Comparison of pedicle screw placement accuracy between two types of imaging support (Artis Zeego versus two-dimensional fluoroscopy): a cross-sectional observational study

Abstract

Background

The pedicle screw system is widely used in spine surgery, and it provides rigid fixation and leads to successful subsequent deformity correction and bony fusion. The standard imaging technique for pedicle screw insertion is two-dimensional images obtained from C-arm-type X-ray fluoroscopy. Artis Zeego is an emerging intraoperative imaging technique that can provide conventional two-dimensional fluoroscopic images and rapid three-dimensional fluoroscopic computed tomography reconstruction imaging. The aim of this study is to compare the insertion accuracies of PS placement using Artis Zeego and conventional 2D X-ray fluoroscopy.

Methods

In this study, we retrospectively reviewed the postoperative images of thoracolumbar fusion patients who underwent surgery using pedicle screws between 2013 and 2018. Pedicle screw malplacement was assessed using a four-grade classification by Rao et al. Misplacement rates were compared between pedicle screws assisted with Artis Zeego and two-dimensional fluoroscopy.

Results

A total of 1107 pedicle screws in 153 patients were inserted using Artis Zeego, and 427 pedicle screws in 80 patients were inserted using fluoroscopy. The overall perforation rate was 4.2% (46 perforations of 1106 pedicle screws) in the Artis Zeego group and 7.7% (33 perforations of 427 pedicle screws) in the fluoroscopy group. In the Artis Zeego group, 43 (3.9%) screws were classified as grade 1, and three (0.3%) screws were classified as grade 2. In the fluoroscopy group, 21 (4.9%) screws were classified as grade 1, 10 (2.3%) screws were classified as grade 2, and 2 (0.5%) screws were classified as grade 3. The use of Artis Zeego was associated with a significantly lower screw malplacement rate than was the use of fluoroscopy (pā€‰<ā€‰0.001).

Conclusions

Our results demonstrated that pedicle screw placement with Artis Zeego was associated with a lower malplacement rate than was conventional two-dimensional fluoroscopy. No severe malplacement was observed in the Artis Zeego group. Thus, Artis Zeego could be a good option for improving pedicle screw accuracy.

Peer Review reports

Background

Instrumented fusion has become an indispensable procedure for spinal surgery and has been used for various spinal conditions. The mainstream of posterior instrumentation for fusion is the pedicle screw (PS) system, which provides rigid fixation and leads to successful deformity correction and bony fusion subsequently. However, the posterior PS insertion technique is inevitably associated with a certain risk of injury to the surrounding structures, such as the great vessels, spinal cord, and nerve roots. The risk is generally higher in patients with severe deformities. Although it depends on the definition of malplacement and patient population, the rate of PS malplacement reported in older studies was up to 40% for the lumbar spine and 55% for the thoracic spine [1, 2], whereas in a more recent study, the misplacement rate was lower at approximately 10% [3].

Previous reports have shown that PS can be safely inserted without imaging support in most patients [4], but recent studies demonstrated that malplacement risk is significantly reduced with appropriate imaging support [5,6,7]. The standard imaging for PS insertion uses two-dimensional (2D) images obtained from C-arm type X-ray fluoroscopy. Recently, various imaging techniques such as three-dimensional (3D) fluoroscopy or computed tomography (CT)-type have been developed and clinically used with or without a navigation system. Studies have shown that these novel techniques lower the rates of screw malplacement relative to those with conventional 2D fluoroscopy and/or free-hand insertion technique [8,9,10]. Intraoperative 3D cone-beam CT (CBCT) is an emerging technique, which was first introduced for the surgical planning of craniofacial reconstruction [11], then expanded its indication to major vascular disease, general surgeries, and orthopedic surgery, including the spine [12]. Artis Zeego is a new CBCT originally developed for angiographical treatment [13]. It consists of a floor-mounted multi-axis robotic C-arm and specialized software for image reconstruction (Fig.Ā 1) and is utilized in two ways: conventional 2D fluoroscopy and rapid 3D fluoroscopic CT-like reconstruction imaging. Setting postoperative CT as the gold standard, one previous study demonstrated that intraoperative images using Artis Zeego showed excellent specificity (98%) and acceptable sensitivity (77ā€“79%) for screw malplacement [14].

Fig. 1
figure 1

Artis Zeego mounted in the hybrid operating room. (Reprinted with permission from Matsuoka A, Significance of Spine Surgery in the Hybrid Operating Room. Journal of the Showa University Society (Showa Gakushikai Zasshi), 2019 Jun; 79 (3):318ā€“322.)

Although studies have demonstrated that CBCTs, including Artis Zeego, improved the accuracy of PS placement. To the best of our knowledge, no study has directly compared CBCT-based PS placement and conventional 2D fluoroscopy. Therefore, in this study, we compared the insertion accuracies of PS placement using Artis Zeego and conventional 2D X-ray fluoroscopy.

Materials

Patient selection

Records and postoperative CT images were reviewed retrospectively from patients who underwent thoracolumbar fusion surgery using percutaneous or open PS, between January 2013 and December 2018 at an academic tertiary care university hospital. Patients with degenerative conditions, trauma, tumor, including metastasis and infections were included, but patients with severe coronal deformity (Cobb angleā€‰>ā€‰20Ā°) were excluded due to measurement difficulty.

The hybrid operating room (OR) and Artis Zeego were installed in April 2015; thus, we collected the data of consecutive patients who underwent surgery at the hybrid OR between April 2015 to December 2018. We also collected data of patients who underwent fusion surgery with 2D fluoroscopic assistance between January 2013 and December 2015 as the control group.

Surgical technique of PS insertion

Two-dimensional fluoroscopy or Artis Zeego without a navigation system was used routinely for all instrumentation patients to check the implant positions during the surgery in this study period. All operations were performed by or under direct supervision of a board-certified spine surgeon who had at least five years of spine surgical experience after basic surgical training. The surgical equipment, operative technique, and surgical experience of the operating surgeons were similar during the study period, except for the use of 2D fluoroscopy or Artis Zeego.

For open surgery, the screws were placed using conventional anatomical landmark techniques, and the positions were confirmed using anteroposterior (AP) and lateral images of 2D fluoroscopy or 3D reconstruction images using Artis Zeego. If there were any concerns about screw placement on the intraoperative images, the screws were redirected or removed. For percutaneous PS (PPS) placement with 2D fluoroscopy, cannulated guide needles were placed first using the AP image. After confirmation of the needle positions on the AP view, lateral images were checked, and guide wires were inserted. Cannulated PSs were inserted through the wires using several lateral fluoroscopic images. After the PS insertion, the AP image was checked again before rod placement. For PPS placement with Artis Zeego, the 2D image mode was used and the technique was similar to that with 2D fluoroscopy. Three-dimensional images were checked twice at the end of needle insertion and screw placement, to secure the needle/screw position. If there were any concerns about the needle/screw position, the needles/screws were redirected or removed before connecting the rod. The decision on whether open or percutaneous screw placement was performed was at the surgeonā€™s discretion.

Evaluation of screw position of postoperative CT images

Based on the institutional protocol, all patients with instrumentation underwent postoperative CT typically within 1ā€“3Ā days after surgery. Screw accuracy was assessed using postoperative CT imaging with axial, sagittal, and coronal views. We utilized the grading system of Rao et al. [4], which consists of 4 grades: ā€œ0ā€ as no evidence of pedicle perforation, ā€œ1ā€ as less than 2Ā mm of pedicle perforation of the pedicle with one screw thread out of the pedicle, ā€œ2ā€ as 2ā€“4Ā mm of perforation of the pedicle, and ā€œ3ā€ as greater than 4Ā mm (Table 1). While this classification system was originally used for medial and lateral breaches on axial images, we applied the same definition to misplacements on sagittal and coronal planes in any direction. In addition, if the screw was too long, this was defined as ā€œanteriorā€ perforation and graded similarly to the perforation in other directions. If a PS was malpositioned in multiple directions, the highest grade was recorded for each pedicle. For instance, if a PS was misplaced 1Ā mm laterally (grade 1) and 3Ā mm caudally (grade 2), we graded the pedicle as ā€œgrade 2.ā€

Table 1 Classification of pedicle and anterior vertebral body perforation

A board-certified spine surgeon who was not directly involved in patient care and was completely blinded for the use of imaging supports, 2D fluoroscopy, and Artis Zeego performed the assessment. In certain situations, the PS was laterally placed intentionally, as per the ā€œinā€“out-inā€ technique usually used for small thoracic pedicles. In such pedicles, lateral pedicle wall perforation was not counted. Additionally, we also defined ā€˜critical perforation,ā€™ (which might cause serious sequalae) as all grade 2 and 3 perforations, except lateral grade 2 and 3 perforations over the thoracic region, as described above.

In the Artis Zeego group, we also investigated the intraoperative revision of the screw, marker, or guide needle, since all intraoperative CT images were stored in the imaging server.

Statistical analysis

Proportion was used to describe the categorical variables. Mean or median was used to describe continuous variables based on the distributionā€™s normality. For comparison of categorical variables, the Fisherā€™s exact or Chi-squared test was used. For comparison of continuous variables, the Studentā€™s t-test or Mannā€“Whitney U test was used. Statistical analyses were performed using JMP software (version 15.0; SAS Institute, Inc., Cary, NC, USA) or R software (R for 4.0.3, r-project.org).

Results

Patient demographics

During the study period, 255 patients underwent spinal fusion under intraoperative imaging using 2D fluoroscopy (80 patients) or Artis Zeego (175 patients). Twenty-two patients were excluded due to a coronal imbalance. A total of 1107 PS in 153 patients were inserted using Artis Zeego, and 427 PS in 80 patients were inserted using 2D fluoroscopy. The mean ages of patients with Artis Zeego and 2D fluoroscopy were similar, 68.5Ā years and 67.4Ā years, respectively (pā€‰=ā€‰0.54). Seventy (45.7%) patients in the Artis Zeego group and 32 (40.0%) in the 2D fluoroscopy group were men (pā€‰=ā€‰0.48). Of the 153 patients with Artis Zeego, the diagnoses indicating surgery in the Artis Zeego and 2D fluoroscopy groups were degenerative spine disease in 101 (66.0%) and 48 (60.0%) patients, spinal trauma in 30 (19.6%) and 15 (18.8%), tumor in 11 (7.2%) and 11 (13.8%) patients, and infection in 11(7.2%) and 6 (7.5%) patients, respectively (pā€‰=ā€‰0.44) (Table 2).

Table 2 Patient demographics

Assessment of screw position

The overall perforation rate was 4.2% (46 perforations of 1106 PS) in the Artis Zeego group and 7.7% (33 perforations of 427 PS) in the 2D fluoroscopy group. In the Artis Zeego group, 43 (3.9%) screws were classified as grade 1, and three (0.3%) screws were classified as grade 2. In the 2D fluoroscopy group, 21 (4.9%) screws were classified as grade 1, 10 (2.3%) screws were classified as grade 2, and 2 (0.5%) screws were classified as grade 3. Regarding the accuracy of PS placement, the use of Artis Zeego was associated with a significantly lower screw malplacement rate than was the use of 2D fluoroscopy (pā€‰<ā€‰0.001). Artis Zeego was also associated with significantly fewer critical perforation (Grade 2 and 3) screws (3/1107ā€‰=ā€‰0.3% vs. 12/427ā€‰=ā€‰2.8%, pā€‰<ā€‰0.001) (Table 3). At the lumbosacral level, misplacement most commonly occurred at the S1 pedicle (10/821ā€‰=ā€‰2.2%) in the Artis Zeego group, whereas the misplacement rate was the highest at L1 (6/30ā€‰=ā€‰20.0%) in the 2D fluoroscopy group (Table 4).

Table 3 Total number of pedicle screws, screws with perforations, and rate of each grade
Table 4 Misplacement rate in both groups stratified by surgical level

Two-thirds of the pedicle perforations occurred on the axial plane, medially or laterally. There were no patients with cranial malposition in either group (Table 5). Both screws with grade 3 perforation were in the 2D fluoroscopy group, in the S1 pedicles, and perforated medially. These mispositioned screws were asymptomatic, and no patient had neurological deficiency, great vessel injuries, or revision surgery directly related to them.

Table 5 Direction and grade of screw perforations

In the Artis Zeego group, a total of 79 PS, markers, or guiding needles (79/1107ā€‰=ā€‰7.1%) were revised intraoperatively based on intraoperative images. Fifty-two (65.8%), 22 (27.8%), and 5 (6.3%) revised pedicle instruments had medial, lateral, or caudal misplacements on the initial CT images, respectively (Fig.Ā 2).

Fig. 2
figure 2

Examples of intraoperative computerized tomography (CT) images. AĀ Initial intraoperative image demonstrating a medial misplacement of guiding needle in the left L5 pedicle. B Second intraoperative scan image showing the redirected needle properly located in the pedicle. C Postoperative CT image showing accurately placed pedicle screws. D, E Sagittal reconstruction images showing accurately placed needles (D) and screws (E)

Discussion

This study compared the accuracy of PS placement with two imaging supports: Artis Zeego versus conventional 2D fluoroscopy. We found that the screw malplacement rate with Artis Zeego was significantly lower than that with the 2D fluoroscopy guided technique. No grade 3 malplacement was observed in the Artis Zeego group, in contrast to the 2D fluoroscopy group.

Pedicle screw malplacement was reportedly associated with various complications, such as dural tear, nerve and vascular injury, and visceral complications due to the implant itself [15,16,17]. Moreover, the PS malposition may cause loss of fixation (especially at the lowermost instrumented vertebra), as well as spinal instability [18]. In previous reports, the accuracy of PS placement varied probably due to the definition of malplacement. It has been described as 28ā€“85% with conventional C-arm 2D fluoroscopy, 88ā€“96% with the CT navigation system, and 93ā€“99% with the O-arm navigation system [2, 9, 19, 20]. A meta-analysis of 130 studies with 37,337 PS assessments demonstrated an accuracy of 95.2% with navigation systems and 90.3% without navigation [21]. To the best of our knowledge, only one study has investigated the accuracy of PS placement with Artis Zeego [14]. The main aim of that study was to compare the PS positions between intraoperative images with Artis Zeego and postoperative CT images. Setting postoperative CT images as the reference, the authors reported that the sensitivity, specificity, positive predictive value, and negative predictive value of the Artis Zeego images were 77%, 98%, 86%, and 96%, respectively. They concluded that the Artis Zeego provided an accurate assessment of PS placement, but there was no direct comparison with other imaging supports in their report. In our study, the insertion accuracy with Artis Zeego was 95.8%, which was significantly higher than that with C-arm fluoroscopy. The results were comparable to those of other CT-based techniques, as well as the navigation system. Our results indicate the clinical usefulness of Artis Zeego in patients undergoing lumbar spine surgery.

In our study, the intraoperative needle/marker/PS revision rate in the Artis Zeego group was 7.1%, which was also comparable to that of other CT-guided systems. Bydon et al. reported that they revised 8.97% of screws based on intraoperative CT images [10]. This intraoperative redirection is likely the key factor for superior accuracy of PS with Artis Zeego over 2D fluoroscopy. This feature is useful not only for patient safety, but also for educational purposes. Studies have demonstrated that 80 to over 300 PS insertions are required for spine surgery trainees to reach an adequate level of PS placement [22, 23]. Although further studies are required, immediate intraoperative feedback of the PS position might facilitate the learning of surgical skills.

Interestingly, the level with the lowest accuracy of PS was different between the two groups. The PS malposition most commonly occurred at L1 in the 2D fluoroscopy group, followed by the S1 level. In the Artis Zeego group, the misplacement of L1 was lower, but that of S1 was still high. The L1 vertebra usually has the smallest pedicle, and it seems reasonable to consider it to be more liable to pedicle perforation. In contrast, the S1 vertebra has wider but more angulated pedicles than do the lumbar vertebrae [24,25,26]. This makes S1 susceptible to lateral and anterior perforations, especially when the medial angulation is insufficient. Additionally, S1 commonly has a ā€œtrefoil canalā€ rather than a cylindrical canal with high lumbar vertebral levels. Reports also mentioned that this characteristic of the spinal canal at S1 could lead to more medial perforation [7]. In our results, the redirection rate was not high at S1 in the Artis Zeego group. This suggests that the guide or marker was placed adequately at the time of intraoperative CT, although the PS could be misplaced more commonly at this level. Since screw insertion instruments are usually larger than the guide or marker, the PS trajectory might have been affected by the iliac crest or surrounding soft tissue. Surgeons should be aware of this, and adequate exposure as well as careful planning of PS insertion, are warranted.

Compared to conventional 2D fluoroscopy, Artis Zeego has more advantages other than the PS insertion accuracy. The main body of Artis Zeego is very stable because it is floor-mounted; moreover, the device has a function that enables it to store data regarding the arm position. In addition, it is easier to operate and requires lesser time than does 2D fluoroscopy to reproduce previously obtained arm positions. This is especially beneficial for percutaneous screw insertion into a highly deformed spine, as it provides optimal image quality. However, the floor-mounted design could also be a disadvantage in certain situations. Unlike conventional 2D fluoroscopy, the relative position of the operating table and Artis Zeego is fixed, and Artis Zeego is always located cephalad to the patient. Therefore, the surgeon needs to stand caudal relative to the arm of Artis Zeego when it is used in the real-time 2D fluoroscopy mode, and this position may not be comfortable for surgeons to perform certain procedures. Additionally, surgeons must use a hybrid OR to use Artis Zeego because of its floor-mounted design. On the other hand, conventional 2D fluoroscopy as well as O-arm and 3D fluoroscopy use portable devices that can be utilized in any OR.

Currently, other CT-based systems are available, and good results have been reported. This study, however, was not a direct comparison study between Artis Zeego and other CT-based systems; therefore, we cannot conclude which is better for patient safety. However, the Artis Zeego demonstrated similar results to previously reported outcomes of other CT-based imaging techniques, and it can be a good option for spine surgery if the system is available to meet the demands of other surgical specialties.

Our study had several limitations. First, this was a retrospective analysis with a relatively small number of patients; thus, the confounding for indication might be an issue. Because of the small number of misplaced screws, statistical power was limited, especially for vertebral level-specific analysis. Additionally, we excluded patients with severe coronal deformity because it was often difficult to assess pedicle perforation using postoperative CT images. With regard to the assessment of CT-based PS misplacement, we used the classification by Rao et al. [4], which has been widely used in similar research. However, Rao et al. reported that the reproducibility of their classification was moderate, and that their CT-based assessment tended to overestimate the degree of misplacement when direct visualization of the specimen was set as the gold standard. Furthermore, because of the unavailability of data, we did not include the details of clinical results, such as pain, radiological fusion, or scores of patient-reported outcome measures. In addition, surgical data such as the screw type, diameter, and length were not available for all cases, and we could not evaluate the influence of parameters such as the screw type on screw misplacement. Lastly, cost-effectiveness and radiation exposure were not analyzed. These points should be addressed in future studies.

Conclusions

In conclusion, our results demonstrated that PS placement with Artis Zeego was associated with a lower malplacement rate than was conventional 2D fluoroscopy. No severe malplacement was observed in the Artis Zeego group. Therefore, Artis Zeego could be a good option for improving the accuracy of PS.

Availability of data and materials

The data that support the findings of this study are not deposited in a public repository due to the data sharing policy of the Department of Orthopaedic Surgery, Showa University, but they are available from the corresponding author on reasonable request.

Abbreviations

2D:

Two-dimensional

3D:

Three-dimensional

AP:

Anteroposterior

CBCT:

Cone-beam computed tomography

CT:

Computed tomography

OR:

Operating room

PPS:

Percutaneous pedicle screw

PS:

Pedicle screw

References

  1. Castro WH, Halm H, Jerosch J, Malms J, Steinbeck J, Blasius S. Accuracy of pedicle screw placement in lumbar vertebrae. Spine. 1996;21:1320ā€“4. https://doi.org/10.1097/00007632-199606010-00008.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  2. Gelalis ID, Paschos NK, Pakos EE, Politis AN, Arnaoutoglou CM, Karageorgos AC, Ploumis A, Xenakis TA. Accuracy of pedicle screw placement: a systematic review of prospective in vivo studies comparing free hand, fluoroscopy guidance and navigation techniques. Eur Spine J. 2012;21:247ā€“55. https://doi.org/10.1007/s00586-011-2011-3.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  3. De la Garza RR, Echt M, Benton JA, Gelfand Y, Longo M, Yanamadala V, Yassari R. Accuracy of Freehand versus Navigated Thoracolumbar Pedicle Screw Placement in Patients with Metastatic Tumors of the Spine. J Korean Neurosurg Soc. 2020;63:777ā€“83. https://doi.org/10.3340/jkns.2020.0001.

    ArticleĀ  CASĀ  Google ScholarĀ 

  4. Rao G, Brodke DS, Rondina M, Dailey AT. Comparison of computerized tomography and direct visualization in thoracic pedicle screw placement. J Neurosurg. 2002;97:223ā€“6. https://doi.org/10.3171/spi.2002.97.2.0223.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  5. Hansen-Algenstaedt N, Chiu CK, Chan CY, Lee CK, Schaefer C, Kwan MK. Accuracy and safety of fluoroscopic guided percutaneous pedicle screws in thoracic and lumbosacral spine: a review of 2000 screws. Spine. 2015;40:E954-963. https://doi.org/10.1097/brs.0000000000000958.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  6. Mason A, Paulsen R, Babuska JM, Rajpal S, Burneikiene S, Nelson EL, Villavicencio AT. The accuracy of pedicle screw placement using intraoperative image guidance systems. J Neurosurg Spine. 2014;20:196ā€“203. https://doi.org/10.3171/2013.11.Spine13413.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  7. Chiu CK, Kwan MK, Chan CY, Schaefer C, Hansen-Algenstaedt N. The accuracy and safety of fluoroscopically guided percutaneous pedicle screws in the lumbosacral junction and the lumbar spine: a review of 880 screws. Bone Joint J. 2015;97-b:1111ā€“7. https://doi.org/10.1302/0301-620x.97b8.35330.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  8. Feng W, Wang W, Chen S, Wu K, Wang H. O-arm navigation versus C-arm guidance for pedicle screw placement in spine surgery: a systematic review and meta-analysis. Int Orthop. 2020;44:919ā€“26. https://doi.org/10.1007/s00264-019-04470-3.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  9. Knafo S, Mireau E, Bennis S, Baussart B, Aldea S, Gaillard S. Operative and Perioperative Durations in O-Arm vs C-Arm Fluoroscopy for Lumbar Instrumentation. Oper Neurosurg (Hagerstown). 2018;14:273ā€“8. https://doi.org/10.1093/ons/opx142.

    ArticleĀ  Google ScholarĀ 

  10. Bydon M, Xu R, Amin AG, Macki M, Kaloostian P, Sciubba DM, Wolinsky JP, Bydon A, Gokaslan ZL, Witham TF. Safety and efficacy of pedicle screw placement using intraoperative computed tomography: consecutive series of 1148 pedicle screws. J Neurosurg Spine. 2014;21:320ā€“8. https://doi.org/10.3171/2014.5.Spine13567.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  11. Tang X, Ning R. A cone beam filtered backprojection (CB-FBP) reconstruction algorithm for a circle-plus-two-arc orbit. Med Phys. 2001;28:1042ā€“55. https://doi.org/10.1118/1.1376444.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  12. Tonetti J, Boudissa M, Kerschbaumer G, Seurat O. Role of 3D intraoperative imaging in orthopedic and trauma surgery. Orthop Traumatol Surg Res. 2020;106:S19-s25. https://doi.org/10.1016/j.otsr.2019.05.021.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  13. Boland JE, Wang LW, Love BJ, Christofi M, Muller DW. Impact of New-generation Hybrid Imaging Technology on Radiation Dose during Percutaneous Coronary Interventions and Trans-femoral Aortic Valve Implantations: A comparison with conventional flat-plate angiography. Heart Lung Circ. 2016;25:668ā€“75. https://doi.org/10.1016/j.hlc.2015.08.023.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  14. Cordemans V, Kaminski L, Banse X, Francq BG, Cartiaux O. Accuracy of a new intraoperative cone beam CT imaging technique (Artis zeego II) compared to postoperative CT scan for assessment of pedicle screws placement and breaches detection. Eur Spine J. 2017;26:2906ā€“16. https://doi.org/10.1007/s00586-017-5139-y.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  15. Gaines RW Jr. The use of pedicle-screw internal fixation for the operative treatment of spinal disorders. J Bone Joint Surg Am. 2000;82:1458ā€“76. https://doi.org/10.2106/00004623-200010000-00013.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  16. Lonstein JE, Denis F, Perra JH, Pinto MR, Smith MD, Winter RB. Complications associated with pedicle screws. J Bone Joint Surg Am. 1999;81:1519ā€“28. https://doi.org/10.2106/00004623-199911000-00003.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  17. Amato V, Giannachi L, Irace C, Corona C. Accuracy of pedicle screw placement in the lumbosacral spine using conventional technique: computed tomography postoperative assessment in 102 consecutive patients. J Neurosurg Spine. 2010;12:306ā€“13. https://doi.org/10.3171/2009.9.Spine09261.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  18. AƧikbaş SC, Arslan FY, Tuncer MR. The effect of transpedicular screw misplacement on late spinal stability. Acta Neurochir. 2003;145:949ā€“54. https://doi.org/10.1007/s00701-003-0116-0 (discussion 954ā€“945.).

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  19. Shin MH, Hur JW, Ryu KS, Park CK. Prospective Comparison Study Between the Fluoroscopy-guided and Navigation Coupled With O-arm-guided Pedicle Screw Placement in the Thoracic and Lumbosacral Spines. J Spinal Disord Tech. 2015;28:E347-351. https://doi.org/10.1097/BSD.0b013e31829047a7.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  20. Scarone P, Vincenzo G, Distefano D, Del Grande F, Cianfoni A, Presilla S, Reinert M. Use of the Airo mobile intraoperative CT system versus the O-arm for transpedicular screw fixation in the thoracic and lumbar spine: a retrospective cohort study of 263 patients. J Neurosurg Spine. 2018;29:397ā€“406. https://doi.org/10.3171/2018.1.Spine17927.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  21. Kosmopoulos V, Schizas C. Pedicle screw placement accuracy: a meta-analysis. Spine. 2007;32:E111-120. https://doi.org/10.1097/01.brs.0000254048.79024.8b.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  22. Gonzalvo A, Fitt G, Liew S, de la Harpe D, Turner P, Ton L, Rogers MA, Wilde PH. The learning curve of pedicle screw placement: how many screws are enough? Spine. 2009;34:E761-765. https://doi.org/10.1097/BRS.0b013e3181b2f928.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  23. Gertzbein SD, Robbins SE. Accuracy of pedicular screw placement in vivo. Spine. 1990;15:11ā€“4. https://doi.org/10.1097/00007632-199001000-00004.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  24. Ebraheim NA, Rollins JR Jr, Xu R, Yeasting RA. Projection of the lumbar pedicle and its morphometric analysis. Spine. 1996;21:1296ā€“300. https://doi.org/10.1097/00007632-199606010-00003.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  25. Xu R, Ebraheim NA, Yeasting RA, Wong FY, Jackson WT. Morphometric evaluation of the first sacral vertebra and the projection of its pedicle on the posterior aspect of the sacrum. Spine. 1995;20:936ā€“40. https://doi.org/10.1097/00007632-199504150-00010.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  26. Acharya S, Dorje T, Srivastava A. Lower dorsal and lumbar pedicle morphometry in Indian population: a study of four hundred fifty vertebrae. Spine. 2010;35:E378-384. https://doi.org/10.1097/BRS.0b013e3181cb7f2b.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

Download references

Acknowledgements

We would like to thank Editage (www.editage.com) for English language editing.

Funding

There is no funding information to declare.

Author information

Authors and Affiliations

Authors

Contributions

AM and TT participated in the design of this study. YK, KI, HM, TO, TS, and KI participated in data acquisition and collection. IO analyzed the data. AM and IO drafted the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Ichiro Okano.

Ethics declarations

Ethics approval and consent to participate

Ethical approval was obtained for this retrospective study (Showa University Ethics Committee, Approval No. 2020ā€“3032). The study was conducted in accordance with the Declaration of Helsinki. The requirement for informed consent from each patient was waived by the Showa University Ethics Committee because of the retrospective and anonymous nature of this study. All methods in this study were performed in accordance with the STROBE guideline for observational studies.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisherā€™s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Matsuoka, A., Toyone, T., Okano, I. et al. Comparison of pedicle screw placement accuracy between two types of imaging support (Artis Zeego versus two-dimensional fluoroscopy): a cross-sectional observational study. BMC Musculoskelet Disord 23, 644 (2022). https://doi.org/10.1186/s12891-022-05602-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-022-05602-4

Keywords