Augmented reality in the operating room: a clinical feasibility study

Abstract

Background

Augmented Reality (AR) is a rapidly emerging technology finding growing acceptance and application in different fields of surgery. Various studies have been performed evaluating the precision and accuracy of AR guided navigation. This study investigates the feasibility of a commercially available AR head mounted device during orthopedic surgery.

Methods

Thirteen orthopedic surgeons from a Swiss university clinic performed 25 orthopedic surgical procedures wearing a holographic AR headset (HoloLens, Microsoft, Redmond, WA, USA) providing complementary three-dimensional, patient specific anatomic information. The surgeon’s experience of using the device during surgery was recorded using a standardized 58-item questionnaire grading different aspects on a 100-point scale with anchor statements.

Results

Surgeons were generally satisfied with image quality (85 ± 17 points) and accuracy of the virtual objects (84 ± 19 point). Wearing the AR device was rated as fairly comfortable (79 ± 13 points). Functionality of voice commands (68 ± 20 points) and gestures (66 ± 20 points) provided less favorable results. The greatest potential in the use of the AR device was found for surgical correction of deformities (87 ± 15 points). Overall, surgeons were satisfied with the application of this novel technology (78 ± 20 points) and future access to it was demanded (75 ± 22 points).

Conclusion

AR is a rapidly evolving technology with large potential in different surgical settings, offering the opportunity to provide a compact, low cost alternative requiring a minimum of infrastructure compared to conventional navigation systems. While surgeons where generally satisfied with image quality of the here tested head mounted AR device, some technical and ergonomic shortcomings were pointed out. This study serves as a proof of concept for the use of an AR head mounted device in a real-world sterile setting in orthopedic surgery.

Peer Review reports

Background

Precise positioning of mechanical elements including screws, surgical guides, prosthetic components and anchors to restore anatomy and function remains a desirable goal in orthopedic surgery [1]. Image-based intraoperative techniques such as two-dimensional and three-dimensional (3D) fluoroscopy or Computed tomography (CT) -based navigation increase the precision but also radiation exposure to the patient and operating room personnel [2,3,4,5,6,7,8]. Other promising methods of intraoperative navigation techniques including mechanical drilling aids or CAD-designed and 3D-printed patient-specific instruments are cost intense and may require prolonged preoperative preparation and planning [9,10,11].

Augmented Reality (AR) is a rapidly emerging technology providing the user with computer-generated information superimposed to real-world environment. The user’s field of view is transformed into a display where real-world objects can be complemented with virtual data. This way information including drilling axis or cutting planes can be projected directly onto the patients’ anatomy within the surgical field. Although its application in orthopedic surgery today remains limited, AR was gradually introduced in different experimental medical and surgical settings [12,13,14,15,16]. Recent studies demonstrated that AR may improve accuracy, safety and efficacy of surgical procedures [17, 18]. In order to ultimately provide a benefit for the patient this novel technology is required to support the surgeon in his decision making and be sufficiently comfortable to be worn over long periods of time. Advancements in information technology and hardware manufacturing transformed former bulky and cable-bound AR headsets into ergonomic devices fulfilling strict requirements of ergonomic design [19]. However, as a result of the novelty of this technology and the lack of widespread application, data considering acceptance of surgeons is missing. We therefore performed a prospective clinical feasibility study in a real world-environment where orthopedic surgeons used a commercially available holographic headset (HoloLens I, Microsoft, Redmond, WA, USA) during surgery (Fig. 1). The surgeon’s experience of using the device during surgery was recorded using a standardized questionnaire.

Fig. 1
figure 1

Augmented reality headset (Hololens I, Microsoft, Redmond, WA, USA)

Methods

Orthopedic surgeons from a single Swiss university clinic where the subject of this prospective study. Surgeons were asked to wear a commercially available holographic AR headset (HoloLens, Microsoft, Redmond, WA, USA) during surgery after receiving an introduction explaining its functionality and time to test voice commands and hand gestures.

For each operation, a three-dimensional triangular surface model of the patients readily available CT data was generated using a commercial software (Siemens syngo.via Frontier 3D printing V 1.0.0, Siemens Healthineers, Erlangen, Germany or Materialise Mimics V 19.0, Leuven, Belgium) [20]. Only already available data was used, and no additional CT scans were performed for this study. The three-dimensional model (Fig. 2) was edited with the Unity software package (Version 5.5, Unity Technologies, San Francisco, CA, USA). An application that permitted interactive rotation and translation of the three-dimension model by voice commands and contactless hand gestures was implemented using Microsoft Visual Studio 2015 (Microsoft, Redmond, WA, USA).

Fig. 2
figure 2

Three-dimensional model of a single vertebra generated from CT data

During the operation, the three-dimensional model of the patients CT-data was displayed holographically in the surgeon’s field of view in addition to routinely mounted radiological imaging on conventional screens. The displayed surface model could be moved, rotated, scaled and placed using contactless hand gestures and voice commands (Fig. 3). The timing and duration of the intraoperative use of the head mounted device was at the surgeon’s discretion (Fig. 4).

Fig. 3
figure 3

Surgeon wearing an augmented reality headset during surgery for shoulder arthroplasty using three-dimensional anatomic data supplied by the augmented reality headset

Fig. 4
figure 4

View from the surgeon’s perspective. Both forearms are displayed as 3D virtual objects during corrective forearm osteotomy

The device was evaluated using a standardized 58-item questionnaire administered as an electronic survey using the REDCap data capture tool [21]. The survey used a 100-point scale with anchor statements (1: not useful at all, to 100: very useful) to grade usability of the device, intraoperative implementation, benefit during the procedure, future potential of AR technology in general and considering orthopedic subspecialties as well as evaluation of the device itself. Further, demographic data considering the participating surgeons were recorded. Surgeons where also asked about the future potential of AR technology. A translated English version of this survey originally supplied to the participants in German is available as a supplementary file.

Statistical analysis

Descriptive analysis was performed using SPSS (IBM Corp. Released 2017. IBM SPSS Statistics for Windows, Version 25.0. Armonk, NY: IBM Corp.). Continuous variables are reported as mean ± standard deviation (SD).

Results

Thirteen orthopedic surgeons from different subspecialties (all male, mean age 40 ± 7 years) participated in this study. Four surgeons performed > 500 cases per year, 3 surgeons 251–500 and 6 surgeons 51–250 surgeries per year. Six surgeons were in the position of chief of service and 7 surgeons were attending physicians or fellows. In total, 25 surgeries (10 spine, 4 shoulder, 5 knee, 1 hand and 5 ft) were included in this study.

The general acceptance of the AR device was good (Fig. 5). Wearing the AR device was rated as fairly comfortable (79 ± 13 points). Weight and size of the device were rated with 64 ± 23 points.

Fig. 5
figure 5

Boxplot showing general acceptance of the AR headset

Considering usability (Fig. 6), surgeons found the device to produce good image quality (85 ± 17 points) and a good accuracy of virtual objects (84 ± 19 point). In contrast, functionality of voice commands (68 ± 20 points) and gestures (66 ± 20 points) provided less favorable results. Learnability of the voice commands (74 ± 23 points) and gestures (73 ± 20 points) were well rated. Voice control most frequently failed due to a lack of understanding of commands by the device (8/25) or because of noise in the surroundings (8/25). Overall satisfaction with the speed of the application was rated 69 ± 18 points.

Fig. 6
figure 6

Boxplot showing the usability of the AR headset within the operating room

Most surgeons used the AR device for less than 60 min (23/25 surgeries) and no surgeon used the device for the entire surgery. All but three surgeons used the AR device only for preoperatively defined surgical steps (Fig. 7).

Fig. 7
figure 7

Bar chart showing wearing time of augmented reality headset during surgery

The most frequent reasons for early termination of use were impaired sight caused by the device (3/25) and surgical steps requiring other instruments (3/25) (Fig. 8). However, most surgeons estimated that they could wear the AR device for more than 60 min.

Fig. 8
figure 8

Bar chart showing reason for not wearing the AR headset during hole procedure

The greatest potential in the use of the AR device was found for surgical correction of deformities (87 ± 15 points), osteotomies (82 ± 17 points), revision surgery (77 ± 22 points), and tumor surgery (77 ± 21 points), whereas lowest potential was found for arthroscopic (32 ± 25 points) and reconstructive surgery (41 ± 22 points) (Fig. 9).

Fig. 9
figure 9

Boxplot showing the future potential for AR technology referring to surgical techniques in orthopedic surgery

Potential for future use within the orthopedic subspecialties was highest estimated for application in spine (86 ± 13 points) and pelvic surgery (88 ± 13 points).

The greatest benefits of AR technology were expected to be increased intraoperative accuracy (78 ± 23 points), improved surgical outcome (77 ± 22 points) and reduction of exposure to radiation (72 ± 26 points) (Fig. 10).

Fig. 10
figure 10

Boxpot showing expectations in augmented reality for surgical applications after usage

Overall, surgeons were satisfied with the application of this novel technology (78 ± 20 points) and future access to it was demanded (75 ± 22 points).

Discussion

This prospective clinical study offers a proof of concept of the clinical feasibility of a wearable augmented reality device for using three-dimensional virtual object reconstructions of individual patients’ anatomy as an intraoperative aid during orthopedic surgical procedures. Thirteen surgeons from different orthopedic subspecialties used an AR headset as an additional display providing complementary three-dimensional, patient specific anatomic information during 25 surgical procedures. The surgeons experience with the device was recorded using a standardized questionnaire.

Augmented reality is rapidly developing technology that can display a vast variety of information within the field of view of the user using a compact device with minimum infrastructure. Spatial mapping and recognition of the physical environment is a crucial difference between augmented reality and other navigations systems using conventional display methods. Projecting the required information directly into the line-of sight of the surgeon is considered the natural progression of these well-established methods mitigating the errors associated with attention shift by directly projecting the navigation guidance onto the surgical field [22, 23]. Incekara et al. evaluated the same AR headset for pre-operative planning of neurosurgical tumor resection in 25 patients. The authors reported that surgeons did benefit in terms of maintaining attention and focus on the patient, improved ergonomics and improved understanding of tumor-brain/ skull relationship due to direct three-dimensional holographic representation. In contrast to current neuronavigational systems, the HoloLens AR head mounted device subjectively improved ergonomics during surgical planning [24]. While projecting information directly into field of view of the surgeon is considered to mitigate errors associated with attention shift, the fixed focal distance of two meters of the virtual objects projected by the HoloLens could induce perceptual conflicts as focal rivalry and vergence-accommodation mismatch [23, 25]. However, because virtual objects projected in the here tested scenario were not directly overlayed with the patient anatomy, focal rivalry did not play an important role in this analysis.

The here used iteration of the HoloLens offers a rather narrow diagonal angle of view of 34° for projection of virtual objects. However, for most surgical tasks requiring high precision, a narrow field of view is sufficient and this study, surgeons were generally satisfied with the quality of the displayed three-dimensional images.

The general acceptance of the device was fairly good. Considering comfort, the mostly frontloaded weight of 579 g and the size of this first iteration of the HoloLens was considered as a shortcoming. During 6 of 25 cases, the device was worn for more than 30 min, and in another 7 of 25 cases, for 15–30 min. However, despite its weight and size, most surgeons participating in this study estimated that the device could be worn for more than an hour when considering comfort. Other authors investigating the usability of the HoloLens during neurosurgical procedures graded the comfort more favorable [24, 26]. This might be attributed to the fact that surgical subspecialties more accustomed to the use of optical aids in their field of view including magnification glasses might also be less disturbed be head mounted AR devices. In this study, the small number of surgeons wearing the device for more than 1 h (n = 2) might also be explained by the fact that participants were only supplied with supplementary information required for essential steps of the surgery. Hence, the threshold for removing the device, in particular while performing the surgical approach and during wound closure was low.

The questionnaire also revealed technical limitations of the device, namely the voice command and gesture function. Using voice commands as a mean of interaction with the headset was less favorably graded. In most cases the recognition of voice commands was impaired by loud environment in the operation room. Other reports analyzing the usage of the HoloLens device for different surgical specialties reported more stable results using voice commands [26]. This difference in outcome might be attributed to the variety of noise emitting tools and instruments utilized in orthopedic surgery. However, further advances in technology will surely overcome these limitations and will lead to improved ergonomics and wearability.

Surgeons saw the biggest potential for this technology in complex surgical procedures including revision surgery, deformity correction, tumor surgery and trauma where typical anatomical landmarks are frequently not easily identifiable. However, a crucial step to provide reliable navigation in complex surgery is the registration process, the automated and precise overlay of virtual information with the real environment. Various methods of registration have been described including ultrasound-based techniques, reflective markers and non-invasive skin placed markers [22, 23, 27]. Automated registration of the patient’s anatomy is currently the focus of various projects in augmented reality research [16, 17, 28, 29]. Intraoperative manual surface digitization or machine learning based object detection offer the possibility to establish a correspondence between preoperatively acquired image data and intraoperative anatomy without further requiring intraoperative imaging [29, 30]. Further, the quality of alignment of the three-dimensional objects to the real world not only relies on the accuracy of the registration but also on the calibration of the optical see-through head-mounted display. Therefore, different calibrations methods to map the physical reality to the virtual scene have been proposed [31, 32]. However, in an experimental setting, navigated pedicle instrumentation with augmented reality-based head mounted devices achieved comparable results to commercially available navigation systems when using 3D image-based registration [33]. While this study is the first to describe ergonomic aspects and feasibility of using the HoloLens in a sterile setting in orthopedic surgery, its main limitation is the small number of procedures performed and the short time each surgeon spent with the device. Also, as a result of technical shortcomings not yet overcome, the full potential of AR as a mean to actually navigate the described procedures with pre-planned cutting planes or screw trajectories could not be demonstrated. However, a clinical study accounting for these deficits is planned by this study group. This study might also be prone to selection bias as recruiting was performed on a voluntary basis possibly including subjects more enthusiastic for novel technology.

Conclusion

In summary, AR is a rapidly evolving technology with large potential in different surgical settings, offering the opportunity to provide a compact, low cost alternative requiring a minimum of infrastructure compared to conventional navigation systems. While surgeons where generally satisfied with image quality of the here tested head mounted AR device, some technical and ergonomic shortcomings were pointed out. This study serves demonstrates the feasibility of the use of an AR head mounted device in a real-world sterile setting in orthopedic surgery.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

AR:

Augmented reality

CT:

Computed tomography

SPSS:

Statistical Package for Social Science

References

  1. Keeney JA. Innovations in total knee arthroplasty: improved technical precision, but unclear clinical benefits. Orthopedics. 2016;39:217–20. https://doi.org/10.3928/01477447-20160628-03.

    Article  PubMed  Google Scholar 

  2. Gelalis ID, Paschos NK, Pakos EE, et al. Accuracy of pedicle screw placement: A systematic review of prospective in vivo studies comparing free hand, fluoroscopy guidance and navigation techniques. Eur Spine J. 2012;21:247–55.

    Article  Google Scholar 

  3. Laudato PA, Pierzchala K, Schizas C. Pedicle screw insertion accuracy using O-arm, robotic guidance, or freehand technique. Spine. 2018;43:E373–8. https://doi.org/10.1097/BRS.0000000000002449.

    Article  PubMed  Google Scholar 

  4. Staartjes VE, Klukowska AM, Schröder ML. Pedicle screw revision in robot-guided, navigated, and freehand thoracolumbar instrumentation: a systematic review and meta-analysis. World Neurosurg. 2018;116:433–443.e8.

    Article  Google Scholar 

  5. Perna F, Borghi R, Pilla F, et al. Pedicle screw insertion techniques: an update and review of the literature. Musculoskelet Surg. 2016;100:165–9.

    Article  CAS  Google Scholar 

  6. Kosmopoulos V, Schizas C. Pedicle screw placement accuracy: a meta-analysis. Spine. 2007;32:1. https://doi.org/10.1097/01.brs.0000254048.79024.8b.

    Article  Google Scholar 

  7. Narain AS, Hijji FY, Yom KH, et al. Radiation exposure and reduction in the operating room: perspectives and future directions in spine surgery. World J Orthop. 2017;8:524–30.

    Article  Google Scholar 

  8. Tian NF, Huang QS, Zhou P, et al. Pedicle screw insertion accuracy with different assisted methods: a systematic review and meta-analysis of comparative studies. Eur Spine J. 2011;20:846–59.

    Article  Google Scholar 

  9. Farshad M, Betz M, Farshad-Amacker NA, Moser M. Accuracy of patient-specific template-guided vs. free-hand fluoroscopically controlled pedicle screw placement in the thoracic and lumbar spine: a randomized cadaveric study. Eur Spine J. 2017;26:738–49. https://doi.org/10.1007/s00586-016-4728-5.

    Article  PubMed  Google Scholar 

  10. Lu S, Xu YQ, Zhang YZ, et al. Rapid prototyping drill guide template for lumbar pedicle screw placement. Chinese J Traumatol English Edition. 2009;12:177–80. https://doi.org/10.3760/cma.j.issn.1008-1275.2009.03.011.

    Article  Google Scholar 

  11. Mac-Thiong JM, Labelle H, Aubin CÉ. Thoracic pedicle screw insertion using a Transpedicular drill guide: a preliminary study. J Spinal Disord Tech. 2004;17:29–32. https://doi.org/10.1097/00024720-200402000-00007.

    Article  PubMed  Google Scholar 

  12. Birkfellner W, Figl M, Huber K, et al. A head-mounted operating binocular for augmented reality visualization in medicine - design and initial evaluation. IEEE Trans Med Imaging. 2002;21:991–7. https://doi.org/10.1109/TMI.2002.803099.

    Article  PubMed  Google Scholar 

  13. Chen X, Xu L, Wang Y, et al. Development of a surgical navigation system based on augmented reality using an optical see-through head-mounted display. J Biomed Inform. 2015;55:124–31. https://doi.org/10.1016/j.jbi.2015.04.003.

    Article  PubMed  Google Scholar 

  14. Gavaghan KA, Peterhans M, Oliveira-Santos T, Weber S. A portable image overlay projection device for computer-aided open liver surgery. IEEE Trans Biomed Eng. 2011;58:1855–64. https://doi.org/10.1109/TBME.2011.2126572.

    Article  PubMed  Google Scholar 

  15. Traub J, Stefan P, Heining SM, et al. Hybrid navigation interface for orthopedic and trauma surgery. Med Image Comput Comput Assist Interv. 2006;9:373–80. https://doi.org/10.1007/11866565_46.

    Article  PubMed  Google Scholar 

  16. Watanabe E, Satoh M, Konno T, et al. The trans-visible navigator: a see-through Neuronavigation system using augmented reality. World Neurosurg. 2016;87:399–405. https://doi.org/10.1016/j.wneu.2015.11.084.

    Article  PubMed  Google Scholar 

  17. Gibby JT, Swenson SA, Cvetko S, et al. Head-mounted display augmented reality to guide pedicle screw placement utilizing computed tomography. Int J Comput Assist Radiol Surg. 2019;14:525–35.

    Article  Google Scholar 

  18. Vávra P, Roman J, Zonča P, et al. Recent development of augmented reality in surgery: a review. J Healthc Eng. 2017;2017:4574172.

    Article  Google Scholar 

  19. Martin JL, Norris BJ, Murphy E, Crowe JA. Medical device development: the challenge for ergonomics. Appl Ergon. 2008;39:271–83.

    Article  Google Scholar 

  20. Gelaude F, vander Sloten J, Lauwers B. Accuracy assessment of CT-based outer surface femur meshes. Computer Aided Surg. 2008;13:188–99. https://doi.org/10.3109/10929080802195783.

    Article  CAS  Google Scholar 

  21. Harris PA, Taylor R, Minor BL, et al. The REDCap consortium: building an international community of software platform partners. J Biomed Inform. 2019;95:103208.

    Article  Google Scholar 

  22. Molina CA, Theodore N, Ahmed AK, et al. Augmented reality-assisted pedicle screw insertion: a cadaveric proof-of-concept study. J Neurosurg Spine. 2019;31:1–8. https://doi.org/10.3171/2018.12.SPINE181142.

    Article  Google Scholar 

  23. Elmi-Terander A, Burström G, Nachabe R, et al. Pedicle screw placement using augmented reality surgical navigation with intraoperative 3D imaging: a first in-human prospective cohort study. Spine. 2019;44:517–25. https://doi.org/10.1097/BRS.0000000000002876.

    Article  PubMed  Google Scholar 

  24. Incekara F, Smits M, Dirven C, Vincent A. Clinical feasibility of a wearable mixed-reality device in neurosurgery. World Neurosurg. 2018;118:e422–7. https://doi.org/10.1016/j.wneu.2018.06.208.

    Article  PubMed  Google Scholar 

  25. Condino S, Carbone M, Piazza R, et al. Perceptual Limits of Optical See-Through Visors for Augmented Reality Guidance of Manual Tasks. IEEE Trans Biomed Eng. 2019;67:411. https://doi.org/10.1109/TBME.2019.2914517.

    Article  PubMed  Google Scholar 

  26. Kubben P, Sinlae R. Feasibility of using a low-cost head-mounted augmented reality device in the operating room. Surg Neurol Int. 2019;10:26. https://doi.org/10.4103/sni.sni_228_18.

    Article  PubMed  PubMed Central  Google Scholar 

  27. Ma L, Zhao Z, Chen F, et al. Augmented reality surgical navigation with ultrasound-assisted registration for pedicle screw placement: a pilot study. Int J Comput Assist Radiol Surg. 2017;12:2205–15. https://doi.org/10.1007/s11548-017-1652-z.

    Article  PubMed  Google Scholar 

  28. Cabrilo I, Bijlenga P, Schaller K. Augmented reality in the surgery of cerebral aneurysms: a technical report. Neurosurgery. 2014;10:252–61. https://doi.org/10.1227/NEU.0000000000000328.

    Article  PubMed  Google Scholar 

  29. Liebmann F, Roner S, von Atzigen M, et al. Pedicle screw navigation using surface digitization on the Microsoft HoloLens. Int J Comput Assist Radiol Surg. 2019;14:1157–65. https://doi.org/10.1007/s11548-019-01973-7.

    Article  PubMed  Google Scholar 

  30. von Atzigen M, Liebmann F, Hoch A, et al. HoloYolo: a proof-of-concept study for marker-less surgical navigation of spinal rod implants with augmented reality and on-device machine learning. Int J Med Robot Comp Assisted Surg. 2021;17:1–10. https://doi.org/10.1002/rcs.2184.

    Article  Google Scholar 

  31. Hu X, Baena FRY, Cutolo F. Alignment-free offline calibration of commercial optical see-through head-mounted displays with simplified procedures. IEEE Access. 2020;8:223661–74. https://doi.org/10.1109/ACCESS.2020.3044184.

    Article  Google Scholar 

  32. Sun Q, Mai Y, Yang R, et al. Fast and accurate online calibration of optical see-through head-mounted display for AR-based surgical navigation using Microsoft HoloLens. Int J Comput Assist Radiol Surg. 2020;15:1907–19. https://doi.org/10.1007/s11548-020-02246-4.

    Article  PubMed  Google Scholar 

  33. Müller F, Roner S, Liebmann F, et al. Augmented reality navigation for spinal pedicle screw instrumentation using intraoperative 3D imaging. Spine J. 2020;20:621–8. https://doi.org/10.1016/j.spinee.2019.10.012.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This project was funded by the University Hospital Balgrist. No other funds, grants, or other support was received.

Author information

Authors and Affiliations

Authors

Contributions

The manuscript was authored by C. D and D.B. The manuscript was critically revised by M.F. and J. S who also approved the final version. A-G.S., P.F. and T.G. conducted the experiment and collected the data. The authors read and approved the final manuscript.

Corresponding author

Correspondence to David E. Bauer.

Ethics declarations

Ethics approval and consent to participate

Approval of the local ethics committee has been obtained, Cantonal Ethics Committee Zurich, BASEC 2018–00456.

Consent for publication

Not applicable.

Competing interests

MF is a board member and shareholder of Increment AG, a University Hospital Balgrist startup developing augmented reality solutions for medical care. Otherwise there are no conflicts of interest to disclose.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Dennler, C., Bauer, D.E., Scheibler, AG. et al. Augmented reality in the operating room: a clinical feasibility study. BMC Musculoskelet Disord 22, 451 (2021). https://doi.org/10.1186/s12891-021-04339-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-021-04339-w

Keywords