Skip to main content

Table 2 Absolute error between preoperative planning and postoperative outcomes

From: Is AI 3D-printed PSI an accurate option for patients with developmental dysplasia of the hip undergoing THA?

 

AIPSI Group

Control Group

P Value

Acetabular Component

 Inclination Error > 5( °) *

5(16.7)

17(46.7)

0.003

 Inclination Error > 10( °) *

0(0)

2(6.7)

0.492

 Anteversion Error > 5°( °) *

3(10.0)

19(63.3)

<.001

 Anteversion Error > 10°( °) *

0(0)

5(16.7)

0.052

Femoral Component

 Normal FO (< ± 5 mm compared to CL) *

25(83.3)

13(43.3)

0.047

Postoperative Leg Length Discrepancy (LLD)

 LLD (mm)

4.5 ± 2.0

6.4 ± 2.9

0.004

 LLD > 10*

0(0)

3(10.0)

0.236

  1. LLD leg length discrepancy, FO Femoral offset CL compared to the contralateral normal hip
  2. Presented as the mean (ranges), except for those with* which are presented as n (%)