Skip to main content
Fig. 5 | BMC Musculoskeletal Disorders

Fig. 5

From: Anterior variable-angle locked plating versus tension band wiring of simple and complex patella fractures – a biomechanical investigation

Fig. 5

Longitudinal (a,b) and shear (c,d) articular displacement at the central aspect of the patella, and interfragmentary rotation around the medio-lateral axis (e,f) in terms of mean value and standard deviation, measured between the corresponding proximal and distal patella fragments after 1000, 2000, 3000, 4000 and 5000 test cycles and featuring simple (a,c,e) and complex (b,d,f) fractures fixed by either anterior variable-angle locked plating (Plate) or TBW.

Back to article page