Skip to main content

Table 2 Results of the 5000 cycles tests for both screws

From: Does thread shape affect the fixation strength of the bioabsorbable interference screws for anterior cruciate ligament reconstructions? A biomechanical study

 

2.5 mm pitch

5 mm pitch

p

Number of valid samples

9

6

 

Maximum insertion torque (N m)

2.01 ± 0.45

1.40 ± 0.33

 

Stiffness at 500th cycle (N/mm)

270.1 ± 55.7

261.0 ± 69.6

.906

Stiffness at 5000th cycle (N/mm)

281.3 ± 66.4

286.1 ± 79.4

.814

Displacement at 500th cycle (mm)

1.56 ± 0.78

1.84 ± 1.20

.814

Displacement at 5000th cycle (mm)

11.95 ± 14.11

6.97 ± 5.74

.724

Yield load (N)

476.4 ± 65.3

494.3 ± 39.2

.391

Pull-out stiffness (N/mm)

259.9 ± 53.4

236.5 ± 48.6

.886

Ultimate failure load (N)

586.8 ± 44.9

619.1 ± 127.6

.568

  1. Data are presented as mean ± SD