Elliptical and spherical heads show similar obligate glenohumeral translation during axial rotation in total shoulder arthroplasty

Abstract

Background

Elliptical shape humeral head prostheses have been recently proposed to reflect a more anatomic shoulder replacement. However, its effect on obligate glenohumeral translation during axial rotation compared to a standard spherical head is still not well understood. The purpose of the study was to compare obligate humeral translation during axial rotation using spherical and elliptical shaped humeral head prostheses. It was hypothesized that the spherical head design would show significantly more obligate translation when compared to the elliptical design.

Methods

Six fresh-frozen cadaveric shoulders were utilized for biomechanical testing of internal (IR) and external (ER) rotation at various levels of abduction (0°, 30°, 45°, 60°) with lines of pull along each of the rotator cuff muscles. Each specimen underwent the following three conditions: (1) native; total shoulder arthroplasty (TSA) using (2) an elliptical and (3) spherical humeral head implant. Obligate translation during IR and ER was quantified using a 3-dimensional digitizer. The radius of curvature of the superoinferior and anteroposterior dimensions of the implants was calculated across each condition.

Results

Posterior and inferior translation as well as compound motion of spherical and elliptical heads during ER was similar at all abduction angles (P > 0.05, respectively). Compared to the native humeral head, both implants demonstrated significantly decreased posterior translation at 45° (elliptical: P = 0.003; spherical: P = 0.004) and 60° of abduction (elliptical: P < 0.001; spherical: P < 0.001). During internal rotation at 0° abduction, the spherical head showed significantly more compound motion (P = 0.042) compared to the elliptical head. The spherical implant also demonstrated increased anterior translation and compound motion during internal rotation at 60° abduction (P < 0.001) compared to the resting state. This difference was not significant for the native or elliptical head design at this angle (P > 0.05).

Conclusion

In the setting of TSA, elliptical and spherical head implants showed similar obligate translation and overall compound motion during axial rotation. A gained understanding of the consequences of implant head shape in TSA may guide future surgical implant choice for better recreation of native shoulder kinematics and potentially improved patient outcomes.

Level of evidence

Controlled Laboratory Study.

Peer Review reports

Background

Unlike other ball and socket joints, the glenohumeral articulation differs in that the glenoid cavity is too shallow to capture the entire humerus during motion [1]. As such, the humeral head undergoes obligate translation at extremes of motions in order to remain centered within the glenoid cavity [1, 2]. According to Harryman et al., glenohumeral translation is considered obligate in that it cannot be prevented by application of an oppositely directed force of 30 to 40 N [1]. Further, increased obligate translation can be a result of an asymmetrically tight capsule [1]. While this stability is often provided by soft tissue restraints, [3] recreation of obligate translation during anatomic total shoulder arthroplasty (TSA) may rely greater on implant design in order to recreate the natural interaction of the glenohumeral joint [4].

Improper glenohumeral translation after TSA may have important implications with regard to implant longevity and patient outcomes [5, 6]. As the shoulder moves to near full extension and external rotation, the native humerus will translate posteriorly in order to remain centered in the glenoid, and similarly will translate anteriorly in cross-body position [1, 2]. Interestingly, Karduna et al. compared translation before and after anatomic arthroplasty, finding that active translation had greater dependence on the articular conformity rather than capsular ligaments [4]. This has led to a greater importance on recreating the anatomic relationship of the humeral joint through the implant interface.

While once thought to be perfectly spherical in shape, several recent anatomic studies of the shoulder have differed from the previous notion that the humeral head is a perfect sphere [7,8,9,10,11]. Previous biomechanical studies have shown that a non-spherical prosthetic head more accurately replicated native glenohumeral joint properties in terms of kinematics, translation, and rotational range of motion [12,13,14]. More importantly, the non-spherical humeral head design resulted in increased glenohumeral translation during humeral axial rotation along with increased micromotion of the glenoid component when compared to the spherical head design [13, 15]. However, these studies were limited to the prior resection of rotator cuff muscles and capsuloligamentous structures, leaving the effect of these soft tissue restraints on glenohumeral translation unknown. As the spherical design is characterized by a greater volume in the anteroposterior dimension compared to the elliptical head, tightening of the anterior part (during external rotation) or posterior part (during internal rotation) of the capsule may push the spherical head more posteriorly or anteriorly during axial rotation. This may especially be observed at higher abduction angles with the glenohumeral joint being more constraint.

Although long-term clinical implications have not been shown, understanding the actual translation that occurs could improve further implant design [16]. Thus, the purpose of the present study was to compare obligate humeral translation during axial rotation using spherical and elliptical shaped humeral head prostheses. The authors hypothesized that the spherical head design would show significantly more obligate translation when compared to the elliptical design.

Methods

Six fresh-frozen, cadaveric shoulders with a mean age of 62.7 ± 9.2 years (range 48–74 years) were used for the study (Science Care Inc., Phoenix, AZ, USA). All specimens underwent visual and radiographic inspection to exclude those with tears of the rotator cuff tendons and capsule, moderate to severe osteoarthritis, bony defects, or joint contractures. As de-identified specimens were not considered to constitute human subjects research, prior Institutional Review Board approval was not required.

Specimen preparation

After having been thawed overnight at room temperature, specimens were dissected free of skin, subcutaneous tissue, and muscles. Rotator cuff muscles, capsule, and the coracoacromial ligament were carefully preserved. Under fluoroscopy control (Mini C-Arm, GE Medical Systems Inc.) a 2.0 mm K-wire was drilled parallel to the glenoid surface from posterior to anterior at the middle of the superior–inferior diameter. A second 2.0 mm k-wire was drilled from inferior to superior parallel to the glenoid. The scapulae were trimmed using an oscillating saw and potted in a custom rectangular box with the glenoid surface being aligned parallel to the floor. After being shortened, the humerus was centered and potted in a poly-vinyl chloride pipe (PVC; diameter, 3.8 cm; length, 7 cm) using bone cement, leaving only 2 cm of the proximal humeral shaft exposed, in order to minimize diaphyseal bending moments [17, 18].

Testing setup

The specimens were mounted to a validated shoulder testing rig as previously described, which allowed for positioning of the glenohumeral joint in 6 degrees of freedom (Fig. 1A) [18,19,20,21,22]. With the glenoid surface being in a horizontal position parallel to the floor, the scapula was fixed to a vertical linear bearing translator and lever arm system on top of an X-Y table, allowing for glenohumeral translation in the anteroposterior and superoinferior direction. The rotation of the humerus was defined as neutral with the bicipital groove being aligned with the anterior margin of the acromion according to Selecky et al. [17, 23]. The rotator cuff muscles were loaded based on physiological cross-sectional area ratios with multiple lines of pull as previously described [24, 25]. Specifically, two lines of pull were used for the supraspinatus, three for the subscapularis, two for the infraspinatus, and one for the teres minor (Fig. 1B) [18, 25]. Each line of pull was loaded with 5 N, resulting in a total load of 40 N [18, 25].

Fig. 1
figure 1

(A) Displaying a right shoulder specimen mounted to the shoulder testing rig. The scapula is fixed to a vertical linear bearing translator and lever arm system on top of an X-Y table, allowing for glenohumeral translation in the anteroposterior and superoinferior direction. During testing, an axial compression load of 40 N is constantly applied via the lever arm of the X-Y table to center the joint. As the humerus is fixed in the testing rig, the oppositely directed force of 30 N is consequently applied to the X-Y table (glenoid) in the posterior direction during external rotation (green arrow) and in the anterior direction (red arrow) during internal rotation. The force is applied via a friction-less cable, which is attached to a servohydraulic testing system or 30 N hanging weight, depending on the direction of force. (B) The rotator cuff muscles are loaded based on physiological cross-sectional area ratios with multiple lines of pull. Specifically, two lines of pull are used for the supraspinatus (orange), three for the subscapularis (blue), two for the infraspinatus (red), and one for the teres minor (yellow). Each line of pull is loaded with 5 N, resulting in a total load of 40 N

Biomechanical Testing

During testing, an axial compression load of 40 N was constantly applied via the lever arm of the X-Y table to center the joint [13]. Each specimen underwent the three following conditions: (1) native, TSA with a (2) matched-fit elliptical head and (3) matched-fit spherical head. According to Jun et al., [13] 50° of internal and 50° of external axial rotation were alternatingly applied to the humerus at 0°, 30°, 45° and 60° of glenohumeral abduction in the scapular plane.

According to Harryman et al., [1] who stated that glenohumeral translation is considered obligate in that it cannot be prevented by application of an oppositely directed force of 30 N, an anterior directed force was applied to the humeral head during external rotation due to the posterior translation of the humeral head. Conversely, a posterior directed force was applied to the humeral head during internal rotation due to the anterior translation of the humeral head [1]. As the humerus was fixed in the testing rig, the force of 30 N was consequently had to be applied to the X-Y table (glenoid) in the posterior direction during external rotation and in the anterior direction during internal rotation. The force was applied via a friction-less cable, which was attached to a servohydraulic testing system (Mini Bionix 858; MTS) or 30 N hanging weight, depending on the direction of force (Fig. 1A).

By means of a 3-dimensional digitizer (MicroScribe G2; Immersion) with a position accuracy of 0.23 mm, the position of the X-Y table was measured by carefully digitizing the center of a defined groove on the X-Y table without relevant influence by touching off with the digitizer. The position of the groove was determined at the beginning (start position) and the end (end position) of each application of internal and external rotation. Changes in the position represented the glenohumeral translation and were given in anteroposterior (x-axis) and superoinferior (y-axis) directions. In addition, overall compound motion during internal and external rotation was calculated as the square root of the sum of the squared anteroposterior (x-axis) and squared superoinferior (y-axis) translation.

Internal and external rotation were alternatingly applied five times for every condition. Values of each specimen were then averaged and presented as the final values. Throughout entire testing, specimens were not removed from the testing rig, nor was the testing rig disassembled. The protruding spike of the trunnion allowed for easily switching between the elliptical and spherical head design during testing. To avoid selection bias, the order of glenohumeral abduction positions (0°, 30°, 45°, 60°) and head designs (elliptical or spherical) were randomly assigned.

Radius of curvature

Prior to testing, the superoinferior and anteroposterior dimension of the native humeral head was carefully digitized in 0.5 mm intervals by using the 3-dimensional digitizer (MicroScribe G2; Immersion) (Fig. 2A). Following completion of translational testing, each specimen was disarticulated, and all remaining soft tissue was resected. The matched-fit elliptical (Fig. 2B) and spherical head were each placed on the trunnion and then digitized as previously described. The data was implemented in a 3D computer graphics software (Rhinoceros 3D, McNeel, Boston, USA) and the radius of curvature of the superoinferior and anteroposterior dimensions were calculated for each condition.

Fig. 2
figure 2

(A) Prior to testing, the superoinferior (red) and anteroposterior (blue) dimension of the native humeral head is carefully digitized in 0.5 mm intervals by using the 3-dimensional digitizer. (B) Following completion of translational testing, each specimen is disarticulated, and all remaining soft tissue is resected. The matched-fit elliptical is placed on the trunnion and then digitized as previously described. The data is implemented in a 3D computer graphics software and the radius of curvature of the superoinferior and anteroposterior dimensions are calculated

Surgical technique

Total shoulder arthroplasty was performed using an anatomic stemless implant (Eclipse system, Arthrex Inc., Naples, FL, USA) according to a previously described technique [26, 27]. Each surgery was performed by the same surgeon (L.N.M.) in order to minimize performance bias. Oriented along the specimen’s anatomic retro-torsion, two 1.6 mm K-wires were pre-drilled in line with the desired resection plane, exiting the opposite cortex at the boundary of the articular cartilage. Guided by the two K-wires, an osteotomy was performed using an oscillating saw. After measuring the anterior-posterior dimension of the resected humeral head, the size of the baseplate (trunnion) was determined. The trunnion was then fixed to the resected humeral neck and a hollow screw was inserted. The custom-made trunnion used for this study was additionally secured with a small, protruding spike, to allow for easily switching the different prosthetic heads during testing.

Glenoid replacement was performed using a keeled glenoid system (Univers II, Arthrex Inc., Naples, FL, USA). A glenoid guide was placed on the central axis of the exposed articular surface of the glenoid, with the guide handle being oriented in line with the anatomic slope of the anterior neck. Following preparation, a keeled glenoid implant was inserted in the created slot and impacted.

Humeral head prosthetic design

Both elliptical and spherical prosthetic humeral heads were custom made (Arthrex Inc., Naples, FL, USA). The designs, including equations for dimension width, radius of curvature, and height, were chosen according to previously published studies [7, 8]. A small hole in the undersurface allowed for securely placing the humeral head prosthesis on the protruding spike of the trunnion, avoiding rotation of the head prosthesis during testing and allowing for easily switching heads between testing conditions.

Statistical analysis

A power analysis was carried out to determine detectable differences in translation, using standard deviations estimated from the literature as well as pilot data from our laboratory [12]. Assuming a common standard deviation of 1.5 mm, a sample size of 6 specimens would provide 80% power to detect a 2.5 mm difference in translation at an α level of 0.05. Descriptive statistics including mean and standard deviation (SD) as well as median and interquartile range (IQR) were calculated to characterize the groups. Differences in translation between implants were assessed using multilevel mixed effects generalized linear models. A random intercept was used to account for specimens in different conditions. For each analysis, the distribution of the residual was examined and found to conform to a normal distribution. Comparisons of marginal mean values were carried out with adjustment for multiple comparisons using the Holm-Bonferroni sequential correction method, in case of initial statistical significance. A p value of 0.05 was set to be statistically significant. All statistical analyses were conducted using Stata 15 software (StataCorp. 2017. Stata Statistical Software: Release 15. College Station, TX: StataCorp LLC).

Results

Obligate Glenohumeral translation

Elliptical and spherical heads showed similar obligate posterior translation (Table 1; Fig. 3), inferior translation (Table 2), and overall compound motion (Table 3; Fig. 4A) during external rotation at all tested abduction angles (P > 0.05, respectively). For both implant groups, obligate posterior translation during external rotation was decreased compared to the native state at all angles, however, this did only reach statistical significance at 45° (elliptical vs. native: P = 0.003; spherical vs. native: P = 0.004) and 60° (elliptical vs. native: P < 0.001; spherical vs. native: P < 0.001). In addition, the native humeral head showed significantly more obligate posterior translation (P = 0.007) and inferior translation (P < 0.001) at 60° of abduction during external rotation when compared to the resting position, whereas this was not observed for either the elliptical or spherical head design (P > 0.05, respectively).

Fig. 3
figure 3

Obligate posterior translation (mm) during external rotation

Fig. 4
figure 4

Overall compound motion (mm) during external (A) and internal (B) rotation

During internal rotation, spherical heads showed significantly more obligate anterior translation (P = 0.024; Fig. 5) and compound motion (P = 0.042; Fig. 4B) in the resting position when compared to elliptical heads. For both implant groups, obligate anterior translation during internal rotation was increased compared to the native state at higher abduction angles, however, this did not reach statistical significance (P > 0.05, respectively). In addition, the spherical design showed significantly more obligate anterior translation (P < 0.001) and compound motion (P < 0.001) at 60° of abduction during internal rotation when compared to the resting position, whereas this was not observed for either the native humeral head or the elliptical head design (P > 0.05, respectively).

Table 1 Obligate posterior glenohumeral translation (mm) during external rotation (ER) and obligate anterior glenohumeral translation (mm) during internal rotation (IR). Abbreviations: IQR = interquartile range; SD = standard deviation
Table 2 Inferior glenohumeral translation during external (ER) and internal (IR) rotation. Abbreviations: IQR = interquartile range; SD = standard deviation
Table 3 Overall compound motion (mm) during external (ER) and internal (IR) rotation. Abbreviations: IQR = interquartile range; SD = standard deviation
Fig. 5
figure 5

Obligate anterior translation (mm) during internal rotation

Radius of curvature

When compared to the native humeral head, only the spherical design allowed for restoration of the radius of curvature in the anteroposterior dimension (sphere: P = 0.05; elliptical: P = 0.01). However, none of the two implant designs achieved a similar radius of curvature in the superoinferior dimension when compared to the native humeral head (sphere: P = 0.03, elliptical: P = 0.02; Tables 4 and 5).

Table 4 Radius of curvature (mm) of the anteroposterior and superoinferior dimension. Abbreviations: IQR = interquartile range, SD = standard deviation
Table 5 Comparison of radius of curvature between the native humeral head and the two implant designs. * Indicates statistical significance. Abbreviations: CI = confidence interval

Discussion

The most important finding of the present study was that spherical and elliptical heads demonstrated similar obligate glenohumeral translation during axial rotation at all abduction angles following TSA. Although similar measurements of posterior, inferior, and compound translational motion for elliptical and spherical heads were generated during external rotation at each tested abduction angle, these implants yielded less posterior translation when compared to the native humeral head. These findings indicate that surgical implants may alter the mechanical environment of the shoulder joint, allowing for less translational movement of the humerus after TSA.

In the context of internal rotation at resting position, the spherical head design demonstrated more anterior translation than the elliptical head. Moreover, the spherical head yielded greater anterior translation and compound motion than both the elliptical and native humeral heads at 60° of abduction with internal rotation. These observations may be explained by the influence of the spherical implant shape, which has more material on the posterior side that may be involved in pushing the humeral head anteriorly during internal rotation. Further, the anterior capsule was surgically opened to perform the arthroplasty, while the posterior capsule was left untouched. This may have inadvertently altered the structure of the joint, allowing the humeral head to translate anteriorly during internal rotation to a greater degree after arthroplasty, independent of the implant shape. Consistently, Massimini et al. demonstrated that elongation of the SGHL, MGHL, and IGHL during shoulder rotational movement influenced the degree of glenohumeral translation [28]. Thus, changes in implant obligate translation may be a result of varying tightness of soft tissue restraints after arthroplasty, reflecting true surgical consequences of capsule and tendon reconstructions during TSA.

Previous biomechanical studies have shown that a non-spherical prosthetic head more accurately replicated native glenohumeral joint properties in terms of kinematics, translation, and rotational range of motion [12,13,14]. More importantly, the non-spherical humeral head design resulted in increased glenohumeral translation during humeral axial rotation when compared to the spherical head design [13, 15]. However, these studies were limited to the prior resection of rotator cuff muscles and capsuloligamentous structures, leaving the effect of these soft tissue restraints on glenohumeral translation unknown. Further, prior work did not prove that the translations occurring at the glenohumeral joint were obligate, as no oppositely directed force was applied [1, 13].

In contrast, the present study showed no statistically significant difference in obligate translation between elliptical and spherical heads during both external and internal rotation. This observation may be explained due to the spherical design containing more physical material in the anteroposterior dimension when compared to the elliptical head [7, 8]. As tightening of the anterior part (during external rotation) or posterior part (during internal rotation) of the capsule pushes the spherical head more posteriorly or anteriorly during axial rotation, this may counterbalance the generally greater translational motion of elliptical heads having been reported in models not considering the effect of soft tissue restraints [13]. Due to these characteristic features of its shape, the spherical design generally has greater magnitude of translation at higher abduction angles, with the glenohumeral joint being more constraint along with increased tightness of the capsule. Thus, forces from the rotator cuff muscles and the capsule pressures itself may induce greater amounts of translation motion either anteriorly or posteriorly. Consequently, this trend was especially noticed during internal rotation, as the posterior part of the capsule was left untouched while performing the TSA. Accordingly, the spherical design showed significantly more obligate anterior translation and compound motion at 60° of abduction during internal rotation when compared to the resting position. At lower abduction angles, however, the capsule is looser with less engagement with the rotator cuff muscles.

Anatomic studies have reported that the humeral head diverges to elliptical in the anterior–posterior dimension at the periphery of the articular margin, having roughly an 8–12% difference in head radius when comparing frontal and sagittal planes [10, 29]. With the development of glenohumeral osteoarthritis, initial pathologic deformations are commonly observed at the humeral side, which can cause further progression of this elliptical shape, especially in cases of severe disease [9, 30, 31]. However, neither the elliptical nor the spherical design allowed for replicating the native radius of curvature. Although Habermeyer et al. found that 82.2% of patients who underwent anatomic shoulder arthroplasty for treatment of primary osteoarthritis had an aspherical humeral head shape, this observation also implied that 17.8% of the evaluated humeral heads were still spherical in shape, bringing greater confusion as to which patient would benefit more from a spherical head design rather than from an elliptical one [9]. Thus, these inconsistencies in the anatomy of each individual specimen, with the humeral head being either more spherical or elliptical in shape and differing tightness of the glenohumeral joint capsule, may further contribute to the relatively large variations of differences between spherical and elliptical heads, without reaching significance in favor of one of the two head designs.

While increased translation may resemble native glenohumeral kinematics more accurately, [12, 13] this may also lead to more eccentric loading on and greater micromotion of the glenoid component in the setting of TSA. As excessive micromotion of the glenoid component has been suggested as a significant risk factor for glenoid loosening in the long-term, [5, 32,33,34] these biomechanical time-zero findings may be of clinical importance, alleviating concerns of implant loosening due to increased rocking motion.

There were several limitations to the study. Humeral head prosthetic design may show a different effect in vivo when compared to observations during laboratory cadaveric testing. Further, the inconsistencies in the anatomy of each individual specimen, with the humeral head either being more elliptical or spherical in shape, may have further influenced the results. In addition, specimens with moderate to severe osteoarthritis were excluded from the study, as varying degrees of osteoarthritic changes may have influenced translational measurements. Moreover, the shoulder model of the present study was not able to account for differing tightness of the glenohumeral joint capsule, which may have caused alterations in translational motion of the humeral head. Lastly, the anteroinferior capsule was opened during surgical replacement while preserving the anterior IGHL, to to accurately visualize the resection plane of the humeral head. As switching of prosthetic head shapes and sizes had to be frequently performed during testing, a capsular repair was infeasible, potentially increasing anterior and decreasing posterior translation during rotational testing.

Conclusion

In the setting of TSA, elliptical and spherical head implants showed similar obligate translation and overall compound motion during axial rotation. A gained understanding of the consequences of implant head shape in TSA may guide future surgical implant choice for better recreation of native shoulder mechanics and potentially improved patient outcomes.

Data Availability

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

IGH:

Inferior glenohumeral ligament

IQR:

Interquartile range

MGHL:

Middle glenohumeral ligament

PVC:

Poly-vinyl chloride

SD:

Standard deviation

SGHL:

Superior glenohumeral ligament

TSA:

Total shoulder arthroplasty

USA:

United States of America

References

  1. Harryman DT, Sidles JA, Clark JM, McQuade KJ, Gibb TD, Matsen FA. Translation of the humeral head on the glenoid with passive glenohumeral motion. J Bone Joint Surg Am. 1990;72(9):1334–443.

    Article  PubMed  Google Scholar 

  2. Howell SM, Galinat BJ, Renzi AJ, Marone PJ. Normal and abnormal mechanics of the glenohumeral joint in the horizontal plane. J Bone Joint Surg Am. 1988;70(2):227–32.

    Article  CAS  PubMed  Google Scholar 

  3. Werner CM, Nyffeler RW, Jacob HA, Gerber C. The effect of capsular tightening on humeral head translations. J Orthop Res. 2004;22(1):194–201.

    Article  PubMed  Google Scholar 

  4. Karduna A, Williams GR, Williams JL, Iannotti JP. Glenohumeral joint translations before and after total shoulder arthroplasty. A study in cadavera. J Bone Joint Surg Am. 1997;79(8):1166–74.

    Article  CAS  PubMed  Google Scholar 

  5. Collins D, Tencer A, Sidles JA, Matsen FA. Edge displacement and deformation of glenoid components in response to eccentric loading. The effect of preparation of the glenoid bone. J Bone Joint Surg Am. 1992;74(4):501–7.

    Article  CAS  PubMed  Google Scholar 

  6. Severt R, Thomas BJ, Tsenter MJ, Amstutz HC, Kabo JM. The influence of conformity and constraint on translational forces and frictional torque in total shoulder arthroplasty. Clin Orthop Relat Res 1993(292):151–158.

  7. Humphrey CS, Sears BW, Curtin MJ. An anthropometric analysis to derive formulae for calculating the dimensions of anatomically shaped humeral heads. J Shoulder Elbow Surg. 2016;25(9):1532–41.

    Article  PubMed  Google Scholar 

  8. Humphrey CS, Gale AL. Spherical versus elliptical prosthetic humeral heads: a comparison of anatomic fit. J Shoulder Elbow Surg. 2018;27(6S):50–S57.

    Article  Google Scholar 

  9. Habermeyer P, Magosch P, Weiss C, Hawi N, Lichtenberg S, Tauber M, Ipach B. Classification of humeral head pathomorphology in primary osteoarthritis: a radiographic and in vivo photographic analysis. J Shoulder Elbow Surg. 2017;26(12):2193–9.

    Article  PubMed  Google Scholar 

  10. Iannotti JP, Gabriel JP, Schneck SL, Evans BG, Misra S. The normal glenohumeral relationships. An anatomical study of one hundred and forty shoulders. J Bone Joint Surg Am. 1992;74(4):491–500.

    Article  CAS  PubMed  Google Scholar 

  11. Harrold F, Wigderowitz C. Humeral head arthroplasty and its ability to restore original humeral head geometry. J Shoulder Elbow Surg. 2013;22(1):115–21.

    Article  PubMed  Google Scholar 

  12. Jun BJ, Iannotti JP, McGarry MH, Yoo JC, Quigley RJ, Lee TQ. The effects of prosthetic humeral head shape on glenohumeral joint kinematics: a comparison of non-spherical and spherical prosthetic heads to the native humeral head. J Shoulder Elbow Surg. 2013;22(10):1423–32.

    Article  PubMed  Google Scholar 

  13. Jun BJ, Lee TQ, McGarry MH, Quigley RJ, Shin SJ, Iannotti JP. The effects of prosthetic humeral head shape on glenohumeral joint kinematics during humeral axial rotation in total shoulder arthroplasty. J Shoulder Elbow Surg. 2016;25(7):1084–93.

    Article  PubMed  Google Scholar 

  14. Muench LN, Otto A, Kia C, Obopilwe E, Cote MP, Imhoff AB, Beitzel K, Mazzocca AD, Mehl J. Rotational range of motion of elliptical and spherical heads in shoulder arthroplasty: a dynamic biomechanical evaluation. Arch Orthop Trauma Surg. 2022 Jan;142(1):67–76.

  15. Muench LN, Kia C, Murphey M, Obopilwe E, Cote MP, Imhoff AB, Mazzocca AD, Berthold DP. Elliptical heads result in increased glenohumeral translation along with micro-motion of the glenoid component during axial rotation in total shoulder arthroplasty. Arch Orthop Trauma Surg. 2023 Jan;143(1):177–187

  16. Egger AC, Peterson J, Jones MH, Miniaci A. Total shoulder arthroplasty with nonspherical humeral head and inlay glenoid replacement: clinical results comparing concentric and nonconcentric glenoid stages in primary shoulder arthritis. JSES Open Access. 2019;3(3):145–53.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Imhoff FB, Camenzind RS, Obopilwe E, Cote MP, Mehl J, Beitzel K, Imhoff AB, Mazzocca AD, Arciero RA, Dyrna FGE. Glenoid retroversion is an important factor for humeral head centration and the biomechanics of posterior shoulder stability. Knee Surg Sports Traumatol Arthrosc. 2019 Dec;27(12):3952–3961.

  18. Mehl J, Otto A, Imhoff FB, Murphy M, Dyrna F, Obopilwe E, Cote M, Ladermann A, Collin P, Beitzel K, et al. Dynamic anterior shoulder stabilization with the long head of the Biceps Tendon: a Biomechanical Study. Am J Sports Med. 2019;47(6):1441–50.

    Article  PubMed  Google Scholar 

  19. Arciero RA, Parrino A, Bernhardson AS, Diaz-Doran V, Obopilwe E, Cote MP, Golijanin P, Mazzocca AD, Provencher MT. The effect of a combined glenoid and Hill-Sachs defect on glenohumeral stability: a biomechanical cadaveric study using 3-dimensional modeling of 142 patients. Am J Sports Med. 2015;43(6):1422–9.

    Article  PubMed  Google Scholar 

  20. Mihata T, Lee Y, McGarry MH, Abe M, Lee TQ. Excessive humeral external rotation results in increased shoulder laxity. Am J Sports Med. 2004;32(5):1278–85.

    Article  PubMed  Google Scholar 

  21. Pauzenberger L, Dyrna F, Obopilwe E, Heuberer PR, Arciero RA, Anderl W, Mazzocca AD. Biomechanical evaluation of Glenoid Reconstruction with an Implant-Free J-Bone Graft for Anterior Glenoid Bone loss. Am J Sports Med. 2017;45(12):2849–57.

    Article  PubMed  Google Scholar 

  22. Schneider DJ, Tibone JE, McGarry MH, Grossman MG, Veneziani S, Lee TQ. Biomechanical evaluation after five and ten millimeter anterior glenohumeral capsulorrhaphy using a novel shoulder model of increased laxity. J Shoulder Elbow Surg. 2005;14(3):318–23.

    Article  PubMed  Google Scholar 

  23. Selecky MT, Tibone JE, Yang BY, McMahon PJ, Lee TQ. Glenohumeral joint translation after arthroscopic thermal capsuloplasty of the posterior capsule. J Shoulder Elbow Surg. 2003;12(3):242–6.

    Article  PubMed  Google Scholar 

  24. Veeger HE, Van der Helm FC, Van der Woude LH, Pronk GM, Rozendal RH. Inertia and muscle contraction parameters for musculoskeletal modelling of the shoulder mechanism. J Biomech. 1991;24(7):615–29.

    Article  CAS  PubMed  Google Scholar 

  25. Barrett Payne W, Kleiner MT, McGarry MH, Tibone JE, Lee TQ. Biomechanical comparison of the Latarjet procedure with and without a coracoid bone block. Knee Surg Sports Traumatol Arthrosc. 2016;24(2):513–20.

    Article  CAS  PubMed  Google Scholar 

  26. Habermeyer P, Lichtenberg S, Tauber M, Magosch P. Midterm results of stemless shoulder arthroplasty: a prospective study. J Shoulder Elbow Surg. 2015;24(9):1463–72.

    Article  PubMed  Google Scholar 

  27. Hawi N, Magosch P, Tauber M, Lichtenberg S, Habermeyer P. Nine-year outcome after anatomic stemless shoulder prosthesis: clinical and radiologic results. J Shoulder Elbow Surg. 2017;26(9):1609–15.

    Article  PubMed  Google Scholar 

  28. Massimini DF, Boyer PJ, Papannagari R, Gill TJ, Warner JP, Li G. In-vivo glenohumeral translation and ligament elongation during abduction and abduction with internal and external rotation. J Orthop Surg Res. 2012;7:29.

    Article  PubMed  PubMed Central  Google Scholar 

  29. Hertel R, Knothe U, Ballmer FT. Geometry of the proximal humerus and implications for prosthetic design. J Shoulder Elbow Surg. 2002;11(4):331–8.

    Article  PubMed  Google Scholar 

  30. Knowles NK, Carroll MJ, Keener JD, Ferreira LM, Athwal GS. A comparison of normal and osteoarthritic humeral head size and morphology. J Shoulder Elbow Surg. 2016;25(3):502–9.

    Article  PubMed  Google Scholar 

  31. Youderian AR, Ricchetti ET, Drews M, Iannotti JP. Determination of humeral head size in anatomic shoulder replacement for glenohumeral osteoarthritis. J Shoulder Elbow Surg. 2014;23(7):955–63.

    Article  PubMed  Google Scholar 

  32. Bohsali KI, Wirth MA, Rockwood CA. Complications of total shoulder arthroplasty. J Bone Joint Surg Am. 2006;88(10):2279–92.

    PubMed  Google Scholar 

  33. Chin PY, Sperling JW, Cofield RH, Schleck C. Complications of total shoulder arthroplasty: are they fewer or different? J Shoulder Elbow Surg. 2006;15(1):19–22.

    Article  PubMed  Google Scholar 

  34. Voss A, Beitzel K, Obopilwe E, Buchmann S, Apostolakos J, Di Venere J, Nowak M, Cote MP, Romeo AA, Mazzocca AD. No correlation between radiolucency and biomechanical stability of keeled and pegged glenoid components. BMC Musculoskelet Disord. 2017;18(1):213.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgments:

None.

Funding

The University of Connecticut Health Center/UConn Musculoskeletal Institute receives funding from Arthrex Inc. (Naples, FL, USA). The company had no influence on the study design, data collection, or interpretation of the results of the final article.

Open Access funding enabled and organized by Projekt DEAL.

Open Access funding enabled and organized by Projekt DEAL.

Author information

Authors and Affiliations

Authors

Contributions

LNM made substantial contributions to the conception and design of the study, data acquisition and analysis, drafting the manuscript. He approves the submitted version and has agreed both to be personally accountable for the author’s own contributions and to ensure that questions related to the accuracy or integrity of any part of the work, even ones in which the author was not personally involved, are appropriately investigated, resolved, and the resolution documented in the literature.

MM made substantial contributions to data acquisition. He approves the submitted version and has agreed both to be personally accountable for the author’s own contributions and to ensure that questions related to the accuracy or integrity of any part of the work, even ones in which the author was not personally involved, are appropriately investigated, resolved, and the resolution documented in the literature.

BO made substantial contributions in data interpretation and drafting the manuscript. She approves the submitted version and has agreed both to be personally accountable for the author’s own contributions and to ensure that questions related to the accuracy or integrity of any part of the work, even ones in which the author was not personally involved, are appropriately investigated, resolved, and the resolution documented in the literature.

CK made substantial contributions in drafting and revising the manuscript as well as interpretation of the data. He approves the submitted version and has agreed both to be personally accountable for the author’s own contributions and to ensure that questions related to the accuracy or integrity of any part of the work, even ones in which the author was not personally involved, are appropriately investigated, resolved, and the resolution documented in the literature.

EO made substantial contributions to the conception and design of the study, data acquisition and analysis. He approves the submitted version and has agreed both to be personally accountable for the author’s own contributions and to ensure that questions related to the accuracy or integrity of any part of the work, even ones in which the author was not personally involved, are appropriately investigated, resolved, and the resolution documented in the literature.

MPC made substantial contributions to data analysis and interpretation of the data. He approves the submitted version and has agreed both to be personally accountable for the author’s own contributions and to ensure that questions related to the accuracy or integrity of any part of the work, even ones in which the author was not personally involved, are appropriately investigated, resolved, and the resolution documented in the literature.

ADM substantial contributions to the conception and design of the study as well as revising the manuscript. He approves the submitted version and has agreed both to be personally accountable for the author’s own contributions and to ensure that questions related to the accuracy or integrity of any part of the work, even ones in which the author was not personally involved, are appropriately investigated, resolved, and the resolution documented in the literature.

DPB made substantial contributions to data acquisition and interpretation as well as drafting the manuscript. He approves the submitted version and has agreed both to be personally accountable for the author’s own contributions and to ensure that questions related to the accuracy or integrity of any part of the work, even ones in which the author was not personally involved, are appropriately investigated, resolved, and the resolution documented in the literature.

Corresponding author

Correspondence to Lukas N. Muench.

Ethics declarations

Ethical approval and consent to participate

This research involves human material. As de-identified specimens were not considered to constitute human subjects research, prior Institutional Review Board (IRB) approval by the University of Connecticut was not required. Further, a waiver of informed consent to participate was granted by the IRB. Please see documents provided in the related files section. The research was performed in accordance with the guidelines and regulations of the Declaration of Helsinki.

Consent for publication

Not applicable.

Competing Interest

ADM receives consulting fees as well as research support from Arthrex Inc. (Naples FL). The company was not involved in the study design, data collection, or final manuscript. LNM, MM, BO, CK, EO, MPC, and DPB report no conflict of interest.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Muench, L.N., Murphey, M., Oei, B. et al. Elliptical and spherical heads show similar obligate glenohumeral translation during axial rotation in total shoulder arthroplasty. BMC Musculoskelet Disord 24, 171 (2023). https://doi.org/10.1186/s12891-023-06273-5