Skip to main content

Table 2 Precision of Measurement--AKDX and Hologic (n = 30)

From: Prescreening for osteoporosis with forearm bone densitometry in health examination population

 

BMD Mean Value±SD

p-value

RMS-SD (g/cm2)

Precision (CV%)

Low

AKDX

0.300 ± 0.001

< 0.01

0.033

0.40

Hologic

0.432 ± 0.003

0.002

0.50

Medium

AKDX

0.492 ± 0.002

< 0.01

0.003

0.20

Hologic

0.671 ± 0.006

0.006

0.85

High

AKDX

0.582 ± 0.002

< 0.01

0.003

0.19

Hologic

0.886 ± 0.007

0.007

0.82

Healthy volunteers

AKDX

0.402 ± 0.044

< 0.01

0.019

1.65

Hologic

0.633 ± 0.072

0.009

1.48

  1. Low, Medium, and High densities gradient phantom, SD standard deviation, RMS-SD root mean square standard deviation, % CV percent coefficient of variation