):1079–89.
Eysturoy NH, Nielsen TG, Lind MC. Anteromedial portal drilling yielded better survivorship of anterior cruciate ligament reconstructions when comparing recent versus early surgeries with this technique. Arthroscopy. 2019;35(1):182–9.
Rahr-Wagner L, Thillemann TM, Pedersen AB, Lind MC. Increased risk of revision after anteromedial compared with transtibial drilling of the femoral tunnel during primary anterior cruciate ligament reconstruction: results from the Danish knee ligament reconstruction register. Arthroscopy. 2013;29(1):98–105.
Acknowledgements
The authors would like to thank statistician Bengt Bengtsson from the Statistiska Konsultgruppen for help with the statistical analyses.
Funding
Open access funding provided by University of Gothenburg.
Author information
Authors and Affiliations
Contributions
All listed authors have contributed substantially to this work: literature search, interpretation of the results, and initial manuscript preparation were performed by JK, NAB, IDE, EMN, and BZ. EHS, VM, and KS assisted with study design and finalizing the manuscript. All authors have read and approved the final manuscript to be submitted and published.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
In Sweden, the law (2008:35) states that written consent is not required for national registries. Patients are provided with information about the registry and possibility to withdraw from participation at any time. This study was approved by the Regional Ethical Board in Stockholm, Sweden (2011/337–31/3), and the Swedish Ethical Review Authority (2022–00913-01) as well as performed in accordance with the Declaration of Helsinki. The Swedish National Knee Ligament Registry complies with the Swedish legislation related to data security and therefore, no written consent is necessary for national databases in Sweden. Investigators had access only to unidentifiable patient data.
Consent for publication
Not applicable.
Competing interests
JK, NAB, IDE, EMN, and BZ have nothing to disclose. EHS is an associate editor of the Journal of Orthopaedic and Sports Physical Therapy. Volker Musahl reports educational grants, consulting fees, and speaking fees from Smith & Nephew plc, educational grants from Arthrex and DePuy/Synthes, is a board member of the International Society of Arthroscopy, Knee Surgery and Orthopaedic Sports Medicine (ISAKOS), and deputy editor-in-chief of Knee Surgery, Sports Traumatology, Arthroscopy (KSSTA). Kristian Samuelsson is a member of the Board of Directors in Getinge AB (publ.)
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
About this article
Cite this article
Kaarre, J., Benvegnu, N.A., Engler, I.D. et al. Metal interference screw fixation combinations show high revision rates in primary hamstring tendon ACL reconstruction. BMC Musculoskelet Disord 25, 2 (2024). https://doi.org/10.1186/s12891-023-07109-y
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12891-023-07109-y