Table 3 Test-retest reliability of the goniometer and ETS for hip ROM measurement in patients with FAI and healthy subjects

From: Validity and test-retest reliability of manual goniometers for measuring passive hip range of motion in femoroacetabular impingement patients.

 

ICC2,1

CV (%)

SEM (°)

LOA (°)

 

Gonio

ETS

Gonio

ETS

Gonio

ETS

Gonio

ETS

Flexion

0.916

0.943

3.12

2.66

3.94

2.96

0.72 ± 11.15

0.27 ± 8.42

Abduction

0.924

0.947

5.84

5.66

2.36

2.01

0.14 ± 6.67

-0.08 ± 5.72

Adduction

0.842

0.823

6.73

6.34

2.36

1.59

-0.54 ± 6.60

1.07 ± 4.01

Internal rotation

0.950

0.902

7.74

10.19

2.42

2.93

-0.24 ± 6.89

1.32 ± 7.90

External rotation

0.914

0.934

5.23

5.10

2.53

1.81

0.02 ± 7.16

0.25 ± 5.14

  1. ETS: electromagnetic tracking system, ROM: range of motion, FAI: femoroacetabular impingement, ICC: intraclass correlation coefficient, CV: coefficient of variation, SEM: standard error of measurement, LOA: limits of agreement, Gonio: goniometer. Note that bias (retest lower than test) was significant only for adduction measured with the ETS (P < 0.01).