Impact of a double-layer cementing technique on the homogeneity of cementation and the generation of loose bone cement fragments in tibial unicompartmental knee arthroplasty

Abstract

Background

The objective of this study was to evaluate the impact of a single- vs. double-layer cementing technique on morphological cementation and the generation of microscopic cement layers or loose cement fragments in unicompartmental knee arthroplasty (UKA).

Methods

UKAs were implanted in 12 cadaver knees. The specimens were divided into two groups of comparable bone mineral density. Six UKAs were implanted using a single-layer cementing technique (group A) and six UKAs were implanted using a double-layer cementing technique (group B). Morphological cementation was assessed on nine cuts through the implant–cement–bone interface in the frontal plane. Loose bone cement fragments and the microscopically quality of layer formation were evaluated.

Results

Contact between bone and prosthesis was observed in 45.4% of interfaces in group A and 27.8% in group B (p = 0.126). The significant increase of areas without visible cement interlocking in the anteroposterior direction in group A (p = 0.005) was not evident in group B (p = 0.262). Penetration around the peg tended to occur more frequently in group B (67.5% vs. 90.6% p = 0.091). Scanning electron microscopy identified no evidence of fissure formations within the bilaminar cement mantle. Free bone cement fragments were documented in 66.7% in both groups with no difference concerning mass (p = 1.0).

Conclusions

This in-vitro study showed a tendency towards a more homogenous cementation of tibial UKAs using a double-layer cementing technique, although most of the differences did not reach the level of significance. However, theoretical downsides of the double-layer cementing technique such as an increased formation of free bone fragments or a microscopically fissure formation within the cement layer could not be detected either.

Peer Review reports

Background

With adequate patient selection, unicompartmental knee arthroplasty (UKA) has become an effective treatment for anteromedial osteoarthritis of the knee, leading to good long-term survival [1, 2] and better clinical outcomes and patient satisfaction than total knee arthroplasty (TKA) [3,4,5]. For both TKA and UKA, cementation is considered as standard method of fixation. However, register studies still reveal higher revision rates for UKA than for TKA [6, 7], with aseptic loosening of the tibial component being the most common cause of failure [8]. Aseptic loosing is associated with mechanical fatigue or collapse of the interface between bone, cement and prosthesis. A complete cement mantle with good interdigitation is therefore regarded as prerequisite for good long-term results of UKA [9,10,11,12].

The objective of this study was to evaluate the morphological effects of a double-layer cementing technique on the homogeneity of the cement mantle and the interlocking of bone cement with trabecular bone. Moreover, potential negative effects of this technique such as the formation of loose cement particles in the posterior aspect of the joint or a microscopic layer formation within the cement mantle were analyzed.

Methods

The study received approval from the local ethics committee. Twelve tibial and femoral UKA components (Univation® XF, Aesculap Tuttlingen, Germany) were implanted in the medial compartment of fresh-frozen human cadaver knees using a minimally-invasive approach. In advance, CT-scans (Sensation 64 Somatom, Siemens AG Munich, Germany) of all specimen were obtained to determine bone density and to exclude specimens with osseous irregularities (Sensation 64 Somatom, Siemens Munich, Germany). Based on bone mineral density (BMD), two groups with comparable bone density were formed (Table 1). The Hounsfield units in the anterior, central and posterior regions of the medial tibial plateau were additionally determined for the local determination of bone density [13,14,15].

Table 1 Overview of the 12 specimen, including sex, age, BMD, cementation technique and tibial component size

All implantations were performed by the same experienced surgeon (PEM) following the guidelines of the manufacturer. The appropriate implant size was selected based on preoperative CT measurement and intraoperative findings. After tibial resection, the cancellous bone was conditioned with pulsed lavage (Pulsavac® Plus, Zimmer, Warsaw USA) and saline solution. The tibial and femoral components were cemented in two stages using manually mixed bone cement (Palacos® R 20 g powder/10 ml monomer, Heraeus Medical Wehrheim, Germany). In group A, approximately 10 g of cement were applied manually to the underside of the prosthesis while the resected tibia remained cementless (single-layer cementing technique). In group B, approximately 10 g of cement were applied in equal parts by hand to the underside of the prosthesis plus the tibia. Each tibial component, was carefully placed on the tibia, beginning with the dorsal edge to prevent dorsal cement extrusion, and impacted using a special instrument (Univation® F instruments, Aesculap, Germany). Manual pressure was maintained until the cement cured completely. The joint was then opened and free cement bodies were collected, conserved and quantified using an analytical balance (Sartorius AG, Goettingen, Germany; Fig. 1).

Fig. 1
figure 1

Collection of free fragments of bone cement after complete arthrotomy of the knee joint and after implantation of medial UKA

For morphological analysis of the cementation, the tibial prosthesis with the cement phase and the adjacent bone were cut out as blocks. These blocks were then embedded in rectangular aluminium tubes with a transparent resin (Technovit 4004, Heraeus Medical Wehrheim, Germany) and sawed into 10 slices in the coronary plane. To ensure that slices of the different component sizes showed the same implant region, we virtually designed a masterplate for the cutting process (Dassault Systèmes, France) that was manufactured in a rapid-prototyping process (Ultimaker Geldermalsen; Netherlands), attached to the embedded implant and guaranteed that all 10 slices of a specimen had the same thickness.

The cut surfaces were cleaned and tiff-images with a resolution of 100 pixel per mm were acquired of all interfaces (HP Scanjet G3110; Hewlett-Packard Palo Alto, USA). The further evaluation of both sides of the nine serial cuts through the implant–cement–bone interface of the 12 specimens was carried out using Adobe Photoshop CS6 (Adobe, San Jose, USA).

In order to adequately capture the special geometry of the tibial tray and the associated penetration behavior, the underside was divided into two areas. Area 1 comprised the horizontal part of the prosthesis. Area 2 related to the area around the anchoring peg (Fig. 2).

Fig. 2
figure 2

Subdivision of the underside of the prosthesis into area 1 (horizontal underside of the prosthesis; black) and area 2 (area around the anchoring peg; grey)

The investigation of area 1 consisted of two indicators. First, the percentage of interfaces with interruptions of the cement mantle, i.e. contact between prosthesis and bone, was determined (Fig. 3). Second, presence and length proportion without cement intrusion into trabecular bone was assessed (Fig. 4). The latter was done by dividing the length without visible cement penetration by the total length of the horizontal bottom side of each tibial tray. Cement intrusion was defined as visible cement penetration into the trabecular bone on the high resolution images. The analysis distinguished between the anterior (cut 9–7), central (cut 6–4) and posterior section (cut 3–1). The investigation in area 2 focused on the proportion of surfaces with visible cement penetration in the area adjacent to the peg (Fig. 5).

Fig. 3
figure 3

Interruption of the cement mantle with contact between prosthesis and bone

Fig. 4
figure 4

Interruption of the (cement) penetration, i.e. no visible interlocking of the cement with the trabecular bone

Fig. 5
figure 5

Morphological evaluation of area 2. Left: no penetration; right: visible penetration into trabecular bone

In addition to the macroscopic assessment of conventional scans of the cut surfaces, a anterior (front of 8th slice), central (front of 5th slice) and posterior (back of 3rd slice) interface was analyzed by a scanning electron microscope (SEM, Zeiss Evo LS 10, Carl Zeiss Microscopy GmbH, Jena, Germany) and the associated software (SmartSEM V05.04.03.00 Carl Zeiss Microscopy GmbH, Jena, Germany), to approximate a quartering of the specimen. The respective cement-bone-interface was put on the stage of the specimen chamber facing up. A low-vacuum of 1.23 × 10− 5 hPa, high voltage of 20.56 kV and magnification of 100 were applied. The surfaces were scanned automatically with an overlap of 15% and the individual tiff-images were manually assembled and analyzed using Adobe Photoshop CS6 (Adobe, San Jose, USA). If fissure formations were apparent, they were defined for the section number, crack number, location of the crack, and length of the crack.

The statistical analysis was performed using GraphPad Prism 5 (GraphPad Software, Inc., La Jolla, CA, USA). A p-value < 0.05 was defined to be statistically significant. For statistical evaluation a repeated-measures analysis of variance (ANOVA) was performed to test for significant differences in the morphologic cementation parameters. Prior to analysis, normal distribution was verified (normal P–P plots; p < 0.05). Differences of the parameters between the groups were evaluated with a Bonferroni post hoc analysis. Additionally, the correlation between penetration and BMD and the two measurement techniques of bone density was determined using the Spearman’s rank correlation.

Results

The mean bone density was 154.6 (SD 48.3) HU or 103.3 (SD 26.0) mg/mm3 BMD in group A and 168.4 (SD 55.9) HU or 113.3 (SD 41.1) mg/mm3 BMD in group B, demonstrating good comparability (p = 0.699 [HU]; p = 0.937 [BMD]) between the groups. The correlation between HU and BMD was rs = 0.90 (P < 0.0001). In group A, the mean HU was 169.8 (SD 73.6) in the anterior, 123.4 (SD 38.9) in the central and 170.5 (SD 53.1) in the posterior section (p = 0.203). In group B, the corresponding figures were 185.8 (SD 78.3) HU anterior, 131.0 (SD 49.4) HU central and 188.4 (SD 57.6) HU posterior (p = 0.191).

In area 1, interruptions of the cement mantle, e.g. direct contact between the bone and the prosthesis were observed in 45.4% of all interfaces in group A and in 27.8% of all interfaces in group B (p = 0.126; Fig. 6).

Fig. 6
figure 6

Interruptions of the cement mantle, e.g. interfaces with direct contact between trabecular bone and prosthesis in area 1

Areas without visible cement intrusion at cement-bone-interface were observed in 87.8% of all interfaces in group A and 72.2% in group B (p = 0.936). The corresponding overall length proportion of area 1 without visible cement intrusion at cement-bone-interface was 37.6% of (SD 10.5%) in group A and 31.0% (SD 21.3%) in group B (p = 0.818). However, in the anterior-posterior development, group A showed a proportion of 14.2% (SD 10.5%) of area 1 without visible intrusion in the anterior section, 42.6% (SD 16.7%) in the central section and 55.2% (SD 17.1%) in the posterior section (p = 0.005). The corresponding figures for group B were 20.2% (SD 16.8%), 33.3% (SD 23.8%) and 40.1% (SD 28.3%; p = 0.262; Fig. 7). The correlation between HU and cement interdigitation was rs = 0.37 (p = 0.50) in group A, rs = 0.66 (p = 0.18) in group B and rs = 0.50 (p = 0.10) with respect to all specimens.

Fig. 7
figure 7

Anteroposterior development of the length portion at the bone-cement-prosthesis interface in area 1 with interruptions of the cement penetration

Visible cement penetration in area 2 was less frequently observed in group A with 67.5% than in group B with 90.6% (Fig. 8). However, this difference was not significant (p = 0.091).

Fig. 8
figure 8

Average proportion of anchoring pegs (area 2) with visible cement penetration

Loose fragments of bone cement could be found in 4 out of 6 cases in both groups (66,7%). Their average weight was 40.1 mg (SD 10.1 mg) in group A and 40.5 mg (SD 10.6 mg) in group B (p = 1.0; Fig. 9).

Fig. 9
figure 9

Weight of the free cement bodies

No evidence of microscopic failure or dissociation within the cement mantle could be detected (Fig. 10; Fig. 11). No cracks were observed in any specimen using both cementing techniques.

Fig. 10
figure 10

SEM image of a test specimen with double-layer cementation (group B) – overview with no layer formation, cracks or fissure formation apparent

Fig. 11
figure 11

SEM image of a test specimen with double-layer cementation (group B) - magnification; no indication of layer formation, cracks or fissure formation apparent within the cement phase

Discussion

Along with other causes such as wear, osteolysis or infection, the main reason for failure in UKA is represented by aseptic loosening of the tibial component [8]. Therefore, a strong interface with cement penetration and mechanical interlocking with trabecular bone is crucial for long-term success [9, 16]. Common clinical practice in cementing UKA is to finger pack bone cement on the prosthesis only. However, former studies using single-layer technique show only limited cement penetration in knee arthroplasty [9, 14, 17]. In particular, a solid fixation in the posterior aspect of the joint seems to be difficult to achieve via a minimally invasive medial approach [13].

With regard to a single- or double-layer cementing technique in knee arthroplasty, only few data is available. Clarius et al. recommended a double-layer technique for femoral UKA in order to ensure a complete filling of drill holes [18]. Furthermore, Vanlommel et al. reported a significantly higher cement penetration depth using a double-layer cementing technique in a study on sawbones [19]. However, according to Hauptmann et al., a thicker cement layer is associated with excess cement bodies [20]. Grupp et al. showed no overall difference in cement penetration depth after implanting tibial UKA in an in vitro cadaver study [21].

This study assesses potential morphologic advantages and disadvantages of a bilaminar cementing technique in UKA. Therefore, the first goal of this study was to assess, whether a double-layer cementation technique improves the homogeneity of the cement mantle and its penetration into trabecular bone. The second goal was to quantify theoretical concerns of the double-layer application of bone cement, like failure at the interface between the two layers of cement and the generation of loose cement fragments in the posterior aspect in the joint.

Concerning the macroscopic examination, there was an overall trend in favor of the double-layer cementing technique, but no significant differences. The proportion of cement-bone-prosthesis interfaces with direct contact between the prosthesis and trabecular bone was 45.6% in group A („single layer“) and 27.8% in group B („double layer“) in area 1 (p = 0.126). Cement penetration into trabecular bone in area 2 was visible on 67.5% of all interfaces in group A and 90.6% in group B (p = 0.096). The significant increase of the length proportion of area 1 without cement interdigitation in the anteroposterior direction in group A (p < 0.005) was not evident in group B (p = 0.262). The latter might be important, as the posterior part of the joint is often regarded as a weak point of cemented fixation [18, 22], especially using a minimally-invasive approach [13]. One reason is the difficulty to reach the posterior aspect of the joint with a conventional jet lavage [14, 22, 23]. Another might be the fact that the component is inserted parallel to the bone surface, which makes orthogonal pressure application on the bone cement challenging, in particular when using a minimally invasive approach [13].

Although there were no significant advantages of double-layer cementation in the macroscopic examination of the interfaces, the concern about potential disadvantages could not be confirmed either. In our study, no microscopic differences between both techniques could be detected. In particular, neither morphological layers, nor the generation of cracks or fissures were recognized using SEM with a low-vacuum and a BSE-detector. It was not possible to identify the cement-on-cement interface. These findings are consistent with earlier observations from cement-within-cement technique in revision hip arthroplasty [24,25,26]. This technique involves the retention of the existing bone-cement-interface and the recementing of the stem within the completely or partially maintained old cement mantle. Ishihara et al. defined cement failure as the coalescence of several cracks, where cracks occur between voids within the cement [27]. According to Weinrauch et al., the absence of a visible layer formation is probably due to a diffusion-based molecular interdigitation and fusion of the two applied cement phases [25]. Wilson et al. investigated the biomechanical properties of a bilaminar cementation on Sawbones under cyclic loading. They could neither detect a macroscopic failure of the bilaminar cementation nor identify a boundary between the cement layers using SEM [28]. Furthermore, the cement layers in the double-layer cementing technique are not hardened during the implantation of the tibial tray and the remaining MMA monomer ensures a chemical bonding without any layer formation after curing. Microscopic inspection of the cement mantle using SEM can be regarded as an established method and was applied by other authors before [10, 26, 29, 30].

Moreover, there was no indication of an increase in the mass of free cement fragments in the posterior aspect of the joint (p = 1.0). Their frequent occurrence in four out of six specimen in both groups is in accordance with a study by Hauptmann et al., according to which free cement fragments after minimally-invasive implantation of UKA represent an underestimated problem [20]. Elmadag et al. documented the radiologic occurrence of posteromedial cement extrusion after UKA in 18,6% of cases [31]. Bhuta et al. reported a case of tibial nerve impingement as a consequence of posterior cement extrusion after a minimally-invasive UKA [32]. Our results are in line with this case and other observation of extruded cement remnants in UKA [28, 33, 34].

In summary, the significant advantage of two-layer cementation on cement penetration described in Sawbones could not be reproduced under minimally-invasive surgical conditions in human tibia. However, no indications of adverse effects of two-layer cementation were observed. These results are consistent with previous results, that described no differences between the single and double-layer cementing technique of unicompartmental tibial components with regard to failure load [21].

This study is subject to potential limitations. First, the present in-vitro examination on cadaver legs did not simulate blood circulation. It was postulated from hip arthroplasty experiments that blood flow might have a negative influence on cement penetration [35, 36]. For example, Greenwald et al. postulated a 37% decrease in shear strength due to blood inclusion [37]. Li et al. [24] quantified the negative influence of a blood-bone marrow layer on tensile strength and shear force at 80 and 85%, respectively, and described a microscopically visible layer of less than 100 μm between the cement phases. However, this view is controversial as MacDonalds et al. did not find any significant traces of blood at the interface between bone and cement or blood inclusions within the cement in an in-vivo animal model [12]. In addition, Rudol et al. stated that only large amounts of highly viscous liquids (e.g. bone marrow) weaken the interface. In contrast, low-viscosity fluids like blood did not significantly influence the interface strength [38]. Moreover, bleeding can be prevented during implantation by the application of a tourniquet. Second, common manual pressure application during curing of the cement was applied in order to achieve the best possible simulation of surgical conditions. Although one experienced surgeon performed all implantations, an absolute consistence of force can therefore not be secured. Third, human specimen were not randomly distributed to the groups due to the rather small sample size. To ensure comparable conditions, their allocation was performed as matched-pairs of comparable bone mineral density.

Conclusions

The present study does not provide any indications of negative effects caused by a double-layer cementing technique such as an increase in the formation of free bone fragments or a microscopically layer formation in tibial UKA. However, improvements in the morphologic homogeneity of the cement mantle in area 1 or the improved average penetration in area 2 do not reach the level of significance. Therefore, the effectiveness of a double-layer cementation technique should be investigated further in randomized controlled clinical trials.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

BMD:

Bone mineral density

BSE:

Back-scatter detector

CT:

Computed tomography

et al.:

et alii

Fig.:

Fig.

HU:

Hounsfield units

ROI:

Region of interest

SD:

Standard deviation

SEM:

Scanning electron microscope

TKA :

Total knee arthroplasty

UKA:

Unicompartmental knee arthroplasty

References

  1. Berger RA, Meneghini RM, Sheinkop MB, Della Valle CJ, Jacobs JJ, Rosenberg AG, Galante JO. The progression of patellofemoral arthrosis after medial unicompartmental replacement: results at 11 to 15 years. Clin Orthop Relat Res. 2004;428:92–9.

    Article  Google Scholar 

  2. Price AJ, Svard U. A second decade lifetable survival analysis of the Oxford unicompartmental knee arthroplasty. Clin Orthop Relat Res. 2011;469(1):174–9.

    Article  Google Scholar 

  3. Laurencin CT, Zelicof SB, Scott RD, Ewald FC. Unicompartmental versus total knee arthroplasty in the same patient. A comparative study. Clin Orthop Relat Res. 1991;273:151–6.

    Google Scholar 

  4. Rougraff BT, Heck DA, Gibson AE. A comparison of tricompartmental and unicompartmental arthroplasty for the treatment of gonarthrosis. Clin Orthop Relat Res. 1991;273:157–64.

    Google Scholar 

  5. Newman J, Pydisetty RV, Ackroyd C. Unicompartmental or total knee replacement: the 15-year results of a prospective randomised controlled trial. J Bone Joint Surg Br. 2009;91(1):52–7.

    Article  CAS  Google Scholar 

  6. Paxton EW, Inacio MC, Khatod M, Yue EJ, Namba RS. Kaiser Permanente National Total Joint Replacement Registry: aligning operations with information technology. Clin Orthop Relat Res. 2010;468(10):2646–63.

    Article  Google Scholar 

  7. SKAR: (Swedish Knee Arthroplasty Register) Annual report 2012. Lund University Hospital, Lund 2012 Available online at wwwkneese/ 2012.

  8. Epinette JA, Brunschweiler B, Mertl P, Mole D, Cazenave A, French Society for H, knee. Unicompartmental knee arthroplasty modes of failure: wear is not the main reason for failure: a multicentre study of 418 failed knees. Orthop Traumatol Surg Res. 2012;98(6 Suppl):S124–30.

    Article  Google Scholar 

  9. Schlegel UJ, Siewe J, Delank KS, Eysel P, Puschel K, Morlock MM, de Uhlenbrock AG. Pulsed lavage improves fixation strength of cemented tibial components. Int Orthop. 2011;35(8):1165–9.

    Article  Google Scholar 

  10. Krause WR, Krug W, Miller J. Strength of the cement-bone interface. Clin Orthop Relat Res. 1982;163:290–9.

    Google Scholar 

  11. Halawa M, Lee AJ, Ling RS, Vangala SS. The shear strength of trabecular bone from the femur, and some factors affecting the shear strength of the cement-bone interface. Arch Orthop Trauma Surg. 1978;92(1):19–30.

    Article  CAS  Google Scholar 

  12. MacDonald W, Swarts E, Beaver R. Penetration and shear strength of cement-bone interfaces in vivo. Clin Orthop Relat Res. 1993;286:283–8.

    Google Scholar 

  13. Scheele C, Pietschmann MF, Schroder C, Grupp T, Holderied M, Jansson V, Muller PE. Effect of minimally-invasive implantation of unicompartmental knee arthroplasty on cement penetration and biomechanical stability. An experimental study in human tibiae. Clin Biomech (Bristol, Avon). 2018;51:34–9.

    Article  CAS  Google Scholar 

  14. Scheele C, Pietschmann MF, Schroder C, Grupp T, Holderied M, Jansson V, Muller PE. Effect of lavage and brush preparation on cement penetration and primary stability in tibial unicompartmental total knee arthroplasty: an experimental cadaver study. Knee. 2017;24(2):402–8.

    Article  Google Scholar 

  15. Scheele CB, Muller PE, Schroder C, Grupp T, Jansson V, Pietschmann MF. Accuracy of a non-invasive CT-based measuring technique for cement penetration depth in human tibial UKA. BMC Med Imaging. 2019;19(1):9.

    Article  Google Scholar 

  16. Clarius M, Hauck C, Seeger JB, James A, Murray DW, Aldinger PR. Pulsed lavage reduces the incidence of radiolucent lines under the tibial tray of Oxford unicompartmental knee arthroplasty: pulsed lavage versus syringe lavage. Int Orthop. 2009;33(6):1585–90.

    Article  Google Scholar 

  17. Maistrelli GL, Antonelli L, Fornasier V, Mahomed N. Cement penetration with pulsed lavage versus syringe irrigation in total knee arthroplasty. Clin Orthop Relat Res. 1995;312:261–5.

    Google Scholar 

  18. Clarius M, Mohr G, Jaeger S, Seeger JB, Bitsch RG. Femoral fixation pattern in cemented Oxford unicompartmental knee arthroplasty--an experimental cadaver study. Knee. 2010;17(6):398–402.

    Article  CAS  Google Scholar 

  19. Vanlommel J, Luyckx JP, Labey L, Innocenti B, De Corte R, Bellemans J. Cementing the tibial component in total knee arthroplasty: which technique is the best? J Arthroplast. 2011;26(3):492–6.

    Article  Google Scholar 

  20. Hauptmann SM, Weber P, Glaser C, Birkenmaier C, Jansson V, Muller PE. Free bone cement fragments after minimally invasive unicompartmental knee arthroplasty: an underappreciated problem. Knee Surg Sports Traumatol Arthrosc. 2008;16(8):770–5.

    Article  CAS  Google Scholar 

  21. Grupp TM, Pietschmann MF, Holderied M, Scheele C, Schroder C, Jansson V, Muller PE. Primary stability of unicompartmental knee arthroplasty under dynamic compression-shear loading in human tibiae. Clin Biomech (Bristol, Avon). 2013;28(9–10):1006–13.

    Article  Google Scholar 

  22. Walker PS, Soudry M, Ewald FC, McVickar H. Control of cement penetration in total knee arthroplasty. Clin Orthop Relat Res. 1984;185:155–64.

    Google Scholar 

  23. Labutti RS, Bayers-Thering M, Krackow KA. Enhancing femoral cement fixation in total knee arthroplasty. J Arthroplast. 2003;18(8):979–83.

    Article  Google Scholar 

  24. Li PL, Ingle PJ, Dowell JK. Cement-within-cement revision hip arthroplasty; should it be done? J Bone Joint Surg Br. 1996;78(5):809–11.

    Article  CAS  Google Scholar 

  25. Weinrauch PC, Bell C, Wilson L, Goss B, Lutton C, Crawford RW. Shear properties of bilaminar polymethylmethacrylate cement mantles in revision hip joint arthroplasty. J Arthroplast. 2007;22(3):394–403.

    Article  Google Scholar 

  26. Wilson LJ, Bell CG, Weinrauch P, Crawford R. In vitro cyclic testing of the Exeter stem after cement within cement revision. J Arthroplast. 2009;24(5):789–94.

    Article  Google Scholar 

  27. Ishihara S, McEvily AJ, Goshima T, Kanekasu K, Nara T. On fatigue lifetimes and fatigue crack growth behavior of bone cement. J Mater Sci Mater Med. 2000;11(10):661–6.

    Article  CAS  Google Scholar 

  28. Kim WY, Shafi M, Kim YY, Kim JY, Cho YK, Han CW. Posteromedial compartment cement extrusion after unicompartmental knee arthroplasty treated by arthroscopy: a case report. Knee Surg Sports Traumatol Arthrosc. 2006;14(1):46–9.

    Article  Google Scholar 

  29. Banwart JC, McQueen DA, Friis EA, Graber CD. Negative pressure intrusion cementing technique for total knee arthroplasty. J Arthroplast. 2000;15(3):360–7.

    Article  CAS  Google Scholar 

  30. Willert HG, Ludwig J, Semlitsch M. Reaction of bone to methacrylate after hip arthroplasty: a long-term gross, light microscopic, and scanning electron microscopic study. J Bone Joint Surg Am. 1974;56(7):1368–82.

    Article  CAS  Google Scholar 

  31. Elmadag M, Imren Y, Erdil M, Bilsel K, Tuncay I. Excess retained cement in the posteromedial compartment after unicondylar knee arthroplasty. Acta Orthop Traumatol Turc. 2013;47(4):291–4.

    Article  Google Scholar 

  32. Bhutta MA, Doorgakant A, Marynissen H. Tibial nerve impingement secondary to posterior cement extrusion after unicompartmental knee arthroplasty. J Arthroplast. 2010;25(7):1168 e1117–1168.

    Article  Google Scholar 

  33. Howe DJ, Taunton OD Jr, Engh GA. Retained cement after unicondylar knee arthroplasty. A report of four cases. J Bone Joint Surg Am. 2004;86-A(10):2283–6.

    Article  Google Scholar 

  34. Karataglis D, Agathangelidis F, Papadopoulos P, Petsatodis G, Christodoulou A. Arthroscopic removal of impinging cement after unicompartmental knee arthroplasty. Hippokratia. 2012;16(1):76–9.

    CAS  PubMed  PubMed Central  Google Scholar 

  35. Benjamin JB, Gie GA, Lee AJ, Ling RS, Volz RG. Cementing technique and the effects of bleeding. J Bone Joint Surg Br. 1987;69(4):620–4.

    Article  CAS  Google Scholar 

  36. Majkowski RS, Miles AW, Bannister GC, Perkins J, Taylor GJ. Bone surface preparation in cemented joint replacement. J Bone Joint Surg Br. 1993;75(3):459–63.

    Article  CAS  Google Scholar 

  37. Greenwald AS, Narten NC, Wilde AH. Points in the technique of recementing in the revision of an implant arthroplasty. J Bone Joint Surg Br. 1978;60(1):107–10.

    Article  CAS  Google Scholar 

  38. Rudol G, Wilcox R, Jin Z, Tsiridis E. The effect of surface finish and interstitial fluid on the cement-in-cement interface in revision surgery of the hip. J Bone Joint Surg Br. 2011;93(2):188–93.

    Article  CAS  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

All funding and the execution of the preparation, implantation und testing was conducted by the Laboratory for Biomechanics and Experimental Orthopaedics of the Ludwig Maximilians University. The was no influence of the manufacturer concerning the design of the study and collection, analysis, and interpretation of data and/ or writing this manuscript.

Author information

Authors and Affiliations

Authors

Contributions

Hereby, I confirm that all authors have made substantial contributions to the manuscript: (1) the conception and design of the study (CBS, MFP, CS1, TMG, PEM), the acquisition of the data (CBS, MFP, CS1, TMG), or analysis and interpretation of the data (CBS, MFP, CS2, TMG, PEM); (2) drafting the article (CBS, MFP) or revising it critically for important intellectual content (CS1, CS2, TMG, PEM); and (3) final approval of the version to be submitted (CBS, MFP, CS1, CS2, TMG, PEM). Furthermore, each of the authors has read and concurs with the content in the final manuscript (CBS, MFP, CS1, CS2, TMG, PEM).

Corresponding author

Correspondence to Christian B. Scheele.

Ethics declarations

Ethics approval and consent to participate

The study concept was presented to the local ethics committee of the Ludwig Maximilians University of Munich (Chairman: Prof. Dr. W. Eisenmenger) and received approval. This approval included consent to the use of human tissue (cadaver legs). Human tissue involved in this study was obtained from the Department of Pathology of the Ludwig Maximilians University of Munich. No tissue was obtained from prisoners.

Consent for publication

Not applicable.

The presented manuscript does not include any details, images, or videos relating to an individual person. Specimen pictured are entirely unidentifiable and there are no details on individuals reported within the manuscript.

Competing interests

One author (PEM) was getting research funding in correlation with Aesculap R&D project and is advising surgeon of Aesculap R&D projects. Two authors (MFP, PEM) received payments for lectures and instructional courses in knee/ shoulder arthroscopy. There are no other relationships/conditions/circumstances that present a potential conflict of interest.

One of the authors (TMG) is an employee of Aesculap (manufacturer of orthopaedic implants).

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Scheele, C.B., Pietschmann, M.F., Schröder, C. et al. Impact of a double-layer cementing technique on the homogeneity of cementation and the generation of loose bone cement fragments in tibial unicompartmental knee arthroplasty. BMC Musculoskelet Disord 20, 539 (2019). https://doi.org/10.1186/s12891-019-2929-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-019-2929-x

Keywords