Table 1 Test-retest reliability measures for isometric knee extensor force (N) and Torque (Nm) (n = 22)

From: A portable isometric knee extensor strength testing device: test-retest reliability and minimal detectable change scores of the Q-Force ӀӀ in healthy adults

Force (N)

Leg

Testa

(mean ± SD)

Retesta

(mean ± SD)

ICC*

(95% CI)

SEM (%)b

MDC (%)c

Peak

Left

448.1 ± 144.5

462.8 ± 148.3

0.978 (0.946-0.991)

30.4 (6.7)

84.1 (18.5)

Right

478.4 ± 141.7

477.8 ± 136.5

0.978 (0.948-0.991)

28.5 (6.0)

79.0 (16.5)

Mean

Left

429.1 ± 138.8

443.6 ± 143.1

0.977 (0.944-0.991)

29.6 (6.8)

82.0 (18.8)

Right

460.4 ± 135.9

457.3 ± 134.1

0.978 (0.947-0.991)

28.0 (6.1)

77.6 (16.9)

Torque (Nm)

Peak

Left

136.1 ± 49.3

140.5 ± 50.0

0.977 (0.946-0.991)

10.4 (7.5)

28.9 (20.9)

Right

145.9 ± 48.5

145.4 ± 47.9

0.981 (0.954-0.992)

9.3 (6.4)

25.7 (17.6)

Mean

Left

130.3 ± 47.1

134.7 ± 48.1

0.977 (0.944-0.990)

10.2 (7.7)

28.3 (21.4)

Right

140.3 ± 46.4

139.1 ± 46.7

0.980 (0.952-0.992)

9.2 (6.6)

25.5 (18.2)

  1. a Mean of 3 trials; b% SEM expressed as a percentage of the average strength measure of the test and retest; c% MDC expressed as a percentage of the average strength measure of the test and retest. *All P-values < 0.001; CI Confidence Interval, ICC Intra Class Correlation, MDC Minimal Detectable Change, SD Standard deviation, SEM Standard Error of Measurement