Assessing component orientation of total hip arthroplasty using the low-dose bi-planar radiographs

Abstract

Background

Three-dimensional computed tomography (3D CT) reconstruction is the reference standard for measuring component orientation. However, functional cup orientation in standing position is preferable compared with supine position. The low-dose bi-planar radiographs can be used to analyze standing cup component orientation. We aimed to assess the validity and reliability of the component orientation using the low-dose bi-planar radiographs compared with the 3D CT reconstruction, and explore the differences between the functional cup orientation in standing radiographs and supine CT scans.

Methods

A retrospective study, including 44 patients (50 hips) with total hip arthroplasty (THA), was conducted. CT scans were taken 1 week after surgery and the low-dose bi-planar radiographs were taken in the follow-up 6 weeks later. Component orientation measurement was performed using the anterior pelvic plane and the radiographic coronal plane as reference, respectively.

Results

The study showed no significant difference in cup anteversion (p = 0.160), cup inclination (p = 0.486), and stem anteversion (p = 0.219) measured by the low-dose bi-planar radiographs and 3D reconstruction. The differences calculated by the Bland–Altman analysis ranged from − 0.4° to 0.6° for the three measured angles. However, the mean absolute error was 4.76 ± 1.07° for functional anteversion (p = 0.035) and 4.02 ± 1.08° for functional inclination (p = 0.030) measured by the bi-planar radiographs and supine CT scans.

Conclusions

The low-dose bi-planar radiographs are the same reliable and accurate as 3D CT reconstruction to assess post-THA patients’ component orientation, while providing more valuable functional component orientation than supine CT scans.

Peer Review reports

Background

Total hip arthroplasty (THA) has achieved great success in relieving pain and restoring function among patients with hip osteoarthritis. However, suboptimal implant placement can lead to impingement, dislocation, and accelerate wear [1,2,3]. The assessment of component orientation is critical for the postoperative evaluation of THA [4, 5]. Three-dimensional computed tomography (3D CT) reconstruction is precise and is not influenced by positional variables, and currently, is the reference standard for measuring component orientation [6, 7]. However, most surgeons do not order CT scans routinely to measure implant position postoperatively due to its high cost and radiation exposure [8]. Additionally, cup orientation measured in the supine position usually differs from the functional orientation in standing position, due to spinal pelvic motion.

The low-dose bi-planar radiographs (EOS imaging, Paris, France) have been developed as a new method for clinical implant position analysis, which allows patients to be evaluated in standing posture [9, 10]. With two perpendicular X-ray beams mounted on a vertically traveling C-arm, the system scans all or part of the body and produces projections in two perpendicular planes simultaneously. Additionally, the system employs dedicated software (sterEOS, EOS imaging, Paris, France) to adjust generic models of the hip and femur, thereby, generating highly accurate 3D models of the patient’s bone and THA prosthesis, which can be used for assessing prosthetic orientation parameters in standing position, including cup anteversion, cup inclination, and stem anteversion [11]. The low-dose bi-planar radiographs can, thus, provide two sets of parameters: one with the anterior pelvic plane (APP) as the reference plane, which is defined by both anterior superior iliac spines (ASIS) and the pubic symphysis (Fig. 1A) [12], while the other with the radiographic coronal plane (functional coronal plane). Some previous studies have already evaluated the accuracy of measurements of femoral, tibial, and femorotibial torsion using the low-dose bi-planar radiographs [13,14,15,16,17]. Demzik et al. [18] analyzed inter-rater and intra-rater repeatability, and the reliability of pelvic parameters.

Fig. 1
figure 1

A-D The reconstruction process is a software-guided step-by-step procedure. A Identifying the sacral plate, the sacroiliac joints, the acetabula, the pubis and the anterosuperior iliac spines. B Adjustment of the 3D ellipse on the border of the acetabular cup. C Identifying the key landmarks on the femur, the position of the trochlear notch and condyles, and adjustment of the prosthetic head. D Adjustment for the position of the prosthetic neck’s landmarks and identifying the inferior extremity of the stem

In addition, it has been reported that the sagittal pelvic tilt (PT) is different between the supine and standing positions, which inevitably leads to a change in cup position, leading to the failure of the traditional Lewinnek safe zone [19,20,21]. The spinal pelvic motion necessitates the evaluation of functional cup orientation in standing position, which is one of the major advantages of the low-dose bi-planar radiographs over the supine CT scans [22, 23]. However, there is scant research on the difference between the functional standing and supine cup orientation as measured by bi-planar radiographs and CT scans, respectively.

Therefore, the purpose of this study was to answer two questions: What is the validity and reliability of the component orientation assessment of the low-dose bi-planar radiographs in comparison with 3D CT reconstruction using the APP as the reference plane? How does the functional component orientation differ, when measured by the low-dose bi-planar radiographs, from that by supine CT scans?

Methods

Patients

Forty-four patients (50 hips, 18 men, 26 women; mean age, 51.3 years, range, 26–78 years, standard deviation [SD], 12.9 years), who underwent robotic-assisted THA between September 2019 and September 2020, were included in this study with approval of our institutional review board. The pre-operative diagnoses were osteoarthritis (18 hips), osteonecrosis of the femoral head (17 hips), developmental dysplasia of the hips (13 hips), old femoral neck fracture (1 hip), and Charcot’s arthropathy (1 hip). 6 patients underwent bilateral THA (2 osteoarthritis; 2 osteonecrosis of the femoral head; 2 developmental dysplasia of the hips). All arthroplasties were performed by an experienced orthopedic surgeon using a modified Gibson approach and Accolade II implants (Stryker, Mahwah, NJ, USA). All patients underwent supine CT scans 1 week after the surgery and the low-dose bi-planar radiographs, in the follow-up 6 weeks later.

EOS measurements

The EOS biplanar radiographs were obtained with the patients in a weight-bearing standing position. The reconstruction process was a software-guided step-by-step procedure, including identifying anatomical landmarks (Fig. 1). Recognition of anatomical landmarks is crucial for accurate measurements. At the same time, we obtained the parameters with the functional coronal plane as the reference plane. The cup anteversion was anatomical, and the inclination was radiographic [24]. The femoral stem neck axis was defined as the axis passing through the center of the femoral head and the mid-point of the stem neck. The posterior condylar axis was defined as the axis passing through the most posterior points of the medial and lateral condyles. Stem anteversion was defined as the angle between the neck axis and the posterior condylar axis projected in the plane orthogonal to the femoral mechanical axis (Fig. 2). PT was defined in the sagittal plane between the line connecting the midpoint of the sacral plate and the midpoint of the acetabular axis, and the vertical axis.

Fig. 2
figure 2

The 3D modeling and radiology parameters of the hip are completed by the sterEOS software

CT reconstruction and measurement

In this study, a spiral CT scanner (80-slice CT-scanner Aquilion Prime, Toshiba) was used for measuring the pelvis, hip joint, and knee joint. All images were digitally acquired using the Rogan-Delft View (Pro-X, Version 3.2.0.12, Veenendaal, The Netherlands). The volume data were stored in Digital Imaging and Communications in Medicine (DICOM) format for later 3D analysis. CT images of the pelvis and femur were segmented and measured using Mimics software 17 (Materialize, Leuven, Belgium) and computer-aided design software (SOLIDWORKS®Premium 2017 SP2.0, USA).

We used the method proposed by Wang RY et al. [25] to measure the anatomical cup anteversion and radiological inclination identical to that measured in the low-dose bi-planar radiographs. The APP determined by the bilateral ASIS, pubic tubercles, and sacral crests, was utilized as the reference coronal plane identical to that in the EOS method (Fig. 3 B-C). The plane perpendicular to the line joining the bilateral ASIS was defined as the sagittal plane, and the plane perpendicular to the above two planes was used as the transverse plane (Fig. 4). The acetabular axis was determined by the edge of the cup (Fig. 3A). Angles were calculated by normal vector projection using a mathematical formula. For example, the anatomical anteversion was the angle between the transverse axis of the transverse plane and the acetabular axis when projected to the transverse plane. Therefore, the vector of the acetabular axis was projected to the transverse plane firstly, and then the angle was calculated by:

$$\uptheta ={\cos}^{-1}\frac{\overrightarrow{n}\cdot \overrightarrow{m}}{\left|\overrightarrow{n}\right|\mid \overrightarrow{m}\mid }$$
Fig. 3
figure 3

A-D The pelvis and femur are 3D reconstructed. A Identification of the acetabular axis by the edge of the cup. B Identification of the sacral crest. C Identification of the bilateral anterior superior iliac spine (ASIS) and pubic tubercles, and the midline of bilateral ASIS. D Definition of the stem anteversion as the angle formed by the axis of the femoral neck and the posterior tangential line of femoral condyles

Fig. 4
figure 4

The APP and the sagittal plane are defined as bony landmarks. The APP is defined as bilateral ASIS and pubic tubercles. The sagittal plane is defined as the midline of bilateral ASIS and sacral crest

\(\overrightarrow{n}\) was the vector of projection of the acetabular axis and \(\overrightarrow{m}\) was the vector of the transverse axis of the transverse plane.

To measure the cup orientation in supine CT scans, we chose the radiographic coronal plane as the reference plane, and the operational steps were the same as above. We also measured PT in supine CT scans.

We employed the method proposed by Yan W et al. [13] to measure stem anteversion with the same definition as that used in the EOS method. The 3D model was rotated so that a horizontal line could connect the most posterior point of the medial and lateral condyles and the lowest point of the greater trochanter, which was located in the middle of the medial and lateral condyles. Another line connected the prosthetic head and the center of the base of the stem’s neck. Stem anteversion was defined as the angle formed by these two lines (Fig. 3D). The angle was positive if the stem neck was anteverted.

Statistical analysis

To assess intra-observer reliability of the low-dose bi-planar radiographs, one examiner (MZY) measured all the hips twice with a 2-week interval and randomly ordered radiographs without knowing the previous results. To assess inter-observer reliability, another examiner (WSY) independently measured all the hips using the low-dose bi-planar radiographs. The intra-class correlation coefficient (ICC) and 95% confidence interval (CI) were calculated for both inter-and intra-observer reliability. To determine the validity of the low-dose bi-planar radiographs with the 3D CT reconstruction as the reference standard, we performed the paired t-test with statistical significance set at p < 0.05. The mean absolute error (MAE) was calculated with a 95% CI, and the Bland and Altman plots were used to assess the discrepancies. Moreover, we compared the functional cup orientation and PT by the bi-planar radiographs versus supine CT scans, respectively. MedCalc (version 19.5.2, Mariakerke, Belgium) and SPSS 19.0 (IBM, Armonk, NY, USA) were used to conduct statistical analysis tests. We also calculated the means and SDs of component orientation and PT measured by the bi-planar radiographs, 3D CT reconstruction, and supine CT scans, respectively. We calculated the sample size from the effect size obtained from Khan M et al. [26] using the Gpower 3.1 software. We calculated that a minimum of 42 hips would be required when α = 0.05 for a power of 0.95.

Results

Our results showed assessing component orientation using the low-dose bi-planar radiographs was accurate. With 3D CT reconstruction as the reference method, there was no significant difference in the cup anteversion (0.62°, 3.05° SD, p = 0.160), cup inclination (0.32°, 3.21° SD, p = 0.486), and stem anteversion (− 0.41°, 2.34° SD, p = 0.219) (Table 1). The 95% CI of MAE in measuring component orientation with the APP as the reference plane was relatively low, as follows: 2.43 ± 0.53° for cup anteversion, 2.48 ± 0.57° for cup inclination, and 2.09 ± 0.30° for stem anteversion (Table 2).

Table 1 Validity of bi-planar radiographs compared with CT scans for component orientation using a paired t-test
Table 2 MAE of the low-dose bi-planar radiographs and CT scans with APP as the reference plane

The Bland-Altman analysis revealed that, in comparison with 3D CT reconstruction, the means of errors and the percentage of agreement in the low-dose bi-planar radiographs with the APP as the reference plane were: 0.6° (range, − 5.4° to 6.6°), 92% for anteversion, 0.3° (range, − 6.0° to 6.6°), 94% for cup inclination, and − 0.4° (range, − 5.0° to 4.2°), 100% for stem anteversion (Fig. 5), indicating that there was no systematical error.

Fig. 5
figure 5

A-C The measured results of bi-planar radiographs and 3D CT reconstruction are examined by Bland-Altman analysis. (A) Cup anteversion with the APP as the reference plane; (B) Cup inclination with the APP as the reference plane; (C) Stem anteversion

With the APP as the reference plane, the inter- and intra-observer reliabilities of the low-dose bi-planar radiographs were good. The ICCs were 0.945 and 0.956 for measuring cup anteversion, 0.923 and 0.928 for measuring cup inclination, and 0.981 and 0.987 for measuring stem anteversion (Table 3).

Table 3 Inter- and intra-observer reliability for component orientation measured using the low-dose bi-planar radiographs

For the functional cup orientation and PT assessment, significant differences were found between the standing radiographs and supine CT scans with the radiographic coronal plane as reference, including the cup anteversion (1.80°, 5.89° SD, p = 0.035), the cup inclination (− 1.69°, 5.35° SD, p = 0.030), and the PT (2.05°, 6.73° SD, p = 0.037), respectively (Table 4). The MAE was relatively high, reaching 4.76 ± 1.07°, 4.02 ± 1.08°, and 5.36 ± 1.25°, respectively (Table 5).

Table 4 Differences between the functional cup orientation and PT in standing images and in supine CT
Table 5 MAE of bi-planar radiographs with functional coronal reference plane and supine CT scans

With the APP as the reference plane, the mean values (measured on the low-dose bi-planar radiographs) were 35.16°(SD, 8.54°)for anatomical anteversion, 42.16°(SD, 5.16°)for radiographic inclination, and 17.68°(SD, 9.16°)for stem anteversion. The mean 3D CT values were 34.55°(SD, 8.30°)for anatomical anteversion, 41.84°(SD, 4.97°)for radiographic inclination, and 18.09°(SD, 8.67°)for stem anteversion. With the radiographic coronal plane as the reference plane, the mean values (measured on the low-dose bi-planar radiographs) were 29.88°(SD, 9.45°)for anatomical functional anteversion, 39.74°(SD, 5.21°)for radiographic functional inclination, and 4.31°(SD, 7.11°)for PT. The mean supine CT values were 28.08°(SD, 7.72°)for anatomical anteversion, 41.43°(SD, 5.52°)for radiographic inclination, and 2.26°(SD, 8.57°)for PT.

Discussion

Component orientation is one of the most important factors determining the long-term outcomes of THA. Although 3D CT reconstruction has been reported to be the most accurate method for measuring prosthetic orientation, CT scans are not routinely used for post-operative and follow-up assessment because of the high cost, high radiation exposure, and supine scanning posture. Our results showed that the low-dose bi-planar radiographs were accurate and reliable in measuring cup anteversion and inclination with the APP as the reference plane compared with 3D CT reconstruction, as well as stem anteversion. Most importantly, we found that the parameters were significantly different from the 3D CT reconstruction when we chose the functional coronal plane as the reference plane.

Our data showed that the validity and reliability of the low-dose bi-planar radiographs were comparable with the CT scans, for assessing cup and stem orientation using the APP as the reference standard. Our research results were consistent with previous literature [13, 14, 27], and the results were better than traditional radiography, which is distorted by magnification and cannot be corrected by single radiograph [28]. In addition, the low-dose bi-planar radiographs have an advantage in the correction of axial rotation in standing posture compared with single anterior-posterior view radiography. This new imaging modality, thus, provides an accurate method to evaluate the orientation of the THA component.

With the accuracy validated, we found a major difference in the functional cup orientation and PT between the low-dose bi-planar radiographs and supine CT scans, which was most likely due to the change of pelvic orientation between these two postures. Dorr LD et al. [21, 29] have reported that spinal pelvic motion is a crucial factor determining the functional component orientation and has rendered the traditional Lewinnek safe zone ineffective in predicting dislocation. Acetabular orientation is not a static parameter, because in the sagittal plane, the pelvis moves due to several factors [30]. Functional cup orientation can be analyzed by measuring the sagittal tilt of the pelvis. Pierrepoint J et al. [22] found that PT in all functional positions showed variations from the supine position. The mean absolute change in sagittal PT moving from supine to standing was 6.0° in their study, which was 5.36° in the current study. This change of PT explained the change in the functional cup anteversion and inclination between the supine and standing postures. We thus recommend routine use of standing view radiographs for evaluation of functional standing cup orientation, especially for those at high risk of dislocation or analyzing the cause of dislocation [31,32,33]. Furthermore, we recommend that preoperative standing view radiography of the pelvis should be evaluated before computer-assisted THA to make accurate planning, as surgeons need to target the cup’s functional orientation of standing position, instead of merely based on supine CT scans [34,35,36].

Different definitions of inclination and anteversion can be easily misused [24]. The current study utilized the radiographic inclination and anatomical anteversion identical to CT measurement to facilitate the comparisons between the CT and the bi-planar radiographs assessment. This should be considered while interpreting the results of the imaging assessment of cup orientation.

Stem anteversion is equally important for the stability of THA. Our data showed that the anatomical stem version measured by the low-dose bi-planar radiographs was as accurate as measured by CT scans. It is also important to consider the functional anteversion of the stem, which may vary with the rotational posture of the lower limb and can be different from anatomical anteversion [37, 38]. Therefore, the clinician must consider how the functional anteversion of the stem changes when the patient is in different postures performing various functional tasks. To study how axial rotation of the femur changed in supine and standing positions, Uemura K et al. [39] employed an intensity-based 2D-3D registration technique to quantify axial rotation of the hip. They found substantial variability in the femoral rotational angle, which confirmed the above view. We recommend further development of new algorithms for this application.

An obvious advantage of the low-dose bi-planar radiographs is that it accommodates the patient’s whole body, which enables the patient to pose differently in weight-bearing positions, including standing, sitting, squatting, and standing on one leg. Moreover, the reduced radiation dose is an additional critical advantage, which is 2.5 times lower than the plain X-ray and 4–8 times lower than that of the CT scanner [28, 40, 41]. In addition, it is also difficult for some patients to lie down on the CT examination bed, such as patients with degenerative joint disease and patients with hunchback, making the low-dose bi-planar radiographs a more convenient method for them [18].

Our study has some limitations. First, this was a non-randomized retrospective study, which might introduce some selection bias into the patient groups. However, all patients were consecutively enrolled from the registration center in our hospital. Second, the CT scans should also be taken in the follow-up 6 weeks later. Although the research results of Dorr LD et al. [42] shown some patients did not have sufficient lower-limb strength to gain balance in the first 6 postoperative weeks, we assumed this would not affect the results of supine CT scans. Third, other imaging methods were not included, such as standing CT scans, which might overcome the shortcomings of supine CT scans. However, radiation exposure is always a major concern, which impedes the routine application of CT in THA patients’ follow-up.

Conclusion

The low-dose bi-planar radiographs is comparable with 3D CT reconstructions to assess post-THA patients’ component orientation. The functional standing component orientation is different from the supine CT measurement, indicating the necessity of assessing the component orientation in the standing position.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

3D CT:

Three-dimensional computed tomography

THA:

Total hip arthroplasty

APP:

Anterior pelvic plane

ASIS:

Anterior superior iliac spines

PT:

Pelvic tilt

SD:

Standard deviation

DICOM:

Digital Imaging and Communications in Medicine

ICC:

Intra-class correlation coefficient

CI:

Confidence interval

MAE:

Mean absolute error

References

  1. Scheerlinck T. Cup positioning in total hip arthroplasty. Acta Orthop Belg. 2014;80(3):336.

    PubMed  CAS  Google Scholar 

  2. Seagrave KG, Troelsen A, Malchau H, Husted H, Gromov K. Acetabular cup position and risk of dislocation in primary total hip arthroplasty. Acta Orthop. 2017;88:10–7.

    Article  Google Scholar 

  3. Shon WY, Baldini T, Peterson MG, Wright TM, Salvati EA. Impingement in total hip arthroplasty a study of retrieved acetabular components. J Arthroplast. 2005;20:427–35.

    Article  Google Scholar 

  4. Murphy WS, Yun HH, Hayden B, Kowal JH, Murphy SB. The safe zone range for cup Anteversion is narrower than for inclination in THA. Clin Orthop Relat Res. 2018;476:325–35.

    Article  Google Scholar 

  5. Harrison CL, Thomson AI, Cutts S, Rowe PJ, Riches PE. Research synthesis of recommended acetabular cup orientations for total hip arthroplasty. J Arthroplast. 2014;29:377–82.

    Article  Google Scholar 

  6. Lin F, Lim D, Wixson RL, Milos S, Hendrix RW, Makhsous M. Validation of a computer navigation system and a CT method for determination of the orientation of implanted acetabular cup in total hip arthroplasty: a cadaver study. Clin Biomech (Bristol, Avon). 2008;23:1004–11.

    Article  Google Scholar 

  7. Olivecrona H, Weidenhielm L, Olivecrona L, Beckman MO, Stark A, Noz ME, et al. A new CT method for measuring cup orientation after total hip arthroplasty: a study of 10 patients. Acta Orthop Scand. 2004;75:252–60.

    Article  Google Scholar 

  8. Lu M, Zhou YX, Du H, Zhang J, Liu J. Reliability and validity of measuring acetabular component orientation by plain anteroposterior radiographs. Clin Orthop Relat Res. 2013;471(9):2987–94.

    Article  Google Scholar 

  9. Lazennec JY, Rousseau MA, Rangel A, Gorin M, Belicourt C, Brusson A, et al. Pelvis and total hip arthroplasty acetabular component orientations in sitting and standing positions: measurements reproductibility with EOS imaging system versus conventional radiographies. Orthop Traumatol Surg Res. 2011;97:373–80.

    Article  CAS  Google Scholar 

  10. Journé A, Sadaka J, Bélicourt C, Sautet A. New method for measuring acetabular component positioning with EOS imaging: feasibility study on dry bone. Int Orthop. 2012;36:2205–9.

    Article  Google Scholar 

  11. Bendaya S, Anglin C, Lazennec JY, Allena R, Thoumie P, Skalli W. Good vs poor results after Total hip arthroplasty: an analysis method using implant and anatomic parameters with the EOS imaging system. J Arthroplast. 2016;31:2043–52.

    Article  Google Scholar 

  12. McKibbin B. Anatomical factors in the stability of the hip joint in the newborn. J Bone Joint Surg Br. 1970;52:148–59.

    Article  CAS  Google Scholar 

  13. Yan W, Xu X, Xu Q, Yan W, Sun Z, Jiang Q, et al. Femoral and tibial torsion measurements based on EOS imaging compared to 3D CT reconstruction measurements. Ann Transl Med. 2019;7:460.

    Article  Google Scholar 

  14. Folinais D, Thelen P, Delin C, Radier C, Catonne Y, Lazennec JY. Measuring femoral and rotational alignment: EOS system versus computed tomography. Orthop Traumatol Surg Res. 2013;99:509–16.

    Article  CAS  Google Scholar 

  15. Mayr HO, Schmidt JP, Haasters F, Bernstein A, Schmal H, Prall WC. Anteversion angle measurement in suspected torsional malalignment of the femur in 3-dimensional EOS vs computed tomography-a validation study. J Arthroplast. 2021;36(1):379–86.

    Article  Google Scholar 

  16. Kato S, Debaud C, Zeller RD. Three-dimensional EOS analysis of apical vertebral rotation in adolescent idiopathic scoliosis. J Pediatr Orthop. 2017;37:e543–7.

    Article  Google Scholar 

  17. Guenoun B, El Hajj F, Biau D, Anract P, Courpied JP. Reliability of a new method for evaluating femoral stem positioning after total hip arthroplasty based on stereoradiographic 3D reconstruction. J Arthroplast. 2015;30:141–4.

    Article  Google Scholar 

  18. Demzik AL, Alvi HM, Delagrammaticas DE, Martell JM, Beal MD, Manning DW. Inter-rater and intra-rater repeatability and reliability of EOS 3-dimensional imaging analysis software. J Arthroplast. 2016;31:1091–5.

    Article  Google Scholar 

  19. Dorr LD, Callaghan JJ. Death of the Lewinnek "safe zone". J Arthroplast. 2019;34:1–2.

    Article  Google Scholar 

  20. Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–20.

    Article  CAS  Google Scholar 

  21. Tezuka T, Heckmann ND, Bodner RJ, Dorr LD. Functional safe zone is superior to the Lewinnek safe zone for Total hip arthroplasty: why the Lewinnek safe zone is not always predictive of stability. J Arthroplast. 2019;34:3–8.

    Article  Google Scholar 

  22. Pierrepont J, Hawdon G, Miles BP, Connor BO, Baré J, Walter LR, et al. Variation in functional pelvic tilt in patients undergoing total hip arthroplasty. Bone Joint J. 2017;99-b:184–91.

    Article  CAS  Google Scholar 

  23. Malik A, Wan Z, Jaramaz B, Bowman G, Dorr LD. A validation model for measurement of acetabular component position. J Arthroplast. 2010;25(5):812–19.

  24. Murray DW. The definition and measurement of acetabular orientation. J Bone Joint Surg Br. 1993;75:228–32.

    Article  CAS  Google Scholar 

  25. Wang RY, Xu WH, Kong XC, Yang L, Yang SH. Measurement of acetabular inclination and anteversion via CT generated 3D pelvic model. BMC Musculoskelet Disord. 2017;18:373.

    Article  CAS  Google Scholar 

  26. Khan M, Beckingsale T, Marsh M, Holland J. Difference in the acetabular cup orientation in standing and supine radiographs. J Orthop. 2016;13(3):168–70.

    Article  Google Scholar 

  27. Buck FM, Guggenberger R, Koch PP, Pfirrmann CW. Femoral and tibial torsion measurements with 3D models based on low-dose biplanar radiographs in comparison with standard CT measurements. AJR Am J Roentgenol. 2012;199:W607–12.

    Article  Google Scholar 

  28. Chiron P, Demoulin L, Wytrykowski K, Cavaignac E, Reina N, Murgier J. Radiation dose and magnification in pelvic X-ray: EOS™ imaging system versus plain radiographs. Orthop Traumatol Surg Res. 2017;103:1155–9.

    Article  CAS  Google Scholar 

  29. Stefl M, Lundergan W, Heckmann N, McKnight B, Ike H, Murgai R, et al. Spinopelvic mobility and acetabular component position for total hip arthroplasty. Bone Joint J. 2017;99-B:37–45.

    Article  CAS  Google Scholar 

  30. Langston J, Pierrepont J, Gu Y, Shimmin A. Risk factors for increased sagittal pelvic motion causing unfavourable orientation of the acetabular component in patients undergoing total hip arthroplasty. Bone Joint J. 2018;100-b:845–52.

    Article  CAS  Google Scholar 

  31. Yang G, Li Y, Zhang H. The influence of pelvic tilt on the Anteversion angle of the acetabular prosthesis. Orthop Surg. 2019;11:762–9.

    Article  Google Scholar 

  32. Yun H, Murphy WS, Ward DM, Zheng G, Hayden BL, Murphy SB. Effect of pelvic tilt and rotation on cup orientation in both supine and standing positions. J Arthroplast. 2018;33:1442–8.

    Article  Google Scholar 

  33. Maillot C, Harman C, Villet L, Cobb J, Rivière C. Modern cup alignment techniques in total hip arthroplasty: a systematic review. Orthop Traumatol Surg Res. 2019;105:907–13.

    Article  Google Scholar 

  34. Wan Z, Malik A, Jaramaz B, Chao L, Dorr LD. Imaging and navigation measurement of acetabular component position in THA. Clin Orthop Relat Res. 2009;467(1):32–42.

    Article  Google Scholar 

  35. Tang H, Li Y, Zhou Y, Wang S, Zhao Y, Ma Z. A modeling study of a patient-specific safe zone for THA: calculation, validation, and key factors based on standing and sitting sagittal pelvic tilt. Clin Orthop Relat Res. 2022;480(1):191–205.

    PubMed  Google Scholar 

  36. Tang H, Zhao Y, Wang S, Ma Z, Li Y, Shi H, et al. Conversion of the sagittal functional safe zone to the coronal plane using a mathematical algorithm: the reason for failure of the Lewinnek safe zone. J Bone Joint Surg Am. 2022. https://doi.org/10.2106/JBJS.21.00840.

  37. Uemura K, Takao M, Hamada H, Sakai T, Sugano N. Change in axial rotation of the femur in the resting supine position following Total hip arthroplasty. Artif Organs. 2018;42:290–6.

    Article  Google Scholar 

  38. Uemura K, Takao M, Sakai T, Nishii T, Sugano N. The validity of using the posterior condylar line as a rotational reference for the femur. J Arthroplast. 2016;31:302–6.

    Article  Google Scholar 

  39. Uemura K, Takao M, Otake Y, Koyama K, Yokota F, Hamada H, et al. Can anatomic measurements of stem Anteversion angle be considered as the functional Anteversion angle? J Arthroplast. 2018;33:595–600.

    Article  Google Scholar 

  40. Dubousset J, Charpak G, Skalli W, Kalifa G, Lazennec JY. EOS stereo-radiography system: whole-body simultaneous anteroposterior and lateral radiographs with very low radiation dose. Rev Chir Orthop Reparatrice Appar Mot. 2007;93:141–3.

    Article  CAS  Google Scholar 

  41. Ben Abdennebi A, Aubry S, Ounalli L, Fayache MS, Delabrousse E, Petegnief Y. Comparative dose levels between CT-scanner and slot-scanning device (EOS system) in pregnant women pelvimetry. Phys Med. 2017;33:77–86.

    Article  CAS  Google Scholar 

  42. Dorr LD, Maheshwari AV, Long WT, Wan Z, Sirianni LE. Early pain relief and function after posterior minimally invasive and conventional total hip arthroplasty. A prospective, randomized, blinded study. J Bone Joint Surg Am. 2007;89:1153.

    Article  Google Scholar 

Download references

Acknowledgements

None.

Funding

The institution of one or more of the authors (HT) has received, during the study period, funding from the National Science Foundation of China (grant number 82002372 and grant number 52035012), Beijing natural science foundation (grant number L212008), the Beijing Municipal Administration of Hospitals Youth Program (grant number QML20210401), and the Beijing Jishuitan Hospital Incubation Program (grant number YGQ-202106).

Author information

Authors and Affiliations

Authors

Contributions

All authors were involved in the drafting of this article or the critical revising of the important intellectual content, and all authors approved the final version to be published. All authors had full access to all the data in the study and were responsible for the integrity of the data and the accuracy of data analysis. Study conception and design: Zhuyi Ma, Hao Tang, Yixin Zhou, and Dejin Yang; Acquisition of data: Zhuyi Ma and Siyuan Wang; Analysis and interpretation of data: Zhuyi Ma and Shaoyi Guo.

Corresponding author

Correspondence to Yixin Zhou.

Ethics declarations

Ethics approval and consent to participate

Implementation of this study was approved by the Institutional Review Board of Beijing Jishuitan Hospital, the informed consent was waived by the IRB considering the retrospective nature of the study. All methods were performed in accordance with the Declaration of Helsinki.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ma, Z., Tang, H., Zhou, Y. et al. Assessing component orientation of total hip arthroplasty using the low-dose bi-planar radiographs. BMC Musculoskelet Disord 23, 886 (2022). https://doi.org/10.1186/s12891-022-05835-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-022-05835-3

Keywords