Consistency of 3D femoral torsion measurement from MRI compared to CT gold standard

Abstract

Background

Several hip and knee pathologies are associated with aberrant femoral torsion. Diagnostic workup includes computed tomography (CT) and magnetic resonance imaging (MRI). For three-dimensional (3D) analysis of complex deformities it would be desirable to measure femoral torsion from MRI data to avoid ionizing radiation of CT in a young patient population. 3D measurement of femoral torsion from MRI has not yet been compared to measurements from CT images. We hypothesize that agreement will exist between MRI and CT 3D measurements of femoral torsion.

Methods

CT and MRI data from 29 hips of 15 patients with routine diagnostic workup for suspected femoroacetabular impingement (FAI) were used to generate 3D bone models. 3D measurement of femoral torsion was performed by two independent readers using the method of Kim et al. which is validated for CT. Inter-modalitiy and inter-reader intraclass correlation coefficients (ICC) were calculated.

Results

Between MRI and CT 3D measurements an ICC of 0.950 (0.898; 0.976) (reader 1) respectively 0.950 (0.897; 0.976) (Reader 2) was found. The ICC (95% CI) expressing the inter-reader reliability for both modalities was 0.945 (0.886; 0.973) for MRI and 0.957 (0.910; 0.979) for CT, respectively. Mean difference between CT and MRI measurement was 0.42° (MRI – CT, SD: 2.77°, p = 0.253).

Conclusions

There was consistency between 3D measurements of femoral torsion between computer rendered MRI images compared to measurements with the “gold standard” of CT images. ICC for inter-modality and inter-reader consistency indicate excellent reliability. Accurate, reliable and reproducible 3D measurement of femoral torsion is possible from MRI images.

Peer Review reports

Background

Femoral torsion was described as the angle between the femoral neck and the femoral condyles by Julius Wolff in 1868 [1, 2]. Several pathologies are associated with aberrant femoral torsion, such as slipped capital femoral epiphysis, developmental dysplasia of the hip and early-onset hip osteoarthritis [3,4,5]. There is an association between reduced femoral antetorsion and cam-type femoroacetabular impingement (FAI) [6,7,8].

Clinical quantification of femoral torsion is not reliable [6, 9, 10]. Initially standard radiographs such as the Dunn and modified Dunn view were used [1, 2, 11]. This has been replaced by more precise computer tomography (CT) and magnetic resonance imaging (MRI) measurements with various differences in measurement techniques for both of them [12, 13]. All these conventional methods that use cross sectional CT, MRI or ultrasound are two-dimensional (2D) imaging methods and encounter problems representing the complex, three-dimensional (3D) structure of the femur. To overcome the limitations of conventional 2D imaging methods, a 3D imaging method was developed by Kim et al. with greatly improved accuracy compared to conventional 2D imaging methods [14, 15]. These 3D measurements require CT imaging which has become the “gold standard” for quantification of 3D bone morphology in patients with structural hip disorders, such as FAI [16,17,18,19,20]. But using CT to characterize 3D hip morphology in the mostly young population of FAI patients is controversial due to potential harmful ionizing radiation exposure (about 4-5 mSv per CT) with a small lifetime attributable risk (0.034–0.177% for a 20-year-old) but a large relative risk (5–17 times) of cancer compared with radiographs alone [21, 22]. On the other hand MRI is useful to evaluate intra- and extra-articular soft tissue structures and the cartilage [23,24,25,26,27,28]. Thereby it provides essential predictors for the benefit of FAI surgery [29].

It would be preferable to measure the femoral torsion in 3D from MRI images without the need of additional CT imaging and consequently avoiding harmful ionizing radiation and save healthcare resources.

To understand complex deformities 3D measurements can be helpful. That is why they are desirable to get, best with as little as possible additional expenses. Correction of such complex deformities in particular may benefit from 3D analysis and 3D planning of the deformity correction to minimize errors which could lead to biomechanical alteration [30,31,32]. The required 3D bone models can be generated from CT or MRI data [33,34,35,36,37].

However, the proof that 3D femoral torsion measurements from MRI generated 3D models correspond with measurements from CT generated 3D models has not occurred yet. The aim of this paper was to investigate if 3D measurement of femoral torsion from MRI yielded comparable results as measurement from CT data. We hypothesized that agreement existed between reconstructed MRI and CT 3D measurements of femoral torsion.

Methods

Patient selection

The local ethical committee approved this study (BASEC Number 2012–02242) and all patients gave their informed written consent for their participation and the publication of this study.

We retrospectively analysed CT and MRI data of 29 hips from 15 patients (7 female, 8 male) who had routine workup for clinical symptoms suggestive for FAI between May and November 2019. The average age at the time of the scans was 32 years (range 21–47 years). Fourteen patients had an MRI and a CT scan of both sides. One patient had imaging only of the right side, which lead to a total of 29 hips (15 right hips, 14 left hips) with complete MRI and CT dataset.

Imaging and segmentation

All CT scans were performed at our institution, using a 64-detector row CT scanner (Somatom Edge Plus, Siemens Healthcare, Erlangen, Germany), slice thickness was 1.0 mm. The protocol was as follows: Feet first, supine; central positioning of the pelvis, slightly internal rotation of the legs; image centered in the middle of the pelvis slightly above the iliac crest. The proximal femur was scanned from just above the iliac crest to the end of the lesser trochanter. The distal femur was scanned from the femoral condyles to the joint line.

MRI were also performed at our institution using a 3.0 Tesla MR scanner (Magnetom Skyra 3.0 T; Siemens Healthcare, Erlangen, Germany). The protocol was as follows: Coronal 3D T1-weighted VIBE-Dixon sequence (femur, bilateral): Slice thickness 1.5 mm, FOV 348x655mm, Echo time 5.7 ms, Repetition Time 2.5 ms, number of images 88.

CT and MRI scans were performed on the same day. No intraarticular contrast was given for CT scans nor for MRI.scans.

3D measurement method

3D bone models of all included femurs were generated from CT and MRI data using the global thresholding and region growing functionality of a standard segmentation software (Mimics Medical 19.0, Materialise NV, Leuven, Belgium). Segmentation of the MRI was performed using the T1-weighted VIBE-Dixon sequence. The bone models were imported into the in-house developed surgical planning software CASPA (Balgrist CARD AG, Zurich, Switzerland) for 3D measurement of the femoral torsion (Figs. 1a and b and 2a and b).

Fig. 1
figure 1

a 3D model of the proximal femur from CT data. b The same 3D model then 1a with the sphere fitted to the femoral head (green), the cross section at the narrowest point of the femoral neck (red) and the femoral neck axis (yellow), connecting the two centers. Both figures were created by the authors using the in-house developed surgical planning software CASPA (Balgrist CARD AG, Zurich, Switzerland)

Fig. 2
figure 2

a 3D model of the proximal femur from MRI data. (Note the much less precise osseous morphology compared to the CT model shown in Fig. 1). b The same 3D model with the sphere fitted to the femoral head (green), the cross section at the narrowest point of the femoral neck (red) and the femoral neck axis (yellow), connecting the two centers. Both figures were created by the authors using the in-house developed surgical planning software CASPA (Balgrist CARD AG, Zurich, Switzerland)

3D femoral torsion measurements were performed subsequently using a method based on Kim et al. [15]. Thereby, femoral torsion is defined as the angle between the femoral neck axis and the tangent to the posterior condyles, both projected to a plane perpendicular to the anatomical axis (Fig. 3). The femoral neck axis is defined as the line connecting the center of the femoral head and the center of the cross-section at the narrowest point of the femoral neck. First, the center of the femoral head was determined by fitting a sphere to the femoral head, minimizing the distance to a user-selected region on the femoral head [38]. Second, a plane was manually fit to the narrowest diameter of the femoral neck, perpendicular to the estimated femoral neck axis. The center point of the intersection between this plane and the femoral bone was connected with the center of the femoral head, resulting in the femoral neck axis.

Fig. 3
figure 3

Femoral condylar axis (blue) and femoral neck axis (yellow) of a CT rendered image. The figure was created by the authors using the in-house developed surgical planning software CASPA (Balgrist CARD AG, Zurich, Switzerland)

The tangent to the posterior condyles was determined by manually fitting a plane to the posterior side of the femur, in a way that the plane passed through the most posterior points of the femoral condyles and the most posterior point of the greater trochanter. The condylar tangent was then defined as a line between most posterior points of the femoral condyles.

Finally, the anatomical femoral axis was defined between the centre of an axial cross-section located in the middle of the tip of the lesser and the greater trochanter and the centre of an axial cross-section just above the femoral condyles. By using the three defined axes (i.e. femoral neck axis, femoral condylar axis, and anatomical femoral axis), the 3D femoral torsion was calculated in MATLAB (Version 2019a, The MathWorks Inc., Natick MA, USA). For inter-reader reliability all 3D femoral torsion measurements were performed by two independent readers (TH and LJ).

Statistical analysis

Inter-reader and inter-modality reliability were assessed with intraclass correlation coefficients (ICC) based on a two-way random effects and a two-mixed effects model, respectively. Absolute agreement based on single measures was analyzed. These analyses were stratified by reader or by modality as applicable. The standard error of measurement was computed to yield an estimate of the expected error associated with a measurement. To test for a systematic difference in angle measurements between the modalities, a paired t-test was conducted. This test was applied on the pooled data from both readers. Intra-observer variability between CT and MRI for both readers were also assessed. Analysis was performed with SPSS (IBM SPSS Statistics for Windows, Version 26.0. Armonk, NY: IBM Corp.). P-values below 0.05 were considered statistically significant.

Results

Individual 3D measurement values of both readers from CT and MRI data for all 29 hips are shown in Table 1. The range of the measured 3D femoral torsion was from  -16.4° to 28.2°.

Table 1 3D torsional measurements from CT and MRI from reader 1 and reader 2. Values in °

Inter-modality reliability

The ICC (95% CI) expressing the inter-modality reliability for both assessed modalities was 0.950 (0.898; 0.976) (reader 1) and 0.950 (0.897; 0.976) (reader 2), respectively. The standard error of measurement (SEM) was 1.97° (reader 1) and 1.92° (reader 2), respectively (Fig. 4).

Fig. 4
figure 4

Comparison of the measurements based on the two imaging modalities assessed, for reader 1 and reader 2

The comparison of angle measurements conducted with either modality yielded a mean difference of 0.42° (MRI – CT, SD: 2.77°, p = 0.253).

Inter-reader reliability

The ICC (95% CI) expressing the inter-reader-reliability for both modalities was 0.945 (0.886; 0.973) for MRI and 0.957 (0.910; 0.979) for CT. The standard error of measurement (SEM) was 2.01° (MRI) and 1.83° (CT), respectively (Fig. 5).

Fig. 5
figure 5

Comparison of the measurements performed by the two readers for both image modalities

Intra-observer variability

The mean intra-observer variability between CT and MRI was 1.51 (reader 1) respectively 1.62 (reader 2).

Discussion

The most important finding of the present study is that consistency exists between 3D measurements of femoral torsion from MRI and CT reconstructed bone models. The mean difference of the modalities was very low with 0.42° (MRI – CT, SD: 2.77°, p = 0.253). Our hypothesis was confirmed further by an ICC for the inter-modality reliability of > 0.9 expressing an excellent reliability [39]. Additionally, inter-reader reliability was estimated to be excellent indicating high reproducibility of the measurements. The SEM was 1.97° (reader 1) respectively 1.92° (reader 2). In our opinion the MRI can still provide valuable information with a measurement error of 2°.

Three dimensional femoral torsion measurements with computer rendered CT images using the 3D modeling method described by Kim et al. is the “gold standard” [15]. On computer rendered MRI data the anatomy may be depicted less precise than on CT data, since some landmarks (i.e. the lesser trochanter) are more difficult to identify (see Figs. 1 and 2), slice thickness of MRI rotational sequences is usually higher (i.e. 1 mm for CT vs. 1.5 mm for MRI in this study) and soft tissue is more difficult to distinguish from bone and cartilaginous tissue during the segmentation process. Surprisingly, these large differences do not seem to have a significant impact on the final result of the 3D torsional measurements. MRI is more time consuming than CT (about 91 seconds vs. 5 seconds) and so there is a potential for movement of the patient’s leg between scanning the proximal femur and the condylar region and thereby influencing the torsional angle. But still the chance of complete failure of the MRI is negliable. Anyhow, this is a hypothesis of the authors and could not be proved by this study. It is therefore crucial that the rotational sequences are scanned as quickly as possible in succession. The great advantage of MRI is that it can analyze cartilage and soft tissue disorders in addition to bone conditions whereas the CT is considered to be more precise for evaluating the bony anatomy. However, previous studies have shown that MRI can be used as a precise alternative to CT for evaluation of 3D osseous morphology in the shoulder and knee [33, 36]. Moreover, the identification and location of cam-type morphology was equivalent using 3D osseous reconstructions of the femur generated from 3 T MRI and CT scans [40]. However, none of these studies could show that 3D measurement of patients’ femoral torsion could be performed from MRI data instead of CT data. To our knowledge this is the first study that calculates 3D femoral torsion using MRI data comparing it to 3D femoral torsion measurements form CT data.

Our results indicate that 3D measurement of femoral torsion can be performed using MRI data, making the routinely performed CT imaging for measuring the 3D torsional angle questionable and, thus, giving a potential to reduce healthcare resources and harmful ionizing radiation.

We acknowledge limitations of our study. The performed measurement was not analyzed for subgroups of patients with specific torsional angles (i.e. normal, too high, too low torsion). Therefore we can only state that the two measurement methods are comparable within the range tested in our collective. The number of patients in this study would be too small to analyze subgroups with “extreme” torsional angles and the goal was primarily to evaluate the accuracy of the measurement from MRI data in general and should be considered as a preliminary study that serves as proof of concept.

Further MR image segmentation was performed by a person without prior academic training in the matter but he was carefully instructed to carry out the task. Individuals with different levels of training may yield varying results.

Conclusions

Accurate 3D measurement of femoral torsion is possible from MRI images. It shows consistency compared to measurements with the “gold standard” of CT images in our collective of 29 hips.

Availability of data and materials

Anonymized source data can be obtained from the corresponding author on reasonable request.

Abbreviations

2D:

Two-dimensional

3D:

Three-dimensional

CT:

Computer Tomography

FAI:

Femoroacetabular impingement

ICC:

Intraclass correlation coefficients

MRI:

Magnetic Resonance Imaging

References

  1. Fabry G. Torsion of the femur. Acta Orthop Belg. 1977;43(4):454–9.

    CAS  PubMed  Google Scholar 

  2. Dunlap K, Shands AR Jr, Hollister LC Jr, Gaul JS Jr, Streit HA. A new method for determination of torsion of the femur. J Bone Joint Surg Am. 1953;35-A(2):289–311.

    Article  CAS  Google Scholar 

  3. Tonnis D, Heinecke A. Diminished femoral antetorsion syndrome: a cause of pain and osteoarthritis. J Pediatr Orthop. 1991;11(4):419–31. https://doi.org/10.1097/01241398-199107000-00001.

    Article  CAS  PubMed  Google Scholar 

  4. Wedge JH, Wasylenko MJ. The natural history of congenital disease of the hip. J Bone Joint Surg (Br). 1979;61-B(3):334–8. https://doi.org/10.1302/0301-620X.61B3.158025.

    Article  CAS  Google Scholar 

  5. Gelberman RH, Cohen MS, Shaw BA, Kasser JR, Griffin PP, Wilkinson RH. The association of femoral retroversion with slipped capital femoral epiphysis. J Bone Joint Surg Am. 1986;68(7):1000–7. https://doi.org/10.2106/00004623-198668070-00006.

    Article  CAS  PubMed  Google Scholar 

  6. Botser IB, Ozoude GC, Martin DE, Siddiqi AJ, Kuppuswami S, Domb BG. Femoral anteversion in the hip: comparison of measurement by computed tomography, magnetic resonance imaging, and physical examination. Arthroscopy. 2012;28(5):619–27. https://doi.org/10.1016/j.arthro.2011.10.021.

    Article  PubMed  Google Scholar 

  7. Sutter R, Dietrich TJ, Zingg PO, Pfirrmann CW. Femoral antetorsion: comparing asymptomatic volunteers and patients with femoroacetabular impingement. Radiology. 2012;263(2):475–83. https://doi.org/10.1148/radiol.12111903.

    Article  PubMed  Google Scholar 

  8. Tibor LM, Liebert G, Sutter R, Impellizzeri FM, Leunig M. Two or more impingement and/or instability deformities are often present in patients with hip pain. Clin Orthop Relat Res. 2013;471(12):3762–73. https://doi.org/10.1007/s11999-013-2918-6.

    Article  PubMed  PubMed Central  Google Scholar 

  9. Souza RB, Powers CM. Concurrent criterion-related validity and reliability of a clinical test to measure femoral anteversion. J Orthop Sports Phys Ther. 2009;39(8):586–92. https://doi.org/10.2519/jospt.2009.2996.

    Article  PubMed  Google Scholar 

  10. Maier C, Zingg P, Seifert B, Sutter R, Dora C. Femoral torsion: reliability and validity of the trochanteric prominence angle test. Hip Int. 2012;22(5):534–8. https://doi.org/10.5301/HIP.2012.9352.

    Article  PubMed  Google Scholar 

  11. Dunn DM. Anteversion of the neck of the femur; a method of measurement. J Bone Joint Surg (Br). 1952;34-B(2):181–6. https://doi.org/10.1302/0301-620X.34B2.181.

    Article  CAS  Google Scholar 

  12. Sutter R, Dietrich TJ, Zingg PO, Pfirrmann CW. Assessment of femoral antetorsion with MRI: comparison of oblique measurements to standard transverse measurements. AJR Am J Roentgenol. 2015;205(1):130–5. https://doi.org/10.2214/AJR.14.13617.

    Article  PubMed  Google Scholar 

  13. Tomczak RJ, Guenther KP, Rieber A, Mergo P, Ros PR, Brambs HJ. MR imaging measurement of the femoral antetorsional angle as a new technique: comparison with CT in children and adults. AJR Am J Roentgenol. 1997;168(3):791–4. https://doi.org/10.2214/ajr.168.3.9057536.

    Article  CAS  PubMed  Google Scholar 

  14. Kim JS, Park TS, Park SB, Kim JS, Kim IY, Kim SI. Measurement of femoral neck anteversion in 3D. Part 1: 3D imaging method. Med Biol Eng Comput. 2000;38(6):603–9. https://doi.org/10.1007/BF02344864.

    Article  CAS  PubMed  Google Scholar 

  15. Kim JS, Park TS, Park SB, Kim JS, Kim IY, Kim SI. Measurement of femoral neck anteversion in 3D. Part 2: 3D modelling method. Med Biol Eng Comput. 2000;38(6):610–6. https://doi.org/10.1007/BF02344865.

    Article  CAS  PubMed  Google Scholar 

  16. Bedi A, Dolan M, Magennis E, Lipman J, Buly R, Kelly BT. Computer-assisted modeling of osseous impingement and resection in femoroacetabular impingement. Arthroscopy. 2012;28(2):204–10. https://doi.org/10.1016/j.arthro.2011.11.005.

    Article  PubMed  Google Scholar 

  17. Heyworth BE, Dolan MM, Nguyen JT, Chen NC, Kelly BT. Preoperative three-dimensional CT predicts intraoperative findings in hip arthroscopy. Clin Orthop Relat Res. 2012;470(7):1950–7. https://doi.org/10.1007/s11999-012-2331-6.

    Article  PubMed  PubMed Central  Google Scholar 

  18. Kang RW, Yanke AB, Espinoza Orias AA, Inoue N, Nho SJ. Emerging ideas: novel 3-D quantification and classification of cam lesions in patients with femoroacetabular impingement. Clin Orthop Relat Res. 2013;471(2):358–62. https://doi.org/10.1007/s11999-012-2693-9.

    Article  PubMed  Google Scholar 

  19. Khan O, Witt J. Evaluation of the magnitude and location of cam deformity using three dimensional CT analysis. Bone Joint J. 2014;96-B(9):1167–71. https://doi.org/10.1302/0301-620X.96B9.33555.

    Article  CAS  PubMed  Google Scholar 

  20. Milone MT, Bedi A, Poultsides L, Magennis E, Byrd JW, Larson CM, et al. Novel CT-based three-dimensional software improves the characterization of cam morphology. Clin Orthop Relat Res. 2013;471(8):2484–91. https://doi.org/10.1007/s11999-013-2809-x.

    Article  PubMed  PubMed Central  Google Scholar 

  21. Amis ES Jr, Butler PF, Applegate KE, Birnbaum SB, Brateman LF, Hevezi JM, et al. American College of Radiology white paper on radiation dose in medicine. J Am Coll Radiol. 2007;4(5):272–84. https://doi.org/10.1016/j.jacr.2007.03.002.

    Article  PubMed  Google Scholar 

  22. Wylie JD, Jenkins PA, Beckmann JT, Peters CL, Aoki SK, Maak TG. Computed tomography scans in patients with young adult hip pain carry a lifetime risk of malignancy. Arthroscopy. 2018;34(1):155–63 e3. https://doi.org/10.1016/j.arthro.2017.08.235.

    Article  PubMed  Google Scholar 

  23. Cibere J, Russell D, Qian H, Forster BB, Wong H, Barber M, et al. Significant association of radiographic femoroacetabular impingement with early MRI features of hip disease: a population-based study. Osteoarthr Cartil. 2016;24:22–3.

    Article  Google Scholar 

  24. Ito K, Minka MA 2nd, Leunig M, Werlen S, Ganz R. Femoroacetabular impingement and the cam-effect. A MRI-based quantitative anatomical study of the femoral head-neck offset. J Bone Joint Surg (Br). 2001;83(2):171–6. https://doi.org/10.1302/0301-620X.83B2.0830171.

    Article  CAS  Google Scholar 

  25. James SL, Ali K, Malara F, Young D, O'Donnell J, Connell DA. MRI findings of femoroacetabular impingement. AJR Am J Roentgenol. 2006;187(6):1412–9. https://doi.org/10.2214/AJR.05.1415.

    Article  PubMed  Google Scholar 

  26. Pfirrmann CW, Mengiardi B, Dora C, Kalberer F, Zanetti M, Hodler J. Cam and pincer femoroacetabular impingement: characteristic MR arthrographic findings in 50 patients. Radiology. 2006;240(3):778–85. https://doi.org/10.1148/radiol.2403050767.

    Article  PubMed  Google Scholar 

  27. Samaan MA, Pedoia V, Zhang AL, Gallo MC, Link TM, Souza RB, et al. A novel mr-based method for detection of cartilage delamination in femoroacetabular impingement patients. J Orthop Res. 2018;36(3):971–8. https://doi.org/10.1002/jor.23667.

    Article  PubMed  Google Scholar 

  28. Tresch F, Dietrich TJ, Pfirrmann CWA, Sutter R. Hip MRI: Prevalence of articular cartilage defects and labral tears in asymptomatic volunteers. A comparison with a matched population of patients with femoroacetabular impingement. J Magn Reson Imaging. 2017;46(2):440–51. https://doi.org/10.1002/jmri.25565.

    Article  PubMed  Google Scholar 

  29. Mygind-Klavsen B, Lund B, Nielsen TG, Maagaard N, Kraemer O, Holmich P, et al. Danish hip arthroscopy registry: predictors of outcome in patients with femoroacetabular impingement (FAI). Knee Surg Sports Traumatol Arthrosc. 2019;27(10):3110–20. https://doi.org/10.1007/s00167-018-4941-3.

    Article  PubMed  Google Scholar 

  30. Chai W, Xu M, Zhang GQ, Zhang LH, Gou WL, Ni M, et al. Computer-aided design and custom-made guide in corrective osteotomy for complex femoral deformity. J Huazhong Univ Sci Technol Med Sci. 2013;33(3):398–405. https://doi.org/10.1007/s11596-013-1131-x.

    Article  Google Scholar 

  31. Fucentese SF, Meier P, Jud L, Kochli GL, Aichmair A, Vlachopoulos L, et al. Accuracy of 3D-planned patient specific instrumentation in high tibial open wedge valgisation osteotomy. J Exp Orthop. 2020;7(1):7. https://doi.org/10.1186/s40634-020-00224-y.

    Article  PubMed  PubMed Central  Google Scholar 

  32. Jud L, Vlachopoulos L, Haller TV, Fucentese SF, Rahm S, Zingg PO. The impact of mal-angulated femoral rotational osteotomies on mechanical leg axis: a computer simulation model. BMC Musculoskelet Disord. 2020;21(1):50. https://doi.org/10.1186/s12891-020-3075-1.

    Article  PubMed  PubMed Central  Google Scholar 

  33. Gyftopoulos S, Yemin A, Mulholland T, Bloom M, Storey P, Geppert C, et al. 3DMR osseous reconstructions of the shoulder using a gradient-echo based two-point Dixon reconstruction: a feasibility study. Skelet Radiol. 2013;42(3):347–52. https://doi.org/10.1007/s00256-012-1489-z.

    Article  Google Scholar 

  34. Liang X, Jacobs R, Hassan B, Li L, Pauwels R, Corpas L, et al. A comparative evaluation of cone beam computed tomography (CBCT) and multi-slice CT (MSCT) Part I. on subjective image quality. Eur J Radiol. 2010;75(2):265–9. https://doi.org/10.1016/j.ejrad.2009.03.042.

    Article  PubMed  Google Scholar 

  35. Malloy P, Gasienica J, Dawe R, Espinoza Orias AA, Nwachukwu BU, Inoue N, et al. 1.5 T magnetic resonance imaging generates accurate 3D proximal femoral models: surgical planning implications for femoroacetabular impingement. J Orthop Res. 2020;38(9):2050–6. https://doi.org/10.1002/jor.24596.

    Article  CAS  PubMed  Google Scholar 

  36. Neubert A, Wilson KJ, Engstrom C, Surowiec RK, Paproki A, Johnson N, et al. Comparison of 3D bone models of the knee joint derived from CT and 3T MR imaging. Eur J Radiol. 2017;93:178–84. https://doi.org/10.1016/j.ejrad.2017.05.042.

    Article  PubMed  Google Scholar 

  37. Rathnayaka K, Momot KI, Noser H, Volp A, Schuetz MA, Sahama T, et al. Quantification of the accuracy of MRI generated 3D models of long bones compared to CT generated 3D models. Med Eng Phys. 2012;34(3):357–63. https://doi.org/10.1016/j.medengphy.2011.07.027.

    Article  PubMed  Google Scholar 

  38. Schneider PJ, Eberly DH. Geometric tools for computer graphics. Burlington: Morgan Kaufmann Publishers; 2003.

    Google Scholar 

  39. Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med. 2016;12(2):155–63.

    Article  Google Scholar 

  40. Samim M, Eftekhary N, Vigdorchik JM, Elbuluk A, Davidovitch R, Youm T, et al. 3D-MRI versus 3D-CT in the evaluation of osseous anatomy in femoroacetabular impingement using Dixon 3D FLASH sequence. Skelet Radiol. 2019;48(3):429–36. https://doi.org/10.1007/s00256-018-3049-7.

    Article  Google Scholar 

Download references

Acknowledgements

The authors would like to thank Tabitha Roth from the ROCS Team for her help with the 3D measurement technique and Reto Sutter, MD from the radiology department for his support.

Funding

No funding was received for conducting this study.

Author information

Authors and Affiliations

Authors

Contributions

TVH, PS and POZ designed the study. TVH, AH and LJ performed the measurements and data acquisition. TVH wrote the manuscript. TVH and AH prepared the graphics. TG performed the statistical analysis. PS, TG and POZ revised the manuscript critically. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Thomas Vincent Häller.

Ethics declarations

Ethics approval and consent to participate

The local ethical committee approved this study (Zurich Cantonal Ethics Commission, BASEC Number 2012–02242). Informed written consent was obtained from all individual participants included in the study.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions