Skip to main content

Table 3 Timing of instrumentation failure in vertical expandable prosthetic titanium rib technique

From: Instrumentation failure following pediatric spine deformity growth-sparing surgery using traditional growing rods or vertical expandable prosthetic titanium ribs

 

Index surgery

Lengthening surgery

P for trend

1st–3rd

4th–6th

7th–9th

≥ 10th

No. of surgeries

31

167

133

85

39

-

No. of IFs (%)

2 (6.5)

12 (7.2)

2 (1.5)

2 (2.4)

0 (0)

0.01*

No. of anchor loosenings/dislocations (%)

2 (6.5)

8 (4.8)

1 (0.8)

0 (0)

0 (0)

< 0.01*

No. of rod breakages (%)

0 (0)

1 (0.6)

0 (0)

2 (2.4)

0 (0)

0.55

No. of implant protrusions (%)

0 (0)

2 (1.2)

1 (0.8)

0 (0)

0 (0)

0.24

No. of proximal junctional kyphoses (%)

0 (0)

1 (0.6)

0 (0)

0 (0)

0 (0)

0.31

  1. IF: instrumentation failure
  2. *Statistically significant association between the number of lengthening surgeries performed and the surgery-based incidence of IF in the Cochran–Armitage test