Accuracy and reproducibility analysis of different reference axes for femoral prosthesis rotation alignment in TKA based on 3D CT femoral model

Abstract

Background

There are many reference axes to determine the rotational positioning of the femoral prosthesis in total knee arthroplasty (TKA), mainly including the surgical transepicondylar axis (sTEA), anatomical transepicondylar axis (aTEA), Whiteside line, and the posterior condylar line (PCL), etc., but there is still no definite conclusion on which is the most accurate reference axis.

Objective

To explore the reproducibility of each reference axis of femoral external osteotomy based on the 3D CT femoral model, compare the deviation of the simulated femoral prosthesis rotation alignment, positioned based on each reference axis, with the gold standard sTEA, and analyze the accuracy of each reference axis.

Methods

The imaging data of 120 patients with knee osteoarthritis who underwent a 3D CT examination of the knee in our hospital from June 2018 to December 2021 were retrospectively collected. The 3D model of the femur was established by Mimics software. The line relative to PCL externally rotated 3° (PCL + 3°), aTEA, and the vertical line of the Whiteside line were constructed and compared with the gold standard sTEA. Intra-observer, as well as inter-observer reproducibility analysis, was performed by the intra-group correlation coefficient (ICC) and Bland-Altman method.

Results

The angle WS, between the vertical line of Whiteside and sTEA, was 2.54 ± 2.30°, with an outlier of 54.2%; the angle aTEA, between aTEA and sTEA, was 4.21 ± 1.01°, with an outlier of 99.1%; the angle PCL, between PCL + 3° external rotation and sTEA, was 0.50 ± 1.06°, with the highest accuracy and an outlier of 5.8%, and the differences among all three were statistically significant, P < 0.05. The intra-observer ICC values of WS, aTEA, and PCL were 0.975 (0.964–0.982), 0.926 (0.896–0.948), and 0.924(0.892,0.946), respectively, and the reproducibility levels were excellent; the inter-observer ICC values of WS, aTEA, and PCL were 0.968(0.955–0.978), 0.906 (0.868–0.934) and 0.970 (0.957,0.979), respectively, with excellent reproducibility levels; Bland-Altman plots suggested that the scatter points of intra-observer and inter-observer measurement differences more than 95% were within the limits of agreement.

Conclusion

The reference axis for locating the distal femoral external rotation osteotomy based on the 3D CT femoral model has good reproducibility. The PCL is easy to operate, has the highest precision, and the lowest outliers among the reference axes is therefore recommended.

Peer Review reports

Introduction

Osteoarthritis of the knee occurs in the elderly, and the advanced stage can lead to severe joint pain and deformity, which can seriously affect the life of patients. Total knee arthroplasty (TKA) is the most effective treatment for advanced knee osteoarthritis, as it can effectively relieve knee pain, improve knee function, and enhance patients’ quality of life [1]. Poor positioning of the rotational alignment of the femoral prosthesis in TKA can lead to a series of complications, including anterior knee pain, limited knee motion, poor patellar trajectory, and prosthesis loosening [2]. Therefore, how to accurately position the femoral prosthesis rotation alignment intraoperatively has been a long-standing concern for orthopedic surgeons.

The surgical transepicondylar axis (sTEA), first defined by Berger et al., is the line connecting the most prominent point of the lateral epicondyle of the femur, with the medial epicondylar sulcus [3]. In knee flexion and extension activities, the sTEA has been shown to be the axis with high overlap with the functional axis of knee flexion and extension, and it is most likely to create a balanced flexion gap when positioning the femoral prosthesis rotation with reference to the sTEA intraoperatively [4, 5]. In addition, the application of surgical computer navigation techniques has further confirmed the consistency of the sTEA with the functional axis of knee motion, and relevant anatomical and biomechanical analyses have revealed no significant differences between the two [6]. Therefore sTEA is considered the gold standard for determining the rotational alignment of the femoral prosthesis. However, intraoperative localization of the sTEA can be difficult because the sulcus of the medial epicondyle is often covered by dense soft tissue such as the deep fibers of the medial collateral ligament [7], which may lead to varying degrees of error based on the visual observation and touch localization [8].

Therefore, alternative reference axes such as the anatomical transepicondylar axis (aTEA), Whiteside line, and posterior condylar line (PCL) have been reported based on the measured resection technique for external rotation osteotomy of the femur. However, which reference axis deviates the least from the sTEA is still controversial [9, 10]. In addition, most of the previous studies were based on two-dimensional CT images, and the accuracy of the reference axes was evaluated at the most obvious CT plane of the studied anatomical axes [11]. In practice, the sTEA and the reference axes are most likely not in the same CT horizontal section, so there is a certain error in measuring in the same CT plane. However, there is a lack of studies based on three-dimensional measurements and large sample sizes to comprehensively compare the accuracy of intraoperative commonly used reference axes for positioning the femoral prosthesis rotation alignment. Therefore, the purpose of this study is to reconstruct a 3D model of the femur based on CT scan data of the knee using Mimics software, precisely locate each reference axis and project them onto the same plane, compare the deviation of the simulated rotational alignment of the femoral prosthesis positioned based on each reference axis to the gold standard sTEA, and analyze the accuracy as well as the reproducibility of each reference axis.

Materials and methods

Research materials

Three-dimensional CT of the knee and full-length anteroposterior X-ray images of both lower limbs were retrospectively analyzed of 186 patients from June 2018 to December 2021 at Fuyang People’s Hospital, affiliated with Anhui Medical University. Finally, 120 cases were enrolled according to the inclusion and exclusion criteria, including 45 males and 75 females; ages ranged from 26 to 77 years, with an average of 50.18 ± 10.92 years. The study was approved by the Ethics Committee of Fuyang People’s Hospital (IRB: [2022]79) and exempted from the requirement for informed consent. The study was conducted in accordance with the ethical norms and guidelines established by this committee.

Inclusion criteria: patients aged 18 years and older; simultaneous 3D CT of the knee and full-length anteroposterior X-ray images of both lower limbs; osteoarthritis of the knee KL (Kellgren-Lawrence, KL) grade II-IV. Exclusion criteria: measurement of the distal femur was affected by previous trauma or surgery; the presence of significant defects, deformities, and other lesions in the femur that affect the measurement; CT scan was not standardized, and image quality was poor; patients with post-traumatic arthritis or rheumatoid arthritis. The screening process is shown in Fig. 1.

Fig. 1
figure 1

The participant selection process

Imaging examination

3D CT examination of the knee

The patient was supine on the CT examination table with both lower limbs rotated to neutrality, and the patella and toes pointed upward. Scanning parameters: tube voltage 120 kV, tube current 150 mA, display field 27.3*27.3 cm, exposure time 1000 ms. The scanning layer is perpendicular to the mechanical axis of the femur, and the layer thickness is 1 mm. The CT image data were stored in DICOM format. Modeling was performed using Mimics Research 19.0 software (Materialize, Belgium), and parameters were measured using the software’s own measurement tool.

Weight-bearing and full-length anteroposterior X-ray examination of both lower limbs

The patient was examined using a digital X-ray imaging camera (GE Healthcare, US) with the patient standing upright on the camera frame with the back pressed against the frame and hands hanging naturally, knees as straight as possible, feet shoulder-width apart, internally rotated about 15°, distance from the X-ray machine flat detector 1.6-1.8 m, keeping the body still, showing a field of 37.0*105.0 cm, the head of the fibula on the X-ray film overlap with the tibia by about 1/3, and the patella points directly anterior.

Determination of femoral valgus angle

The patient’s full-length radiographs of both lower limbs were shown in the Pacs image viewing system (Infinitt Healthcare, Korea), and a circle was drawn to fit the edge of the femoral head, with the center of the circle being the center of the hip joint; the center of the most distal cortex of the intercondylar notch of the femur was selected as the center of the knee joint, and the line connecting the two centers was the mechanical axis (axis A) of the femur; the medullary canal of the femur was fitted by drawing a circle 20 cm above the knee joint line [12], and the line connecting the center of this circle to the center of the knee joint was the anatomic axis (axis B) of the femur [13]. The angle between the mechanical axis and the anatomical axis of the femur is known as the femoral valgus angle, denoted as α., noted as α. The valgus angle α in this case is 7.9°. Figure 2.

Fig. 2
figure 2

Measurement of femoral valgus angle in full-length radiographs of both lower limbs. Note: A: mechanical axis of the femur. B: anatomical axis of the femur. α: femoral valgus angle

Modeling of the femur and determination of the measurement plane

Construction of the femoral 3D model

The stored CT image data in DICOM format were imported into Mimics software, and the 3D model of the distal femur was reconstructed by marking, creating contour lines, and filling the cavity, as shown in Fig. 3.

Determination of the measurement plane for measuring the deviation of the rotational alignment of the femoral component

We draw a circle in the cross section of the femoral shaft to make it fit the medullary canal of the femur to the maximum, and the center of the circle is the center of the medullary canal. At the same time, find the last cross section of the distal femur containing the cortex of the vertex of the intercondylar notch of the femur, and draw a circle here to make the front and rear edges of the circle tangent to the anterior and posterior cortex of the vertex. The center of the circle is the apex of the intercondylar notch. At the same time, the line between the two centers of the two circles is the anatomical axis of the femur. Then the angle α(7.9°) between the anatomical axis and mechanical axis measured in the full-length radiographs of both lower limbs was used to determine the mechanical axis of the femur model, and finally, the vertical plane of the mechanical axis was determined, which was defined as the transverse section “o” [14]. Each reference axis was projected onto the transverse plane “o”, and then the deviation between the rotational alignment of the femoral component simulated based on each reference axis and the sTEA was measured. Figure 3.

Fig. 3
figure 3

Construction of the 3D model of the femur and determination of the transverse section of the mechanical axis of the femur. Note: Fig. 2A: 3D model of the reconstructed femur in Mimics software. Fig B ~ D: positioning of the anatomical axis of the femur a: most vertex of the intercondylar notch. b: center of the medullary canal of the femoral shaft. c: anatomical axis of the femur. Fig E: locating the femoral mechanical axis by 7.9° varus relative to the anatomical axis. c: the anatomical axis of the femur. d: the femoral mechanical axis. Fig F: Determination of the transverse section of the femoral mechanical axis. Gray plane o: transverse section of the femoral mechanical axis

Positioning of anatomical landmarks of distal femur and simulated femoral prosthesis rotation alignment

In the 3D model of the distal femur reconstructed by Mimics software, the bony landmark points were marked following previously published studies, and then each reference axis was determined. sTEA is the line between the most prominent point of the lateral epicondyle and the sulcus of the medial epicondyle [7]; aTEA is the line between the most prominent point of the medial and lateral epicondyles [7]; Whiteside line, also known as the anterior-posterior axis (AP axis) is the line connecting the deepest points of the anterior and posterior intercondylar notch of the femur when the femur is viewed in the axial position [15]; PCL is the line connecting the lowest points of the medial and lateral posterior condyles of the femur. Finally, all the reference axes are projected into the transverse section “o”. The rotational alignment of the femoral component simulated based on the Whiteside line is the perpendicular line to the Whiteside line in the transverse section “o”; The femoral component rotational alignment simulated based on the aTEA is the line that corresponds to the aTEA in the plane “o”; The femoral component rotational alignment simulated based on the PCL reference is the line that corresponds to the PCL + 3° external rotation in the plane “o”. Figure 4.

Fig. 4
figure 4

Positioning of anatomic landmarks of the distal femur and the simulated femoral component rotational alignment in transverse section. Note: Fig. A: The positioning of anatomical landmarks. a: the deepest point of femoral trochlear groove. b: the most prominent point of the medial epicondyle. c: the medial epicondylar sulcus. d: the lowest point of the medial posterior condyle. e: the deepest point of femoral posterior intercondylar notch. f: the lowest point of the lateral posterior condyle. g: the most prominent point of the lateral epicondyle. Fig. B: Positioning of the reference axis of the rotational osteotomy of the femur. L1: Whiteside line. L2:aTEA. L3:sTEA. L4:PCL Fig. C: Positioning of the transverse section of femoral mechanical axis. Gray plane "o": the transverse section plane. M: the anatomical axis of the femur. N: the mechanical axis of the femur. Fig. D: Positioning of the simulated rotation alignment of the femoral prosthesis in transverse section “o”. L’1: projection of L1 on the transverse section. L’2: projection of L2 on the transverse section. L’3: projection of L3 on the transverse section L’4: projection of L4 on the transverse section. L5: the external rotation of L’4 by 3°. L6: the vertical line of L’1. Gray plane “o”: the transverse section plane

Measurement indexes

All reference axes and sTEA were projected onto the transverse section “o”, and the measurement of the angle between the simulated femoral prosthesis rotationl alignment positioned by each reference axis and sTEA was performed on the transverse section “o”. The deviation of the simulated rotational alignment of the femoral component based on the Whiteside line reference is the angle between the perpendicular line to the Whiteside line and the sTEA, denoted as WS; The deviation of the simulated rotational alignment of the femoral component based on the aTEA reference is the angle between the aTEA and the sTEA, denoted as aTEA; The deviation of the simulated rotational alignment of the femoral component based on the PCL reference is the angle between the PCL + 3° external rotation and the sTEA, denoted as PCL. The angles of the deviation of the simulated femoral prosthesis rotation alignment were recorded as positive when it was externally rotated to the sTEA and negative when it was internally rotated. These angles were measured and compared, and deviations of >±2° [16] were categorized as outliers to evaluate the data of this study. Measurements were retaken two weeks later [17]by the same measurer and another measurer to assess intraobserver and interobserver reproducibility. All measurers who performed the measurements were well-trained for the relevant measures, Fig. 5.

Fig. 5
figure 5

Measurement of the angle between the simulated femoral prosthesis rotation alignment positioned by reference axis and the sTEA on the transverse section of the femoral mechanical axis. Note: L’1: projection of whiteside line on the transverse section L’2: projection of aTEA on the transverse section. L’3: projection of sTEA on the transverse section. L’4: projection of PCL on the transverse section. L5: the external rotation of L’4 by 3° on the transverse section. L6: the vertical line of L’1 on the transverse section. Gray plane o: the transverse section plane

Statistical analysis

SPSS 25.0 software was used for analysis. Quantitative data conforming to normal distribution or approximately normal distribution were described by \(\stackrel{-}{x}\)±\(S\). The overall comparison between groups of each reference axis was performed by ANOVA, Welch test was used when the variance was not equal, and the LSD method was used for pairwise comparison between groups. The intraclass correlation coefficient (ICC) and Bland-Altman method were used to assess intra-observer and inter-observer repeatability. ICC evaluation criteria ICC < 0.4 indicated poor consistency, 0.40 to 0.75 indicated good consistency, and > 0.75 showed high consistency [18]. Evaluation criteria for Bland Altman analysis: at least 95% of the differences between measurements lie within the 95% limits of consistency, i.e., the limits of consistency contain more than 95% scatter points [19]. P < 0.05 was considered statistically significant.

Results

Consistency test for deviations between femoral prosthesis rotational alignment positioned by three reference axes and sTEA

Intra-observer consistency

The ICC values of WS, aTEA, and PCL measured twice by the same measurer were 0.975 (0.964–0.982), 0.926 (0.896–0.948), and 0.924 (0.892,0.946), respectively, all with high levels of consistency, as shown in Table 1.

The scatter of the difference between the two measurements of WS, aTEA, and PCL within the observer all lie within the consistency limits by more than 95%, Fig. 6.

Table 1 ICC values of the intra-observer angle between the femoral component rotational alignment and the sTEA (\(\stackrel{-}{x}\)±\(S\))
Fig. 6
figure 6

Bland-Altman plot of the angle between the femoral prosthesis rotational alignment and sTEA measured twice by the same measurer. Note: WS: angle between the perpendicular line of Whiteside line and sTEA. aTEA: angle between aTEA and sTEA. PCL: angle between PCL + 3° external rotation and sTEA

Inter-observer consistency

The ICC values of WS, aTEA, and PCL measured by different observers were 0.968 (0.955–0.978), 0.906 (0.868–0.934), and 0.970 (0.957,0.979), respectively, all with high levels of agreement, Table 2.

The scatter of the difference between the two measurements of WS, aTEA, and PCL within the observer all lie within the consistency limits by more than 95%, Fig. 7.

Table 2 ICC values of the inter-observer angle between the femoral component rotational alignment and the sTEA (\(\stackrel{-}{x}\)±\(S\))
Fig. 7
figure 7

Bland-Altman plot of the angle between the femoral prosthesis rotational alignment and sTEA measured twice by the different measurers. Note: WS: angle between the perpendicular line of Whiteside line and sTEA. aTEA: angle between aTEA and sTEA. PCL: angle between PCL + 3° external rotation and sTEA

Comparison of the accuracy of femoral prosthesis rotational alignment positioned by the three reference axes

The angles of each femoral prosthesis rotational alignment with sTEA, i.e., WS, aTEA, and PCL were 2.54 ± 2.30°, 4.21 ± 1.01°, and 0.50 ± 1.06°, respectively. PCL + 3° external rotation with sTEA error was smaller relative to the Whiteside line as well as aTEA, and the difference was statistically significant, P < 0.05, Fig. 8.

Fig. 8
figure 8

Comparison of the deviations of the femoral component rotational alignment positioned by the three reference axes with the sTEA. Note: WS: angle between the perpendicular line of Whiteside line and sTEA. aTEA: angle between aTEA and sTEA PCL: angle between PCL + 3° external rotation and sTEA. *difference was statistically significant

Distribution of outliers of deviation from the gold standard sTEA for the femoral prosthesis rotational alignment positioned by the three reference axes

The outliers for the angle of rotation alignment of the femoral prosthesis positioned by each reference axis to the sTEA are 65 (54.2%) (WS), 119 (99.1%)(aTEA) and 7 (5.8%) (PCL), respectively, and the scatter distribution is shown in Fig. 9.

Fig. 9
figure 9

Distribution of outliers of the difference between the rotational the three femoral rotational prostheses and the gold standard sTEA. Note: ±2° is the error boundary WS: angle between the perpendicular line of Whiteside line and sTEA. aTEA: angle between aTEA and sTEA. PCL: angle between PCL + 3° external rotation and sTEA

Discussion

Poor rotation of the femoral prosthesis can lead to a series of complications [20]. The choice of the most optimal reference axis to position the femoral rotational alignment is still controversial [21]. Considering the limitations of localizing and comparing the precision of each reference axis in the same plane of 2D CT slice and the fact that no study to date has proposed an ideal method for localizing the reference axis of the femoral prosthesis rotational alignment using 2D CT. In this study, based on 3D CT data of the knee, a 3Dmodel of the femur was reconstructed using Mimics software to achieve accurate positioning of various reference axes, which were then projected onto the same transverse section, and the accuracy and reliability of the reference aTEA, whiteside line, and PCL in determining the femoral prosthesis rotational alignment were compared comprehensively based on a relatively large sample size.

In this study, each reference axis was located in the 3D model of the femur. Its deviation from sTEA was measured, excluding the influence of soft tissue coverage and intraoperative visual field limitation on positioning [7], allowing for relatively accurate positioning. The ICC values within and between observers are greater than 0.75, which can be considered good repeatability according to the evaluation criteria. At the same time, combined with the Bland-Altman plots, the scatter points of the difference of the angle between each reference axis and sTEA within and between observers are more than 95% within the consistency limit, which indicates that the positioning and measurement of the deviation between each reference axis and sTEA in the 3D model with robust repeatability. Our results are consistent with the results reported by Lei et al. [10], who positioned and measured the deviation of the Whiteside line and the PCL positioned femoral prosthesis rotational alignment from sTEA in a reconstructed 3D CT model of the femur, with intra- and inter-observer ICC > 0.9, with good reproducibility. The results of the study conducted by Robertson et al. [22] also indicated that compared to 2D CT, the accuracy and repeatability of locating reference axes were higher in 3D CT models. However, many studies [10, 23] have performed CT scans of the entire femur to locate the mechanical axis of the femur in the 3D model, which can increase the exposure time to radiation and additional financial burden. Compared to previous studies, the present study only utilized 3D CT data of the knee joint and combined it with the femoral valgus angle in full-length films to locate the transverse section of the femoral mechanical axis, ensuring accurate measurement of all indicators while also considering patient safety and economy.

In TKA, the measured resection technique mainly uses distal femoral anatomical landmarks to locate the distal femoral rotational osteotomy line. The primary reference axes commonly used during operation are aTEA, sTEA, PCL and the Whiteside line [24]. The sTEA has been widely accepted as the gold standard for rotational positioning of the femoral prosthesis [3, 7, 25].

Earlier studies have shown that the Whiteside line is stable to individual differences in femoral morphology and is not affected by gender or ethnicity [15]. Because the Whiteside line is based on the non-weight-bearing area of the distal femoral surface, it is relatively stable to changes in cartilage thickness due to knee degeneration. However, the deviation of the femoral prosthesis rotational alignment positioned by the Whiteside line from the sTEA in this study was 2.54 ± 2.30°, an outlier of 54.2%, indicating that the Whiteside line is not the ideal reference axis, which is in line with the findings of most previous studies [26, 27]. To analyze the reason, it may be due to more osteophyte hyperplasia in some patients, and the femoral intercondylar fossa is closed by osteophytes. Before positioning the Whiteside line, it is often necessary to remove the osteophytes in the intercondylar fossa, which may change the shape of the intercondylar fossa, resulting in inaccurate positioning of the Whiteside line [28, 29]. Nagamine et al. [27] found that the vertical line of the Whiteside line was 1.4 ± 3.3 ° internal rotation relative to sTEA through CT scanning of 84 knee joints and concluded that the reliability of determining the femoral prosthesis rotational alignment with reference to Whiteside line was lower than that of PCL when performing TKA for medial tibiofemoral arthritis.

The transepicondylar axis (TEA) is the line connecting the most prominent points of the medial and lateral epicondyles of the femur. It can be divided into the sTEA (the line connecting the most prominent points of the lateral epicondyle and medial epicondyle sulcus of the femur) and the aTEA (the line connecting the most prominent points of the lateral and medial epicondyle of the femur). Compared to the Whiteside line and sTEA, aTEA is relatively easy to position intraoperatively. The deviation of the femoral prosthesis rotational alignment with reference to aTEA from sTEA in this study was 4.21 ± 1.01°, an outlier of 99.1%, indicating that intraoperative parallel aTEA positioning of the external rotation osteotomy line of the femur is subject to significant error. This is similar to the findings of Jabalameli et al. [30], who performed CT scans of the knee in 108 patients with osteoarthritis of the knee and showed that aTEA showed 4.3° of internal rotation compared to sTEA. In addition, the results of this study found that the variability of aTEA was the smallest among the three (SD = 1.01°), which suggests that the deviation between aTEA and sTEA is more constant, which can be used to evaluate the femoral prosthesis rotation concerning aTEA when sTEA cannot be accurately identified.

PCL is easily identified during operation and is a relatively reliable reference axis. Many total knee replacement instruments currently use PCL + 3 ° external rotation to determine the femoral prosthesis rotational alignment. This value is based on the angle between PCL and sTEA measured by Berger et al. on the cadaver specimen; the posterior condylar angle (PCA) is 3.5 ± 1.2 °, which can be approximately parallel to sTEA with the conventional 3 ° external rotation osteotomy relative to PCL [7]. In addition, the proximal tibia physiologically varus articular surface was resected perpendicular to the tibial mechanical axis during osteotomy. An externally rotated 3 ° osteotomy would also be beneficial to create a stable flexion gap and maintain the consistency of tibiofemoral kinematics [31]. Our results indicate that PCL + 3 ° external rotation is closest to the physiological axis of knee flexion-extension movements, sTEA, with a mean difference of 0.5 ° and an outlier of 5.8%, which is significantly better than aTEA and Whiteside line. This is consistent with the findings of Jang et al. [9], who reconstructed 2128 femoral models and found a minor deviation from sTEA relative to the PCL + 3° external rotation osteotomy with an average of 0.60° and concluded that the PCL was the most accurate compared to other reference axes. But their included were all normal femora, and the included femora in this study were all KL grade II and above, so this study was closer to clinical reality, and the conclusions were more reliable. However, many studies that consider performing routine external rotation 3° osteotomy on patients do not take into account the anatomical variation of the distal femur [22, 32], the asymmetric wear, varus deformity, etc., which can make PCA individualized [33], and routine external rotation 3° osteotomy is not accurate. In a cadaveric study, Mantass et al. [34] reported that the sTEA relative to PCL external rotation angle ranged from − 1° to 7°, with a mean external rotation of 5°. Yip et al. [35] found a mean PCA of 5.8° in women and 5.1° in men by measuring 82 femoral specimens. Griffin et al. [36] used MRI to measure PCA in 104 knees. The results showed an increase in PCA with age. They considered that this was associated with wear of the posterior lateral femoral cartilage, leading to an increase in PCA at the time of measurement. The reasons for this may be related to differences in the anatomy of the distal femur in different ethnic groups and people with different characteristics [35, 37], in addition to the fact that different measurement methods may have an impact on the results [22]. In our opinion, although the deviation of PCL + 3° external rotation from sTEA has clear outliers, it is accurate in most cases. In addition, many studies with larger sample sizes confirm that the deviation of PCL + 3° external rotation from sTEA is more accurate relative to other reference axes [9, 27] and that the PCL is readily available intraoperatively and more reproducible, so we believe that reference to the PCL is still a reliable method for determining the femoral prosthesis rotational alignment when the medial and lateral posterior condyles are not severely disrupted.

The strength of this study was that the three-dimensional model of the distal femur was reconstructed using Mimics software, precisely positioned, and the deviation of each reference axis from the sTEA was measured and compared. In addition, some studies do not consider the effect of osteoarthritis on the anatomical structure of the knee [9]; all included in our study were knees with osteoarthritis KL grading, grade II or above, which is closer to clinical reality and more instructive for clinical practice.

However, this study also has shortcomings. Firstly, the reference axes analyzed in our study were only those commonly used intraoperatively; other reference axes were not included in the analysis; secondly, the reconstructed 3D model did not include cartilage structures and did not consider the influence of factors such as race, gender, and age on the anatomy of the distal femur; thirdly, the exclusion criteria may affect the generalizability of this study, but they also ensure the reliability and accuracy of the results; finally, our study did not compare the deviation of the rotational alignments of the femoral prosthesis positioned by each reference axis postoperatively, which could further justify our conclusions. The results of this study need to be further validated by multicenter clinical studies with large sample size. In the future, studies on the accuracy of the reference axis of the femoral prosthesis rotational alignment in TKA still need to use more precise methods and consider the influence of demographic characteristics and other factors on the anatomy of the femur.

Conclusions

Femoral component rotational alignment has an essential impact on the prognosis of TKA, but a consensus reference axis with high precision has not been reached. The deviations of the femoral component rotational alignment positioned by each reference axis from the sTEA could be accurately and reliably positioned on the reconstructed 3D model of the distal femur, and the PCL + 3 ° external rotation was still closest to the physiological axis of knee flexion and extension.

Data Availability

On reasonable request, the datasets generated and analyzed during the current study are available from the corresponding author.

References

  1. Young SW, Saffi M, Spangehl MJ, Clarke HD. Unexplained pain following total knee arthroplasty: is rotational malalignment the problem. Knee. 2018;25(2):329–34.

    Article  PubMed  Google Scholar 

  2. Huang T, Long Y, George D, Wang W. Meta-analysis of gap balancing versus measured resection techniques in total knee arthroplasty. Bone Joint J. 2017;99–B(2):151–8.

    Article  PubMed  Google Scholar 

  3. Asano T, Akagi M, Nakamura T. The functional flexion-extension axis of the knee corresponds to the surgical epicondylar axis: in vivo analysis using a biplanar image-matching technique. J Arthroplasty. 2005;20(8):1060–7.

    Article  PubMed  Google Scholar 

  4. Olcott CW, Scott RD. The Ranawat Award. Femoral component rotation during total knee arthroplasty. Clin Orthop Relat Res. 1999. (367): 39–42.

  5. Churchill DL, Incavo SJ, Johnson CC, Beynnon BD. The transepicondylar axis approximates the optimal flexion axis of the knee. Clin Orthop Relat Res. 1998. (356): 111–8.

  6. Oussedik S, Scholes C, Ferguson D, Roe J, Parker D. Is femoral component rotation in a TKA reliably guided by the functional flexion axis. Clin Orthop Relat Res. 2012;470(11):3227–32.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Berger RA, Rubash HE, Seel MJ, Thompson WH, Crossett LS. Determining the rotational alignment of the femoral component in total knee arthroplasty using the epicondylar axis. Clin Orthop Relat Res. 1993. (286): 40–7.

  8. Yoshino N, Takai S, Ohtsuki Y, Hirasawa Y. Computed tomography measurement of the surgical and clinical transepicondylar axis of the distal femur in osteoarthritic knees. J Arthroplasty. 2001;16(4):493–7.

    Article  CAS  PubMed  Google Scholar 

  9. Jang ES, Connors-Ehlert R, LiArno S, Geller JA, Cooper HJ, Shah RP. Accuracy of reference axes for femoral component rotation in total knee arthroplasty: computed tomography-based study of 2,128 femora. J Bone Joint Surg Am. 2019;101(23):e125.

    Article  PubMed  Google Scholar 

  10. Lei K, Liu LM, Fan HQ, Chen X, Yang L, Guo L. Comparison of the accuracy of three rotational femoral osteotomy reference axes in total knee arthroplasty. J Third Military Med Univ. 2021;43(03):261–6.

    Google Scholar 

  11. Konigsberg B, Hess R, Hartman C, Smith L, Garvin KL. Inter- and intraobserver reliability of two-dimensional CT scan for total knee arthroplasty component malrotation. Clin Orthop Relat Res. 2014;472(1):212–7.

    Article  PubMed  Google Scholar 

  12. Ou YL, Li PY, Xia H. Optimal sagittal insertion depth and direction of femoral intramedullary rod in total knee arthroplasty in chinese osteoarthritis patients. Orthop Surg. 2020;12(4):1238–44.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Lei K, Liu LM, Xiang Y, et al. Clinical value of CT-based patient-specific 3D preoperative design combined with conventional instruments in primary total knee arthroplasty: a propensity score-matched analysis. J Orthop Surg Res. 2020;15(1):591.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Victor J, Van Doninck D, Labey L, Van Glabbeek F, Parizel P, Bellemans J. A common reference frame for describing rotation of the distal femur: a ct-based kinematic study using cadavers. J Bone Joint Surg Br. 2009;91(5):683–90.

    Article  CAS  PubMed  Google Scholar 

  15. Arima J, Whiteside LA, McCarthy DS, White SE. Femoral rotational alignment, based on the anteroposterior axis, in total knee arthroplasty in a valgus knee. A technical note. J Bone Joint Surg Am. 1995;77(9):1331–4.

    Article  CAS  PubMed  Google Scholar 

  16. Michaut M, Beaufils P, Galaud B, Abadie P, Boisrenoult P, Fallet L. [Rotational alignment of femoral component with computed-assisted surgery (CAS) during total knee arthroplasty]. Rev Chir Orthop Reparatrice Appar Mot. 2008;94(6):580–4.

    Article  CAS  PubMed  Google Scholar 

  17. Lohman M, Tallroth K, Kettunen JA, Remes V. Changing from analog to digital images: does it affect the accuracy of alignment measurements of the lower extremity. Acta Orthop. 2011;82(3):351–5.

    Article  PubMed  PubMed Central  Google Scholar 

  18. Jiao H, He Q, Veldkamp BP, Editorial. Process data in Educational and Psychological Measurement. Front Psychol. 2021;12:793399.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Sajian GL. Consistency evaluation of quantitative measurement results and the application of the Bland-Altman method. Chin J Health Stat. 2011;28(04):409–11.

    Google Scholar 

  20. Pietsch M, Hofmann S. Early revision for isolated internal malrotation of the femoral component in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2012;20(6):1057–63.

    Article  PubMed  Google Scholar 

  21. Talbot S, Dimitriou P, Mullen M, Bartlett J. Referencing the sulcus line of the trochlear groove and removing intraoperative parallax errors improve femoral component rotation in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2017;25(9):2743–50.

    Article  PubMed  Google Scholar 

  22. Robertson EL, Hengherr M, Amsler F, Hirschmann MT, Mathis DT. A comparison of femoral component rotation after total knee arthroplasty in Kanekasu radiographs, axial CT slices and 3D reconstructed images. Skeletal Radiol. 2021;50(7):1389–97.

    Article  PubMed  PubMed Central  Google Scholar 

  23. Jiawei L, Canran L, Yujie1 H, et al. Digital measurement of femoral valgus angle and femoral posterior condyle angle in elderly people over 60 years old in northern China and comparison with young and middle-aged population. Chin J Tissue Eng Res. 2023;27(27):4291–6.

    Google Scholar 

  24. Patel AR, Talati RK, Yaffe MA, McCoy BW, Stulberg SD. Femoral component rotation in total knee arthroplasty: an MRI-based evaluation of our options. J Arthroplasty. 2014;29(8):1666–70.

    Article  PubMed  Google Scholar 

  25. Kobayashi H, Akamatsu Y, Kumagai K, Kusayama Y, Aratake M, Saito T. Is the surgical epicondylar axis the center of rotation in the osteoarthritic knee. J Arthroplasty. 2015;30(3):479–83.

    Article  PubMed  Google Scholar 

  26. Victor J, Van Doninck D, Labey L, Innocenti B, Parizel PM, Bellemans J. How precise can bony landmarks be determined on a CT scan of the knee. Knee. 2009;16(5):358–65.

    Article  CAS  PubMed  Google Scholar 

  27. Nagamine R, Miura H, Inoue Y, et al. Reliability of the anteroposterior axis and the posterior condylar axis for determining rotational alignment of the femoral component in total knee arthroplasty. J Orthop Sci. 1998;3(4):194–8.

    Article  CAS  PubMed  Google Scholar 

  28. Poilvache PL, Insall JN, Scuderi GR, Font-Rodriguez DE. Rotational landmarks and sizing of the distal femur in total knee arthroplasty. Clin Orthop Relat Res. 1996. (331): 35–46.

  29. Katz MA, Beck TD, Silber JS, Seldes RM, Lotke PA. Determining femoral rotational alignment in total knee arthroplasty: reliability of techniques. J Arthroplasty. 2001;16(3):301–5.

    Article  CAS  PubMed  Google Scholar 

  30. Jabalameli M, Moradi A, Bagherifard A, Radi M, Mokhtari T. Evaluation of distal femoral rotational alignment with spiral CT scan before total knee arthroplasty (a study in iranian population). Arch Bone Jt Surg. 2016;4(2):122–7.

    PubMed  PubMed Central  Google Scholar 

  31. Bin L, Wolong Y, Shengshan G, et al. Combination of measured resection and gap balancing technique used in femoral component rotation in total knee arthroplasty. Chin J Tissue Eng Res. 2020;24(15):2323–8.

    Google Scholar 

  32. Talbot S, Bartlett J. The anterior surface of the femur as a new landmark for femoral component rotation in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2008;16(3):258–62.

    Article  PubMed  Google Scholar 

  33. Fujii T, Kondo M, Tomari K, Kadoya Y, Tanaka Y. Posterior condylar cartilage may distort rotational alignment of the femoral component based on posterior condylar axis in total knee arthroplasty. Surg Radiol Anat. 2012;34(7):633–8.

    Article  PubMed  Google Scholar 

  34. Mantas JP, Bloebaum RD, Skedros JG, Hofmann AA. Implications of reference axes used for rotational alignment of the femoral component in primary and revision knee arthroplasty. J Arthroplasty. 1992;7(4):531–5.

    Article  CAS  PubMed  Google Scholar 

  35. Yip DK, Zhu YH, Chiu KY, Ng TP. Distal rotational alignment of the chinese femur and its relevance in total knee arthroplasty. J Arthroplasty. 2004;19(5):613–9.

    Article  PubMed  Google Scholar 

  36. Griffin FM, Math K, Scuderi GR, Insall JN, Poilvache PL. Anatomy of the epicondyles of the distal femur: MRI analysis of normal knees. J Arthroplasty. 2000;15(3):354–9.

    Article  CAS  PubMed  Google Scholar 

  37. Koh YG, Nam JH, Chung HS, Kim HJ, Lee HY, Kang KT. Gender differences exist in rotational anatomy of the distal femur in osteoarthritic knees using MRI. Knee Surg Sports Traumatol Arthrosc. 2020;28(9):2990–7.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

Project of the Orthopedic of Fuyang People’s Hospital, National Clinical Key Specialist.

Natural Science Project of Bengbu Medical College (2022byzd171).

Author information

Authors and Affiliations

Authors

Contributions

Kun Liu, Xuande Liu, and Donglin Fu: participated in the study design and implementation; Kun Liu: article writing; Yujun Guan, Haotong Ma: participated in data collection and statistical analysis; Donglin Fu, Zongqing Fan: responsible for the study design and implementation, and review of the content of the article.

Corresponding authors

Correspondence to Donglin Fu or Zongqing Fan.

Ethics declarations

Ethics approval and consent to participate

The study was approved by the Ethics Committee of Fuyang People’s Hospital (IRB: [2022]79) and exempted from the requirement for informed consent. The study was conducted in accordance with the ethical norms and guidelines established by this committee.

Consent for publication

Not applicable.

Competing interests

All authors declare that no conflict of interest in this study.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article