Skip to main content

Is TKA femoral implant stability improved by pressure applied cement? a comparison of 2 cementing techniques

Abstract

Background

The majority of knee endoprostheses are cemented. In an earlier study the effects of different cementing techniques on cement penetration were evaluated using a Sawbone model. In this study we used a human cadaver model to study the effect of different cementing techniques on relative motion between the implant and the femoral shaft component under dynamic loading.

Methods

Two different cementing techniques were tested in a group of 15 pairs of human fresh frozen legs. In one group a conventional cementation technique was used and, in another group, cementation was done using a pressurizing technique. Under dynamic loading that simulated real life conditions relative motion at the bone-implant interface were studied at 20 degrees and 50 degrees flexion.

Results

In both scenarios, the relative motion anterior was significantly increased by pressure application. Distally, it was the same with higher loads. No significant difference could be measured posteriorly at 20°. At 50° flexion, however, pressurization reduced the posterior relative motion significantly at each load level.

Conclusion

The use of the pressurizer does not improve the overall fixation compared to an adequate manual cement application. The change depends on the loading, flexion angle and varies in its proportion in between the interface zones.

Peer Review reports

Background

Osteoarthritis is the most common joint disease worldwide, affecting 344 million people [1]. The knee is the most frequently involved joint and accounts for 50% of cases [2]. Total knee arthroplasty is one of the most successful interventions for restoring knee joint function and reducing pain after nonsurgical treatment options have been exhausted and the patient’s quality of life has been permanently impaired [3]. In 2020 there were 111,365 primary knee replacements and 13,767 revisions registered in the German Arthroplasty Registry [4]. Cementation remains the gold standard in knee arthroplasty [5], even though numerous publications have reported comparable survival and functional outcome for both cemented and cementless fixation methods [3, 6, 7]. This trend is also confirmed in reports from several prosthesis registries, in which cementation was used in 68% to 94% of their cases [4, 8, 9].

Cementing technique has been the subject of scientific research for many years because it affects the crucial interface between prosthesis and bone and is intended to enable long term survival [10]. There are many factors influencing good cementation results including the type of cement, viscosity, volume used, mixing procedures, temperature, humidity, jet lavage, timing, speed and force during impaction of the components and handling of the cement [11,12,13,14]. Although loosening of the femoral components accounts for only 4.6% of revisions [4], the continuously increasing number of implantations and revisions, in addition to earlier reports of significantly higher loosening rates of modern high-flex prostheses has justified detailed investigations of methods for optimizing prosthesis fixation [15,16,17,18]. Also, recent publications have reported a higher incidence of radiolucent lines accompanying a new prosthesis design when compared to its predecessor, although the clinical and biomechanical significance of this is currently unclear [19, 20].

Studies were able to indicate that applying cement to both the bone and implant prior to implantation is advantageous in TKA [13, 14, 21, 22]. Also, several authors have demonstrated that using a cement gun is advantageous [14, 23, 24]. In a publication from 2019, Schwarze et al. were able to demonstrate in a Sawbone® model of knee arthroplasty that cement application with a pressurizer creates a more homogeneous cement coating and adequate cement penetration [12], thereby confirming positive results from prior studies [23,24,25]. However, the used Sawbone® model only partially reflects the in-vivo scenario in reference to the cement-bone inferface. The aim of the current study was to further investigate the effects of pressurized application of cement to human femoral cadaver specimens, in particular the effects on implant stability and relative motion of the implant/bone interface during loading.

Methods

General

This study was performed in accordance with the Declaration of Helsinki and approved by the local Ethics Committee (Ethikkommission der Medizinischen Fakultät Heidelberg, reference S-351/2018). The tissue samples were obtained from Science Care (Phoenix, AZ, USA), which is accredited by the American Association of Tissue Banks. All donors and/or their legal guardian(s) provided informed consent prior to sample acquisition.

In 15 pairs of fresh frozen human legs the Attune total knee replacement system (DePuy Synthes, Warsaw, IN, USA) was implanted by a surgeon experienced in the surgical technique. Preservation of biomechanical properties prior to the experimental period was ensured by frozen storage [26]. In a randomized manner, two different cementation techniques (Groups A and B) were used for the implantation of the femoral component of the Attune system.

Group A consisted of specimens in which the articular surfaces of the femoral components and of the femoral condyles all had conventional cement application using a cement gun. Both surfaces were covered with cement, as this has been shown to provide the best results [13]. Also, the cement gun has been shown to provide superior cementation of the bone to finger packing [14, 23, 24].

Group B had the cement applied to the distal femur with a pressurizing nozzle added to the cement gun. The cement was applied to the femoral component in the conventional manner as in group A. (Fig. 1).

Fig. 1
figure 1

Femoral component and femur with red and blue colored areas on which cement has been applied. In Group A both areas were covered in the conventional manner; in Group B the red area indicates the femoral surface cemented using a no-touch technique implanted by use of a cement gun with a pressurizer. (a) is the anterior, (d) the distal and (p) the posterior regions

More details on the cementation technique are provided below.

The right and left sides of the 15 leg pairs were randomly allocated to group A or B by means of a computer-generated list (Randlist 1.2; Datinf GmbH, Tübingen, Germany). The mean donor data showed a mean age of 68.3 ± 11.5 years, a mean height of 174.4 ± 10.9 cm, a mean weight of 75.1 ± 16.4 kg, and a mean body mass index of 24.6 ± 4.7 kg/m2.

The bone mineral density (BMD) was assessed for both groups to improve the comparability. Franck et al. showed a high correlation between standard dual-energy absorptiometry (DXA) at the hip and various locations such as the extremities [27]. Therefore, we measured bone mineral density using DXA with standard hip parameters (Hologic QDR-2000, Marlborough, Massachusetts, USA). For all 30 knee joints, native radiographs in anterior–posterior (a.p.) and lateral projections were obtained to exclude bone pathology that would preclude a knee prosthesis and to determine prosthesis size using TraumaCad software (Voyant Health, Ltd., Brainlab AG, Munich, Germany). The same prosthesis size was planned and implanted on the right and left side of each leg pair. The following prosthesis sizes were used: 5 × size 5, 3 × size 6, 3 × size 7, 4 × size 8. Postoperative radiographs were performed to verify the implantation result and to exclude intraoperative fractures.

Cementing procedure

Prior to surgery, the human legs were thawed to room temperature. To standardize the experimental conditions and the surgical steps, all adjustments and resection measurements were documented and repeated on the contralateral side. Bone stock preparation and implantations were performed according to the prosthesis manufacturer's surgical instructions. The entire prosthesis was implanted and the femur and tibia were subsequently separated for testing. Prior to cementation, the cancellous bone was cleaned of lipid deposits, blood and bone debris using the OptiLavage system (Zimmer Biomet Holdings, Warsaw, Indiana, USA) and superficially dried with a compress until immediate cement application. The implantation of the femoral components for both Groups A and B was performed with a vacuum mixed high viscosity bone cement (Optipac 40 Refobacin Bone Cement R, Zimmer Biomet Holdings, Warsaw, Indiana, USA). The cement was applied early (in other words directly after the waiting phase) using cement timing for vacuum mixed cement at a room temperature of 21.2 ± 0.2 °C. We applied the bone cement to the non-articulating surface of the femoral components in Groups A and B 80 s after starting the mixing process. In the next step for Group A, the cement was applied to the prepared bone at 110 s using the above described cementing technique. In Group B, a cement gun with cement cartridge was also used, but a pressurizer nozzle was attached to the conventional nozzle to apply the cement to the bone in a no-touch technique (no manual manipulation of the cement after application) at 110 s (Fig. 2).

Fig. 2
figure 2

shows the standard nozzle short (A), the 23° pressurizer add-on (B), the attached 23° pressurizer on the standard nozzle short (C) and the contact surface between pressurizer and bone (D)

In group A, a homogeneous and uniform layer of PMMA cement was applied to the femoral bone stock with a cement gun (Optigun, Zimmer Biomet, Warsaw, Indiana, USA) medial and lateral from anterior to posterior for complete coverage. In addition, a homogeneous and uniform layer of cement was applied to the entire inner surface of the femoral knee component medially, laterally, and anteriorly transversely with the cement gun. On both the component and the bone, manual modelling with clean medical gloves was performed to ensure even coverage. In group B the cement gun was modified to deliver cement at an increased pressure by the attachment of a pressurizing nozzle with 23-degree angled tip. The cement was applied to the femoral surface under pressure, and the amount of cement was standardized to assess the influence of the cementing technique. An identical amount of cement was used in Groups A and B. The cement was applied to the femoral component in the same way in both groups [12].

The impaction of the femoral component was performed 140 s after start of mixing. The femoral component was impacted until the edges of the cement pockets were in contact with the distal bony resection surface. Excess cement was removed, and the trial liner insert was placed on the previously implanted tibial component. Subsequently, the leg was placed in extension position at 240 s after start of mixing, where the cement was allowed to harden.

Load simulation and determination of relative motion

After the cementing procedure, the tibia and femur were separated, and the soft tissues removed. Afterwards, the specimens were cast in a mold using synthetic resin (Rencast FC 53, Huntsman Advanced Materials GmbH, Germany), in order to secure the specimens into the material testing machine. For the assessment of implant stability, an incremental dynamic load was applied at 1 Hz for the axial force with simultaneous extension-flexion between 20° and 50°, as had been done in a prior study [28]. The load maxima occurred at the time of extension and flexion, respectively [29]. A force representing daily stair climbing [30,31,32] was applied using a servo-hydraulic testing machine (MTS 858 Mini Bionix II, MTS Systems Corporation, Eden Prairie, USA) (Fig. 3). A preload of 200 N was applied before cyclic loading was started with the four load levels 1200 N, 1500 N, 1800 N, and 2100 N. The maximum load level corresponded to the force exerted on the knee of a person with a body weight of 75 kg during stair climbing [32]. The selected body weight for the load simulation corresponded to the average donor body weight. Optical markers were placed on the bone and the adjacent implanted component as shown in Fig. 3. The determination of the three-dimensional relative motion between the femoral component and bone was performed using an optical, camera-based system (PONTOS- GOM – Gesellschaft für Optische Messtechnik mbH, Braunschweig, Germany). Figure 3 shows the implant and bone markers (A: anterior, B: distal, C: posterior) of the three analyzed zones. The system is calibrated to a measurement volume of 250 × 200 mm2. The markers on the object to be measured were located in the center of this defined volume. Each of the markers were detected in greyscale by a stereo camera system, and a 3D point triangulation was done to calculate the 3D marker position and displacement vector in the defined coordinate system. The relative motion was calculated from the corresponding implant and bone markers. The two cameras of the stereo system operate each with a resolution of 2448 × 2050 pixels and a measuring accuracy of 1 µm according to the manufacturer's specifications [33]. However, the measuring accuracy of an optical measuring system depends strongly on the environmental conditions. Under laboratory conditions, we achieved a measuring accuracy of ± 4.9 µm for the test setup used. All measurements were done at the medial side. The calculated results of the resultant maximum relative motion were normalized to the right femur for maximum extension and flexion (20°, 50°).

Fig. 3
figure 3

dynamic loading scenario of the distal femoral component in 20° (left) and 50° (center) of flexion articulating with the tibial liner in a servo-hydraulic testing machine. The casting resin in seen proximal to the femoral component. Correspondingly applied implant and bone markers (right) of the three analyzed zones (A: anterior, B: distal, C: posterior)

Statistical analysis

Prior to the start of the experimental study, a sample size calculation was performed using G*Power 3.1 (University of Kiel, Germany) [34] based on the reported data by Schwarze et al. [12]. The sample size differed (7, 9, 11, 16, 12 per group) depending on the cement penetration zone analyzed. The calculation of the sample size estimation with the largest number of cases per group was based on the following input parameters, tails: two, effect size d: 1.33, α err prob: 0.05 and power (1-β err prob): 0.95. This results in the output parameters sample size of 16 for each group. A sample size of 15 paired fresh frozen human legs was chosen for feasibility reasons. The data were evaluated descriptively using the arithmetic mean, standard deviation, minimum and maximum. Prior to analysis, the normal distribution of the data was evaluated using a Shapiro–Wilk-test and the homogeneity of variance was verified using the Levene-test. We conducted a two tailed t-test for independent samples to assess effects between both groups on the parameters BMD and relative motion within each load level, flexion angle and fixation zone. All data were analyzed using SPSS 25 (IBM, Armonk, NY, USA) with a significance level of p < 0.05.

Results

15 fresh frozen pairs were acquired to carry out the experiments. During the radiographic evaluation, one pair was excluded due to a bone lesion and subsequent fracture during experimentation. We evaluated the remaining 14 pairs.

Bone density

Testing the density differences between the two test groups using the Shapiro–Wilk test resulted in a p-value of 0.4. Thus, a normal distribution of the difference in bone density in both groups was confirmed. The following paired t-test showed no significant difference in bone density between the two groups (t(14) = -0.449, p = 0.66, d = 0.12).

Relative motion

The determination of the resulting (XYZ) maximum relative motion between implant and bone was analyzed for all four load levels with a total of 4000 cycles. The points in time of the maximum extension-flexion values (20° and 50°) with simultaneous maximum axial load were analyzed. When analyzing the femoral components, the evaluation was also divided into anterior, distal and posterior regions.

Femur 20º

The check for normal distribution with the Shapiro–Wilk test yielded (α = 0.05) normally distributed data. Therefore, a t-test was used for dependent samples. The test showed the following values for the load levels examined (Table. 1).

Table 1 Anterior, distal and posterior relative motion at 20° flexion of the femur depending on the load

Femur 50º

The check for normal distribution with the Shapiro–Wilk test yielded.

(α = 0.05) normally distributed data. Therefore, a t-test was used for dependent samples. The test showed the following values for the load levels examined (Table. 2).

Tab 2 Anterior, distal and posterior relative motion at 50° flexion of the femur depending on the load

Figures 4, 5 and 6 illustrate the anterior (Fig. 4), distal (Fig. 5) and posterior (Fig. 6) relative motion as a function of the degree of flexion (20˚ or 50˚), group (A or B) and incrementally increased load (1200 – 2100 N).

Fig. 4
figure 4

anterior relative motion as a function of the degree of flexion, group and incrementally increased load

Fig. 5
figure 5

distal relative motion as a function of the degree of flexion, group and incrementally increased load

Fig. 6
figure 6

posterior relative motion as a function of the degree of flexion, group and incrementally increased load

Discussion

Although femoral component loosening is a relatively uncommon occurrence and constitutes only 4.6% of all total knee arthroplasty (TKA) revisions, it is a relevant complication based on the constantly increasing number of implantations and revisions [4, 12]. Studies have shown that cementation technique has a significant impact on femoral component loosening[14, 22]. Our results with human fresh-frozen specimens and cement pressurization show that at 20-degrees and 50-degrees of flexion and incrementally increasing load there is an increase in the relative motion between implant and bone for both cementation techniques. In addition, the increase in relative motion both anteriorly and distally with higher loads is significantly higher in the pressurized cementation group compared to non-pressurized cementation. In contrast, posteriorly at 50 degrees flexion, the relative motion between implant and bone is significantly reduced with cement pressurization compared to non-pressurized cementation. Our measured values of relative micromotion between the bone and the components fell between 7 and 46 µm, which are comparable to values recorded in prior studies [35,36,37].

There have been numerous prior in-vitro and in-vivo studies of the primary stability of hip and knee arthroplasties, using radiostereometric analyses (RSA) and optical measurements as utilized in our current study[37,38,39,40,41,42]. These have shown that many factors influence primary stability, including implant design, bone density, surgical technique and cement penetration [36, 43, 44].

Pressure application during cementation has been shown to be effective in increasing cement penetration [12, 23,24,25]. In 2019, Schwarze et al. published data on 3 different cementation techniques in a Sawbone® model [12] and showed that pressure application improved cement penetration in all zones of the Knee Society Scoring System (KSSS). Although the Sawbone® model had a bone structure similar to human cancellous bone [45], it differed from physiological bone in other properties and therefore may not reflect cement penetration in the clinical setting.

This study demonstrates that the cementation technique can significantly influence the degree of relative motion at the bone/femoral component interface under differing loading conditions. We found a significant reduction of relative motion posteriorly only at 50 degrees flexion at all load levels with pressurization, that we attribute to increased stability resulting from increased compression load during this degree of flexion. All other values of relative motion at both 20 and 50 degrees of flexion showed decreased relative motion in the non-pressurized samples, mostly of a significant degree. These latter findings were unexpected, and we have no good explanation. This may be related to differing elastic and plastic deformation in different areas of the bone implant construct. Less movement in the posterior region may result in more pronounced movement in the other regions. Further investigations would be helpful in this regard.

Improved distributions of bone cement using pressure application [12] can significantly affect force transmission at the cement-host bone interface [46]. In a finite element analysis, Schultze et al. described the influence of cement thickness and prosthesis positioning, with the highest von Mises stresses anteriorly [46]. Our results showed that using a pressurizer only achieved a significant reduction of the relative motion between the implant and bone in the posterior region. The very narrow posterior intercondylar space precludes accurately controlled manual cement application into cancellous bone during the surgical procedure. Our results suggest that pressurized clinical application might be helpful for improved cementation in the posterior region only. The clinical significance of our documented differences in relative motion is unclear.

We cannot determine any clear association between our results and the occurrence of radiolucent lines noted radiographically. Hoskins et al. reported the majority of radiolucent lines distally (34.5%) and anteriorly (6.9%) while Staats et al. described them as being predominantly posteriorly located (12%) [19, 20]. The authors therefore do not see any noticeable association with the results of the current study.

Limitations

Although our experimental set-up mimicked the clinical situation as much as possible, the physiological effect of the surrounding soft tissues, differences in bone density and occurrence of bleeding could not be reproduced, limiting the extrapolation of our results to the clinical scenario.

Only two flexion angles were tested, unlike the physiological state which has a much greater range of motion.

The Attune knee replacement system was the only system tested and the results may vary with other systems.

Our incrementally increasing loads for 1000 cycles represent the immediate post-operative period only and does not reflect micromotion that could occur over the long-term postoperatively.

Conclusions

Pressure application of bone cement changes the relative motion at the implant-bone interface in all areas. The change varied with the degree of loading and the joint flexion angle and differed in the anterior, distal and posterior bone/component interface zones (seen Table 2). Our results suggest that the use of the pressurizer did not improve the overall fixation compared to an adequate application using a cement gun, with the possible exception of the posterior zone. The posterior region was the only area that displayed a significant reduction of micromotion with pressurized cement application during flexion. Therefore, we suggest that application of cement with a pressurizer may be advantageous in this region only, where the narrow intercondylar space makes satisfactory manual application or use of a cement gun without a pressurizer difficult. An improved cementation technique may further decrease the component loosening seen clinically. Additional studies are suggested to investigate this further.

Availability of data and materials

The datasets used and/or analysed during the current study available from the corresponding author on reasonable request.

References

  1. Cieza A, Causey K, Kamenov K, Hanson SW, Chatterji S, Vos T. Global estimates of the need for rehabilitation based on the Global Burden of Disease study 2019: a systematic analysis for the Global Burden of Disease Study 2019. Lancet. 2021;396(10267):2006–17.

    Article  Google Scholar 

  2. Luong ML, Cleveland RJ, Nyrop KA, Callahan LF. Social determinants and osteoarthritis outcomes. Aging health. 2012;8(4):413–37.

    Article  CAS  Google Scholar 

  3. Batailler C, Malemo Y, Demey G, Kenney R, Lustig S, Servien E. Cemented vs Uncemented Femoral Components: A Randomized, Controlled Trial at 10 Years Minimum Follow-Up. J Arthroplasty. 2020;35(8):2090–6.

    Article  Google Scholar 

  4. Grimberg A, Jansson V, Lützner J, Melsheimer O, Morlock M, Steinbrück A. EPRD Jahresbericht 2021; 2021. ISBN: 978-3-9817673-9-1. https://doi.org/10.36186/reporteprd042021.

  5. Nugent M, Wyatt MC, Frampton CM, Hooper GJ. Despite Improved Survivorship of Uncemented Fixation in Total Knee Arthroplasty for Osteoarthritis, Cemented Fixation Remains the Gold Standard: An Analysis of a National Joint Registry. J Arthroplasty. 2019;34(8):1626–33.

    Article  Google Scholar 

  6. Quispel CR, Duivenvoorden T, Beekhuizen SR, Verburg H, Spekenbrink-Spooren A, Van Steenbergen LN, Pasma JH, De Ridder R. Comparable mid-term revision rates of primary cemented and cementless total knee arthroplasties in 201,211 cases in the Dutch Arthroplasty Register (2007–2017). Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA. 2021;29(10):3400–8.

  7. Kim YH, Park JW, Jang YS. The 22 to 25-Year Survival of Cemented and Cementless Total Knee Arthroplasty in Young Patients. J Arthroplasty. 2021;36(2):566–72.

    Article  Google Scholar 

  8. Australian Orthopaedic Association National Joint Replacement Registry Hip, Knee & Shoulder Arthroplasty. Annual Report 2021. 2021. ISN 1445-3657.

  9. Robertsson O, Lidgren L, Sundberg M, W-Dahl A. The Swedish Knee Arthroplasty Register - Annual report 2020; 2020. ISBN 978-91-88017-32-1.

  10. Markolf KL, Amstutz HC. Penetration and flow of acrylic bone cement. Clin Orthop Relat Res. 1976;121:99–102.

    Google Scholar 

  11. Nilsson KG, Dalen T. Inferior performance of Boneloc bone cement in total knee arthroplasty: a prospective randomized study comparing Boneloc with Palacos using radiostereometry (RSA) in 19 patients. Acta Orthop Scand. 1998;69(5):479–83.

    Article  CAS  Google Scholar 

  12. Schwarze M, Schonhoff M, Beckmann NA, Eckert JA, Bitsch RG, Jager S. Femoral Cementation in Knee Arthroplasty-A Comparison of Three Cementing Techniques in a Sawbone Model Using the ATTUNE Knee. J Knee Surg. 2021;34(3):258–66.

    Article  CAS  Google Scholar 

  13. Refsum AM, Nguyen UV, Gjertsen JE, Espehaug B, Fenstad AM, Lein RK, Ellison P, Høl PJ, Furnes O. Cementing technique for primary knee arthroplasty: a scoping review. Acta Orthop. 2019;90(6):582–9.

    Article  Google Scholar 

  14. Han HS, Lee MC. Cementing technique affects the rate of femoral component loosening after high flexion total knee arthroplasty. Knee. 2017;24(6):1435–41.

    Article  Google Scholar 

  15. Cho SD, Youm YS, Park KB. Three- to six-year follow-up results after high-flexion total knee arthroplasty: can we allow passive deep knee bending? Knee Surg, Sports Traumatol, Arthrosc : official journal of the ESSKA. 2011;19(6):899–903.

  16. Choi YJ, Lee KW, Ha JK, Bae JY, Lee SK, Kim SB, Seo DK. Comparison of Revision Rates Due to Aseptic Loosening between High-Flex and Conventional Knee Prostheses. Knee Surg Relat Res. 2018;30(2):161–6.

    Article  Google Scholar 

  17. Han HS, Kang SB, Yoon KS. High incidence of loosening of the femoral component in legacy posterior stabilised-flex total knee replacement. J Bone Joint Surg Br. 2007;89(11):1457–61.

    Article  CAS  Google Scholar 

  18. Jung WH, Jeong JH, Ha YC, Lee YK, Koo KH. High early failure rate of the Columbus posterior stabilized high-flexion knee prosthesis. Clin Orthop Relat Res. 2012;470(5):1472–81.

    Article  Google Scholar 

  19. Hoskins W, Gorup P, Claireaux H, Stokes C, Bingham R. High incidence of radiolucent lines at the implant-cement interface of a new total knee replacement. ANZ J Surg. 2020;90(7–8):1299–302.

    Article  Google Scholar 

  20. Staats K, Wannmacher T, Weihs V, Koller U, Kubista B, Windhager R. Modern cemented total knee arthroplasty design shows a higher incidence of radiolucent lines compared to its predecessor. Knee Surg, Sports Traumatol, Arthrosc : official journal of the ESSKA. 2019;27(4):1148–55.

  21. Wetzels T, van Erp J, Brouwer RW, Bulstra SK, van Raay J. Comparing Cementing Techniques in Total Knee Arthroplasty: An In Vitro Study. J Knee Surg. 2019;32(9):886–90.

    Article  Google Scholar 

  22. Vaninbroukx M, Labey L, Innocenti B, Bellemans J. Cementing the femoral component in total knee arthroplasty: which technique is the best? Knee. 2009;16(4):265–8.

    Article  Google Scholar 

  23. Labutti RS, Bayers-Thering M, Krackow KA. Enhancing femoral cement fixation in total knee arthroplasty. J Arthroplasty. 2003;18(8):979–83.

    Article  Google Scholar 

  24. Lutz MJ, Pincus PF, Whitehouse SL, Halliday BR. The effect of cement gun and cement syringe use on the tibial cement mantle in total knee arthroplasty. J Arthroplasty. 2009;24(3):461–7.

    Article  Google Scholar 

  25. Krause WR, Krug W, Miller J. Strength of the cement-bone interface. Clin Orthop Relat Res. 1982;163:290–9.

    Article  Google Scholar 

  26. Linde F, Sorensen HC. The effect of different storage methods on the mechanical properties of trabecular bone. J Biomech. 1993;26(10):1249–52.

    Article  CAS  Google Scholar 

  27. Franck H, Munz M. Total body and regional bone mineral densitometry (BMD) and soft tissue measurements: correlations of BMD parameter to lumbar spine and hip. Calcif Tissue Int. 2000;67(2):111–5.

    Article  CAS  Google Scholar 

  28. Jaeger S, Eissler M, Schwarze M, Schonhoff M, Kretzer JP, Bitsch RG. Does tibial design modification improve implant stability for total knee arthroplasty? An experimental cadaver study. Bone Joint Res. 2022;11(4):229–38.

    Article  Google Scholar 

  29. Grupp TM, Holderied M, Pietschmann MF, Schroder C, Islas Padilla AP, Schilling C, Jansson V, Muller PE. Primary stability of unicompartmental knee arthroplasty under dynamic flexion movement in human femora. Clin Biomech (Bristol, Avon). 2017;41:39–47.

    Article  Google Scholar 

  30. Kutzner I, Bender A, Dymke J, Duda G, von Roth P, Bergmann G. Mediolateral force distribution at the knee joint shifts across activities and is driven by tibiofemoral alignment. Bone Joint J. 2017;99-B(6):779–87.

    Article  CAS  Google Scholar 

  31. Kutzner I, Heinlein B, Graichen F, Bender A, Rohlmann A, Halder A, Beier A, Bergmann G. Loading of the knee joint during activities of daily living measured in vivo in five subjects. J Biomech. 2010;43(11):2164–73.

    Article  CAS  Google Scholar 

  32. Bergmann G, Bender A, Graichen F, Dymke J, Rohlmann A, Trepczynski A, Heller MO, Kutzner I. Standardized loads acting in knee implants. PLoS ONE. 2014;9(1):e86035.

    Article  Google Scholar 

  33. Nadorf J, Kinkel S, Gantz S, Jakubowitz E, Kretzer JP. Tibial revision knee arthroplasty with metaphyseal sleeves: The effect of stems on implant fixation and bone flexibility. PLoS ONE. 2017;12(5):e0177285.

    Article  Google Scholar 

  34. Faul F, Erdfelder E, Lang AG, Buchner A. G*Power 3: a flexible statistical power analysis program for the social, behavioral, and biomedical sciences. Behav Res Methods. 2007;39(2):175–91.

    Article  Google Scholar 

  35. Miskovsky C, Whiteside LA, White SE. The cemented unicondylar knee arthroplasty An in vitro comparison of three cement techniques. Clin Orthop Relat Res. 1992;284:215–20.

    Article  Google Scholar 

  36. Cristofolini L, Affatato S, Erani P, Leardini W, Tigani D, Viceconti M. Long-term implant-bone fixation of the femoral component in total knee replacement. Proc Inst Mech Eng H. 2008;222(3):319–31.

    Article  CAS  Google Scholar 

  37. Reiner T, Jaeger S, Schwarze M, Klotz MC, Beckmann NA, Bitsch RG. The stability of the femoral component in the Oxford unicompartmental knee replacement: a comparison of single and twin peg designs. Bone Joint J. 2014;96-B(7):896–901.

    Article  CAS  Google Scholar 

  38. Grewal R, Rimmer MG, Freeman MA. Early migration of prostheses related to long-term survivorship Comparison of tibial components in knee replacement. J Bone Joint Surg Br. 1992;74(2):239–42.

    Article  CAS  Google Scholar 

  39. Karrholm J, Borssen B, Lowenhielm G, Snorrason F. Does early micromotion of femoral stem prostheses matter? 4-7-year stereoradiographic follow-up of 84 cemented prostheses. J Bone Joint Surg Br. 1994;76(6):912–7.

    Article  CAS  Google Scholar 

  40. Ryd L, Albrektsson BE, Carlsson L, Dansgard F, Herberts P, Lindstrand A, Regner L, Toksvig-Larsen S. Roentgen stereophotogrammetric analysis as a predictor of mechanical loosening of knee prostheses. J Bone Joint Surg Br. 1995;77(3):377–83.

    Article  CAS  Google Scholar 

  41. Seehaus F, Sonntag R, Schwarze M, Jakubowitz E, Sesselmann S, Kretzer JP, Hurschler C. Early assessment of the risk of later implantloosening using Roentgen Sterophotogrammetric Analysis (RSA). Orthopade. 2020;49(12):1042–8.

    Article  CAS  Google Scholar 

  42. Selvik G. Roentgen stereophotogrammetry A method for the study of the kinematics of the skeletal system. Acta Orthop Scand Suppl. 1989;232:1–51.

    Article  CAS  Google Scholar 

  43. Clarius M, Hauck C, Seeger JB, James A, Murray DW, Aldinger PR. Pulsed lavage reduces the incidence of radiolucent lines under the tibial tray of Oxford unicompartmental knee arthroplasty: pulsed lavage versus syringe lavage. Int Orthop. 2009;33(6):1585–90.

    Article  Google Scholar 

  44. MacDonald W, Swarts E, Beaver R. Penetration and shear strength of cement-bone interfaces in vivo. Clin Orthop Relat Res. 1993;286:283–8.

    Article  Google Scholar 

  45. Schwarze M, Borsai V, Bitsch RG, Eckert JA, Jager S: Letter to the Editor - "Cementing the femoral component in total knee arthroplasty: Which technique is the best?" by Michael Vaninbroukx et al. Knee 2017, 24(2):491–493.

  46. Schultze C, Kluss D, Martin H, Hingst V, Mittelmeier W, Schmitz KP, Bader R. Finite element analysis of a cemented ceramic femoral component for the assembly situation in total knee arthroplasty. Biomed Tech (Berl). 2007;52(4):301–7.

    Article  Google Scholar 

Download references

Acknowledgements

none

Funding

Open Access funding enabled and organized by Projekt DEAL. This research was funded by Stiftung Endoprothetik. For the publication fee we acknowledge financial support by Deutsche Forschungsgemeinschaft within the funding programme “Open Access Publikationskosten” as well as by Heidelberg University.

Author information

Authors and Affiliations

Authors

Contributions

Conceptualization, Mareike Schonhoff, Nicholas Beckmann and Martin Schwarze; Data curation, Mareike Schonhoff, Nicholas Beckmann, Martin Schwarze and Marvin Eissler; Formal analysis, Mareike Schonhoff, Nicholas Beckmann, Martin Schwarze and Sebastian Jaeger; Funding acquisition, Nicholas Beckmann, Martin Schwarze and Sebastian Jaeger; Investigation, Mareike Schonhoff, Nicholas Beckmann, Martin Schwarze, Marvin Eissler and Sebastian Jaeger; Methodology, Mareike Schonhoff, Nicholas Beckmann, Martin Schwarze, J. Philippe Kretzer and Sebastian Jaeger; Project administration, Nicholas Beckmann and Sebastian Jaeger; Resources, Nicholas Beckmann and Sebastian Jaeger; Software, Mareike Schonhoff, Martin Schwarze and Sebastian Jaeger; Supervision, Nicholas Beckmann, J. Philippe Kretzer, Tobias Renkawitz and Sebastian Jaeger; Validation, Mareike Schonhoff, Nicholas Beckmann and Sebastian Jaeger; Visualization, Nicholas Beckmann, Martin Schwarze and Sebastian Jaeger; Writing – original draft, Nicholas Beckmann and Martin Schwarze; Writing – review & editing, Mareike Schonhoff, Nicholas Beckmann, Martin Schwarze, Marvin Eissler, J. Philippe Kretzer, Tobias Renkawitz and Sebastian Jaeger. The author(s) read and approved the final manuscript.

Corresponding author

Correspondence to Nicholas A. Beckmann.

Ethics declarations

Ethics approval and consent to participate

This study was performed in accordance with the Declaration of Helsinki and approved by the local Ethics Committee (Ethikkommission der Medizinischen Fakultät Heidelberg, reference S-351/2018). The tissue samples were obtained from Science Care (Phoenix, AZ, USA), which is accredited by the American Association of Tissue Banks, and receives informed donor consent from all subjects and/or their legal guardian(s) prior to sample acquisition.

Consent for publication

Not applicable.

Competing interests

N.A.B. has received research support from the Deutsche Arthrose Hilfe e.V. for an unrelated project. T.R. has received research support and personal fees from Arbeitsgemeinschaft Endoprothetik (AE), DGOU, DGOOC; BVOU, DePuy International, Otto Bock Foundation, Deutsche Arthrose Hilfe, Aesculap, Zimmer, Stiftung Oskar Helene Heim Berlin, Vielberth Foundation Regensburg, the German Ministry of Education and Research as well as the German Federal Ministry of Economic Cooperation and Development. He is Medical Director and Chair at the Orthopaedic Department at Heidelberg University Hospital, board member of the German Society for Orthopaedics and Trauma (DGOOC) and vice president of the Professional Association of Orthopaedic Specialists and Trauma Surgeons (BVOU). S.J. reports grants from B Braun Aesculap, Johnson & Johnson Depuy Synthes, Heraeus Medical, Waldemar Link, Peter Brehm, Implantcast, Ceramtec, Mathys, Tornier/Stryker and Zimmer Biomet that are not related to the current study. The remaining authors have no conflicts to declare. The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Schonhoff, M., Beckmann, N.A., Schwarze, M. et al. Is TKA femoral implant stability improved by pressure applied cement? a comparison of 2 cementing techniques. BMC Musculoskelet Disord 24, 51 (2023). https://doi.org/10.1186/s12891-023-06151-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-023-06151-0

Keywords