Fig. 3From: Anterior prominence of the femoral condyle varies among prosthesis designs and surgical techniques in total knee arthroplastyComparison of medial, central, and lateral prominences (mm) using the anterior reference method with different implants. *Steel–Dwass test. The medial and central prominences were significantly lower with Journey II than with the other models (p < 0.001). The medial prominences were significantly lower with Attune and Persona than with Legion, but there was no significant difference between Attune and Persona. The central prominences were significantly lower in the order of Journey II, Attune, Persona, and Legion. The lateral prominence was significantly lower with Attune than with the other models (p < 0.04). The lateral prominences were significantly lower in the order of Attune, Journey II, Persona, and LegionBack to article page