Effect of the core bone engaged length on the BASHTI fixation strength, an in-vitro study on bovine tendons using identical-density surrogate bones

Abstract

Background

BASHTI is an implant-less anterior cruciate ligament (ACL) reconstruction technique, which resolves the problems caused by implants such as interference screws. This study aims to investigate the effect of the drill bit and tendon’s diameter on the Core Bone Engaged Length (CBEL) and the fixation strength. CBEL is the length of core bone which has a full engagement with both tunnel and graft at the same time.

Methods

60 in-vitro tests were conducted for 6, 7, 8, and 9 mm tendon sizes with a 10 mm bone tunnel. In this study bovine tendons and dummy bone blocks were used to model the fixation. Drill bits were used to extract the core bone for securing the auto-graft. A three-stage tensile test including a force-controlled cyclical preloading of 10–50 N with a frequency of 0.1 Hz for 10 cycles, followed by the main force-controlled cyclical loading of 50–200 N with a frequency of 0.5 Hz for 150 cycles, and immediately a displacement-controlled single cycle pull-out load with a rate of 20 mm/min were carried out to discover the fixation strength of each sample.

Results

The 6 mm group had the greatest CBEL. However, all cases in this group failed in loadings below 200 N, which is the minimum required strength after ACL reconstruction. The fixation strength of cases with more than 200 N fixation strength for 7, 8, and 9 mm tendon diameters were 275 ± 42, 330 ± 110, and 348 ± 93 N, respectively, showing insignificant difference between groups (P-value = 0.45). Nevertheless, CBELs for these groups were 16.6 ± 3.4, 9.6 ± 2.4, and 11.7 ± 3.8 mm, respectively, implying a significant increase in CBEL in the 7 mm group than that for 8 and 9 mm groups (P-value = 0.002 and 0.049, respectively).

Conclusion

Results showed that CBEL could assess the quality of BASHTI technique. However, CBEL was an inverse function of tendon compression, so it was not an independent parameter to determine BASHTI strength. Also, the CBEL of 7 mm group which fulfilled the 200 N threshold was higher than that of 8 and 9 mm groups, so its healing process speed may be higher, which is recommended for a future study in this field.

Peer Review reports

Background

The anterior cruciate ligament (ACL) has an essential role in stabilizing the rotational movements of the knee [1]. ACL injuries, including partial or complete ligament tear, might occur under extreme sports activities, mostly as a result of a non-contact injury [2]. In a complete ACL tear, due to a lack of self-healing process, surgical reconstruction is required [3]. Using an interference screw is the most frequent fixation method in these surgeries [4, 5]. Nevertheless, this fixation method can be criticized for requiring expensive equipment and problems such as bone tunnel enlargement [6], inflammatory responses [7], and tendon rotation [8]. Consequently, implant-free techniques have been presented as a substitute for conventional techniques to reconstruct the ACL. The best-known implant-free technique is the press-fit method which uses bone plugs at the two ends of the patellar graft tissue for fixation [9]. This technique, however, is criticized due to problems such as the constrained length of the auto-graft [10].

BASHTI technique is an implant-free fixation method that uses a hamstring tendon and a core bone harvested during bone tunneling. The fixation strength of BASHTI and interference screw techniques is compared in both artificial bone and bovine models, indicating no significant differences between these two methods [11, 12]. Also, it is shown that increasing the geometric parameters (e.g., tendon and core bone diameter) to a critical value would increase BASHTI fixation strength [13,14,15]. In addition, the sheathed core bone can be inserted into the tunnel with less hammer impact force and reduce the risk of cracking on the core bone during the insertion process and increase the fixation strength [16, 17]. Moreover, it is shown that BASHTI fixation strength is significantly affected by the bone density and core bone insertion frequency (i.e., hammer strike rate during the insertion process) [18,19,20].

In the insertion process of a core bone into a bony tunnel using a hammer, only a portion of its length would be inserted into the tunnel. This is mainly because of the local fracture of the core bone due to the hammer impacts. The inserted length of the core bone is called the core bone engaged length (CBEL). It is believed that as the contact area between the core bone and the bone tunnel increases, the healing process speed, as well as friction and impact forces, would increase [21]. So, it is important to investigate the effect of involved parameters on the amount of CBEL. This study aims to examine the effect of CBEL on the BASHTI fixation strength and discusses its relationship with the cannulated drill bit and tendon diameter.

Methods

This experimental in-vitro study uses digital tendons harvested freshly from bovine hoofs. These hoofs were bought freshly from a licensed butchery. For consistency of the results, the hoofs were selected from the same breed and close age bovines. It has been already confirmed that the property of the grafts made from bovine digital extensor tendons is similar to human hamstring tendon [22]. These tendons were precisely trimmed to 6, 7, 8, and 9 mm diameters using laboratory sizing equipment (Fig. 1). Tendon trimming was done using a precision cutter by laboratory operator. The tendon was placed on a cutting board and then trimmed to the desired size. To verify the tendon size, the tendon is passed through the related gauge’s hole (Fig. 1). If the tendon is perfectly fitted through the hole, the corresponding bore size is considered as the size of the tendon; otherwise, the sample is failed and a new one must be prepared. The gauge had a range of 6–12 mm diameter bore sizes with a 0.5 mm interval. Accordingly, the measurement accuracy would be 0.5 mm. Tendons were stored at -20 °C for less than 48 h so their mechanical properties did not change [23]. Moreover, the Sawbones artificial bone blocks (Pacific Research Laboratories, Malmo, Sweden) with the same size and the same density of 320 kg/m3 – which is proofed to have similar mechanical properties with femoral/tibial cancellous bones - were used as an alternative for the cancellous bone of a young human [20, 24]. In addition, BASHTI’s cannulated drill bit was used to extract a ~ 30 mm core bone safely from the Sawbones (Fig. 2). These drill bits are used to extract the desired cylindrical core bones from the cancellous bone site while drilling the fixation bone tunnel. The outer diameter of the drill bits is set to 10 mm, which is appropriate for ACL reconstruction surgeries [25]. The inner diameter was made in different sizes to be suitable for the fixation of 6, 7, 8, and 9 mm tendon grafts [26]. Table 1 shows the drill bits and tendon diameters used in this study.

Fig. 1
figure 1

A typical double-strand tendon prepared for the test (A), tendon trimming equipment including surgical blade, gauge, and forceps (B)

Fig. 2
figure 2

A BASHTI’s cannulated drill bit (A) and its extracted core bone (B)

Table 1 Dimensions of the tendon, core bones, and inner size of drill bits

To secure the tendon graft into the tunnel, the tendon was doubled and passed through the tunnel, maintaining a gauge length of 30 mm tendon graft left free outside of the tunnel (Fig. 3a). This gauge length corresponded to the length of natural intact ACL [23]. Following the tendon insertion, the core bone was inserted into the tunnel. The insertion process was conducted using a hand-powered hammer by applying a frequency lower than 300 beats per minute on the top of the core bone in line with its central axis (Fig. 3b) [19]. The tendon was kept moist during the insertion process. The results of a power analysis (considering alpha = 0.05 and effect size = 0.5) indicated that the minimum number of repetitions required for this study was five (statistical power > 0.99). To check the repeatability of the results, therefore, five of each sample was built. Finaly, 60 different in-vitro BASHTI ACL fixation samples (4 different tendon diameters × 3 different dore bone diameters for each tendon diameter × 5 test repeats) were built for this study.

Fig. 3
figure 3

(a) The 30 mm gauge length of the tendon graft left free outside of the tunnel, (b) Core bone insertion process: securing the tendon (A) into the tunnel with a core bone (B) using a hand-powered (manual) hammer (C)

Shortly after the insertion process, the specimen was mounted into a servo-hydraulic machine (Amsler HCT 25–400; Zwick/Roell AG, Germany) to assess the mechanical properties of the fixation. The Sawbones block was mounted on the testing machine using a custom-made rig (Fig. 4a). Also, a custom-made hanger was used to hold the specimen (Fig. 4b). Initially, a cyclical preconditioning load of 10–50 N with a frequency of 0.1 Hz for 10 cycles was applied to the specimen. The preconditioning load was applied to eliminate the tendon graft’s loose length and prepare it for the next loading steps [27]. Immediately after that, the main force-controlled cyclical loading was applied to the specimen. This was set to 50–200 N with a frequency of 0.5 Hz for 150 cycles. This step was planned to simulate the ACL passive flexion-extension loading forces applied to the knee during the early rehabilitation process of a reconstructed graft [28, 29]. Following the main cyclical test, a displacement-controlled single cycle pull-out load with a rate of 20 mm/min was applied to the specimens to measure the failure strength of the fixation [30]. Moreover, after the failure of each sample, the CBEL was measured for each sample by measuring the size of the final pulled-out core bone (Fig. 5). The fixation failure was monitored, either considering an elongation of more than 10 mm or a visible tendon rupture [14]. To simplify the fixation geometrical parameters (i.e. tendon, core bone, and tunnel diameters), the tendon compression (TC) was defined using Eq. (1) [13].

$${\rm{TC}} = \frac{{{{\rm{S}}_{{\rm{tendon}}}} + {{\rm{S}}_{{\rm{core}}}} - {{\rm{S}}_{{\rm{tunnel}}}}}}{{{{\rm{S}}_{{\rm{tendon}}}}}}$$
(1)

Where TC is a dimensionless parameter representing the tendon compression. Stendon, Score, and Stunnel are the cross-section areas of the tendon, core bone, and tunnel, respectively (Fig. 5).

Fig. 4
figure 4

a. Testing setup: Sawbones block (A) fixed using a custom-made holder (B) on the servo-hydraulic machine platform (C), b. The double-strand portion of the tendon was hung on the gripper

Fig. 5
figure 5

Two parameters of tendon compression and CBEL

A Student’s t distribution was used to calculate the 95% Confidence Interval (CI) of the results. The ANOVA method was used to analyze if the difference between the results of different tendon groups were significant. A P-value less than 0.05 was considered to be statistically significant.

Results

According to the outcome obtained in current study, the tendon size had a significant effect on the CBEL, but it did not affect the fixation strength of the samples significantly. The average CBELs of the 6, 7, 8, and 9 mm tendon diameter groups were 31.7 ± 7.2, 21.1 ± 8.4, 14.7 ± 7.8, and 14.6 ± 5.9 mm (95% CI) respectively (Fig. 6). Previous studies on postoperative ACL reconstruction have shown that a load with a magnitude of at least 200 N would be applied to the fixation during a rehabilitation process [31]. Therefore, the value of 200 N was defined as the threshold to quantify CBEL. With the 6 mm tendon group, none of the samples could endure more than 200 N loadings. As a result, the 6 mm tendon was not considered for further development. The average fixation strength of samples enduring more than 200 N failure load with tendon diameters of 7, 8, and 9 mm were 275 ± 42 N, 330 ± 110 N, and 348 ± 93 N (95% CI), respectively (Fig. 7a). So, no significant difference was observed between fixation strengths (P-value = 0.45). On the other hand, the CBELs of these samples were 16.6 ± 3.4 mm, 9.6 ± 2.4 mm, and 11.7 ± 3.8 mm, respectively. Thus, the CBEL of the 7 mm tendon group was significantly more than that of 8 and 9 mm tendon groups (P-value = 0.002 and 0.049, respectively) (Fig. 7b).

Fig. 6
figure 6

Relation between CBEL and Tendon diameter. Error bars shown at 95% CI

Fig. 7
figure 7

(a) Fixations strength and (b) CBEF of samples with failure loads of more than 200 N as a function of core bone diameter for different tendon size groups. Error bars shown at ± 95% CI. Groups without any bars didn’t have any samples with more than 200 N failure loads

Also, it was observed that the CBEL does not affect the fixation strength independently. The measurement results of CBELs for each tendon diameter group demonstrated an inverse relation between TC and CBEL (Fig. 8). Previous studies introduced TC as an effective parameter for the fixation strength of ACL reconstruction [13]. Given this inverse relationship between TC and CBEL, the latter is not an independent parameter affecting the fixation strength.

Fig. 8
figure 8

An inverse relation between CBEL and TC for each tendon’s diameter group. Error bars shown at 95% CI

Discussion

Primary aim of this study was to investigate the effect of the CBEL on the BASHTI fixation strength and its relationship with the cannulated drill bit and tendon diameters. Current study testified that the geometrical parameters changed the amount of the CBEL significantly. Although, the CBEL had an inverse relation with the TC, but it was not an independent parameter to consider its effect on the fixation strength. Significant differences in fixation strength were observed between TC values in each tendon diameter group (P-value = 0.01, 0.00006, 0.0003, and 0.012 for 6, 7, 8, and 9 mm diameters, respectively). This was in agreement with the obtained results in previous studies [13]. In this study, the effect of TC on BASHTI fixation strength was not investigated. Therefore, in each diameter group, only cases with a strength of more than 200 N were considered. According to the obtained results, the 7 mm tendon group was suggested. The higher the CBEL, the more contact area between the core bone and the tunnel surface. This might improve the healing process. Almost all of the other reconstruction methods are non-organic and there is not any bone-to-bone contact in their fixations. Due to the existence of this contact in BASHTI fixation technique, investigating the effect of geometrical parameters was a great novelty in this study.

As it was observed, during a conventional BASHTI fixation process using a hand-powered hammer with a specific insertion frequency (less than 300 beats per minute [19]), CBEL was an inverse function of TC, so it was not an independent parameter to determine the BASHTI fixation strength. Previous studies reported that the insertion frequency affects the fixation strength of BASHTI fixation [19]. This insertion frequency may change this relationship between CBEL and TC and make the CBEL an effective parameter in fixation strength. An auto-hammer with an adjustable impact frequency might improve the CBEL and TC combination, resulting in a higher fixation strength and healing process speed.

Due to the limitation of providing human cadaver bone and tendon samples in a large number (i.e., more than 60 samples), digital bovine tendon samples and Sawbones blocks were used to mimic the human bone and tendon fixation. Although the tendons used in this study were stiffer than human models, the tendon sizes were according to the sizes used in actual reconstructive ACL surgery. However, the use of the human cadaveric specimens is suggested for future study. Finally, future works should investigate the effect of insertion frequency on the CBEL and its relationship with TC and fixation strength. The core bone density was kept constant, as it was assumed that the core bone will be extracted from the same recipient source during its surgical procedure. Obtaining the core bone from other sources e.g. femoral or tibial sites with various densities may requires a further investigation to ensure the effectiveness of the described technique.

Conclusion

The CBEL was proved to be a quality indicator influenced by the geometrical parameters, but not an independent factor on fixation strength of the BASHTI technique. The 6 mm tendons did not satisfy the 200 N fixation strength threshold. Besides, there was no significant difference in the strength of the cases that satisfied the threshold (7, 8, and 9 mm tendon groups). However, the CBEL of the 7 mm group was considerably higher than the other two groups, indicating a better healing speed for the reconstructed ACL. Consequently, the 7 mm tendon group is suggested for clinical purpose and future application of BASHTI technique.

Data availability

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

References

  1. Amin I, Moroz A. Anterior Cruciate Ligament and Posterior Cruciate Ligament Tears. Musculoskeletal Sports and Spine Disorders. Cham:Springer International Publishing; 2017.pp. 265–7.

  2. Fleming BC, Hulstyn MJ, Oksendahl HL, Fadale PD. Ligament injury, reconstruction and osteoarthritis. Curr Opin Orthop. 2005;16:354–62.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Kruse LM, Gray B, Wright RW. Rehabilitation after anterior cruciate ligament reconstruction: a systematic review. J Bone Joint Surg Am. 2012;94:1737–48.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  4. Daneshvarhashjin N, Chizari M, Mortazavi J, Rouhi G. Can the body slope of interference screw affect initial stability of reconstructed anterior cruciate ligament?: an in-vitro investigation. BMC Musculoskelet Disord. 2021;22:556.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Daneshvarhashjin N, Chizari M, Mortazavi SMJ, Rouhi G. Early post-operative performance of an anatomically designed hybrid thread interference screw for ACL reconstruction: A comparative study.J Biomech. 2022;:111031.

  6. Borjali A, Mohseni M, Chizari M. Biomechanical Modeling of a Bone Tunnel Enlargement Post ACL Reconstruction. bioRxiv. 2020;:2020.09.03.281915.

  7. Barber FA. Complications of biodegradable materials. Sports Med Arthrosc. 2015;23:149–55.

    Article  PubMed  Google Scholar 

  8. Saithna A, Chizari M, Morris G, Anley C, Wang B, Snow M. An analysis of the biomechanics of interference screw fixation and sheathed devices for biceps tenodesis. Clin Biomech. 2015;30:551–7.

    Article  Google Scholar 

  9. Malek MM, DeLuca JV, Verch DL, Kunkle KL. Arthroscopically assisted ACL reconstruction using central third patellar tendon autograft with press fit femoral fixation. Instr Course Lect. 1996;45:287–95.

    CAS  PubMed  Google Scholar 

  10. Boszotta H. Arthroscopic anterior cruciate ligament reconstruction using a patellar tendon graft in press-fit technique: Surgical technique and follow-up. Arthrosc J Arthrosc Relat Surg. 1997;13:332–9.

    Article  CAS  Google Scholar 

  11. Bashti K, Tahmasebi MN, Kaseb H, Farahmand F, Akbar M, Mobini A. Biomechanical Comparison between Bashti Bone Plug technique and biodegradable screw for fixation of grafts in ligament surgery. Arch Bone Jt Surg. 2015;3:29–34.

    PubMed  PubMed Central  Google Scholar 

  12. Borjali A, Nourani A, Moeinnia H, Mohseni M, Korani H, Ghias N, et al. Comparison of mechanical properties in interference screw fixation technique and organic anterior cruciate ligament reconstruction method: a biomechanical study. BMC Musculoskelet Disord. 2021;22:1047.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Moeinnia H, Nourani A, Borjali A, Mohseni M, Ghias N, Korani H, et al. Effect of geometry on the fixation strength of Anterior Cruciate Ligament Reconstruction using BASHTI technique. J Knee Surg. 2022;35:539–47.

    Article  PubMed  Google Scholar 

  14. Mohseni M, Nourani A, Korani H, Moeinnia H, Borjali A, Ghias N, et al. Core bone diameter in an organic implant-less technique affecting the biomechanical properties of the anterior cruciate ligament fixation; an in-vitro study. Arch Clin Exp Surg. 2021;10:1–7.

    Google Scholar 

  15. Ghias N, Nourani A, Mohseni M, Korani H, Moeinnia H, Borjali A et al. The use of BASHTI as an organic ACL fixation technique. In: Proceedings of Abstracts Physics, Engineering and Computer Science Research Conference. 2022. p. 84–6.

  16. Borjali A, Farrahi G, Jafarzade H, Chizari M. Experimental study of a sheathed core bone plug in Bashti ACL reconstructive method. In: The 27th Annual International Conference of Iranian Society of Mechanical Engineers-ISME2019. 2019.

  17. Borjali A, Farrahi G, Chizari M. Sheathed fixation improves BASHTI technique in an anterior cruciate ligament reconstruction.Proc Inst Mech Eng Part H J Eng Med. 2023.

  18. Nourani A, Mohseni M, Korani H, Ghias N, Chizari M. Reconstruction of a long head biceps using Bashti method; comparison between two different insertion techniques. In: The 27th Annual International Conference of Iranian Society of Mechanical Engineers-ISME2019. 2019.

  19. Mohseni M, Nourani A, Ghias N, Borjali A, Chizari M. Effect of insertion process on biceps tendon reconstruction in BASHTI technique; an in-vitro study.Sci Iran. 2022;29.

  20. Dehestani P, Farahmand F, Borjali A, Bashti K, Chizari M. Bone density may affect primary stability of anterior cruciate ligament reconstruction when organic core bone plug fixation technique used. J Exp Orthop. 2022;9:5.

    Article  PubMed  PubMed Central  Google Scholar 

  21. Biazzo A, Manzotti A, Motavalli K, Confalonieri N. Femoral press-fit fixation versus interference screw fixation in anterior cruciate ligament reconstruction with bone-patellar tendon-bone autograft: 20-year follow-up. J Clin Orthop Trauma. 2018;9:116–20.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  22. Donahue TLH, Gregersen C, Hull ML, Howell SM. Comparison of viscoelastic, structural, and Material Properties of double-looped anterior cruciate ligament grafts made from bovine Digital Extensor and Human Hamstring Tendons. J Biomech Eng. 2001;123:162–9.

    Article  CAS  PubMed  Google Scholar 

  23. Beynnon BD, Amis AA. Vitro testing protocols for the cruciate ligaments and ligament reconstructions. Knee surgery. Sport Traumatol Arthrosc. 1998;6:70–6.

    Article  Google Scholar 

  24. Snow M, Cheung W, Mahmud J, Evans S, Holt C, Wang B et al. Mechanical assessment of two different methods of tripling hamstring tendons when using suspensory fixation. Knee Surgery, Sport Traumatol Arthrosc. 2012;20:262–7.

  25. Forsythe B, Collins MJ, Arns TA, Zuke WA, Khair M, Verma NN, et al. Optimization of Anteromedial Portal femoral tunnel drilling with flexible and straight Reamers in Anterior Cruciate Ligament Reconstruction: a cadaveric 3-Dimensional Computed Tomography Analysis. Arthrosc J Arthrosc Relat Surg. 2017;33:1036–43.

    Article  Google Scholar 

  26. Kousa P, Järvinen TLN, Vihavainen M, Kannus P, Järvinen M. The fixation strength of six hamstring tendon graft fixation Devices in Anterior Cruciate Ligament Reconstruction: part I: femoral site. Am J Sports Med. 2003;31:174–81.

    Article  PubMed  Google Scholar 

  27. Aga C, Rasmussen MT, Smith SD, Jansson KS, LaPrade RF, Engebretsen L, et al. Biomechanical Comparison of Interference Screws and Combination Screw and Sheath Devices for Soft tissue Anterior Cruciate Ligament Reconstruction on the tibial side. Am J Sports Med. 2013;41:841–8.

    Article  PubMed  Google Scholar 

  28. Kousa P, Järvinen TL, Pohjonen T, Kannus P, Kotikoski M, Järvinen M. Fixation strength of a biodegradable screw in anterior cruciate ligament reconstruction. J Bone Joint Surg Br. 1995;77:901–5.

    Article  CAS  PubMed  Google Scholar 

  29. Markolf KL, Gorek JF, Kabo JM, Shapiro MS. Direct measurement of resultant forces in the anterior cruciate ligament. An in vitro study performed with a new experimental technique. J Bone Joint Surg Am. 1990;72:557–67.

    Article  CAS  PubMed  Google Scholar 

  30. Chizari M, Wang B, Barrett M, Snow M, Biomechanical, testing procedures in tendon graft reconstructive ACL surgery. Biomed Eng Appl Basis Commun. 2010;22:427–36.

    Article  Google Scholar 

  31. Belbasis A, Fuss FK, Sidhu J. Estimation of Cruciate Ligament Forces Via Smart Compression garments. Procedia Eng. 2015;112:169–74.

    Article  Google Scholar 

Download references

Acknowledgements

The authors would like to thank Professor Farzam Farahmand, for his helpful and valuable guidance during our research.

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

All authors contributed to the study conception and design. Material preparation, data collection, and analysis were performed by HM, MM, NG, HK and AB. The first draft of the manuscript was written by HM and MM and all authors commented on previous versions of the manuscript. All authors read and approved the final manuscript. Finally, the study supervised by MC and AN.

Corresponding author

Correspondence to Amir Nourani.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Moeinnia, H., Nourani, A., Mohseni, M. et al. Effect of the core bone engaged length on the BASHTI fixation strength, an in-vitro study on bovine tendons using identical-density surrogate bones. BMC Musculoskelet Disord 24, 226 (2023). https://doi.org/10.1186/s12891-023-06311-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-023-06311-2

Keywords