Skip to main content

Table 1 Biomechanical comparison of range of motion, gap formation and load to failure between DTG array and traditional flexor tendon suture

From: Dynamic Tendon Grip (DTG™) novel knot array compared to traditional sutures for zone two flexor tendon injury – a biomechanical feasibility study

 

Combined Range of motion (deg)

Gap formation (mm)

Load to Failure (N)

Pre-operative

Post-operative

  

DTG suture arraya

244.0 ± 9.9

235.62 ± 9.4

0.31 ± 0.48

76.51 ± 23.15

Traditional flexor tendon sutureb

234.67 ± 6.51

211.67 ± 10.50

0

66.31 ± 40.22

  1. aDynamic Tendon Grip
  2. bTraditional repair was performed with a 4-core suture Kessler array, with 3–0 FiberWire® (2-core sutures) and 3–0 PROLENE™ (2-core sutures) and a peripheral running 6–0 PROLENE™ suture