Long-term survival of hybrid total hip replacement for prior failed proximal femoral nail antirotation: a retrospective study with a median 10-year follow-up

Abstract

Background

Hybrid total hip replacement (THR) is commonly used in the management of proximal femur fractures in elderly individuals. However, in the context of the revision, the literature on hybrid THR is limited, and differences in the long-term survival outcomes reported in the literature are obvious. This retrospective study aimed to evaluate the long-term survival of hybrid THR for failed proximal femoral nail antirotation (PFNA) in elderly individuals aged ≥ 75 years.

Methods

An observational cohort of 227 consecutive individuals aged ≥ 75 years who experienced hybrid THRs following prior primary PFNAs was retrospectively identified from the Joint Surgery Centre, the First Affiliated Hospital, Sun Yat-sen University. Implant survival was estimated using the Kaplan–Meier method. The primary end point was the implant survivorship calculated using the Kaplan–Meier method with revision for any reason as the end point; secondary end points were the function score measured using the modified Harris Hip Score (mHHS) and the incidence of main orthopaedic complications.

Results

In total, 118 individuals (118 THRs) were assessed as available. The median follow-up was 10 (3–11) years. The 10-year survivorship with revision for any reason as the endpoint was 0.914 (95% confidence interval [CI], 0.843–0.960). The most common indication for revision was aseptic loosening (70.0%), followed by periprosthetic fracture (30.0%). At the final follow-up, the median functional score was 83.6 (79.0–94.0). Among the 118 patients included in this study, 16 experienced 26 implant-related complications. The overall incidence of key orthopaedic complications was 13.5% (16/118).

Conclusion

For patients aged ≥ 75 years old with prior failed PFNAs, hybrid THR may yield satisfactory long-term survival, with good functional outcomes and a low rate of key orthopaedic complications.

Peer Review reports

Background

The challenge of dealing with failed proximal femoral nail antirotation (PFNA) is a well-established climb and continues [1,2,3]. As longevity has increased in China, the incidence of failed PFNA is bound to increase, especially in individuals characterised by advancing age, osteoporosis, and multiple comorbidities, which already poses a great threat to the patient's quality of life, despite the lack of Chinese-specific data [2, 3]. Accordingly, revision procedures for prior failed PFNA may present an increasing trend [1]. Recently, several studies [1,2,3] have shown a growing rate of revisions for failed PFNA. Disappointingly, however, options including hybrid, cemented, and uncemented total hip replacements (THRs) used to revise a failed PFNA remain controversial [3,4,5]. Furthermore, different THRs used in this context tend to result in specific complications [6]. Advocates of hybrid THR perceive benefits regarding function score and orthopaedic complications when compared to cemented or uncemented THR [7,8,9]. Hybrid THR increases the early stability of the femoral component after revision surgery and has frequently been used in an attempt to increase early weight bearing, with apparently encouraging results [7, 10]. Nevertheless, the cemented femoral component of hybrid THR may have a great destructive effect on the surrounding bone tissue [11]. Furthermore, there are growing concerns that cemented femoral components have greater associated cement-related complications than uncemented femoral components and could increase the risk of cement-induced osteolysis, which frequently results in prosthesis failure [7, 12]. Although failed PFNA converted to THR has been a recognised treatment strategy, the inappropriate decision as to which type of THR (hybrid, cemented, and uncemented) is the optimum treatment in elderly individuals may lead to a marked difference in treatment outcomes [13, 14].

Currently, there is an ongoing debate on the utilisation of hybrid vs. uncemented THRs [7, 15, 16]. With the development of cement fixation techniques, there has been a growing use of hybrid THR in China, particularly in individuals aged ≥ 75 years. However, concerns related to the long-term implant survival of hybrid THRs remain [1, 3]. Furthermore, there remain no definite long-term follow-up results for patients in China experiencing a hybrid THR following prior PFNA failures that are sufficient to show the superiority of hybrid THR, especially for individuals aged ≥ 75 years. Hence, we executed this retrospective review to assess the long-term outcomes of individuals aged ≥ 75 years old with prior PFNA failures who underwent a hybrid THR.

Methods

Study population

From August 2010 to December 2019, consecutive elderly individuals aged ≥ 75 years old experiencing a hybrid THR following prior failed PFNA were retrospectively reviewed from three joint trauma centres. The median volume of THR revisions per year at each centre was 20 procedures (range, 11–32). The product details of PFNA and hybrid THR are presented in Table 1. The type of and reason for revision PFNA were identified based on electronic medical records. The comorbidities of the included patients were assessed using the Charlson comorbidity index (CCI). Key inclusion criteria consisted of elderly individuals aged ≥ 75 years and individuals experiencing initial PFNA fixation, followed by hybrid THR revision. Key exclusion criteria were as follows: lacking demographic data (i.e., diagnosis, type of fixation, implant details); lack of follow-up data; hip deformity; loss of independent athletic ability; previous contralateral intertrochanteric fractures; disorders of the nervous system of the lower extremities; advanced tumours; active infectious diseases (i.e., sepsis, interstitial pneumonia, osteomyelitis, and meningitis); mental abnormalities (i.e., schizophrenia, mental retardation, severe depression, attention deficit hyperactivity disorder); long-term dialysis or drug therapy (i.e., renal failure, immunosuppressants, antithyroid drugs); and failure to abide by the follow-up plan.

Table 1 Product details of PFNA and hybrid THR

Surgical procedures

We used the previous surgical incision to remove the PFNA device. After the removal of the PFNA device, we routinely measured the length between the lesser trochanter and the distal tip of the main nail after the removal of the PFNA device. The length of the stem was greater than or equal to the length of the main nail measured previously, avoiding the phenomenon of periprosthetic fracture associated with local stress concentration of the stem. All conversion procedures were executed through a lateral approach. A uncemented cup was inserted in accordance with manufacturer’s instructions. We performed two truncations of the femoral neck at the junction of the femoral head and neck and at the base of the femoral neck, removed the glenoid labrum, and polished the acetabulum until the bone surface oozed blood. With a valgus angle of 40° to 45° and an anteversion angle of 15°, the acetabulum cup was punched into the acetabulum, and the liner was installed.

Cemented stems were inserted using third-generation cementing techniques. During stem insertion, subperiosteal osteolysis, resorption of the femoral calcar, proximal femoral bone defect, and disruption of the integrity of the greater trochanter are of concern. There are the following technical points: long-stem prosthesis is selected; wedge-shaped bone masses of the femoral neck can be used to reconstruct the integrity of the greater trochanter; the cancellous bone of the femoral head and the cortical bone of the femoral neck are made into 2–3 mm bone fragments for bone grafting; cortical defects can be bound with steel cables and covered with metal mesh; the proximal femoral medullary cavity should be thoroughly cleaned; long guide needle is located in the center of the medullary cavity in the distal femur and near the lateral cortical bone in the proximal femur; the prepared 2–3 mm bone fragments were implanted into the proximal femoral medullary cavity and the residual lateral screw holes after PFNA removal (due to the coverage of periosteum or musculoaponeurotic layer, the medial screw holes rarely had cement leakage).

Outcomes and variables

The primary end point was the implant survivorship calculated using the Kaplan–Meier method with revision for any reason as the endpoint. Revision was defined as exchange or removal of full or partial implants including the acetabulum, cup and liner for any reason [17], irrespective of component adjustment. Each revision was symptomatic. Indications for conversion to hybrid THR involved instability, mechanical failure, and both. Instability was defined as screw loosening, unacceptable displacement of the fracture site, nonunion, tendency of dislocation. The secondary endpoints were the function score measured using the modified Harris Hip Score (mHHS: range, 0–100, with higher scores indicating better function) and the incidence of main orthopaedic complications (aseptic loosening, dislocation, and periprosthetic fracture). mHHS = 70 was regarded as the threshold for failure [18]. The cemented stem was classified as definite loosening if radiolucent lines with distinct migration measured on two consecutive radiographs were present [19]. Loosening of the acetabular component was defined as a continuous radiolucent line greater than 2 mm in width on both the anteroposterior and the lateral radiographs compared with the immediate postoperative images [20]. Image data collected at every follow-up were reviewed centrally. At the end of the study, the secondary endpoints were confirmed by the two coauthors (WY and MX). Main orthopaedic complications and deaths at any time during the follow-up period were recorded. Patients underwent evaluation on the day following revision PFNA and continued until the occurrence of revision THR, death, or study deadline, whichever came first. Patients were followed up at an interval of 2–3 months, either in person or by telephone.

Statistical analysis

The survival for hybrid THRs used in this study was calculated using the Kaplan–Meier survival with 95% confidence interval (CI). The primary endpoint analysis using death as a competing risk was conducted with revision THR for any reason as an endpoint. Considering that the functional outcome data calculated using the mHHS were normally distributed during follow-up, the preoperative and postoperative mHHS were compared using a paired Student’s t test. The alpha level was set at 0.05 when comparing differences between function scores. Data analyses were performed using SAS 9.4 (SAS Institute).

Results

In total, 227 elderly individuals aged ≥ 75 years old were included in this study. Among them, 109 individuals were identified as unavailable according to our inclusion criteria, and 118 individuals (118 THRs) were assessed as available, as detailed in Fig. 1. Table 2 shows the baseline data. The age of the patients was primarily concentrated in the range of 75–80 years for 76.3%. In this cohort, there were no noteworthy distinctions in the sex of the patients. The most common mechanism of injury was falling (61.0%), followed by tamp (25.4%) and traffic (13.6%). Most patients had a medium CCI at the time of prosthesis revision, accounting for 58.5%. Indications for conversion to hybrid THR were primarily attributed to instability, accounting for 63.6%. The majority of individuals (57.6%) had an ASA status of 2. The median mHHS prior to conversion was 55.0 (46.7–68.9).

Fig. 1
figure 1

Flow diagram showing the method for the identification of study subjects to estimate the long-term survival of hybrid THRs following prior PFNA failure

Table 2 Patient characteristics at baseline

Primary outcome

The median follow-up was 10 (3–11) years. Figure 2 demonstrates the Kaplan–Meier survival curve with revision THR executed for any reason as the end point. The 5-year survivorship with revision for any reason as the endpoint was 0.983 (95% CI, 0.915–0.992). The 8-year survivorship was 0.949 (95% CI, 0.882–0.975). The 10-year survivorship was 0.914 (95% CI, 0.843–0.960). Of 118 hybrid THRs, 98 (83.1%) were functioning at the end of the study. The most common indication for revision was aseptic loosening (70.0%), followed by periprosthetic fracture (30.0%). In this study, dislocations, even frequent dislocations, did not involve revision, which did not affect the prosthesis survival rate.

Fig. 2
figure 2

Kaplan–Meier survival curve with revision THR for any reason as the endpoint

Secondary outcomes

Figure 3 provides the variation trend of mHHS after conversion to hybrid THRs. In total, 108 individuals underwent functional outcome assessment, and 10 individuals did not experience a final functional outcome assessment because they underwent revision THR surgery. At the final follow-up, the median function score was 83.6 (79.0–94.0). From the first follow-up after the revision until the 4th year, the curve basically showed an upwards trend. From the 5th year to the 7th year, the curve basically exhibited a horizontal trend. From the 7th year until the final follow-up, the curve presented a downwards trend.

Fig. 3
figure 3

The variation trend of mHHS after conversion to hybrid THRs

Table 3 shows key implant-related complications. At the final follow-up, 10 (8.4%) individuals underwent a conversion of hybrid THR to revision surgery. The most frequent revision was stem revision (5.1%), followed by acetabular revision (2.5%) and both (0.8%). The most common key complication related to hybrid THR was aseptic loosening (9.3%), followed by dislocation (6.8%) and periprosthetic fracture (5.9%). Figure 4 shows the Kaplan–Meier survival curve with aseptic loosening as the end point, and Fig. 5 shows the Kaplan–Meier survival curve with periprosthetic fractures as the end point. Among the 118 patients included in this study, 16 experienced 26 implant-related complications. The overall incidence of key orthopaedic complications was 13.5% (16/118).

Table 3 Key complications related to hybrid THR
Fig. 4
figure 4

Kaplan–Meier survival curve with aseptic loosening as the endpoint

Fig. 5
figure 5

Kaplan–Meier survival curve with periprosthetic fracture as the endpoint

Discussion

Among individuals aged ≥ 75 years old with prior failed PFNA, we found promising long-term results using hybrid THR in revision PFNA surgery. The rate of 10-year survival (based on the need for revision) was 91.4%, which was comparable to previous studies [21, 22]. Our findings may extend those of a limited body of previous studies that have shown a growing rate of 10-year survival in the application of hybrid THR [23]. This increase may be driven by a combination of factors, comprising improvements in bone cement technology and an understanding of the indications for cementing arthroplasty [6, 11]. Hybrid THR may result in a low incidence of studied hip-related complications and substantially good functional outcomes.

Consistent with recent reports [24,25,26], we did not detect marked distinctions in the 5-year survival rate. The lack of distinctions was most likely attributable to similar follow-up times. However, at the 10-year follow-up, the survival rate was slightly lower than that in previous reports [7, 27,28,29]. This may be attributed to the fact that most of our subjects had high CCI scores and were afflicted with bone and soft-tissue abnormalities attributed to prior failed PFNA surgery. A high CCI score may reflect a combination of factors that may increase competing risk, such as death [3, 6]. Furthermore, poor survival of hybrid THR was related in part to differences in prosthetic design and bone cement technology used [1, 5, 6, 11]. Although our follow-up span is long, the present study may, to some extent, confirm the advantages of hybrid THR. To date, few studies on converting failed PFNAs provide effective consensus recommendations for reducing or avoiding mechanical failures [30, 31]. Lack of the design and material characteristics of THR implants can lead to remarkable distinctions in comparisons between studies [32, 33]. The long-term outcomes of hybrid THR for elderly individuals remain controversial [21, 27].

It is possible that the most concerning complication is aseptic loosening in the revision PFNA cohort [2, 3]. It is not clear what triggers the aseptic loosening of cement-fixed femoral components, which may involve enlargement of the medullary cavity, metal fatigue, aseptic inflammation at the cement–bone interface, or changes in compressive stress in the proximal femoral cortex [1, 2, 5, 29]. Studies [5, 7, 11, 27] have shown that the acetabulum components and the femoral components differ greatly in the mechanism by which aseptic loosening occurs. The loosening of acetabulum components fixed with bone cement is mostly related to biological effects [3, 21, 27]. Wear debris related to bone cement and polyethylene activate macrophages, release a large number of cytokines, and ultimately result in cytokine-induced osteolysis, thus affecting the stability of acetabular components [34, 35]. In contrast, the loosening of cement-fixed femoral components is mostly related to mechanical effects [36]. Unevenly filled cement and high porosity may affect the riveting force between bone cement and bone [37, 38]. Currently, numerous studies [7, 13, 15, 23] have made hybrid THR the preferred mode of arthroplasty. The femoral components are immobilised with bone cement, irrespective of the effect of enlargement of the medullary cavity over time, which relieves postoperative thigh pain and effectively combats the early sinking and loosening of prostheses, while the acetabulum is fixed with uncemented prosthesis, which can reduce the rate of postoperative loosening [5, 9, 11, 14, 34].

Not all THR prostheses function properly for a long time, and not all individuals benefit from THR revision [1,2,3]. Any effort to address the increasing incidence of main orthopaedic complications raises the issues of biological characteristics and limited indications of hybrid THR [9, 11, 14, 15]. Implementing a national survey on indications of hybrid THR can be a relatively tricky issue, although there have been encouraging recommendations recently to advocate that large-scale research should be executed to improve this dilemma [12, 13, 16]. However, it is difficult to determine to what extent the growth in THR applications reflects an increase in THR indications or reflects a clinician's personal preference, and current indications for THR may still depend on the clinician's experience [7, 17, 28].

Several limitations should be acknowledged. First, the retrospective design has inherent limitations, such as susceptibility to selection bias and recall bias, the potential for changes in the definition of symptoms and diseases, and the relatively limited data collected (i.e., prefracture functional status). The deviation related to the retrospective design is large and rough, and often due to incomplete data records, it is impossible to explore and discover some relevant factors in depth, or the data records are not sufficiently accurate, resulting in an increase in the error of the obtained data. The present physical and mental state of study subjects may affect the authenticity and accuracy of past data reports, which may be an important defect of retrospective studies. Second, after 5 years of follow-up, the incidence of key orthopaedic complications increased significantly, which is likely to reflect the measurement bias caused by the stricter definition of the main complication variable, rather than a true increase. However, the incidence of aseptic loosening is likely to increase, which may be the result of an interaction between weight gain attributed to the patient's reduced exercise and the biological properties of bone cement. Third, there was at least one competing risk in this study. The use of the Kaplan–Meier method to estimate the risk of cumulative revision may lead to an overestimation of implant survival. Fourth, the nature of the observations makes it impossible for our study to draw reliable causality. Nonetheless, the current study provides an estimate of the long-term survival and complications of hybrid THR, which may be necessary for surgeons' decision-making.

Conclusion

Hybrid THR may have a remarkable statistical benefit on long-term prosthesis survival, which may reflect the durability of hybrid THR implants, with appropriate mHHS and a low rate of main implant-related complications in the revision setting. We found that the increase in major orthopaedic complications over time may reflect a superposition of a number of underlying factors, including a deterioration in bone quality and an increasing prevalence of cement-related complications. Our findings have contributed to some extent to resolving the debate about the decision-making process in individuals aged ≥ 75 years.

Availability of data and materials

The datasets generated during and analyzed during the current study are not publicly available due to the protection of patient privacy but are available from the corresponding author on reasonable request.

Abbreviations

THR:

Total hip replacement

PFNA:

Proximal femoral nail anti-rotation

mHHS:

Modified Harris Hip Score

CIs:

Confidence intervals

CCI:

Charlson comorbidity index

References

  1. Shi WB, Zhang YD, Xu YK, Zeng XS, Fu HJ, Yu WG. Conversion of failed proximal femoral nail antirotation to uncemented or cemented femoral component fixation: a multicentre retrospective study with a median 10-year follow-up. BMC Musculoskelet Disord. 2022;23(1):375.

    Article  CAS  Google Scholar 

  2. Yu WG, Han XL, Chen WL, Mao S, Zhao MD, Zhang XC, Han GW, Ye JX, Chen MJ, Zhuang JT. Conversion from a failed proximal femoral nail anti-rotation to a cemented or uncemented total hip arthroplasty device: a retrospective review of 198 hips with previous intertrochanteric femur fractures. BMC Musculoskelet Disord. 2020;21(1):791.

    Article  Google Scholar 

  3. Zeng XS, Zhan K, Zhang LL, Zeng D, Yu WG, Zhang XC, Zhao MD. Conversion to total hip arthroplasty after failed proximal femoral nail antirotations or dynamic hip screw fixations for stable intertrochanteric femur fractures: a retrospective study with a minimum follow-up of 3 years. BMC Musculoskelet Disord. 2017;18(1):38.

    Article  Google Scholar 

  4. Tetsunaga T, Fujiwara K, Endo H, Noda T, Tetsunaga T, Sato T, Shiota N, Ozaki T. Total hip arthroplasty after failed treatment of proximal femur fracture. Arch Orthop Trauma Surg. 2017;137(3):417–24.

    Article  Google Scholar 

  5. Morice A, Ducellier F, Bizot P, Orthopaedics, Traumatology Soc W. Total hip arthroplasty after failed fixation of a proximal femur fracture: Analysis of 59 cases of intra- and extra-capsular fractures. Orthop Traumat-Surg Res. 2018;104(5):681–6.

    Google Scholar 

  6. Evans JT, Blom AW, Timperley AJ, Dieppe P, Wilson MJ, Sayers A, Whitehouse MR. Factors associated with implant survival following total hip replacement surgery: A registry study of data from the National Joint Registry of England, Wales, Northern Ireland and the Isle of Man. PLoS Med. 2020;17(8):e1003291.

    Article  Google Scholar 

  7. Fowler AK, Gray AR, Gwynne-Jones DP. Hybrid fixation for total hip arthroplasty showed improved survival over cemented and uncemented fixation: a single-center survival analysis of 2156 hips at 12–18 years. J Arthroplasty. 2019;34(11):2711–7.

    Article  Google Scholar 

  8. Davis ET, Pagkalos J, Kopjar B. Polyethylene manufacturing characteristics have a major effect on the risk of revision surgery in cementless and hybrid total hip arthroplasties an analysis of the national joint registry of England, wales, northern Ireland and the isle of man. Bone Joint J. 2020;102B(1):90–101.

    Article  Google Scholar 

  9. Lindberg-Larsen M, Petersen PB, Jorgensen CC, Overgaard S, Kehlet H, Madsen F, Hansen TB, Gromov K, Laursen M, Hansen LT, Kjaersgaard-Andersen P, Solgaard S, Krarup NH, Bagger J, Lundbeck Fdn Ctr Fast-track H. Postoperative 30-day complications after cemented/hybrid versus cementless total hip arthroplasty in osteoarthritis patients > 70 years A multicenter study from the Lundbeck Foundation Centre for Fast-track Hip and Knee replacement database and the Danish Hip Arthroplasty Register. Acta Orthop. 2020;91(3):286–92.

    Article  Google Scholar 

  10. Reina RJ, Rodriguez JA, Rasquinha VJ, Ranawat CS. Fixation and osteolysis in plasma-sprayed hemispherical cups with hybrid total hip arthroplasty. J Arthroplasty. 2007;22(4):531–4.

    Article  Google Scholar 

  11. Lubbeke A, Garavaglia G, Barea C, Stern R, Peter R, Hoffmeyer P. Influence of patient activity on femoral osteolysis at five and ten years following hybrid total hip replacement. J Bone Joint Surg-Bri Volume. 2011;93B(4):456–63.

    Article  Google Scholar 

  12. Della Valle AG, Slullitel G, Piccaluga F, Salvati EA. The precision and usefulness of preoperative planning for cemented and hybrid primary total hip arthroplasty. J Arthroplasty. 2005;20(1):51–8.

    Article  Google Scholar 

  13. Bolland B, Culliford DJ, Langton DJ, Millington JPS, Arden NK, Latham JM. High failure rates with a large-diameter hybrid metal-on-metal total hip replacement clinical, radiological and retrieval analysis. J Bone Joint Surg-Br Volume. 2011;93B(5):608–15.

    Article  Google Scholar 

  14. Rasquinha VJ, Dua V, Rodriguez JA, Ranawat CS. Fifteen-year survivorship of a collarless, cemented, normalized femoral stem in primary hybrid total hip arthroplasty with a modified third-generation cement technique. J Arthroplasty. 2003;18(7):86–94.

    Article  Google Scholar 

  15. Jain S, Magra M, Dube B, Veysi VT, Whitwell GS, Aderinto JB, Emerton ME, Stone MH, Pandit HG. Reverse hybrid total hip arthroplasty a survival analysis of 1082 consecutive cases with minimum five-year follow-up. Bone Joint J. 2018;100B(8):1010–7.

    Article  Google Scholar 

  16. Wangen H, Havelin LI, Fenstad AM, Hallan G, Furnes O, Pedersen AB, Overgaard S, Karrholm J, Garellick G, Makela K, Eskelinen A, Nordsletten L. Reverse hybrid total hip arthroplasty results from the nordic arthroplasty register association (NARA). Acta Orthop. 2017;88(3):248–54.

    Article  Google Scholar 

  17. Gwam CU, Mistry JB, Mohamed NS, Thomas M, Bigart KC, Mont MA, Delanois RE. Current epidemiology of revision total hip arthroplasty in the united states: national inpatient sample 2009 to 2013. J Arthroplasty. 2017;32(7):2088–92.

    Article  Google Scholar 

  18. Chahal J, Van Thiel GS, Mather RC, Lee S, Song SH, Davis AM, Salata M, Nho SJ. The patient acceptable symptomatic state for the modified harris hip score and hip outcome score among patients undergoing surgical treatment for femoroacetabular impingement. Am J Sports Med. 2015;43(8):1844–9.

    Article  Google Scholar 

  19. Barrack RL, Mulroy RD Jr, Harris WH. Improved cementing techniques and femoral component loosening in young patients with hip arthroplasty. A 12-year radiographic review. J Bone Joint Surg Br volume. 1992;74(3):385–9.

    Article  CAS  Google Scholar 

  20. Kim YH, Oh SH, Kim JS, Koo KH. Contemporary total hip arthroplasty with and without cement in patients with osteonecrosis of the femoral head. J Bone Joint Surg-Am Volume. 2003;85A(4):675–81.

    Article  Google Scholar 

  21. Trebse R, Valic M, Savarin D, Milosev I, Levasic V. Survival rate of total hip replacements with matched and with mixed components with 10.7 years mean follow-up. Hip Int. 2022;32(1):32–8.

    Article  Google Scholar 

  22. Hemmila M, Karvonen M, Laaksonen I, Matilainen M, Eskelinen A, Haapakoski J, Puhto AP, Kettunen J, Manninen M, Makela KT. Survival of 11,390 continuum cups in primary total hip arthroplasty based on data from the Finnish arthroplasty register. Acta Orthop. 2019;90(4):312–7.

    Article  Google Scholar 

  23. Erivan R, Villatte G, Lecointe T, Mulliez A, Descamps S, Boisgard S. Long-term survival of hybrid total hip arthroplasty with the uncemented CLS cup, cemented Muller cobalt-chromium stem, and 28-mm Metasul (TM) bearings: Retrospective review of 115 hips after a minimum of 178 years. Orthop Traumatol-Surg Res. 2019;105(7):1289–95.

    Article  Google Scholar 

  24. Gregory JJ, Starks I, Aulakh T, Phillips SJ. Five-year survival of nonagenerian patients undergoing total hip replacement in the United Kingdom. J Bone Joint Surg-Br Volume. 2010;92B(9):1227–30.

    Article  Google Scholar 

  25. Jackson MP, Sexton SA, Yeung E, Walter WL, Walter WK, Zicat BA. The effect of obesity on the mid-term survival and clinical outcome of cementless total hip replacement. J Bone Joint Surg-Br Volume. 2009;91B(10):1296–300.

    Article  Google Scholar 

  26. Mao S, Chen B, Zhu Y, Qian L, Lin J, Zhang X, Yu W, Han G. Cemented versus uncemented total hip replacement for femoral neck fractures in elderly patients: a retrospective, multicentre study with a mean 5-year follow-up. J Orthop Surg Res. 2020;15(1):447.

    Article  Google Scholar 

  27. Gwynne-Jones DP, Gray AR. Cemented or uncemented acetabular fixation in combination with the Exeter Universal cemented stem long-term survival to 18 years. Bone Joint J. 2020;102B(4):414–22.

    Article  Google Scholar 

  28. Hailer NP, Lazarinis S, MaKela KT, Eskelinen A, Fenstad AM, Hallan G, Havelin L, Overgard S, Pedersen AB, Mehnert F, Karrholm J. Hydroxyapatite coating does not improve uncemented stem survival after total hip arthroplasty! An analysis of 116,069 THAs in the Nordic Arthroplasty Register Association (NARA) database. Acta Orthop. 2015;86(1):18–25.

    Article  Google Scholar 

  29. Hauer G, Heri A, Klim S, Puchwein P, Leithner A, Sadoghi P. Survival rate and application number of total hip arthroplasty in patients with femoral neck fracture: an analysis of clinical studies and national arthroplasty registers. J Arthroplasty. 2020;35(4):1014–22.

    Article  Google Scholar 

  30. Zeng X, Zhan K, Zhang L, Zeng D, Yu W, Zhang X, Zhao M. Conversion to total hip arthroplasty after failed proximal femoral nail antirotations or dynamic hip screw fixations for stable intertrochanteric femur fractures: a retrospective study with a minimum follow-up of 3 years. BMC Musculoskelet Disord. 2017;18(1):38.

    Article  Google Scholar 

  31. Yu W, Han X, Chen W, Mao S, Zhao M, Zhang X, Han G, Ye J, Chen M, Zhuang J. Conversion from a failed proximal femoral nail anti-rotation to a cemented or uncemented total hip arthroplasty device: a retrospective review of 198 hips with previous intertrochanteric femur fractures. BMC Musculoskelet Disord. 2020;21(1):791.

    Article  Google Scholar 

  32. Zhang C, Chen M, Yu W, Han X, Ye J, Zhuang J. Long-term survival after cemented versus uncemented total hip arthroplasty for treatment of acute femoral neck fracture: a retrospective study with a mean 10-year follow-up. J Int Med Res. 2020;48(9):300060520941974.

    CAS  PubMed  Google Scholar 

  33. Liu T, Hua X, Yu W, Lin J, Zhao M, Liu J, Zeng X. Long-term follow-up outcomes for patients undergoing primary total hip arthroplasty with uncemented versus cemented femoral components: a retrospective observational study with a 5-year minimum follow-up. J Orthop Surg Res. 2019;14(1):371.

    Article  Google Scholar 

  34. Hirakawa K, Bauer TW, Stulberg BN, Wilde AH, Secic M. Characterization and comparison of wear debris from failed total hip implants of different types. J Bone Joint Surg Am. 1996;78(8):1235–43.

    Article  CAS  Google Scholar 

  35. Altenburg AJ, Callaghan JJ, Yehyawi TM, Pedersen DR, Liu SS, Leinen JA, Dahl KA, Goetz DD, Brown TD, Johnston RC. Cemented total hip replacement cable debris and acetabular construct durability. J Bone Joint Surg Am. 2009;91(7):1664–70.

    Article  Google Scholar 

  36. Keeling P, Howell JR, Kassam AM, Sathu A, Timperley AJ, Hubble MJW, Wilson MJ, Whitehouse SL. Long-term survival of the cemented exeter universal stem in patients 50 years and younger: an update on 130 hips. J Arthroplasty. 2020;35(4):1042–7.

    Article  Google Scholar 

  37. Ceynowa M, Zerdzicki K, Klosowski P, Zrodowski M, Pankowski R, Roclawski M, Mazurek T. The cement-bone bond is weaker than cement-cement bond in cement-in-cement revision arthroplasty A comparative biomechanical study. PLoS One. 2021;16(2):e0246740.

    Article  CAS  Google Scholar 

  38. Robb C, Harris R, O’Dwyer K, Aslam N. Radiographic assessment of biomechanical parameters following hip resurfacing and cemented total hip arthroplasty. Hip Int. 2009;19(3):251–6.

    Article  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This study failed to receive any grant from funding agencies in the public, commercial, or not-for-profit sectors.

Author information

Authors and Affiliations

Authors

Contributions

LW and MJY performed the data collection and participated in manuscript writing. YZ, MLY, SW, MX, and MZ performed statistical analysis. SG, MX, and WY participated in the study design. All authors have read and approved the final manuscript.

Corresponding authors

Correspondence to Songtao Gao, Min Xiong or Weiguang Yu.

Ethics declarations

Ethics approval and consent to participate

All methods were carried out according to the relevant guidelines and regulations of the institutional and/or national research committee and the 1964 Helsinki declaration. This study was approved by both Medical Ethics Committees (Henan Provincial People's Hospital and Wuhan Fourth Hospital), and informed consent was waived by the Medical Ethics Committees.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Wang, L., Yu, M., Zhang, Y. et al. Long-term survival of hybrid total hip replacement for prior failed proximal femoral nail antirotation: a retrospective study with a median 10-year follow-up. BMC Musculoskelet Disord 23, 868 (2022). https://doi.org/10.1186/s12891-022-05827-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-022-05827-3

Keywords