Do metal implants for total hip arthroplasty affect bioelectrical impedance analysis? A retrospective study
BMC Musculoskeletal Disorders volume 24, Article number: 763 (2023)
Abstract
Background
Evaluation of body composition after total hip arthroplasty (THA) is essential because it can be used to evaluate muscle and functional recovery. Muscle volume and degeneration are evaluated using computed tomography (CT). However, CT evaluation of muscle volume has several limitations, such as radiation exposure and high medical cost. Bioelectrical impedance analysis (BIA) has gained attention for resolving these limitations of CT. BIA takes advantage of the microelectric current; thus, metal implants may affect the results. Therefore, this study aimed to elucidate the effects of metal implants on BIA after THA.
Methods
Two groups of patients were assessed (Group 1: 70 patients who underwent unilateral THA and BIA; Group 2: 35 patients who underwent THA and BIA before and after THA). Electric impedance (resistance and reactance) of the operated and non-operated lower limbs was compared in Group 1. The pre- and post-operative impedances of the torso and operated ipsilateral limbs were compared in Group 2.
Results
Regarding electric impedance in Group 1, no significant differences were observed in electrical resistance and reactance between the operated and non-operated lower limbs. Concerning electric impedance in Group 2, postoperative electric resistance of the torso was significantly lower than that preoperatively. However, no significant difference was seen in electric resistance and reactance of the operated ipsilateral limbs preoperatively and postoperatively.
Conclusions
Electrical resistance and reactance of the limbs did not change significantly after THA. BIA is useful for measuring body composition after THA.
Background
The number of total hip arthroplasties (THAs) has been increasing. THA is one of the most developed surgeries in the orthopedic field and can not only relieve hip pain but also improve gait motion [1]. Muscle strength substantially affects gait, and postoperative muscle evaluation is important in patients after THA. The effectiveness of muscle volume evaluation by using computed tomography (CT) has been previously reported [2,3,4]. Hounsfield units (HUs) vary depending on the tissue (bone, 400–100 HU; muscle, 30–50 HU; water, 0 HU; fat, -100 HU; air, -1000 HU) [5]. Taking advantage of this feature, physicians have evaluated muscle volume and adipose degeneration using CT [4]. However, CT has several disadvantages such as radiation exposure and increased medical cost.
Bioelectrical impedance analysis (BIA) has attracted considerable attention for resolving these problems associated with CT. Body composition is strongly affected by physical function [1, 6,7,8,9] and correlated with osteoarthritis [10], the Barthel index [11], and quadriceps strength [10, 12]. With aging of the population, the number of arthroplasties is also increasing. It is estimated that people have a 7–12% lifetime risk for THA and 8–11% for total knee arthroplasty in the United Kingdom [13]. Therefore, assessment of body composition after arthroplasty will become more important in the near future. Electric impedance consists of resistance and reactance, and BIA devices estimate body composition. Resistance is defined as the power that counteracts electric flow through both direct and alternating currents, and reactance is defined as the power that counteracts the electric flow using only an alternating current. An electric current can easily flow through a fat-free mass containing large amounts of water and electrolytes, but it does not flow through fatty tissue, which contains less water. BIA takes advantage of this principle to estimate the impedance of the body.
BIA enables the measurement of various body conditions such as the fat mass index, free fat mass index, and skeletal muscle volume by applying a harmless electric current to the body. Although BIA can measure body composition easily and rapidly and is more affordable than CT, metal implants such as titanium, steel, and cobalt-chrome can decrease the electrical impedance and affect the results of BIA. However, no studies have elucidated the effects of metal implants on BIA. The primary purpose of this study was to answer the following questions: (1) Is there any difference in electrical impedance between the operated and non-operated limbs? (2) Is there any difference in electrical impedance between the preoperative and postoperative ipsilateral limbs? (3) Is there any difference in electrical impedance between the preoperative and postoperative torso?
Methods
Study design and patients
To elucidate the effects of metal implants on BIA, this retrospective study investigated two groups of patients between 2019 and 2023. Group 1 included 70 patients who underwent unilateral THA and BIA postoperatively; Group 2 included 35 patients who underwent unilateral THA and BIA preoperatively and postoperatively. Patients in Group 1 underwent BIA 6 months after THA. Patients in Group 2 underwent BIA 1 month before and 6 months after THA. Patients who had pacemakers or could not remain in the standing position independently were excluded.
Surgical procedures
Total hip arthroplasty
Metal implants (acetabular shell and stem) were placed during THA. The acetabular shells were fixed using two or three screws. Polyethylene liners and ceramic heads were used in all THAs (Fig. 1). The metal compositions of the implants are listed in Table 1.
Bioelectrical impedance analysis
BIA was performed using mBCA515 (Seca, Hamburg, Germany). This device was selected as it was constructed for medical application and has been used for evaluation in various research fields such as that for metabolic diseases, cancer, and heart and lung transplantation [14,15,16,17]. Moreover, mBCA515 was validated by magnetic resonance imaging [18] and could measure the impedance of each limb and trunk separately at 19 frequencies ranging from 5 to 75 kHz. This device also has eight electrodes, and each handle and footplate has detector and current electrode pairs (Fig. 2). These electrodes were connected to a laptop that could calculate impedance. By constructing a circuit with these electrodes, we could calculate bioelectrical impedance for each body part, including the right arm, right limb, left arm, left limb, both lower limbs, and torso (Fig. 3). The within-day and between-day impedance were measured. Eight participants (four men and four women) underwent BIA three times within the same day to evaluate the within-day impedance variation. They also underwent BIA for three consecutive days to evaluate between-day impedance variation. Participants were instructed to maintain a standing position by attaching all electrodes for 17 s. The bioelectrical resistances of the torso and lower limbs were recorded. BIA was performed under stable temperature and humidity.
Statistical analyses
G*Power software (version 3.1.9.2, Germany) was used to calculate the required sample sizes. The parameters of the Mann–Whitney U-test were as follows: two groups; moderate effect size, d = 0.5; alpha error, 0.5; and power, 80%. The required sample size was 106 limbs. The parameters of the Wilcoxon signed-rank test were as follows: one-sample case; moderate effect size, d = 0.5; alpha error, 0.5; and power, 80%. The required sample size was 35.
The Mann–Whitney test was used to compare bioelectrical resistance and reactance between the operated and non-operated lower limbs of patients with THA. The Wilcoxon signed-rank test was used to compare pre- and post-operative bioelectrical resistance and reactance of the torso and ipsilateral lower limbs. All tests were performed at a significance level of P < 0.05. Analyses were performed using SPSS statistical software, version 26 (IBM Corp., Armonk, NY, USA).
Results
Demographic data are presented in Table 2. The lower limb and torso within-day impedance variation was 0.3–1.5% and 0.7–2.6% respectively. The lower limb and torso between-day impedance variation was 0.9–2.7% and 0.8–2.8%, respectively. Regarding the comparison of electrical resistance and reactance between the operated and non-operated lower limbs, slight differences were observed; however, they were not significant. (Table 3). Additionally, in comparing ipsilateral lower limbs preoperatively and postoperatively, slight differences were observed; however, they were not significant (Table 4). In comparing preoperative and postoperative torso, the resistance of the postoperative torso was significantly lower; however, no significant difference was observed between the reactance of the torso preoperatively and postoperatively (Table 5).
Discussion
This study revealed that neither electrical resistance nor reactance of the lower limbs significantly impacted THA. Few reports have evaluated the electrical impedance of metal implants and its effect on BIA measurements [19]. Wagner compared the operated tibial nail side with the non-operated side in a case report [19]. According to the report, the reactance of the operated side decreased by 9.2% and its resistance decreased by 5.2% compared with the non-operated side [19]. Although the electrical resistance of fatty tissue is very high, those of muscle tissue and metal are low. Thus, the postoperative muscle mass may be overestimated after THA. However, only slight difference was observed in electrical resistance and reactance, and no significant difference was observed in this study between the operated and non-operated limbs as well as between the lower limbs preoperatively and postoperatively. Thus, the length of the implant may affect impedance because the resistance of implants varies based on its length. The tibia nail is longer than the stem, and the proportion of the nail implant on the lower limbs is higher than that of arthroplasty. Thus, the implant length and proportion of the implant in the lower limbs may affect impedance. We used a short stem compared with other stems in this study that may have also affected the result; hence, the impedance of a long stem should be elucidated in future research.
Implants are set between the torso and the upper end of the thigh. Although the resistance of the postoperative torso was significantly decreased compared with the preoperative torso, the mean difference value was only 1 ohm; this slight change may be due to alterations in hydration or body composition. One viewpoint is that the torso contains only a part of the screws and does not contain most of the acetabular shell. Another viewpoint is that the postoperative reactance of the torso was not significant. We believe that implants have little effect on bioelectrical impedance. We consider that the resistance and reactance value of the torso was a tenth of that of the lower limbs; thus, only slight change was significant in the torso.
Other studies also performed BIA in patients with fracture [20,21,22]. Gonzalez-Montalvo et al. performed BIA in patients with hip fractures in the first 72 h after admission and at discharge [21]. However, body composition after a fracture may be quite different from that of degenerative diseases because bleeding directly affects the water balance in the body and the results of BIA. Steihaug et al. evaluated the effect of surgical implants in patients with hip fractures [22]. They compared the electrical impedance between the hospital stay and 3 months after surgery. Their results revealed that although the resistance of the fractured side was significantly decreased in the hospital, there was no significant difference in resistance and reactance between the operated and non-operated limbs at the final follow-up [22]. In addition, resistance and reactance did not vary depending on the surgical implant, such as cannulated screw, hemiarthroplasty, hip compression screw, or THA [22]. This result indicates that BIA is more affected by water balance factors, such as edema, bleeding, and dehydration, than metal implants. However, these studies only performed BIA in patients with hip fractures and only after surgery. We believe that body composition may be affected by fractures and that electrical impedance should be evaluated in non-fractured individuals. Therefore, we included non-fractured individuals and performed postoperative BIA after 6 months to minimize the effect of the operation.
Our report is the first to evaluate the electric impedance preoperatively and postoperatively, and between the operated side and non-operated sides simultaneously. Our results revealed that the resistance and reactance of the limbs did not significantly change after THA. The impedance of the internal body, such as skeletal muscle, cardiac muscle, kidneys, liver, lungs, and spleen, was 171 ohm, 175 ohm, 211 ohm, 342 ohm, 157 ohm, and 405 ohm, respectively [23]. The device used in this study can detect body impedance ranging from 10 to 1000 ohm. However, the impedance of the metal implant was extremely low compared to the detection range of BIA. For example, the impedance of titanium, cobalt, and chrome in THA were 42.7 × 10− 8, 6.24 × 10− 8, and 12.9 × 10− 8 ohm, respectively. Therefore, we consider that BIA cannot detect metal implants and that metal implants do not affect BIA. Additionally, most implants that we used in this study were poor electrical conductors (titanium) and non-conductors (ceramic and polyethylene). Impedance highly varies on the metal type, and the metal length as well as the metal volume may affect the impedance too. Thus, the impedance of the metal type and patients with a metal liner and metal head should be evaluated. The results of this study indicate that BIA is useful for evaluating body composition, even in patients who have undergone THA.
This study had several limitations. First, this study was retrospective, and the sample size was small. To our best knowledge, this is the first study to assess electrical resistance and reactance in patients with THA. Thus, we believe that this study provides new insights into the usefulness of BIA in patients undergoing THA. Second, the metal implants were not unified. However, the composition of the acetabular shell and stem was the same (mainly cobalt, chrome, and titanium), and the electrical resistance and reactance did not vary among implant types [22]. Third, BIA is affected by several factors, such as temperature, exercise, and meals. Therefore, we performed BIA under the same conditions (temperature and humidity) to minimize environmental effects.
Conclusions
Electrical resistance and reactance did not significantly change between the operated and non-operated sides or between the lower limbs preoperatively and postoperatively. Thus, BIA may be useful for evaluating body composition even after THA.
Data Availability
The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.
Abbreviations
- BIA:
-
Bioelectrical impedance analysis
- THA:
-
Total hip arthroplasty
- CT:
-
Computed tomography
- HU:
-
Hounsfield unit
References
Majewski M, Bischoff-Ferrari HA, Grüneberg C, Dick W, Allum JHJ. Improvements in balance after total hip replacement. J Bone Joint Surg Br. 2005;87:1337–43.
Rasch A, Byström AH, Dalén N, Martinez-Carranza N, Berg HE. Persisting muscle atrophy two years after replacement of the hip. J Bone Joint Surg Br. 2009;91:583–8.
Uemura K, Takao M, Sakai T, Nishii T, Sugano N. Volume increases of the gluteus maximus, gluteus medius, and thigh muscles after hip arthroplasty. J Arthroplasty. 2016;31:906–912e1.
Ukai T, Ebihara G, Omura H, Watanabe M. Evaluation of muscle volume and degeneration after total hip arthroplasty: a comparison of the posterolateral approach and the anterolateral supine approach. J Orthop Surg Res. 2021;16:145.
Goodpaster BH, Kelley DE, Thaete FL, He J, Ross R. Skeletal muscle attenuation determined by computed tomography is associated with skeletal muscle lipid content. J Appl Physiol. (1985). 2000;89:104–10.
Norman K, Stobäus N, Pirlich M, Bosy-Westphal A. Bioelectrical phase angle and impedance vector analysis-clinical relevance and applicability of impedance parameters. Clin Nutr. 2012;6:854–61.
Merchant RA, Seetharaman S, Au L, Wong MWK, Wong BLL, Tan LF, et al. Relationship of fat mass index and fat free mass index with body mass index and association with function, cognition and sarcopenia in pre-frail older adults. Front Endocrinol. 2021;12:765415.
Gunn SM, Halbert JA, Giles LC, Stepien JM, Miller MD, Crotty M. Bioelectrical phase angle values in a clinical sample of ambulatory rehabilitation patients. Dyn Med. 2008;7:14.
Tanaka S, Ando K, Kobayashi K, Hida T, Seki T, Suzuki K, et al. Relationship between locomotive syndrome and body composition among community-dwelling middle-age and elderly individuals in Japan: the Yakumo study. Mod Rheumatol. 2019;29:491–5.
Wada O, Kurita N, Yamada M, Mizuno K. Structural severity, phase angle, and quadriceps strength among patients with knee osteoarthritis: the SPSS-OK study. Clin Rheumatol. 2020;39:3049–56.
Norman K, Stobäus N, Zocher D, Bosy-Westphal A, Szramek A, Scheufele R, et al. Cutoff percentiles of bioelectrical phase angle predict functionality, quality of life, and mortality in patients with cancer. Am J Clin Nutr. 2010;92:612–9.
Yamada M, Kimura Y, Ishiyama D, Nishio N, Otobe Y, Tanaka T, et al. Phase angle is a useful indicator for muscle function in older adults. J Nutr Health Aging. 2019;23:251–5.
Culliford DJ, Maskell J, Kiran A, Judge A, Javaid MK, Cooper C, et al. The lifetime risk of total hip and knee arthroplasty: results from the UK general practice research database. Osteoarthritis Cartilage. 2012;20:519–24.
Domaszewski P, Konieczny M, Dybek T, Łukaniszyn-Domaszewska K, Anton S, Sadowska-Krępa E, et al. Comparison of the effects of six-week time-restricted eating on weight loss, body composition, and visceral fat in overweight older men and women. Exp Gerontol. 2023;174:112116.
Durkin K, Heetun A, Ewings S, Munday R, Wootton SA, Turner L, et al. Body composition and chemotherapy toxicity in women with early breast cancer (CANDO-3): protocol for an observational cohort study. BMJ Open. 2022;12:e054412.
Jastrzębska W, Radtke-Łysek A, Bohdan M, Frankiewicz A, Siondalski P, Żegleń S et al. Body composition before and after heart or lung transplantation. Preliminary results. Transplant Proc. 2022;54:1070–3.
Gierach M, Junik R. Assessment of metabolic parameters in female triathletes with Hashimoto’s thyroiditis in Poland. Biomedicines. 2023;11:769.
Bosy-Westphal A, Jensen B, Braun W, Pourhassan M, Gallagher D, Müller MJ. Quantification of whole-body and segmental skeletal muscle mass using phase-sensitive 8-electrode medical bioelectrical impedance devices. Eur J Clin Nutr. 2017;71:1061–7.
Wagner DR. Case study: effect of surgical metal implant on single frequency bioelectrical impedance measures of an athlete. Physiol Rep. 2020;8:e14464.
Villani AM, Miller M, Cameron ID, Kurrle S, Whitehead C, Crotty M. Body composition in older community-dwelling adults with hip fracture: portable field methods validated by dual-energy X-ray absorptiometry. Br J Nutr. 2013;109:1219–29.
González-Montalvo JI, Alarcón T, Gotor P, Queipo R, Velasco R, Hoyos R, et al. Prevalence of sarcopenia in acute hip fracture patients and its influence on short-term clinical outcome. Geriatr Gerontol Int. 2016;16:1021–7.
Steihaug OM, Bogen B, Kristoffersen MH, Ranhoff AH. Bones, blood and steel: how bioelectrical impedance analysis is affected by hip fracture and surgical implants. Electr Bioimpedance. 2019;8:54–9.
Faes TJ, van der Meij HA, de Munck JC, Heethaar RM. The electric resistivity of human tissues (100 Hz–10 MHz): a meta-analysis of review studies. Physiol Meas. 1999;20:R1–10.
Acknowledgements
Not applicable.
Funding
This research received no specific grant from any funding agency in the public, commercial, or not-for-profit sectors.
Author information
Authors and Affiliations
Contributions
TU conceptualized and designed this study, acquired and analyzed the data, and drafted the article; MW critically revised the important intellectual content of the manuscript; and all authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
All procedures were performed in accordance with the guidelines of the Declaration of Helsinki and approved by the Institutional Review Board of Tokai University Hospital (approval number: 22R177). The ethics committee of the University of Tokai waived the need for informed consent.
Consent for publication
Not applicable.
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
About this article
Cite this article
Ukai, T., Watanabe, M. Do metal implants for total hip arthroplasty affect bioelectrical impedance analysis? A retrospective study. BMC Musculoskelet Disord 24, 763 (2023). https://doi.org/10.1186/s12891-023-06893-x
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12891-023-06893-x