Table 2 Comparison of the restoration of anatomical parameters between the CoreHip stem and a standard anatomical stem (Δ = difference between the contralateral, non-operated side and the same parameter on the operated side after total hip arthroplasty, LLD = leg length discrepancy). Mean ± standard deviation (minimum – maximum) is shown

From: Reconstruction of the anatomy of the hip in total hip arthroplasty with two different kinds of stems

 

CoreHIP

Anatomical stem

p-value

  (Mann-Whitney-U-Test)

Δ offset horizontal

2.5 ± 2.8 mm

(0 – 14 mm)

5.4 ± 4.1 mm

(0 – 17 mm)

p < 0.001

Δ femoral neck-shaft-angle

1.7 ± 1.6 degrees

(0 – 6.7 degrees)

5.6 ± 3.4 degrees

(0.2 – 17.1 degrees)

p < 0.001

Δ offset vertical

4.1 ± 3.5 mm

(0 - 15 mm)

5.0 ± 3.8 mm

(0 – 25 mm)

p = 0.024

Δ LLD

4.0 ± 3.9 mm

(0 - 15 mm)

4.5 ± 3.8 mm

(0 – 18 mm)

p = 0.10