Novel MR imaging method – MAVRIC – for metal artifact suppression after joint replacement in musculoskeletal tumor patients

Abstract

Background

Standard imaging modality for the follow-up after prosthetic replacements for musculoskeletal tumor patients has been conventional radiography. This technique is effective in detecting subtle changes in bone adjacent to metal implants, but in many cases, radiographs do not lead to definitive diagnosis of postoperative adverse events such as acute infection, local recurrence of soft tissue tumor or soft tissue local recurrence of osseous sarcoma. Conventional MRI sequences have not been effective due to metal artifacts. In this study, we tried to elucidate the effectiveness of metal artifact suppression using novel sequence, multiacquisition variable-resonance image combination (MAVRIC), after musculoskeletal tumor surgeries.

Methods

We retrospectively analyzed 5 cases of malignant bone and soft tissue sarcoma patients who were reconstructed with metal prosthesis after wide resection of tumors. Images obtained using MAVRIC and short tau inversion recovery (STIR) were compared side by side. The paired MAVRIC and STIR images were qualitatively compared independently by two specialists for 4 parameters: visualization of bone - implant interface, visualization of surrounding soft tissues, image blurring, and overall image quality. Quantitatively, paired images were reviewed to identify the slice where the metal artifact was maximal, and a region of interest encompassing the implant and surrounding artifact was drawn using Advantage Workstation (GE Healthcare, Japan).

Results

There were no local recurrences that were detected. By utilizing MAVRIC, visualization of the bone - implant interface and visualization of the surrounding soft tissue were significantly improved in MAVRIC compared to STIR. Although blurring was worse on the MAVRIC acquisitions, the overall image quality was still better on MAVRIC. Quantitatively, the area of metal artifact measured using MAVRIC was markedly less compared to STIR (61.4 cm2 vs 135.9 cm2).

Conclusion

Despite the relatively small number of cases in the present study, our observation strongly suggests that MAVRIC is able to improve the quality of images by decreasing the artifact caused by endoprosthesis, frequently utilized in reconstruction of musculoskeletal tumor patients. Further installments of conventional imaging sequences with the addition of gadolinium - enhancement will enable increased accuracy in diagnosing local recurrences of sarcoma patients.

Peer Review reports

Background

Multimodal therapy for sarcoma patients has improved the prognosis in the past decades, but surgery is still the mainstay of treatment for malignant tumors. Surgery for sarcoma of both bone and soft tissue usually consists of wide resection including the cuff of normal tissue, which leads to large defects. When skeletal structure is compromised, especially near the joint, endoprosthesis is frequently utilized for the reconstruction. Traditionally, radiograph was utilized to detect the change in bone, but it has not enabled the detection of devastating postoperative events such as soft tissue recurrences. The usage of massive metal implants have prevented the usage of magnetic resonance imaging (MRI) which has high spatial resolution and tissue contrasts compared to plain radiographs due to metal artifacts.

Recently, improvements in metal composition of prosthesis [1, 2] and artifact-reducing MRI sequences have been reported [35]. Multiacquisition variable-resonance image combination (MAVRIC) was first reported in 2009, where it uses multiple excitations to excite the overall volume being imaged [3]. MAVRIC excites a series of limited spectral distortion and uses the three-dimensional (3D) spin-echo acquisition to resolve the profile of each excited slice in the region of interest. MAVRIC has been reported to reduce the metal artifacts around smaller metal implants usually reserved for conventional orthopaedic procedure [4, 6, 7]. In this study, we analyzed the effectiveness of MAVRIC for suppression of metal artifacts around larger tumor endoprosthesis for early possible detection of tumor recurrences.

Methods

Since 2014, 4 megaprosthesis (1 total femur, 1 proximal femur, 1 distal femur, and 1 proximal humerus) and 1 augmented proximal femoral reconstruction stem were utilized for reconstructions and underwent MR imaging on a 3T scanner (Discovery 750, GE Healthcare, Japan). Histological diagnosis included 2 osteosarcomas, 2 chondrosarcomas and 1 synovial sarcoma. There were 4 female and1 male. Proximal humeral implant and reconstruction stem were composed from titanium alloy and all other implants were made up of cobalt-chrome. Conventional short tau inversion recovery (STIR) and MAVRIC sequences were acquired in the coronal planes using similar scan parameters and readout bandwidth. MAVRIC sequences were obtained with a 3T scanner (GE SIGNA discovery MR 750, GE Healthcare, Milwaukee, WI) in the supine position. The protocol included coronal, STIR images (TR/TE: 4500–6700/6.3–6.8 ms; inversion time 175 ms; slice thickness 4.5 mm; gap 0 mm; flip angle 65°; field of view 280 mm, matrix 320 × 320, 24 slices, band width 125 kHz, acquisition time from 4 min 15 s to 6 min 32 s) and coronal proton density weighted images (TR/TE: 2400/6.3–7.1 ms; slice thickness 4.5 mm; gap 0 mm; flip angle 75°; field of view 280 mm, matrix 384 × 320, 24 slices, band width 125 kHz, acquisition time from 4 min 5 s to 6 min 7 s) with an 32 – channel torso surface coil. These MR images were retrospectively analyzed between two musculoskeletal tumor specialists (one radiologist and one orthopaedic oncologist) in a matching anatomic plane with the maximal artifact for both qualitative and quantitative variables. Artifact was defined as areas of signal void pile-up or geometric distortion. Images were graded on a five-point scale from −2 to +2, where lower score suggested better outcome for MAVRIC acquisitions, which was previously reported by Gutierrez et al. [8]. −2 indicate that MAVRIC is significantly better, and −1 meant somewhat better compared to STIR. Readers were blinded as to the type of sequence that they were grading. Quantitatively, plane with the maximal artifact was determined, and region of interest was measured using Advantage Workstation (GE Healthcare, Japan) (Fig. 1). This study was conducted with the approval of Institutional Review Board of Keio University and all participants gave their informed consent to assessing and using their data.

Fig. 1
figure 1

a. Conventional STIR sequence obained after distal femoral endoprosthesis reconstruction after chondrosarcoma resection in a 46 year - old female. b. MAVRIC image. The maximum area of metal artifact was measured in a matching plane for both images. Area of the artifact was 130.3 cm2 for STIR and 107.0 cm2 for MAVRIC

Results

There were no local recurrences that were detected and no complications such as local hotness during the examination. The quality of images were compared and scored independently between the 2 observers. By utilizing MAVRIC, visualization of the bone - implant interface (average grade: −1.88) (Fig. 2) and visualization of the surrounding soft tissue (average grade: −1.75) (Fig. 3) were significantly improved in MAVRIC compared to STIR. Although blurring was worse on the MAVRIC acquisitions with lower resolution, lower contrasts and reduced fat saturation (average grade: +1.12) (Fig. 4), the overall image quality was still better on MAVRIC (average grade: −1.88). Quantitatively, the area of metal artifact measured using MAVRIC (range: 43 – 107 cm2, average: 61.4 cm2) was less compared to STIR (range: 76.3 – 246 cm2, average: 135.9 cm2). Clinically, MAVRIC was able to detect fluid collection surrounding the implant in one patient. Although the entire body of the patient was swollen and covered with eruption which was suggestive of toxic shock syndrome, thrombotic thrombocytopenia purpura/hemolytic uremic syndrome, or adult stills syndrome, MAVRIC acquisition revealed fluid collection surrounding the endoprosthesis (Fig. 3b), which lead to local puncture and diagnosis of methicillin-sensitive Staphylococcus aureus infection. The patient underwent irrigation and at the final follow-up, the patient is well with no symptom.

Fig. 2
figure 2

a. Conventional STIR image and b. MAVRIC image were compared for bone and implant interface. MAVRIC was able to depict the bone immediately adjacent to the endoprosthesis. Additionally, because of the reduced artifact, the prosthesis itself is now visible as signal void especially around the knee

Fig. 3
figure 3

a. A 27 year - old female received a distal femoral endoprosthetic reconstruction after wide resection of an osteosarcoma 6 years prior to the exmaination. b. Conventional STIR image and c. MAVRIC image were compared for the depiction of surrounding soft tissue adjacent to the endoprosthesis. Joint effusion is observed surrounding the metal implant in MAVRIC image (red arrow). Due to early detection of the periprosthetic infection, thorough irrigation and debriedement was performed, and the patient is well without recurrence

Fig. 4
figure 4

a. STIR image of a proximal humeral endoprosthetic reconstruction and b. MAVRIC image were compared side by side for the qualitaive assesment of the blurring. Image quality of the MAVRIC is reduced compared to STIR with blurry contours, lower contrast, and lower resolution as depicted by the ribs (average grade: +1.12). The fat saturation is clearly reduced with the MAVRIC as seen in the subcutaneous fat tissue

Discussion

There are many histological subtypes of sarcoma of both bone and soft tissues, and treatment is increasingly becoming case-specific with the advent of genetic testing. Several molecular targeting agents have been reported to improve the prognosis and many more are set to be reported. Although, these medical therapies are important, almost all sarcomas need some kind of surgical intervention to eradicate the tumor. It is especially important to perform a wide resection of the lesion to prevent local recurrence from the residual sarcoma cells. To reconstruct large defects, several methods have been proposed including endoprosthesis, allografts, autografts, and recycled bone by pasteurization or liquid nitrogen. When the defect is situated around the joint, endoprosthesis is often utilized due to the poor outcome using other reconstruction methods.

In order to improve the outcome of sarcoma patients, it is important to control the various postoperative adverse events such as acute infection and local recurrence as well as the distant metastasis which is a devastating event both physically and psychologically. Infection leads to multiple salvage operations and delay of chemotherapy which is a poor prognosticator of survival. Local recurrences have been reported to cause higher morbidity due to consequent distant metastasis [9]. One of the reasons for poor outcome after local recurrence is the difficulty of additional surgery. It is difficult to ascertain the tumor infiltration after multiple operations and it is even more difficult after endoprosthesis placement. One of the method recently reported utilize positron emission tomography (PET) - CT to localize the recurrence in the vicinity of metal implants [10]. We have also reported on the possibility of local recurrence detection using PET-CT, where SUVmax of 5.0 or greater might differentiate between tumor and fibrosis (data not shown). But, utilization of CT still causes artifacts due to beam hardening which pose significant challenge in delineating the tumor for successful surgery. Novel method for early diagnosis of postoperative adverse event is imperative to improve the outcome of sarcoma patients.

MRI has improved the prognosis of sarcoma patients by accurately detecting the localization and spread of the tumor prior to operation. It has the highest spatial resolution and tissue contrasts compared to other imaging modalities, but metal implants after surgery have impeded its accuracy by causing artifacts that results from metal disturbing the main magnetic field and inducing strong and spatially - varying local gradients [11]. MRI implementation on tumor endoprosthesis including expandable prosthesis has been reported to be safe without local hotness or unintentional lengthening [12]. MAVRIC and the slice encoding for metal artifact attenuation technique has shown promising results by reducing artifacts. MAVRIC has been reported in several reports after orthopaedic implant surgery as a possible solution for suppression of artifacts [4, 6, 7, 13]. MAVRIC has been able to detect joint abnormalities such as joint effusion and bursitis in painful hip [4] and shoulder [7], and showed clinical relevance after total knee replacements [14]. Results from this study also demonstrate that MAVRIC correction for metal induced artifacts improved postoperative visualization around the endoprosthesis. Fortunately, no local recurrence was detected in this case series, but imaging finding of local fluid collection was confirmed by subsequent invasive treatment in one case.

A limitation of this study is its relatively small sample size, which is consistent with other past small case series. Although the preparation of images is still time consuming, it is well suited for dedicated application in areas under suspicion. Further improvement, such as the introduction of MAVRIC - T1WI will enable the use of gadolinium - enhancement and increase the sensitivity for small lesions near the metal implants.

Conclusion

Several variations of metal artifact suppression technique such as FSE, patient positioning, and use of high bandwidth have been reported [5, 1517], but there is still significant challenge in detecting small lesions near metal implants. MAVRIC significantly improved the image quality and has the potential to improve patient managements. However, further prospective randomized studies are needed to establish the optimum use of MAVRIC for early diagnosis of local recurrences after sarcoma surgeries.

References

  1. Wang JC, Yu WD, Sandhu HS, Tam V, Delamarter RB. A comparison of magnetic resonance and computed tomographic image quality after the implantation of tantalum and titanium spinal instrumentation. Spine (Phila Pa 1976). 1998;23(15):1684–8.

    Article  CAS  Google Scholar 

  2. Lee MJ, Kim S, Lee SA, Song HT, Huh YM, Kim DH, et al. Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multi-detector CT. Radiographics. 2007;27(3):791–803.

    Article  PubMed  Google Scholar 

  3. Koch KM, Lorbiecki JE, Hinks RS, King KF. A multispectral three-dimensional acquisition technique for imaging near metal implants. Magn Reson Med. 2009;61(2):381–90.

    Article  PubMed  Google Scholar 

  4. Kretzschmar M, Nardo L, Han MM, Heilmeier U, Sam C, Joseph GB, et al. Metal artefact suppression at 3 T MRI: comparison of MAVRIC-SL with conventional fast spin echo sequences in patients with Hip joint arthroplasty. Eur Radiol. 2015;25(8):2403–11.

    Article  PubMed  Google Scholar 

  5. Vandevenne JE, Vanhoenacker FM, Parizel PM, Butts Pauly K, Lang RK. Reduction of metal artefacts in musculoskeletal MR imaging. JBR-BTR. 2007;90(5):345–9.

    CAS  PubMed  Google Scholar 

  6. Choi SJ, Koch KM, Hargreaves BA, Stevens KJ, Gold GE. Metal artifact reduction with MAVRIC SL at 3-T MRI in patients with hip arthroplasty. AJR Am J Roentgenol. 2015;204(1):140–7.

    Article  PubMed Central  PubMed  Google Scholar 

  7. Hayter CL, Koff MF, Shah P, Koch KM, Miller TT, Potter HG. MRI after arthroplasty: comparison of MAVRIC and conventional fast spin-echo techniques. AJR Am J Roentgenol. 2011;197(3):W405–11.

    Article  PubMed  Google Scholar 

  8. Gutierrez LB, Do BH, Gold GE, Hargreaves BA, Koch KM, Worters PW, et al. MR imaging near metallic implants using MAVRIC SL: initial clinical experience at 3T. Acad Radiol. 2015;22(3):370–9.

    Article  PubMed  Google Scholar 

  9. Trovik CS, Gustafson P, Bauer HC, Saeter G, Klepp R, Berlin O, et al. Consequences of local recurrence of soft tissue sarcoma: 205 patients from the Scandinavian Sarcoma Group Register. Acta Orthop Scand. 2000;71(5):488–95.

    Article  CAS  PubMed  Google Scholar 

  10. Gelfand MJ, Sharp SE. [F-18]2-fluoro-2-deoxyglucose (FDG) positron emission tomography after limb salvage surgery: post-surgical appearance, attenuation correction and local complications. Pediatr Radiol. 2015;45(8):1182–8.

    Article  PubMed  Google Scholar 

  11. Wendt 3rd RE, Wilcott 3rd MR, Nitz W, Murphy PH, Bryan RN. MR imaging of susceptibility-induced magnetic field inhomogeneities. Radiology. 1988;168(3):837–41.

    Article  PubMed  Google Scholar 

  12. Ogg RJ, McDaniel CB, Wallace D, Pitot P, Neel MD, Kaste SC. MR safety and compatibility of a noninvasively expandable total-joint endoprosthesis. Magn Reson Imaging. 2005;23(7):789–94.

    Article  PubMed  Google Scholar 

  13. Sutter R, Ulbrich EJ, Jellus V, Nittka M, Pfirrmann CW. Reduction of metal artifacts in patients with total hip arthroplasty with slice-encoding metal artifact correction and view-angle tilting MR imaging. Radiology. 2012;265(1):204–14.

    Article  PubMed  Google Scholar 

  14. Chen CA, Chen W, Goodman SB, Hargreaves BA, Koch KM, Lu W, et al. New MR imaging methods for metallic implants in the knee: artifact correction and clinical impact. J Magn Reson Imaging. 2011;33(5):1121–7.

    Article  PubMed Central  PubMed  Google Scholar 

  15. Butts K, Pauly JM, Gold GE. Reduction of blurring in view angle tilting MRI. Magn Reson Med. 2005;53(2):418–24.

    Article  PubMed  Google Scholar 

  16. Cho ZH, Kim DJ, Kim YK. Total inhomogeneity correction including chemical shifts and susceptibility by view angle tilting. Med Phys. 1988;15(1):7–11.

    Article  CAS  PubMed  Google Scholar 

  17. Walde TA, Weiland DE, Leung SB, Kitamura N, Sychterz CJ, Engh Jr CA, et al. Comparison of CT, MRI, and radiographs in assessing pelvic osteolysis: a cadaveric study. Clin Orthop Relat Res. 2005;437:138–44.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Michiro Susa.

Additional information

Competing interests

The authors declare that there is no conflict of interests.

Authors’ contributions

MS carried out the data analysis and manuscript writing. SO and KK carried out the collection and assembly of data. KN and KH contributed to the manuscript writing. MJ, MN, KC and MM provided administrative support. HM was responsible for the provision of study material of patients and final approval of the manuscript. All authors read and approved the final manuscript.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions