Minimally invasive transiliac anatomical locking plate for posterior pelvic ring injury: a retrospective analysis of clinical outcomes and radiographic parameters for the gull wing plate

Abstract

Background

Posterior pelvic ring injuries are challenging for surgeons to treat adequately due to difficulties with reduction and stabilization. Surgical intervention is a beneficial option to protect neurological structures and provide sufficient stability for early mobilization. The gull wing plate (GWP) is a pre-contoured anatomical locking plate with six screws, and its design is unique among posterior transiliac tension-band plates. The purpose of this study was to investigate clinical results of the GWP.

Methods

Patients who had an unstable posterior pelvic ring injury and underwent internal fixation with GWP were retrospectively analyzed at a trauma center. Demographic data, fracture type, perioperative data, and radiological evaluation with computed tomography (CT) were collected. Clinical outcomes were graded using the functional independence measure (FIM) and Majeed outcome score.

Results

Twenty-six patients were enrolled (mean age, 54 years), and the mean follow-up period was 23 months. The mean Injury Severity Score was 24 points, and internal fixation was performed 6.6 days post-trauma. CT evaluation showed the lateral surface angle of the uninjured ilium was approximately 68°. The GWP pre-contoured anatomical design closely matched this angle. The mean FIM and Majeed score were 119 and 76 points, respectively, which were graded as excellent (n = 14), good (n = 9), or fair (n = 3).

Conclusions

With the retrospective single-center data available, the GWP seems to be a minimally-invasive alternative, provides reliable stability of the posterior pelvic ring and allows for rehabilitation within normal ranges.

Peer Review reports

Background

Posterior pelvic ring disruptions are commonly caused by sacral fractures or iliosacral joint dislocation. To avoid long-term bedrest and complications, surgical intervention is critical to achieve sufficient stability for early mobilization [1]. However, these injuries are still challenging for surgeons to treat adequately due to difficulty with reduction and stabilization. Various methods for internal fixation have been reported, such as percutaneous iliosacral screws, transiliac bars, spinopelvic instrumentation, sacral plates, or transiliac plates [2].

Percutaneous iliosacral screws seem to be the least invasive and the most promising method. However, the surgery is technically demanding, and complications such as iatrogenic neurovascular injury and postoperative screw loosening cannot be overlooked [3,4,5,6]. Minimally invasive transiliac tension-band plates may be an alternative because of their safety and mechanical stability. The gull wing plate (GWP; OMIC Corporation, Shiga, Japan) is a pre-contoured anatomical locking plate with two cancellous screws (φ6.5 mm) and four locking screws (φ5.0 mm), and it is widely accepted and used as a posterior transiliac plate in Japan (Fig. 1). Clinical outcomes of posterior transiliac plates with a similar design have not been reported. The purpose of this study was to investigate our results using the GWP.

Fig. 1
figure 1

Characteristics of the gull wing plate. a The GWP design takes into account the bilateral ilia, S1 spinal process, and paravertebral muscle. The angle between the posterior surface and the bilateral wings is 70°. Two longer cancellous screws (φ6.5 mm; gold) toward the AIIS and four tri-cortical locking screws (φ5.0 mm; silver) penetrating the iliosacral joint can be inserted. Scale bar: 30 mm. b 3D reconstruction of the postoperative CT scan from the posterior side. Bilateral cancellous bone screws are inserted as far as possible (yellow arrows). GWP, gull wing plate; AIIS, anterior inferior iliac spine; CT, computed tomography

Methods

The study protocol and research were performed in accordance with the Ethics Committee at our institution. Between July 2018 and August 2021, patients who had a posterior pelvic ring injury and underwent GWP were enrolled into this retrospective study. Patients who required posterior spinal instrumentation for vertically displaced or bilateral severe sacroiliac joint injuries were excluded.

Surgical technique and postoperative protocol

Internal fixation with the GWP was performed as reported precisely in the previous study [7]. Briefly, before surgery, the optimal plate size can be planned with reference to the distance between bilateral posterior superior iliac spine (PSIS) on the axial computed tomography (CT) slice. Fixation of an unstable anterior pelvic ring is performed in the supine position, if necessary. In the prone position, approximately 5-cm longitudinal incisions are designed at the bilateral PSIS at the level of the first sacral vertebra (S1). After creating a 1.5 cm × 1.5 cm bone groove at the top of the PSIS, the paravertebral muscle (PVM) is elevated from the posterior surface of S1, and the spinal process is resected to create a tunnel connecting the bilateral incision. If reduction for a rotation deformity is necessary, a Schanz pin can be inserted from the PSIS, and the displaced ilia can be reduced by tilting that pin medially or laterally using the joy stick technique. For reducing a vertical deformity, a Schanz screw at the femoral neck or Kirschner wire direct traction from the distal femoral can be useful. The selected plate is inserted beneath the PVM, and bilateral cancellous bone screws (φ6.5 mm) are inserted as far as possible towards the anterior inferior iliac spine (AIIS). After tri-cortical drilling through locking screw sleeves at the plate wings, locking screws (φ5.0 mm) of a selected length are inserted using the torque lenti driver. The incised common fascia is completely repaired to cover the plate and reduce postoperative tenderness. For the postoperative rehabilitation protocol, all patients were mobilized under physiotherapeutic supervision with half weight bearing from post-operative days (POD) 1 to 13 and full weight bearing after 14 POD as tolerated.

Case example

A female in her twenties was involved in a motor vehicle accident (Fig. 2). The patient had a pelvic ring injury (AO/OTA 61B2), right sacral fracture (Denis zone I), and bilateral superior and inferior pubic rami. She was initially managed with transcatheter arterial embolization (TAE) and pelvic external fixator at iliac crests. Internal fixation with a gull wing plate was performed on post-accident day 3. The operation time was 80 min, with an intraoperative blood loss of 20 ml. The patient had an excellent pelvic outcome (Majeed score: 87 points) at 35 months follow-up.

Fig. 2
figure 2

Case example. a Plain AP radiograph at the emergency room. b 3D reconstruction of the CT scan (AP view). c 3D reconstruction of the CT scan (inlet view). d 3D reconstruction of the CT scan (outlet view). e Postoperative radiograph (AP view). f Postoperative radiograph (inlet view). g Postoperative radiograph (outlet view). h Plain AP radiograph at 26 months postoperatively (after implant removal)

Data collection

Patient demographic data included the following: age, gender, trauma mechanism, Injury Severity Score (ISS), associated lesion other than pelvis, AO/OTA classification, information about the ramus and sacral fracture (Denis classification) [8], preoperative neurological deficit (Gibbon classification) [9], the presence of the preoperative external fixator and TAE, trauma-to-surgery time, admission period, and follow-up period.

Clinical outcome measurement

The following intraoperative information was recorded: surgical time (only for GWP, excluding fixation for anterior component), blood loss, plate size, iliac screw length, and complementary fixation for anterior component. Functional independence measure (FIM) at admission, discharge, and final follow-up were evaluated [10]. Majeed score at the final follow-up was recorded [11]. Postoperative complications and implant removal were investigated.

Radiographic parameters

Pre- and post- operative CT evaluation (Fig. 3) was performed to ensure that the wings of GWP coincided with the lateral surface of the ilium. The axial slice where the S1 foramen ends at the anterior surface of the sacrum was selected. The horizontal plane of the sacrum was defined on the basis of the spinal canal and neural foramina (white dotted line). The lateral surface of the ilium was defined at the deep portion of the PSIS (yellow line). The angle between both lines was measured bilaterally. Next, the absolute value of difference between lateral surface angles of injured and uninjured ilium were calculated. Pre- and post-operative values were compared using the Student’s t-test. Values of p < 0.05 were considered to be statistically significant.

Fig. 3
figure 3

a Pre- and b Post-operative CT evaluation. The angle between the horizontal plane of the sacrum (white dotted line) and the lateral surface of the ilium (yellow line) were measured bilaterally

Results

During the study period, 26 patients underwent GWP for posterior pelvic ring injury. Patient demographics are shown in Table 1. Eighteen men and eight women were enrolled, with a mean age of 54 years. The mean follow-up period was 23 months. The most common mechanism of injury was a traffic accident in 16 patients. The mean ISS was 24 points. The AO/OTA classification showed the following results: B1, six patients; B2, nine patients; B3, four patients; C1, two patients; and C2, five patients. Eighteen patients had a displaced anterior arch. Sacral fractures were classified as follows: Denis 1, six patients; Denis 2, 14 patients; and Denis 3, three patients. Although neurological deficits were recorded at the initial examination in six patients, further assessment was impossible in four patients due to other injuries, such as cervical cord injury, direct blow to the common peroneal nerve, and sciatic nerve injury followed by displaced distal femur fracture. Almost 70% of patients underwent external fixator and TAE on the day of injury. The mean timing of surgery was 6.6 days after trauma. The mean admission period was 34 days.

Table 1 Patient demographics

Clinical results are shown in Table 2. The mean surgical time was 92 min, and the mean blood loss was 162 mL. The mid-sized (115 mm) GWP was most commonly selected (16 patients). The mean iliac screw length was 72 mm. Nine patients underwent fixation of the anterior pelvic ring before GWP. CT evaluation showed that the lateral surface angle of the uninjured ilium was approximately 68°, while that of the injured ilium was 65.9° ± 6.9° preoperatively and 67.9° ± 4.5° postoperatively. In terms of the absolute value of difference between injured and uninjured ilium, post-operative values (2.7° ± 3.1°) were significantly smaller than pre-operative values (6.3° ± 3.7°) (p < 0.01) (Fig. 4). No infections were recorded. One locking screw came out probably due to interference with the iliac screw. In accordance with the patients’ wishes, implant removal was performed in nine cases. The FIM improved over time, and it was 119 points at the final follow-up. The mean Majeed score at the final follow-up was 76 points, which were graded as excellent in 14 patients, good in nine patients, and fair in three patients. All three patients who were graded as fair, had associated spinal cord injury or neurological deficits and their symptoms were treated with medication or a prosthetic appliance.

Table 2 Clinical results
Fig. 4
figure 4

The absolute value of difference between lateral surface angles of injured and uninjured ilium. Post-operative values were demonstrated to be smaller than pre-operative values (p < 0.01)

Discussion

Management of a pelvic ring injury poses significant challenges for orthopaedic surgeons [12]. An initial X-ray or CT scan in the emergency department might not reflect the potential displacement or instability of the injured pelvic ring due to the recoil capacity of the pelvis [13]. The posterior part of the pelvic ring is the key structure in terms of hemorrhage control and load transmission from the lower limbs to the lumbar spine. The sacrum forms joints with the bilateral ilia and contains the central canal and neural foramina. Because of its thin cortex, sacral fracture is often comminuted, and this often causes neurological deficits. Perfect reduction and rigid fixation, which provide sufficient stability for early mobilization, are essential to improve the patient’s quality of life.

Among several fixation methods, the percutaneous iliosacral screw or trans-iliac screw is the first choice in most cases. These screws can be used only for a minimal displacement, but a short surgical time, less blood loss, and lower risk of soft tissue complications are attractive features. However, there are some concerns related with percutaneous screw fixation. First, as reported in most studies, it requires extensive expertise, thorough knowledge of anatomical structure around the pelvis, and careful recognition of dysmorphic sacrum. Screw malpositioning, which directly leads to serious neurologic and vascular injury, should be avoided. Matiyahu et al. demonstrated that the rate of dysmorphism affecting the iliosacral screw trajectory was 17% in their randomized multicenter study [14]. Thus, with conventional fluoroscopy methods, malposition occurred in 12% (9/73) of screws in the normal sacral group and in up to 32% (7/22) of screws in the dysmorphic sacral group [14]. Sagi et al. presented a case report involving screw malpositioning that was apparently an appropriate position on intraoperative fluoroscopy [4]. They emphasized awareness of the tangential or oblique nature of the sacral foramina and concluded that the screw should be placed in the anterior portion and not just above the foramen in the outlet view [4]. Zwingmann et al. performed a systematic review and meta-analysis, and they estimated rate of malposition to be 2.6% (1832 screws) using the conventional fluoroscopy method, and this was significantly higher than using the CT navigation method, which had a malposition rate of 0.1% (262 screws) [6]. Therefore, recent studies recommended an intraoperative CT scan or a navigation system for optimal screw positioning [6, 14]. A second concern is related to percutaneous screw fixation and its biomechanical stability and postoperative implant difficulties. A biomechanical study revealed that bilateral iliosacral screws (8 mm in diameter) were inferior to the posterior tension band plate (4.5 mm, non-locking) [15]. The rate of implant failure, such as screw loosening or loss of reduction, was 7%–13% in clinical studies [3, 5]. Kim et al. demonstrated in their systematic review and meta-analysis that implant loosening was more common in the percutaneous screw group than in the posterior plate group [16]. Third, intraoperative radiation exposure for the percutaneous screw is higher than for other surgical methods. For accurate and safe positioning of the iliosacral screw or transiliac screw, the iliosacral screw was demonstrated to require 7.6-times more radiation exposure during surgery than that of the posterior transiliac plates [17].

Currently, several researchers have re-evaluated and verified the effectiveness and safety of posterior transiliac plates [18,19,20,21,22]. Among them, large-sized plates (4.5 mm) are preferred to small-sized plates (3.5 mm), and locking compression plates seem to have replaced conventional non-locking plates [22]. We suggest that the GWP three-dimensional design provides even greater stability compared with that of other plates because two longer cancellous screws (φ6.5 mm) toward the AIIS pass through a corridor with a high bone density and four tri-cortical angular-stable screws (φ5.0 mm) fix the iliosacral joint. The wing angle is 70°, which closely matches the angle of the uninjured positional relationship between the sacrum and ilium (approximately 68°) in the current study. The pre-contoured anatomical design is beneficial because bending the thick straight plates during surgery takes time, reduces the plate strength, and breaks the screw hole locking mechanism. While Ayoub et al. and Suzuki et al. reported a relatively low rate of infection of 7.5% (3/42 cases) and 10.5% (2/19 cases), respectively, there were no infections in our study cohort [18, 19]. The GWP was slipped beneath the PVM through bilateral longitudinal 5 cm incisions, and it was only minimal invasive in the soft tissue. The wavelike form of GWP is also gentle on the PVM, PSIS, and the S1 spinal process, which is partially resected by the osteotome. Compared with the open method from a midline incision, the shorter surgical time and reduced blood loss might contribute to the low rate of postoperative infection. However, before selecting GWP, soft tissue conditions such as a degloving injury should be carefully investigated. Another advantage of GWP is less radiation exposure. Because image intensifier is necessary only when checking the trajectory of iliac screw and the tip of the locking screw, a radiolucent fracture table is not required for GWP. Direct accessibility to the posterior surface of the sacrum is also a benefit of GWP. Autologous bone grafting using the resected PSIS can be easily applied to a sacral fracture site. When nerve decompression is required, a small, medialized skin incision enables a sacral laminectomy. However, in this case, another stab incision is necessary to set the locking screw sleeve. GWP clinical outcomes are satisfactory, and equal to or better than previous studies. FIM in the current study improved over time after admission. Near full activities of daily living (FIM, 119 points) was achieved at the final follow-up visit. Our patient cohort showed a Majeed score of 76 points and 88.5% had a satisfactory (excellent and good) result, while previous studies showed 71.8–78.5 points and 72.2%–80.0% with satisfactory results [18, 20, 21]. The achieved rigidity of GWP provides sufficient stability to allow early mobilization, weightbearing, and activities of daily living. There are only two contraindications for using GWP. The first contraindication is the fracture type, such as cranially/caudally displaced fractures and bilateral severe injuries around the sacroiliac joints, which should be suitable for fixation using spinal instruments that place anchors at the lumbar spine. The second contraindication is severely damaged soft tissue, such as a Morel–Lavallée lesion. When suspected, careful evaluation and judgment is essential to avoid surgical site infection. The only shortcoming of the GWP is that it requires that the patient is in the prone position, which seems more disadvantageous for reduction than the supine position. However, if necessary, a Schanz screw from the PSIS can help to reduce the displaced ilia [23].

There are several limitations that could affect the results in the current study. First, this study is a retrospective analysis from our institution, and a control group was not included. Because of the low incidence of unstable pelvic ring injuries in our country, a randomized controlled trial is difficult to perform even at a level 1 trauma center. Further controlled clinical study with a sufficient number of patients is required. Second, selection bias in the outcome measurement may occur in our patient cohort because vertically displaced or bilateral severe sacroiliac joint injury patients who may have a poor outcome were excluded in the current study. Third, radiographic assessment of vertical deformity is missing in the current study, although a rotational deformity parameter was measured on the axial CT slice. Because GWP was not actually applied to severe vertical deformity fractures, it was difficult to precisely measure the difference between pre- and post-operative displacement.

Conclusions

The GWP represents a simple, safe, minimally invasive procedure with fewer complications and less radiation exposure. It can be used for many fracture types, achieve reliable stability of the posterior pelvic construct, and accelerate rehabilitation. Further study is necessary to investigate the long-term outcomes and compare them to those of alternative surgical techniques.

Availability of data and materials

The datasets of this study are available from the corresponding author upon reasonable request.

Abbreviations

GWP:

Gull wing plate

CT:

Computed tomography

FIM:

Functional independence measure

PSIS:

Posterior superior iliac spine

S1:

The first sacral vertebra

PVM:

Paravertebral muscle

AIIS:

Anterior inferior iliac spine

POD:

Post-operative days

ISS:

Injury severity score

AO/OTA:

AO Foundation/Orthopaedic Trauma Association

TAE:

Transcatheter arterial embolization

References

  1. Hoch A, Ozkurtul O, Pieroh P, Josten C, Bohme J. Outcome and 2-year survival rate in elderly patients with lateral compression fractures of the pelvis. Geriatr Orthop Surg Rehabil. 2017;8:3–9.

    Article  PubMed  Google Scholar 

  2. Tonetti J. Management of recent unstable fractures of the pelvic ring. An update Conference supported by the Club Bassin Cotyle (Pelvis-Acetabulum Club). Orthop Traumatol Surg Res. 2013;99S:S77-86.

    Article  Google Scholar 

  3. Routt C, Peter S. Closed reduction and percutaneous skeletal fixation of sacral fractures. Clin Orthop Relat Res. 1996;329:121–8.

    Article  Google Scholar 

  4. Sagi HC, Lindvall EM. Inadvertent intraforaminal iliosacral screw placement despite apparent appropriate positioning on intraoperative fluoroscopy. J Orthop Trauma. 2005;19:130–3.

    Article  CAS  PubMed  Google Scholar 

  5. Schweitzer D, Zylberberg A, Cordova M, Gonzalez J. Closed reduction and iliosacral percutaneous fixation of unstable pelvic ring fractures. Injury. 2008;39:869–74.

    Article  PubMed  Google Scholar 

  6. Zwingmann J, Hauschild O, Bode G, Sudkamp NP, Schmal H. Malposition and revision rates of different imaging modalities for percutaneous iliosacral screw xation following pelvic fractures: a systematic review and meta-analysis. Arch Orthop Trauma Surg. 2013;133:1257–65.

    Article  PubMed  Google Scholar 

  7. Mitsuzawa S, Kusakabe K, Matsuda S. Minimally invasive transiliac anatomical locking plate for posterior pelvic ring injury: a technical trick of the gull wing plate. J Clin Orthop Trauma :doi. 2022. https://doi.org/10.1016/j.jcot.2022.101991.

    Article  Google Scholar 

  8. Denis F, Davis S, Comfort T. Sacral fractures: an important problem. Retrospective analysis of 236 cases. Clin Orthop Relat Res. 1988;227:67–81.

    Article  CAS  PubMed  Google Scholar 

  9. Gibbons KJ, Soloniuk DS, Razack N. Neurological injury and patterns of sacral fractures. J Neurosurg. 1990;72:889–93.

    Article  CAS  PubMed  Google Scholar 

  10. Keith RA, Granger CV, Hamilton BB, Sherwin FS. The functional independence measure: a new tool for rehabilitation. Adv Clin Rehabil. 1987;1:6–18.

    CAS  PubMed  Google Scholar 

  11. Majeed SA. Grading the outcome of pelvic fractures. J Bone Joint Surg. 1989;71(2):304–6.

    Article  CAS  Google Scholar 

  12. Halawi MJ. Pelvic ring injuries: Surgical management and long-term outcomes. J Clin Orthop Trauma. 2016;7:1–6.

    Article  PubMed  Google Scholar 

  13. Grewal IS, Mir HR. Assessment of instability in type B pelvic ring fractures. J Clin Orthop Trauma. 2020;11:1009–15.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Matityahu A, Kahler D, Krettek C, Stockle U, Grutzner PA, Messmer P, et al. Three-dimensional navigation is more accurate than two-dimensional navigation or conventional fluoroscopy for percutaneous sacroiliac screw fixation in the dysmorphic sacrum: a randomized multicenter study. J Orthop Trauma. 2014;28:707–10.

    Article  PubMed  Google Scholar 

  15. Berber O, Amis AA, Day AC. Biomechanical testing of a concept of posterior pelvic reconstruction in rotationally and vertically unstable fractures. J Bone Joint Surg Br. 2011;93(2):237–44.

    Article  CAS  PubMed  Google Scholar 

  16. Kim CH, Kim JW. Plate versus sacroiliac screw fixation for treating posterior pelvic ring fracture: a systematic review and meta-analysis. Injury. 2020;51:2259–66.

    Article  PubMed  Google Scholar 

  17. Chen H, Liu G, Fei J, Yi X, Pan J, Ou S, et al. Treatment of unstable posterior pelvic ring fracture with percutaneous reconstruction plate and percutaneous sacroiliac screws: a comparative study. J Orthop Sci. 2012;17:580–7.

    Article  PubMed  Google Scholar 

  18. Suzuki T, Hak DJ, Ziran BH, Adams SA, Stahel PF, Morgan SJ, et al. Outcome and complications of posterior transiliac plating for vertically unstable sacral fractures. Injury. 2009;40:405–9.

    Article  PubMed  Google Scholar 

  19. Ayoub MA, Gad HM, Seleem OA. Standalone percutaneous transiliac plating of vertically unstable sacral fractures: outcomes, complications, and recommendations. Eur Spine J. 2016;25:1153–62.

    Article  PubMed  Google Scholar 

  20. Khaleel VM, Pushpasekaran N, Prabhu N, Pandiyan A, Koshy GM. Posterior tension band plate osteosynthesis for unstable sacral fractures: a preliminary study. J Clin Orthop Trauma. 2019;10:S106-111.

    Article  PubMed  PubMed Central  Google Scholar 

  21. Boudissa M, Saad M, Kerschbaumer G, Ruatti S, Tonetti J. Posterior transiliac plating in vertically unstable sacral fracture. Orthop Traumatol Surg Res. 2020;106:85–8.

    Article  PubMed  Google Scholar 

  22. Schmerwitz IU, Jungebluth P, Lehmann W, Hockertz TJ. Minimally invasive posterior locked compression plate osteosynthesis shows excellent results in elderly patients with fragility fractures of the pelvis. Eur J Trauma Emerg Surg. 2021;47:37–45.

    Article  PubMed  Google Scholar 

  23. Martin MP, Rojas D, Mauffrey C. Reduction and temporary stabilization of Tile C pelvic ring injuries using a posteriorly based external xation system. Eur J Orthop Surg Traumatol. 2018;28:893–8.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

No funding.

Author information

Authors and Affiliations

Authors

Contributions

SMi conceived the research, performed data collection, and wrote the draft of the manuscript. KK, SN, TM, TY and SMa substantively revised it, and gave final approval of the submitted manuscript. The author(s) read and approved the final manuscript.

Corresponding author

Correspondence to Sadaki Mitsuzawa.

Ethics declarations

Ethics approval and consent to participate

This study was performed in accordance with the Declaration of Helsinki (as revised in 2013) and approved by the Ethics Committee at our institution (RGMC 2021–022). Informed consent was obtained from all patients/parents before inclusion in the study and anonymous publication of the results.

Consent for publication

Not applicable.

Competing interests

The authors have no conflicts of interest to disclose.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Mitsuzawa, S., Kusakabe, K., Nakao, S. et al. Minimally invasive transiliac anatomical locking plate for posterior pelvic ring injury: a retrospective analysis of clinical outcomes and radiographic parameters for the gull wing plate. BMC Musculoskelet Disord 23, 880 (2022). https://doi.org/10.1186/s12891-022-05829-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-022-05829-1