Reliability of two different measuring techniques with computer tomography for penetration and distribution of cement in the proximal tibia after total knee arthroplasty

Abstract

Background

To evaluate the reliability of two different techniques for measuring penetration and distribution of the cement mantle in the proximal tibia after total knee arthroplasty (TKA) with Computer Tomography (CT) in vivo.

Methods

Standardized CT scans of the proximal tibia were taken 1 to 2 years after total knee arthroplasties implanted with a surface cementing technique. These prospectively acquired transversal CT images of the surface of the proximal tibia were divided into four quadrants and were assessed once manually and once with a numerical computing program (MATLAB® Update 2, The MathWorks, Inc.) based on Hounsfield Units by one of the researchers. The assessments were repeated by the same and a second researcher. The ratio cement/trabecular bone was calculated 1, 3 and 5 mm distal of the tibia tray per quadrant. Kruskall-Wallis tests with multiple pairwise comparisons (Dunn’s test) were used to determine differences between the quadrants. Intra- and inter-rater reliability as well as the inter method reliability were assessed with the Intraclass Correlation Coefficient (ICC) per level of depth and with Bland-Altman plots.

Results

A total of 92 CT scans were included. The intra- and inter-rater reliability of the manual method ranged from 0.22 and 0.52. The intra- and inter-rater reliability of the matlab method varied between 0.98 to 0.99.

The median percentage cement measured with the matlab method 1 mm underneath the tibial tray varied between 82 and 88%; at 3 mm depth between 38 and 54% and at 5 mm between 15 and 25%. There was significantly (p < 0.05) less cement in the antero-medial quadrant compared to the antero-lateral and postero-lateral quadrant at 3 mm and 5 mm depth.

Conclusions

Distribution and penetration of cement in the proximal tibia in a total knee arthroplasty can be measured reliably with CT in combination with the matlab method presented in this manuscript. This method can be used for clinical purposes as well as for scientific research.

Trial registration

METC-nr: 06–104 Dossier NL14807.098.06/versie 06.

Peer Review reports

Background

Total Knee Arthroplasty (TKA) is a very successful orthopedic procedure. Numerous studies have confirmed excellent results, with 10 to 15 year survivorship of more than 90% [1,2,3,4]. Nevertheless, there are still patients with complaints, and in a large number of cases these complaints are due to aseptic loosening of the prosthesis [5, 6]. Aseptic loosening occurs mostly at the tibia component and might be caused by suboptimal fixation of the prosthesis. Knee prostheses fixated with bone cement (Polymethylmethacrylaat = PMMA) have equally good or even better results regarding aseptic loosening and clinical outcome than knee prostheses fixated without bone cement [7, 8]. The key to optimizing the interfacial strength is achieving and maintaining maximal infiltration of cement into the bone to obtain large inter-digitation and a large contact area [9, 10].

Several studies concluded that optimal fixation requires penetration of cement into the proximal tibia of 3 to 4 mm [10,11,12]. By contrast, recent in vitro research stated that a mean penetration of 1.1 mm is sufficient and that it is not necessary to achieve a penetration of more than 2 mm to have a good pull-out strength [13]. In vivo, however, the tibial tray is not only subject to pull-out forces but also and may be even more so to compression and shearing forces. The volume of cement penetrating into the bone is less predictive of failure than the mean penetration, which indicates the importance of a good distribution of the cement [13].

So far, there is no good method for evaluating the penetration and distribution of the cement in vivo. Until now, the penetration depth of the cement is measured on conventional radiographs with a ruler. This is a rough method and it only allows measurements in the coronal or sagittal plane and not in the transverse plane. The distribution of the cement cannot be visualized with conventional radiography, nor is it possible to obtain a three-dimensional image with this technique. By contrast, in a cadaver model, Computer Tomography (CT) proved to be a reliable procedure to assess the cement mantle in the proximal tibia, with the metal prosthesis in situ [14, 15]. So far, however, this procedure has not been evaluated in vivo. To our knowledge, no clinical studies about the validation of CT-based cement penetration depth and distribution assessment in TKA have been published yet.

We hypothesized that Computer Tomography (CT) is a reliable and adequate method for in vivo assessment of the penetration and distribution of the cement in the proximal tibia after cementing a total knee arthroplasty. The secondary aim was to study if assessment of the cement with a numerical computing method is just as reliable as a manual method. Furthermore, the third aim of the study was to study the cement penetration and cement distribution underneath the tibial tray in vivo.

Methods

Two hundred fifty seven patients were initially asked, prior to surgery, to participate in a prospective clinical study that evaluated the clinical and radiographic results of two different approaches for primary TKA, the mini mid-vastus and the conventional approach [16]. In total 97 patients (100 TKA) were included in that clinical study. Part of the radiographic follow-up of that clinical study was a CT scan 1 to 2 years post-operative. For this CT scan study, we included 92 CT scans (Fig. 1). We used these CT scans in this study for assessing the cement mantle of the tibial tray of the total knee prosthesis.

Fig. 1
figure 1

Flow diagram of included patients. BMI = body mass index, TKA = total knee arthroplasty

Inclusion criteria of the prospective clinical study were patients from 18 to 75 years of age with a Body Mass Index (BMI) of less than 35 kg/m2 who were diagnosed with osteoarthritis and indicated for a primary TKA. Exclusion criteria were the need of another joint replacement within 6 months, severe varus or valgus malalignment (> 15 degrees), extreme pre-existing knee deformities, previous knee surgery except arthroscopy, and unwillingness to cooperate in this study.

We started this CT scan study between 1 and 2 years after patients underwent surgery for primary TKA in the prospective clinical study. The clinical and the CT scan study were both approved by the local Medical Ethics Committee (NL14807.098.06) and all patients gave written consent.

Surgery

The surgeries were performed by two experienced orthopedic knee surgeons. All patients were operated under general or spinal anesthesia. Antibiotics were administered for 24 h and a dynamic leg holder and a tourniquet were used. All patients received the NexGen® LPS- High Flex Fixed Bearing Knee (Zimmer Biomet®, Warsaw, IN, USA). A surface cementing technique was applied using Palacos R + G® bone cement (Zimmer Biomet®, Warsaw, IN, USA). In this technique, the cement is applied at the undersurface of the tibial tray only, and the stem is placed press fit to achieve fixation (see Fig. 2).

Fig. 2
figure 2

Tibial tray just before introducing into the proximal tibia with cement applied by surface cementing technique

Computer tomography

Standardized CT scans were performed 1 to 2 years post-operatively with a Toshiba Aquilion® 32 slice scanner. The CT scanner was set at 135 kV, 150 mA and a slice thickness of 0.5 mm using an osseous mixed filter. Reconstructions were made with a helical pitch 21 and image thickness reconstruction of 2 mm. To do all measurements in the same way the CT images were re-sampled perfectly parallel to the tibia baseplate.

Data analysis

The CT reconstructions were assessed using the graphics module CARESTREAM Vue PACS (Picture Archiving and Communication System, Carestream.com). This PACS graphics module was used to draw the outlines of the Range of Interest by the researchers (see for illustration the orange line in Fig. 3). The assessments were done in the CT-slices 1, 3 and 5 mm below the slice with the last visual metal from the tibial baseplate.

Fig. 3
figure 3

a CT scan 1 mm below the tibial tray. There is an equal distribution of the cement. The yellow line is the projection of the circumference of the tibial tray. The orange line is the circumference of the area with cemented trabecular bone in the Postero-Medial quadrant. A-M = Antero-Medial, A-L = Antero-Lateral, P-M = Postero-Medial, P-L = Postero-Lateral. b CT scan 3 mm below the tibial tray. There is less cement in the periphery of all the three quarters than in the CT slice 1 mm below the tibial tray (Fig. 3a). The yellow line is the projection of the circumference of the tibial tray. The orange line is the circumference of the area with cemented trabecular bone in the Postero-Medial quadrant. A-M = Antero-Medial, A-L = Antero-Lateral, P-M = Postero-Medial, P-L = Postero-Lateral. c CT scan 5 mm below the tibial tray. There is overall less cement in the trabecular bone than in the CT slices 1 mm and 3 mm below the tibial tray. There is less cement in the anterior quadrants than in the posterior quadrants. The yellow line is the projection of the circumference of the tibial tray. The orange line is the circumference of the area with cemented trabecular bone in the Postero-Medial quadrant. A-M = Antero-Medial, A-L = Antero-Lateral, P-M = Postero-Medial, P-L = Postero-Lateral

After drawing purely visual and with manual contouring (manual method) the different Ranges of Interest (prosthesis, trabecular bone and cemented trabecular bone), the PACS graphics module measured these areas in square millimeters.

Also, the CT reconstructions were assessed with a numerical computing program (MATLAB® Update 2, The MathWorks, Inc.) based on Hounsfield Units (HU) (matlab method, see additional files for the matlab scripts). The HU was set between 400 and 550 HU for trabecular bone, between 550 and 1600 HU for cement or cemented trabecular bone, 1600–2110 for cortical bone and > 2100 HU for prosthesis [14, 15].

The area of the proximal tibia was divided into 4 quadrants, the antero-medial (A-M), the antero-lateral (A-L), the postero-medial (P-M) and the postero-lateral (P-L) quadrant to examine the distribution of the cement in those quadrants in particular for all levels of dept.

The matlab method subtracted the area of the prosthesis and the area of cortical bone from the total area. The area of cemented and uncemented trabecular bone was defined as 100%. Of this surface we calculated the areas of trabecular bone with or without cement as a part of it in percentages. In this way, we could examine the cement penetration per quadrant into the trabecular bone of the proximal tibia at a depth of 1, 3 and 5 mm in percentages.

The measurements, both the manual method and the matlab method, were done by one experienced researcher. The first 50 consecutive measurements were repeated by the same researcher, who was blinded with regard to the first assessments and also by a second researcher.

Statistics

Analyses were performed using SPSS® Statistics for Windows Version 21, Chicago, IL, USA. Intra- and inter-rater reliability as well as the inter-method reliability were assessed with the Intraclass Correlation Coefficient (ICC) per level of depth and with Bland-Altman plots. Kruskal-Wallis tests with multiple pairwise comparisons (Dunn’s test) were used to determine differences between the quadrants. Significance values for the pairwise comparisons were adjusted by the Bonferroni correction for multiple tests. The level of significance was set at p < 0.05.

Results

From August 2007 to June 2009, 257 patients were assessed for eligibility, 97 of whom were included in the prospective clinical study (see Fig. 1). Four of the included patients could not be included in the CT scan study: 2 patients had died from cancer, 1 patient moved abroad and in 1 patient the TKA had been revised for persistent pain below the medial tibia plateau. Consequently, 93 patients were asked to participate in the CT scan study, 3 of whom had a bilateral TKA, and 89 (92 TKAs) gave informed consent. Of these patients, 66% (59 patients) were female. Mean age was 67 years (SD 6.6), and the mean BMI was 28.6 (SD 3.2).

The intra- and inter-rater variability of the manual method ranged from 0.22 and 0.52 for the various depths. The intra- and inter-rater reliability of the matlab-method varied between 0.98 to 0.99 and the inter-method reliability ranged between 0.19 and 0.24 (Table 1). Bland-Altman plots of both methods as well as of the inter-method reliability are shown in Fig. 4. Since the reliability of the matlab method was much better than the manual method we performed the statistical analysis with the results of the matlab method.

Table 1 The intra-rater reliability and inter-rater reliability of the manual and the matlab method
Fig. 4
figure 4

a CI = Confidence Interval. b CI = Confidence Interval. c CI = Confidence Interval. d CI = Confidence Interval. e CI = Confidence Interval

The median percentages cement at 1, 3 and 5 mm beneath the tibial tray are shown in Table 2 and Fig. 5. There was significantly less cement in the antero-medial quadrant compared to the antero-lateral and postero-lateral quadrant at 3 mm and 5 mm depth.

Table 2 The median percentages cement at 1, 3 and 5 mm beneath the tibial tray
Fig. 5
figure 5

a: Measurements with the Matlab Method. Cement distribution 1 mm underneath the tibial tray over the 4 quadrants. AL = Antero-Lateral quadrant, AM = Antero-Medial quadrant, PL = Postero-Lateral quadrant, PM = Postero-Medial quadrant. b: Measurements with the Matlab Method. Cement distribution 3 mm underneath the tibial tray over the 4 quadrants. AL = Antero-Lateral quadrant, AM = Antero-Medial quadrant, PL = Postero-Lateral quadrant, PM = Postero-Medial quadrant. c: Measurements with the Matlab Method. Cement distribution 5 mm underneath the tibial tray over the 4 quadrants. AL = Antero-Lateral quadrant, AM = Antero-Medial quadrant, PL = Postero-Lateral quadrant, PM = Postero-Medial quadrant

Discussion

The aim of this study was to assess the accuracy of two different measuring methods for cement penetration and cement distribution of the tibia using CT scan images in patients with a total knee arthroplasty.

Intra-rater variability and inter-rater variability

The variability of the manual method was very low. Bland Altman plots showed that the variability of the manual method can be explained by the lack of agreement at 3 mm depth underneath the tibial tray. Cement is more scattered at that depth and therefore difficult to obtain visually. However, the ICC at 1 and 5 mm depth was also low. Which was in contrast to the variability of the matlab method. Results of the intra-rater variability and inter-rater variability were highly superior over the manual method at all depths. Portney and Watkins [17] suggested that ICC values of > 0.90 represent an excellent reliability. Therefore, we think that, with this user-friendly matlab approach, we present a reliable method for assessing the cement mantle, which can be used for clinical evaluation and for scientific studies.

Cement penetration

Walker et al.(1984) already stated that optimal fixation requires penetration of cement into the proximal tibia of 3 to 4 mm. In recent years a good penetration of cement into the trabecular bone is still a requirement for a solid fixation strength of the prosthesis [12, 15].

In this study we used the common surface cementing technique for the tibial baseplate. With this method the cement penetration at a depth of 1 mm was very good. In contrast, at 3 mm depth measured with the matlab method a median of 39% in the antero-medial quarter to 53% in the antero-lateral quarter of the trabecular bony surface was penetrated with cement. At 5 mm depth only 15% of the antero-medial and 25% of the postero-lateral quarter was filled with cement. Further research is necessary to optimize the cementing technique for a better fixation of the tibia baseplate with the trabecular bone. If these cementing techniques are evaluated to examine the cement penetration, then this CT scan method provides a reliable measurement method for penetration in vivo.

Cement distribution

The volume of cement penetrating into the bone is less predictive of failure than the mean penetration, which indicates the importance of a good distribution of the cement [13].

At 1 mm beneath the tibial tray, all four quadrants had an equal distribution of cement over an equal percentage of the trabecular bone surface.

At 3 and 5 mm depth, there was significantly less cement penetration in the anterior-medial quadrant. Perhaps this is due to the holes of the pins that were used for fixation of the tibial cutting guide and for the tibial sizing plate. Both instruments were fixated temporarily on the anterior-medial side, and these holes frequently penetrates the cortex of the proximal tibia. It seems possible that by compressing the cement, an amount of the cement is pressed through these holes out of the tibia. More research is needed to support the explanation for these divergent results.

Limitations

A limitation of our study is that with this CT technique, it is impossible to differentiate between bulk cement, such as filled bony defects or cysts, and trabecular bone that is interlocked with cement. When cement is pressurized into trabecular bone, the cement fills up the space between the trabeculae, which enlarges the trabecular bone’s density and improves its homogeneity. As a result, the homogeneity value of cement-penetrated trabecular bone is almost the same as the homogeneity value of pure cement. With this result, we can assume that the density of cement-penetrated trabecular bone equals the density of pure cement. To differentiate between them, we would have to use micro-CT, which cannot be used in vivo. We therefore accepted this limitation, since this study aimed to develop and validate a measuring technique that can be used in clinical practice.

A second limitation is that it is not possible to assess the space just surrounding the stem of the prosthesis due to the scattering of the metal. This even was the case with the prosthesis that we used in this study, which has a tibial component of titanium alloy and allowed post-surgery CT scanning of the cement mantle without excessive metal scattering effects of the prosthesis [13].

However, the scattering had the same influence on the data at the three different levels, and hence it did not influence the results of this study.

Conclusions

Distribution and penetration of cement in the proximal tibia in a total knee arthroplasty can be measured reliably with CT in combination with the matlab method presented in this manuscript. This method can be used for clinical purposes as well as for scientific research.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

A-L:

Antero-lateral

A-M:

Antero-medial

BMI:

Body mass index

CT:

Computer tomography

HU:

Hounsfield Unit

ICC:

Intraclass correlation coefficient

IQR:

Interquartile range

kV:

Kilovolt

mA:

Milliampere

mm:

Millimeter

PACS:

Picture archiving and communication system

P-L:

Postero-lateral

P-M:

Postero-medial

PMMA:

Polymethylmethacrylaat

SD:

Standard deviation

SPSS:

Statistical package for the social sciences

TKA:

Total knee arthroplasty

References

  1. Carr AJ, Robertsson O, Graves S, Price AJ, Arden NK, Judge A, et al. Knee replacement. Lancet. 2012;379:1331–40.

    Article  Google Scholar 

  2. Meding JB, Keating EM. Overview: long-term followup: predicting the success of a total knee arthroplasty. Clin Orthop Relat Res. 2001;388:7–9.

    Article  Google Scholar 

  3. Buechel FF Sr, Buechel FF Jr, Pappas MJ, D’Alessio J. Twenty-year evaluation of meniscal bearing and rotating platform knee replacements. Clin Orthop Relat Res. 2001;388:41–50.

    Article  Google Scholar 

  4. Keating EM, Meding JB, Faris PM, Ritter MA. Long-term followup of nonmodular total knee replacements. Clin Orthop Relat Res. 2002;404:34–9.

    Article  Google Scholar 

  5. Arsoy D, Pagnano MW, Lewallen DG, Hanssen AD, Sierra RJ. Aseptic tibial debonding as a cause of early failure in a modern total knee arthroplasty design. Clin Orthop Relat Res. 2013;471(1):94–101.

    Article  Google Scholar 

  6. Khan M, Osman K, Green G, Haddad FS. The epidemiology of failure in total knee arthroplasty: avoiding your next revision. Bone Joint J. 2016;98-b(1 Suppl A):105–12.

    Article  CAS  Google Scholar 

  7. Nakama GY, Peccin MS, Almeida GJ, Lira Neto Ode A, Queiroz AA, Navarro RD. Cemented, cementless or hybrid fixation options in total knee arthroplasty for osteoarthritis and other non-traumatic diseases. Cochrane Database Syst Rev. 2012;10:Cd006193.

    PubMed  Google Scholar 

  8. Gandhi R, Tsvetkov D, Davey JR, Mahomed NN. Survival and clinical function of cemented and uncemented prostheses in total knee replacement: a meta-analysis. J Bone Joint Surg Br. 2009;91(7):889–95.

    Article  CAS  Google Scholar 

  9. Waanders D, Janssen D, Mann KA, Verdonschot N. The mechanical effects of different levels of cement penetration at the cement-bone interface. J Biomech. 2010;43(6):1167–75.

    Article  Google Scholar 

  10. Janssen D, Mann KA, Verdonschot N. Micro-mechanical modeling of the cement-bone interface: the effect of friction, morphology and material properties on the micromechanical response. J Biomech. 2008;41(15):3158–63.

    Article  Google Scholar 

  11. Walker PS, Soudry M, Ewald FC, McVickar H. Control of cement penetration in total knee arthroplasty. Clin Orthop Relat Res. 1984;185:155–64.

    Google Scholar 

  12. Cawley DT, Kelly N, McGarry JP, Shannon FJ. Cementing techniques for the tibial component in primary total knee replacement. Bone Joint J. 2013;95-B(3):295–300.

    Article  CAS  Google Scholar 

  13. Nagel K, Bishop NE, Schlegel UJ, Püschel K, Morlock MM. The influence of cement morphology parameters on the strength of the cement-bone interface in tibial tray fixation. J Arthroplasty. 2017;32(2):563–9.

    Article  Google Scholar 

  14. Verburg H, van de Ridder LC, Verhoeven VW, Pilot P. Validation of a measuring technique with computed tomography for cement penetration into trabecular bone underneath the tibial tray in total knee arthroplasty on a cadaver model. BMC Med Imaging. 2014;14:29.

    Article  Google Scholar 

  15. Scheele CB, Müller PE, Schröder C, Grupp T, Jansson V, Pietschmann MF. Accuracy of a non-invasive CT-based measuring technique for cement penetration depth in human tibial UKA. BMC Med Imaging. 2019;19:9.

    Article  Google Scholar 

  16. Verburg H, Mathijssen NM, Niesten DD, Verhaar JA, Pilot P. Comparison of mini-midvastus and conventional total knee arthroplasty with clinical and radiographic evaluation: a prospective randomized clinical trial with 5-year followup. J Bone Joint Surg Am. 2016;98(12):1014–22.

    Article  Google Scholar 

  17. Portney LG, Watkins MP. Foundations of clinical research: applications to practice. 3rd ed. New Jersey: Pearson Education; 2007.

    Google Scholar 

Download references

Acknowledgements

We thank Brechtje Hesseling for her support with the MATLAB® -analysis and interpretation of the data. Olivier Hiemstra and Rianne Oomen for their support in writing and optimizing the MATLAB® -Script. Ian Blom and Willem Rasenberg for their support in making and re-sample the CT-scans.

Funding

There was no funding for this study and there was no influence of the manufacturer concerning the design of the study and collection, analysis and interpretation of data and/ or writing this manuscript.

Author information

Authors and Affiliations

Authors

Contributions

Hereby, I confirm that all authors have made substantial contributions to all of the following: (1) the conception and design of the study (HV, PP, and NM), or acquisition of data (HV, LvZ-K and DDN), or analysis and interpretation of data (HV, LvZ-K, PP, and NM), (2) drafting the article (HV) or revising it critically for important intellectual content (HV, LvZ-K, DDN, PP and NM), (3) final approval of the version to be submitted (HV, LvZ-K, DDN, PP and NM). Furthermore, each of the authors has read and concurs with the content in the manuscript. HV: Hennie Verburg

Corresponding author

Correspondence to Hennie Verburg.

Ethics declarations

Ethics approval and consent to participate

The study concept was presented to the local ethics committee of the Medical Ethics Committee of South-West Holland, The Netherlands and received approval (METC-nr: 06–104 Dossier NL14807.098.06/versie 06). This approval included consent to the use of CT scan for the purpose of measuring the quality of the cementing technique in total knee arthroplasty. Written informed consent was obtained from all the participants in this study.

Consent for publication

Not applicable.

The presented manuscript does not include any details, images, or videos relating to an individual person. CT scan pictures are entirely unidentifiable and there are no details on individuals reported within the manuscript.

Competing interests

One of the authors (PP) is an employee of Zimmer Biomet, Warsaw, IN, USA, a manufacturer of orthopedic implants.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary information

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Verburg, H., van Zeeland Koobs, L., Niesten, D.D. et al. Reliability of two different measuring techniques with computer tomography for penetration and distribution of cement in the proximal tibia after total knee arthroplasty. BMC Musculoskelet Disord 21, 374 (2020). https://doi.org/10.1186/s12891-020-03390-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-020-03390-3

Keywords