Intra-articular temperature monitoring during radiofrequency ablation in ex-vivo bovine hip joints via Fiber Bragg grating sensors

Abstract

Purpose

Radiofrequency ablation is an increasingly used surgical option for ablation, resection and coagulation of soft tissues in joint arthroscopy. One of the major issues of thermal ablation is the temperature monitoring across the target areas, as cellular mortality is a direct consequence of thermal dosimetry. Temperatures from 45 °C to 50 °C are at risk of damage to chondrocytes. One of the most reliable tools for temperature monitoring is represented by fiber optic sensors, as they allow accurate and real-time temperature measurement via a minimally invasive approach. The aim of this study was to determine, by fiber Bragg grating sensors (FBGs), the safety of radiofrequency ablation in tissue heating applied to ex-vivo bovine hip joints.

Methods

Ex vivo bovine hips were subjected to radiofrequency ablation, specifically in the acetabular labrum, for a total of two experiments. The WEREWOLF System (Smith + Nephew, Watford, UK) was employed in high operating mode and in a controlled ablation way. One optical fiber embedding seven FBGs was used to record multipoint temperature variations. Each sensor was 1 mm in length with a distance from edge to edge with each other of 2 mm.

Results

The maximum variation was recorded in both the tests by the FBG1 (i.e., the closest one to the electrode tip) and was lower than to 2.8 °C. The other sensors (from FBG2 to FBG7) did not record a significant temperature change throughout the duration of the experiment (maximum up to 0.7 °C for FBG7).

Conclusions

No significant increase in temperature was observed at any of the seven sites. The sensor nearest to the radiofrequency source exhibited the highest temperature rise, but the variation was only 3 °C. The minimal temperature increase registered at the measurement sites, according to existing literature, is not expected to be cytotoxic. FBGs demonstrate the potential to fulfil the strict requirements for temperature measurements during arthroscopic surgery.

Peer Review reports

Introduction

Radiofrequency ablation in arthroscopy is an increasingly used surgical option for ablation, resection, and coagulation of soft tissues [4, 24, 36]. The development of the radiofrequency technique was applied in orthopedics to decrease laxity of connective tissues around joints, at first in the context of shoulder instability. The rationale behind it was to obtain temperatures between 70 °C and 80 °C, which would cause collagen to shrink and trigger a healing response [2, 12, 15, 21, 33]. This method involves inserting electrodes into a target site to generate an electrical current, causing frictional agitation at the ionic level and a consequent heat generation [1, 26].

However, radiofrequency energy also presents some hazards, as well as time-dependent consequences that clinicians must consider [28, 36].

One of the main challenges of thermal ablation is the measurement of temperature across the target sites, because cellular mortality is a direct consequence of thermal dosimetry. This task is challenging as all thermal procedures induce heat patterns in the tissue with high spatial and temporal gradients, frequently exceeding 50 °C/mm and 5 °C/s [4, 32, 35].

Several publications reported the risk of complications following thermal ablation during arthroscopic procedures. These devices raise the temperature of joint fluid, potentially leading to skin burns, nerve lesions, capsular damage and extensive chondrolysis. These complications may be the result of an excessive transmission of heat in the nearby tissues [7, 13, 14, 24, 25, 27].

The risk for chondrocyte damage is particularly relevant when it comes to the hip joint. In the hip’s central compartment, factors predisposing to chondrocyte damage comprise insufficient saline volume, extensive cartilage coverage, risk for poor saline flow, and direct proximity of weight-bearing cartilage to the site where the radiofrequency is applied [24].

Chondrocytes are damaged at temperatures ranging from 45 °C to 50 °C and damage arises after an elevation of only 8 °C [5, 24, 34].

For these reasons, temperature monitoring during radiofrequency ablation treatments may become a key aspect of minimally invasive procedures in arthroscopic surgeries, since it allows the surgeon to adjust temperature distribution in the tissue proximal to the radiofrequency electrodes and it diminishes the risk of potentially adverse effects to the nearby healthy tissue [26].

The use of thermocouples is the standard procedure for in-situ temperature measurement. These sensors, however, present two significant weaknesses: thermocouples may interfere with heating propagation since they consist of two metallic wires, and they are only able to measure the temperature in a single point, while they are unable to define a spatial profile of temperature [22, 31].

An alternative to thermocouples is thermal imaging, which exploits magnetic resonance imaging (MRI) or computed tomography (CT). However, these techniques also come with major limitations in terms of costs, exposure to X-rays and presence of artefacts [4, 9, 30].

Fiber optic sensors constitute the primary alternative to these techniques and have been recently applied for real-time temperature detection during radiofrequency ablation. Most importantly, specific fiber optic sensors (i.e., fiber Bragg grating sensors, FBGs) can outperform thermocouples due to their unique property of detecting temperature patterns on a single fiber with resolutions ranging from 0.1 to 10 mm, and the potential to multiplex across multiple fibers to measure the temperature pattern in one- or two-dimensional geometries [17].

Given the challenges posed by obtaining instantaneous and precise temperature data via a minimally invasive approach, multi-point temperature measurements obtained via FBGs appear to be a promising answer to these challenges [9].

Currently, some studies have tried to assess the variations in joint temperature during arthroscopic procedures on the shoulder and knee, both in human and animal models [2, 7, 10, 23, 24]. However, there is still a lack of evidence determining variations in joint and tissue temperature during radiofrequency ablation treatments in the hip joint. Moreover, none of the available studies used FBG technology.

The aim of the current study was to evaluate, by FBGs, the safety of radiofrequency ablation in tissue heating applied to ex-vivo bovine hip joints.

Materials and methods

Ex vivo bovine hips were subjected to radiofrequency ablation during hip arthroscopy, specifically in the acetabular labrum. The WEREWOLF System (Smith + Nephew, Watford, UK) was employed in high operating mode and in a controlled ablation way (also referred as coablation). No limits to the temperature sensor were set, in order to assess the maximal excursion that is reached in the presented settings and decrease potential bias in the experiment. The joint perfusion pump used during hip arthroscopy was set at 50 mmHg. The water temperature used during hip arthroscopy typically ranges from 20 °C to 25 °C.

The co-ablation was activated with no arthroscopic manipulation. Aspiration via the probe was connected to the operating room aspiration system all the time and at maximum intensity. This modality allows the achievement of lower temperatures with respect to the traditional radiofrequency ablation method, thus reducing the risk of energy dissipation in the neighboring structures. This technology involves a generator through which the desired settings can be set and an electrode responsible for conveying the energy to the tissue. The tests were executed for around 20 s at room temperature (i.e., 25 °C), assessed through a thermocouple proximally to the joint surface before starting.

Standard hip arthroscopy was performed in a adult bovine hips.

The optical fiber was positioned in the acetabular labrum by means of an 18 gauge biopsy needle and in the proximity of the RF electrode. The optical fiber embedding seven FBGs (FBG1 closer to the electrode tip and the others accordingly to Fig. 1) from AtGrating Technologies was used to record temperature variations in the acetabular labrum. This solution based on multiple sensors allowed us to perform multi-point measurements (in 7 sites) within the acetabular labrum. Each FBG was 1 mm in length with a distance from edge to edge with each other of 2 mm. They were characterized by reflectivity values around 65% and λB ones ranging from 1512 to 1588 nm.

Fig. 1
figure 1

FBGs and electrode positioning in the acetabular labrum

Basically, an FBG consists of an optical fiber portion inscribed in its core. The so-called grating represents a periodic perturbation of the refractive index of the fiber core. So, when a broadband light spectrum is transmitted inside the fiber, a narrow spectrum is reflected backward in correspondence with the grating. Such retroreflected signal is centered around the Bragg wavelength (\({{\uplambda }}_{\text{B}}\)), satisfying the following Eq. [1] :

$${\lambda }_{B}= 2\cdot {n}_{eff}\cdot {\Lambda }$$
(1)

where \({n}_{eff}\) is the effective refractive index and \({\Lambda }\) is the grating period. When a fiber is exposed to strain (\(\epsilon\)) or temperature variation (\(\varDelta \text{T}\)), a Bragg wavelength shift (\({\varDelta {\uplambda }}_{\text{B}}\)) occurs. In this study, FBGs was used as temperature sensors only considering a strain-free configuration. Thus, by tracking \({\varDelta {\uplambda }}_{\text{B}}\) over time, \(\varDelta \text{T}\) can be derived using the following relationship:

$${\varDelta {\uplambda }}_{\text{B}}= \text{S}\text{T}\cdot \varDelta \text{T}$$
(2)

where ST is the thermal sensitivity (i.e., 10 pm\(\cdot\)C−1 in the case of the sensors used in this study). During the experiments, FBGs’ response was collected by means of an optical interrogator (si255 based on Hyperion Platform, Micron Optics, Atlanta, GA, USA) at a sampling frequency of 100 Hz.

Results

After experiments, FBGs data were examined in MATLAB (Mathworks, Natick, MA, USA) environment to retrieve the temperature trends recorded during the RF discharge. Figure 2a and b report the results obtained for the two experiments carried out with the WEREWOLF System. As shown, in this case, maximum \(\varDelta \text{T}\) was recorded in both cases by the FBG1 (i.e., the one close to the electrode tip) and was lower than to 2.8 °C. The other sensors (from FBG2 to FBG7) did not record a significant temperature change throughout the duration of the experiment (maximum up to 0.7 °C for FBG7).

Fig. 2
figure 2

Temperature variations collected during the first (a) and the second (b) RF experiments carried out with the WEREWOLF System and measured by the seven FBGs of the optical fiber employed

Discussion

Radiofrequency ablation is increasingly used for soft tissue ablation, resection, and coagulation in arthroscopy. However, it presents hazards and time-dependent consequences that clinicians must consider. Precise temperature measurement is crucial to avoid complications and tissue damage. FBGs offer a promising alternative to standard thermocouples and thermal imaging due to their ability to detect temperature patterns on a single fiber with high resolutions. While some studies have explored joint temperature variations during arthroscopic procedures, there is a lack of evidence for hip joint radiofrequency ablation, especially using FBGs. This study aimed to assess the safety of radiofrequency ablation in tissue heating applied to ex-vivo bovine hip joints using FBGs.

The main finding of this study is that the slight temperature increase detected at the measurement sites, based on the literature data, should not be cytotoxic [5, 24, 34].

The sensors covered a surface area ranging from 3 to 22 mm from the radiofrequency source. The sensor closest to the radiofrequency source resulted in measuring the highest temperature increase. Nonetheless, the measured increase was of only three degrees, and, therefore, not cytotoxic for the surrounding tissue.

The focus of the current study on the hip joint is of significant relevance, given hip-specific characteristics that warrant consideration. The significant reduction in joint space and the absence of fluid-heating capacity buffer in the hip could potentially amplify the increase in temperature caused by the application of radiofrequencies, making the hip joint more susceptible to thermal damage [24, 29, 37].

McKormick et al. had previously investigated the role of joint fluid lavage in preserving safe temperatures (< 50°) in the hip joint and showed that the constant application of radiofrequency for 90 s at the capsulolabral junction can considerably increase intra-articular temperatures near (1 and 2 mm) and far (5 and 1 mm) from the radiofrequency source to values linked with irreversible chondrocyte toxicity [24].

Recent interests have been poured in newer devices that exploit a plasma bubble that allows for more effective debridement and decreased heat transmission to the joint fluid and nearby tissue. However, their benefits over standard radiofrequency protocols are yet to be verified. Faruque et al. in their randomized control trial compared intra-articular temperature profile in standard ablation versus plasma ablation radiofrequency tools for rotator cuff repair. 17.5% of their patients showed temperatures above 45 °C but no significant variations where observed between standard and plasma radiofrequency devices [10].

Moreover, the safety of both bipolar and coablation systems in contexts of subacromial decompression, rotator cuff surgery and impingement syndrome has been demonstrated, showing similar results for both devices [7, 8, 11, 16].

For these reasons, it is important to consider all the possible variables that may influence intra-articular temperature during arthroscopic procedures. According to Zoric et al., the three primary factors that affect the quantity and effect of heat produced during arthroscopic surgery are: the flow rate of the irrigation fluid, the duration of application of radiofrequency ablation, and the distance between the probe tip and target tissue [23]. The most significant effect came from the flow, and it was crucial to understand that in situations when the flow was restricted or stopped entirely, temperatures might exceed 50 °C after only 5 s of ablation [38].

On the other hand, the temperature of the irrigation fluid has been proposed as the most influential variable in intra-articular temperature variation during arthroscopic procedures [3, 6, 7].

Regarding knee joint arthroscopy, no difference were found in temperature generation between standard radiofrequency ablation and plasma ablation [23].

The current study has several points of strength. To the author’s knowledge, this is the first study assessing intra-articular temperature variations upon radiofrequency treatments in the hip joint. Also, FBGs were exploited for temperature sensing and, given their unique property of detecting temperature patterns on a single fiber with resolutions ranging from 0.1 to 10 mm, and the potential to multiplex across several fibers to catch the temperature pattern in one- or two-dimensional geometries, they can be considered among the most reliable tools for temperature detection within arthroscopic procedures.

The present article also has some limitations. First, the findings of this study are limited to the hip joint and cannot be extrapolated to other joints due to the peculiar characteristics of the joint in question. The authors are fully aware of the fact that the volume of the cow’s hip joint is not directly comparable to that of a human. One of our future objectives will be to replicate similar experiments in different joints, both in ex vivo and in vivo models. The volume of the joint can indeed influence the temperature of the fluid and co-ablation.

Moreover, the radiofrequency devices were tested using ex-vivo bovine tissues and, despite the fact that these modalities represent reasonable surrogates, the characteristics of these environments differ from the environment of real, human intra-operative procedures [19, 20].

In this study, ex vivo bovine hips were utilized for radiofrequency ablation procedures targeted at the acetabular labrum during hip arthroscopy. It is important to note that a ex vivo bovine hip lacks both circulation and body temperature, which significantly differs from the conditions present within a living organism. Despite maintaining a constant room temperature, the dynamics of water temperature in an ex vivo setting are expected to diverge from those observed in a living body. Therefore, a future endeavor will be to conduct future experiments in an in vivo setting, where at least some circulatory dynamics are present to better simulate real-life conditions.

On the other hand, to our knowledge, this is the first radiofrequency ablation study using the fiber bragg grating sensors, providing preliminary results of feasibility in this rapidly evolving and highly relevant research field.

Additionally, the intrinsic anatomical differences in soft tissue thickness between human and bovine models represent a limitation of the current article. The aim of the current study was to determine the feasibility of the exploited system in a cadaveric model, not to determine a specific temperature threshold after which cytotoxic damage would occur.

Additionally, the reported findings are limited to coablation radiofrequency systems, and do not consider temperature variations that may be caused by standard radiofrequency devices, although similar outcomes have been reported among the different systems [18].

The technology used in this pilot study is promising as testified by the broad acceptance in other medical and bioengineering applications. It can represent the future gold standard to evaluate the temperature increase in nearby tissue during radiofrequency treatments.

Given the growing use of arthroscopy all over the world and consequently of radiofrequencies, it will be important to evaluate the cytotoxicity values ​​for each of the tissues around the joints (tendons, ligaments, muscles, cartilages, bones). FBGs may fulfil the strict requirements for temperature measurements during arthroscopic surgery.

Conclusions

No significant increase in temperature was observed at any of the seven sites. The sensor nearest to the radiofrequency source exhibited the highest temperature rise, but the variation was only 3 °C. The minimal temperature increase registered at the measurement sites, according to existing literature, is not expected to be cytotoxic. FBGs demonstrate the potential to fulfil the strict requirements for temperature measurements during arthroscopic surgery.

Availability of data and materials

The data presented in this study are available on request from the corresponding author.

References

  1. Ahmed M, Brace CL, Lee FT, Goldberg SN. Principles of and advances in percutaneous ablation. Radiology. 2011;258:351–69.

    Article  PubMed  Google Scholar 

  2. Anderson SR, Faucett SC, Flanigan DC, Gmabardella RA, Amin NH. The history of radiofrequency energy and coblation in arthroscopy: a current concepts review of its application in chondroplasty of the knee. J Exp Orthop. 2019;6:1.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  3. Barker SL, Johnstone AJ, Kumar K. In vivo temperature measurement in the subacromial bursa during arthroscopic subacromial decompression. J Shoulder Elbow Surg. 2012;21:804–7.

    Article  PubMed  Google Scholar 

  4. Beisenova A, Issatayeva A, Sovetov S, Korganbayev S, Jelbuldina M, Ashikbayeva Z, et al. Multi-fiber distributed thermal profiling of minimally invasive thermal ablation with scattering-level multiplexing in MgO-doped fibers. Biomed Opt Express. 2019;10:1282–96.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  5. Carnevale A, Longo UG, Schena E, Massaroni C, Lo Presti D, Berton A, et al. Wearable systems for shoulder kinematics assessment: a systematic review. BMC Musculoskelet Disord. 2019;20:546.

    Article  PubMed  PubMed Central  Google Scholar 

  6. Cheng SC, Jou IM, Chern TC, Wang PH, Chen WC. The effect of normal saline irrigation at different temperatures on the surface of articular cartilage: an experimental study in the rat. Arthroscopy. 2004;20:55–61.

    Article  PubMed  Google Scholar 

  7. Chivot M, Airaudi S, Galland A, Gravier R. Analysis of parameters influencing intraarticular temperature during radiofrequency use in shoulder arthroscopy. Eur J Orthop Surg Traumatol. 2019;29:1205–10.

    Article  PubMed  Google Scholar 

  8. Davies H, Wynn-Jones H, De Smet T, Johnson P, Sampath S, Sjølin S. Fluid temperatures during arthroscopic subacromial decompression using a radiofrequency probe. Acta Orthop Belg. 2009;75:153–7.

    PubMed  Google Scholar 

  9. Fani F, Schena E, Saccomandi P, Silvestri S. CT-based thermometry: an overview. Int J Hyperthermia. 2014;30:219–27.

    Article  CAS  PubMed  Google Scholar 

  10. Faruque R, Matthews B, Bahho Z, Doma K, Manoharan V, Wilkinson M, et al. Comparison between 2 types of radiofrequency ablation Systems in Arthroscopic Rotator Cuff Repair: a Randomized Controlled Trial. Orthop J Sports Med. 2019;7:2325967119835224.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Gereli A, Kocaoglu B, Guven O, Turkmen M. Warm irrigation fluid does not raise the subacromial temperature to harmful levels while using radiofrequency device. Int J Shoulder Surg. 2015;9:99–100.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Goldberg SN, Gazelle GS, Compton CC, Mueller PR, Tanabe KK. Treatment of intrahepatic malignancy with radiofrequency ablation: radiologic-pathologic correlation. Cancer. 2000;88:2452–63.

    Article  CAS  PubMed  Google Scholar 

  13. Good CR, Shindle MK, Kelly BT, Wanich T, Warren RF. Glenohumeral chondrolysis after shoulder arthroscopy with thermal capsulorrhaphy. Arthroscopy. 2007;23:797e791–795.

    Article  Google Scholar 

  14. Hanypsiak BT, Faulks C, Fine K, Malin E, Shaffer B, Connell M. Rupture of the biceps tendon after arthroscopic thermal capsulorrhaphy. Arthrosc 20 Suppl. 2004;2:77–9.

    Article  Google Scholar 

  15. Hayashi K, Markel MD. Thermal capsulorrhaphy treatment of shoulder instability: basic science. Clin Orthop Relat Res. 2001;390:59–72.

  16. Huynh V, Barbier O, Bajard X, Bouchard A, Ollat D, Versier G. Subacromial temperature profile during bipolar radiofrequency use in shoulder arthroscopy. Comparison of Coblation Orthop Traumatol Surg Res. 2017;103:489–91.

    Article  CAS  PubMed  Google Scholar 

  17. Jelbuldina M, Korganbayev S, Korobeinyk AV, Inglezakis VJ, Tosi D. Temperature profiling of ex-vivo Organs during Ferromagnetic Nanoparticles-Enhanced Radiofrequency ablation by Fiber Bragg grating arrays. Annu Int Conf IEEE Eng Med Biol Soc. 2018;2018:1–4.

    PubMed  Google Scholar 

  18. Longo UG, Franceschetti E, Maffulli N, Denaro V. Hip arthroscopy: state of the art. Br Med Bull. 2010;96:131–57.

    Article  PubMed  Google Scholar 

  19. Longo UG, Forriol F, Campi S, Maffulli N, Denaro V. Animal models for translational research on shoulder pathologies: from bench to bedside. Sports Med Arthrosc Rev. 2011;19:184–93.

    Article  PubMed  Google Scholar 

  20. Longo UG, Ciuffreda M, Candela V, Berton A, Maffulli N, Denaro V. Hip scores: a current concept review. Br Med Bull. 2019;131(1):81–96.

    Article  PubMed  Google Scholar 

  21. Lopez MJ, Hayashi K, Fanton GS, Thabit G, Markel MD. The effect of radiofrequency energy on the ultrastructure of joint capsular collagen. Arthroscopy. 1998;14:495–501.

    Article  CAS  PubMed  Google Scholar 

  22. Manns F, Milne PJ, Gonzalez-Cirre X, Denham DB, Parel JM, Robinson DS. In situ temperature measurements with thermocouple probes during laser interstitial thermotherapy (LITT): quantification and correction of a measurement artifact. Lasers Surg Med. 1998;23:94–103.

    Article  CAS  PubMed  Google Scholar 

  23. Matthews B, Wilkinson M, McEwen P, Hazratwala K, Doma K, Manoharan V, et al. In vivo arthroscopic temperatures: a comparison between 2 types of radiofrequency ablation Systems in Arthroscopic Anterior Cruciate Ligament Reconstruction-A Randomized Controlled Trial. Arthroscopy. 2017;33:165–72.

    Article  PubMed  Google Scholar 

  24. McCormick F, Alpaugh K, Nwachukwu BU, Xu S, Martin SD. Effect of radiofrequency use on hip arthroscopy irrigation fluid temperature. Arthroscopy. 2013;29:336–42.

    Article  PubMed  Google Scholar 

  25. Obrzut SL, Hecht P, Hayashi K, Fanton GS, Thabit G, Markel MD. The effect of radiofrequency energy on the length and temperature properties of the glenohumeral joint capsule. Arthroscopy. 1998;14:395–400.

    Article  CAS  PubMed  Google Scholar 

  26. Palumbo G, De Vita E, Schena E, Massaroni C, Verze P, Carlomagno N, et al. Multidimensional thermal mapping during radiofrequency ablation treatments with minimally invasive fiber optic sensors. Biomed Opt Express. 2018;9:5891–902.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  27. Pell RF, Uhl RL. Complications of thermal ablation in wrist arthroscopy. Arthrosc 20 Suppl. 2004;2:84–6.

    Article  Google Scholar 

  28. Peng L, Li Y, Zhang K, Chen Q, Xiao L, Geng Y, et al. The time-dependent effects of bipolar radiofrequency energy on bovine articular cartilage. J Orthop Surg Res. 2020;15:106.

    Article  PubMed  PubMed Central  Google Scholar 

  29. Ponce BA, Rosenzweig SD, Thompson KJ, Tokish J. Sequential volume reduction with capsular plications: relationship between cumulative size of plications and volumetric reduction for multidirectional instability of the shoulder. Am J Sports Med. 2011;39:526–31.

    Article  PubMed  Google Scholar 

  30. Rieke V, Butts Pauly K. MR thermometry. J Magn Reson Imaging. 2008;27:376–90.

    Article  PubMed  PubMed Central  Google Scholar 

  31. Saccomandi P, Schena E, Silvestri S. Techniques for temperature monitoring during laser-induced thermotherapy: an overview. Int J Hyperthermia. 2013;29:609–19.

    Article  PubMed  Google Scholar 

  32. Schena E, Majocchi L. Assessment of temperature measurement error and its correction during nd:YAG laser ablation in porcine pancreas. Int J Hyperthermia. 2014;30:328–34.

    Article  PubMed  Google Scholar 

  33. Schena E, Tosi D, Saccomandi P, Lewis E, Kim T. Fiber Optic Sensors for temperature monitoring during thermal treatments: an overview. Sens (Basel). 2016;16(7):1144.

    Article  Google Scholar 

  34. Taffoni F, Formica D, Saccomandi P, Di Pino G, Schena E. Optical fiber-based MR-compatible sensors for medical applications: an overview. Sens (Basel). 2013;13:14105–20.

    Article  Google Scholar 

  35. Tosi D, Macchi EG, Gallati M, Braschi G, Cigada A, Rossi S, et al. Fiber-optic chirped FBG for distributed thermal monitoring of ex-vivo radiofrequency ablation of liver. Biomed Opt Express. 2014;5:1799–811.

    Article  PubMed  PubMed Central  Google Scholar 

  36. Vij N, Liu JN, Amin N. Radiofrequency in arthroscopic shoulder surgery: a systematic review. Clin Shoulder Elb; 2022. Epub ahed of print.

  37. Yen CH, Leung HB, Tse PY. Effects of hip joint position and intra-capsular volume on hip joint intra-capsular pressure: a human cadaveric model. J Orthop Surg Res. 2009;4:8.

    Article  PubMed  PubMed Central  Google Scholar 

  38. Zoric BB, Horn N, Braun S, Millett PJ. Factors influencing intra-articular fluid temperature profiles with radiofrequency ablation. J Bone Joint Surg Am. 2009;91:2448–54.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

None.

Funding

The authors received no financial or material support for the research, authorship, and/or publication of this article.

Author information

Authors and Affiliations

Authors

Contributions

Conceptualization, U.G.L.; methodology, F.D.T., C.M.; validation, U.G.L., E.S.; formal analysis, F.D.T.; investigation, A.L., C.M.; data curation, A.L., D.L.P.; writing—original draft preparation, A.L, D.L.P.; writing—review and editing, E.S., U.G.L.; visualization, A.L.; supervision, U.G.L., E.S.; project administration, U.G.L. All authors have read and agreed to the published version of the manuscript.

Corresponding author

Correspondence to Umile Giuseppe Longo.

Ethics declarations

Ethics approval and consent to participate

The Institutional Animal Care and Use Committee at Campus Bio-Medico University approved the study (Permission number: PT-O 051.23). All animal procedures were performed under the approval and guidance of the Institutional Review Board of Campus Bio-Medico University of Rome. The experiments were conducted in accordance with the ARRIVE guidelines. All methods were carried out in accordance with relevant guidelines and regulations.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Longo, U.G., De Tommasi, F., Salvatore, G. et al. Intra-articular temperature monitoring during radiofrequency ablation in ex-vivo bovine hip joints via Fiber Bragg grating sensors. BMC Musculoskelet Disord 24, 766 (2023). https://doi.org/10.1186/s12891-023-06836-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-023-06836-6

Keywords