Myth or fact: 3D-printed off-the-shelf prosthesis is superior to titanium mesh cage in anterior cervical corpectomy and fusion?

Abstract

Background

To find out if three-dimensional printing (3DP) off-the-shelf (OTS) prosthesis is superior to titanium mesh cages in anterior cervical corpectomy and fusion (ACCF) when treating single-segment degenerative cervical spondylotic myelopathy (DCSM).

Methods

DCSM patients underwent ACCF from January 2016 to January 2019 in a single center were included. Patients were divided into the 3DP group (28) and the TMC group (23). The hospital stays, operation time, intraoperative blood loss, and the cost of hospitalization were compared. The Japanese Orthopedic Association (JOA) scores and Neck Disability Index (NDI) were recorded pre-operatively, 1 day, 3, 6, 12, and 24 months post-operatively. Radiological data was measured to evaluate fusion, subsidence, and cervical lordosis. Patients were sent with SF-36 to assess their health-related quality of life (HRQoL).

Results

The differences in operative time, intraoperative blood loss, and hospital stay were not statistically significant between groups (p > 0.05). Postoperative dysphagia occurred in 2 cases in the 3DP group and 3 cases in the TMC group, which all relieved one week later. The difference in improvement of JOA and NDI between the two groups was not statistically significant (p > 0.05). No hardware failure was found and bony fusion was achieved in all cases except one in the 3DP group. The difference in cervical lordosis (CL), fused segmental angle (FSA), mean vertebral height (MVH), and subsidence rates between groups at each follow-up time point was not statistically significant and the results of the SF-36 were similar (p > 0.05). The total cost was higher in the 3DP group with its higher graft cost (p < 0.05).

Conclusion

In treating single-segment DCSM with ACCF, both 3DP OTS prosthesis and TMC achieved satisfactory outcomes. However, the more costly 3DP OTS prosthesis was not able to reduce subsidence as it claimed.

Peer Review reports

Background

Degenerative cervical spondylotic myelopathy (DCSM) is a debilitating condition characterized by compression of the spinal cord due to degenerated intervertebral discs or osteophytes, resulting in symptoms such as upper arm pain, numbness, muscle weakness, and impaired mobility. Consequently, this disorder significantly diminishes the patient's overall quality of life [1, 2]. The symptoms of DCSM can be significantly alleviated through surgical intervention involving neurological decompression and stability reconstruction. Among the various surgical approaches, anterior cervical corpectomy and fusion (ACCF) has gained widespread recognition since its initial description in the 1950s [3, 4]. However, the optimal implant for the post-decompression gap remains a subject of debate due to the persisting complications associated with current implants, such as subsidence or non-union [5,6,7]. Autologous iliac bone grafts and titanium mesh cages (TMC) have been extensively utilized in ACCF. However, the occurrence of postoperative iliac pain, infection, and the high subsidence rate of implants have prompted surgeons to explore novel implant alternatives [8,9,10]. With the rapid advancement of the Three-dimensional printing (3DP) technique [11,12,13,14], the 3DP trabecular structured prosthesis is purported to enhance bone ingrowth and mitigate stress shielding effects by incorporating appropriate trabecular spacing, thereby ensuring enhanced initial stability [15, 16]. Anterior cervical spinal surgery has emerged as a pioneering adopter of 3DP technology, encompassing both patient-specific (PS) and market-available “Off-The-Shelf” (OTS) implants [13, 14, 16,17,18,19]. Only a limited number of studies reported sporadic cases of PS 3DP implants with a short-term follow-up [20, 21]. Nevertheless, there is a paucity of clinical studies to verify the efficacy of 3DP OTS prosthesis in ACCF. This study aimed to find out if 3DP OTS prosthesis is superior to TMC in ACCF as it claimed.

Methods

Patient population and indications

Fifty-one patients with DCSM who received ACCF from January 2016 to January 2019 were enrolled in this study with an informed consent agreement. The inclusion criteria were: (1) patients with symptoms and signs of DCSM in which conservative treatment was ineffective; (2) Patients with single or two-level disc degeneration, where the herniated disc or osteophyte extends to the posterior margin of the vertebral body, or focal ligaments are thickened, resulting in spinal cord compression, ACDF is not expected to provide complete decompression. The exclusion criteria were: (1) Continuous ossification of the posterior longitudinal ligament (OPLL); (2) developmental spinal stenosis; (3) preexisting dysphagia; (4) cervical scoliosis or kyphosis; (5) history of rheumatoid arthritis; (6) history of cervical spine tumor; (7) history of allergy to materials used in the procedure; (8) history of cervical spine trauma with/without cervical spine surgery; (9) evidence of systemic or local infection; (10) Osteoporosis was diagnosed by DXA (T-score is less than -2.5). Patients were divided into two groups: The 3DP group consisted of 28 patients (16 males and 12 females), aged 53.2 ± 6.4 years, who underwent ACCF with a 3DP OTS prosthesis; The TMC group consisted of 23 patients (12 males and 11 females), aged 53.3 ± 6.5 years, who underwent ACCF with TMC. All patients were followed up for at least 24 months postoperatively. None of these patients were lost to follow-up.

Surgical procedures

All surgeries were performed by the same senior surgeon (Kuang) using the standard Smith-Robinson method. The two intervertebral discs of the diseased segment are first removed, carefully removing the cartilage endplates and avoiding excessive damage to the endplates, followed by the removal of the middle vertebral body and removal of the posterior osteophytes with scrapers and Kerrison Rongeurs. After the decompression of the spinal cord and nerve roots, the cervical curvature was adjusted from a hyper-extended state to a physiological lordotic state. Subsequently, the Caspar retractor was released to maintain the gap at a similar height as the normal upper or lower segment. Lateral X-ray fluoroscopy was also used to check the height and position. In comparison with the adjacent normal vertebral space, there was no significant opening observed in the facet joint space, indicating an absence of excessive distraction [22]. The Casper retractor is also deployed upon implantation of the prosthesis to effectively restore the "natural height" of the segment to ensure congruity between the surgical space and implant dimensions. This reduces the occurrence of postoperative pain caused by over-extension and prevents rapid cage subsidence or loss of cervical curvature [23]. The width of the implant was determined by the distance between the two Luschka's joints. In the 3DP group, an appropriately sized OTS 3D-printed prosthesis (TITAN, AK Medical, Beijing, China) was implanted under fluoroscopy guidance. In the TMC group, titanium mesh cages (WEGO titanium mesh cage, WEGO, Weihai, China) were inserted with autogenous bone blended with porous bio-ceramic artificial bone (Dragonbio, Hubei, China) before implanted into the intervertebral space. The prostheses utilized in this study have diameter of either 12 or 14 mm. The upper and lower cover plates are absent in TMC. The inserted substitute, surrounded by the sharp ring of TMC, directly contacted with the vertebral endplates. In contrast, the exterior and interior of the 3DP prosthesis exhibit a porous trabecular structure with a flat surface on both ends without any sharp protrusion penetrating into the vertebral endplates. The same type of titanium anterior cervical plate (ATLANTIS, Medtronic Sofamor Danek, Memphis, USA) was attached and screw-locked to the front of the adjacent vertebrae of the surgical segment to stabilize the cervical spine. Drainage was placed after flushing and the wound was closed by layers.

Outcome assessment

The hospital stays, operation time, and intraoperative blood loss were recorded. The Japanese Orthopedic Association (JOA) scores, and Neck Disability Index (NDI) were measured preoperatively and recorded at 1 day, 3, 6, 12, and 24 months postoperatively. The results of the procedures were evaluated by two independent investigators (Wang and Chen), and the procedures were graded according to Odom’s criteria. Disagreements were resolved by discussion and consensus with another independent expert (Y Li). The interspinous motion (ISM) < 1 mm and superjacent interspinous motion ≥ 4 mm confirm the fusion diagnosis on the 150% mag magnified flexion and extension radiographs [24] (Fig. 1). The cervical lordosis (CL) fused segmental angle (FSA) and the mean vertebral height (MVH) of the surgical segment were taken pre-operatively and at 1 day, 3, 6, 12, and 24 months post-operatively (Fig. 2). The MVH was defined as the average height of the anterior border (HAB) and height of the posterior border (HPB). Subsidence was defined as loss of more than 3 mm in any of the HAB and HPB measured heights compared with the 1 day after operation [25]. Two independent researchers evaluated the radiographs (Pan and Yuan), and the results of their measurements were analyzed by intra-class correlation coefficient for data consistency. Disagreements on fusion were discussed with another independent expert (Lü) and a consensus was reached to minimize the observer bias. The Chinese version of Short Form-36 (SF-36) was used to assess the health-related quality of life (HRQoL) of patients(X Li).

Fig. 1
figure 1

Measurement of interspinous movement (ISM) at superjacent level (C4-5) and operated levels (C5-7) on the 150% magnified flexion and extension radiographs. The superjacent ISM at C4-5 (A and a) was 6.3 mm, which indicated adequate dynamic motion (≥ 4 mm). ISM at C5-6 (B and b) and C6-7 (C and c) were 1.5 mm and 1.4 mm, which was inconsistent with the definition of fusion (< 4 mm)

Fig. 2
figure 2

A The measurement method for cervical Cobb’s lordosis (CL): the angle formed by the upper end plate of C2 and lower end plate of C7 in neutral position (line a and line b); B the measurement method of the fused segment angle (FSA): the angle formed by the upper endplate of the superior vertebrae body and the lower endplate of the inferior vertebrae body in neutral position (line c and line d); C the measurement method for mean vertebral height (MVH): the mean value of the height of anterior (HAB) and the posterior border (HPB), The HAB and HPB were measured as the distance between the anterior and posterior points of the upper endplate of the superior vertebra and the lower endplate of the inferior vertebra

Statistical analysis

Statistical analysis was performed using Statistical Product and Service Solutions 28.0 statistical software (SPSS Inc, Chicago, Illinois, USA). The mean and standard deviation (x ± S) were used to express the measurement data, and the count data was expressed by the number of cases and percentages. The comparison of measurement data between groups was performed by student’s t-test, and the Paired t-tests were used to compare differences in preoperative and postoperative measurement data within groups; the non-parametric test was used for comparison of count data, and the difference was statistically significant at p < 0.05. The consistency of data collected by different researchers was analyzed by intra-class correlation coefficient (ICC) (ICC < 0.4, poor consistency; ICC > 0.75, good consistency).

Results

Clinical outcomes

There were no statistically significant differences in general information of patients between the 3DP group and the TMC group (Table 1). There were no statistically significant differences observed in terms of operation time, blood loss, and hospital stays between the 3DP group and the TMC group (Table 2). A total of five cases of postoperative dysphagia were observed (2 cases in the 3DP group and 3 cases in the TMC group), and symptoms were alleviated one week later following symptomatic treatment. Both groups showed improvement in NDI and JOA scores after surgery (p < 0.05), but differences at each time point were not statistically significant (Table 3). The mental health status and social roles of the HRQoL scores in both groups were nearly equivalent to those of healthy individuals. Additionally, the SF-36 results also revealed there was no significant difference between the two groups in HRQoL (Table 4).

Table 1 General information of patients (mean ± SD)
Table 2 Clinical results of patients (mean ± SD)
Table 3 Preoperative and postoperative (NDI, JOA) (mean ± SD)
Table 4 Results of the Short Form -(SF-)36 (Mean ± SD)

Radiological outcomes

The radiological data have good consistency derived from measurements by Wang and He. No instances of screw loosening or plate breakage were observed in either group, and successful bone fusion was achieved in all cases except for one patient within the 3DP group. The incidence of subsidence in the 3DP group at 3, 6, 12, and 24 months postoperatively was as follows: 7 cases (25.0%), 9 cases (32.1%), 10 cases (35.7%), and 10 cases (35.7%) respectively; while in the TMC group, it was observed as: 6 cases (26.1%), 7 cases (30.4%), 8 cases (34.8%), and 8 cases (34.8%). The observed differences in subsidence rate between the two groups did not reach statistical significance (p >0.05). The radiological parameters showed improvement at the last follow-up compared to the pre-operative measurements in each group (p <0.05) (Table 5). The radiological parameters did not exhibit statistically significant differences between the two groups at each follow-up time point. (p >0.05) (Fig. 3). The results showed that the intraobserver ICCs for CL, FSAs, and MVH were 0.990, 0.982, and 0.984, respectively, while the interobserver ICCs were 0.976, 0.963, and 0.968, respectively.

Table 5 Preoperative and postoperative radiological data (CL, FSA, MVH) (mean ± SD)
Fig. 3
figure 3

The radiological outcomes. There was no statistically significant difference between the two groups at each follow-up time point. pre-OP pre-operation, 1d P.O 1 day post-operatively, 3 m P.O 3 months post-operatively, 6 m P.O 6 months post-operatively, 12 m P.O 12 months post-operatively, 24 m P.O 24 months post-operatively

Costs

The costs of both groups were shown in Table 6. The total cost was higher in the 3DP group (p < 0.05). The cost of the prosthesis was higher in the 3DP group (p < 0.05). Differences in internal fixation and other costs between the two groups were not statistically significant (p > 0.05).

Table 6 Hospitalization cost of the patients

Discussion

The anterior cervical corpectomy and fusion achieve satisfactory decompression efficiency in the treatment of DCSM by relieving the compression of the spinal cord directly [1, 2]. It not only provides preliminary stability of the cervical spine but also plays a significant role in the maintenance of cervical lordosis. However, concerns have arisen regarding complications associated with the implant, including subsidence, non-union, screw migration, or implant breakage, which can ultimately lead to fixation failure [5,6,7]. Spine surgeons have been seeking implants with fewer implant-related complications [5,6,7,8,9,10]. The titanium mesh cage has replaced the autologous iliac crest bone for its initial stability and satisfactory fusion rate while avoiding harvest-related infection and injury, which facilitates pain relief and patient recovery [10]. However, most implants that are made of titanium alloy (Ti-6Al-4 V) have a high radiological opacity, which makes it difficult to evaluate their fusion on CT. Thus, fusion must be evaluated by observing the angle changes of the fused segment on the lateral flexion–extension radiograph. Besides, a high risk of subsidence of the titanium mesh cage has been reported due to its high stiffness [9]. Riew et al. evaluated the accuracy of four radiographic fusion criteria for determining anterior cervical fusion status [24]. These criteria include interspinous motion (ISM) < 1 mm with superjacent ISM ≥ 4 mm on dynamic radiographs [26], conventional bridging bone on CT scans [27], extra-graft bridging bone (ExGBB) and intra-graft bridging bone (InGBB) observed on multi-directional reconstructed CT scans [28]. He found that the ExGBB has the highest accuracy, sensitivity, and specificity compared to the other criteria, but the ISM criterion also demonstrated a similar accuracy to that of the conventional bridging bone criteria on CT scans. Despite the advantages of CT scans in detecting bridging bone and arthrodesis, the ISM criteria have advantages over CT scan in terms of cost and radiation exposure and also showed high inter- and intra-observer reliability and accuracy, as suggested in the article, making it the diagnostic standard for our patients.

The development of 3DP technology has led to the advancement of 3DP prostheses [11,12,13]. The 3DP prosthesis, characterized by its exquisite microstructure and excellent mechanical properties, has been developed with the aim of reducing complications associated with implants and shortening surgical duration [13]. The 3DP prosthesis currently used are divided into patient-specific (PS) and market-available “Off-The-Shelf” (OTS) ones according to the design requirements [13, 16,17,18,19]. The PS prosthesis is designed using computer-aided design (CAD) software, utilizing anatomical data obtained from the patient's CT scans. This CAD modeling technique ensures precise reconstruction of the patient's unique anatomy and enables the fabrication of a customized prosthesis tailored to meet specific surgical requirements and anticipated intraoperative needs. The clinical efficacy and radiological results of PS 3DP prosthesis used in single-segment ACCF were reported sporadically. Amelot et al. used PS 3DP prosthesis for vertebral reconstruction in ACCF and achieved satisfactory results [20]. Li et al. [21] and Lu et al. [29] compared the PS 3DP prosthesis with the titanium mesh cage (TMC) and concluded that PS-3DP prosthesis achieved comparable results in cervical lordosis reduction, neurological function, and HRQoL score, and the subsidence rate. Compared to the OTS implants, PS implants not only required less operative time but also exhibited reduced subsidence rates, potentially attributed to their enhanced geometric consistency achieved through mimicking the surface morphology of the endplate [30,31,32,33]. However, the extended production time and high manufacturing costs impose limitations on its practical application. Consequently, a 3DP OTS prosthesis was developed based on anatomical data from a large population, offering the potential to replace PS 3DP implants in most spine surgeries due to reduced costs and production time. Nevertheless, there is a scarcity of studies comparing clinical and radiological outcomes between the 3DP OTS prosthesis and commonly used TMCs, leaving the efficacy of the 3DP OTS prosthesis unknown [34]. In contrast to the PS 3DP prosthesis, our study found that the 3DP OTS prosthesis did not exhibit superior clinical or radiological results compared to conventional TMC.

Subsidence was defined as the migration of the intervertebral fusion device into the adjacent vertebral body, caused by stress concentration on the contact surface between the implant and the bone, which leads to implant failure, as well as undesirable consequences such as non-fusion and cervical kyphosis [35, 36]. Several risk factors may be associated with graft subsidence, including graft-related factors such as material composition, structure, shape, size, and dimension. Currently, both the 3DP OTS prosthesis and TMC are made of titanium alloy (Ti6Al-4 V) for their load-carrying capacity, cellular adhesion, and corrosion resistance [37, 38]. The stress shielding effect resulting from the mismatch in elastic modulus between titanium alloy and bone tissue induces higher local stresses at the edges, leading to bone resorption, non-union, and subsidence [39,40,41,42]. In our study, a total of 18 cases (35.3%) of subsidence were observed, with 10 cases (35.7%) in the 3DP group and 8 cases (38.1%) in the TMC group, respectively. However, no statistically significant difference was found between the two groups. Therefore, it can be concluded that the use of 3DP OTS prosthesis does not demonstrate superiority over TMC in preventing subsidence in ACCF.

Previous researches have demonstrated a significant correlation between implant shape, size, and dimensions and the occurrence of subsidence [43,44,45]. Furderer et al. [46] said that rectangular implants have a better-carrying capacity and could reduce the occurrence of subsidence. The reduction in contact area has been demonstrated by numerous biomechanical studies to result in stress concentration, which is implicated in the development of subsidence [43, 44, 46,47,48,49]. A study reported that the implementation of porous structures in 3DP prosthesis facilitates improved stress distribution, mitigates stress shielding effects, and reduces subsidence [50]. Some researchers compared the mechanical performance of individual PS designs with generic OTS designs and reported that the geometric conformity of the bone-implant interface of PS designs minimizes point-loading and resulting stress rise through a more uniform load distribution [34, 51]. The design of the OTS prosthesis failed to achieve similar geometric conformity as the PS ones. This may result in stress concentration, which may contribute to the result in our study that the subsidence rate of the 3DP OTS prosthesis was not superior to TMC.

Another reason is that although the 3D-printed prosthesis mimics the structure of bone trabeculae to facilitate bone ingrowth, it does not stimulate osteogenesis due to the absence of graft bone. During the fusion process between the implant and the vertebral body, it has been reported that osseointegration at the cortical and cancellous sites of long bones does not directly translate into intervertebral fusion because the endplate is a laminar porous structure composed of fused bone trabeculae, and fusion of the intervertebral fusion device to the endplate is similar to osseointegration at cancellous sites, even when implanted in a press-fit fashion. Consequently, the fusion between the implant and the vertebral body became non-robust, and subsidence occurred naturally. This may also be the reason for the non-fusion case in the 3DP group although her age was only 32 years and had no history of smoking, drinking, osteoporosis, obesity, or other comorbidity.

The bone mineral density (BMD) plays an important role in preventing subsidence [52,53,54,55]. The decreases in BMD aggravate the degree of elastic modulus mismatch between the vertebral body and the implant. Smoking, age, and sex may be involved in the occurrence of subsidence by affecting bone healing and BMD [52, 55, 56]. In our study, we excluded patients with osteoporosis to ensure that there would be no differences in bone quality that could affect the results.

The vertebral bone quality is a major factor affecting fixation. In terms of BMD assessment, dual-energy X-ray absorptiometry (DEXA) is considered the “gold standard” due to its simplicity and cost-effectiveness, with low-level radiation exposure. The t-value of DEXA less than 2.5 was used to diagnose osteoporosis. Although DEXA is evidently effective, there exist several methodological constraints in accurately quantifying BMD in patients with degenerative spine disorders. The presence of osteophytes, articular hypertrophy, and soft tissue deterioration, such as abdominal vascular wall calcification, can affect the lumbar BMD value and cause it to be exaggerated [57]. Besides, the measurement sites of DEXA are the lumbar spine and femur, not the cervical spine. It is still unclear whether there is a correlation between the actual value of the cervical spine and the lumbar spine, and how much bias there is. Therefore, there may be no overall osteoporosis but low bone mass density locally. Therefore, this study may not be able to completely rule out osteoporosis in the cervical spine. Recently, BMD assessment using computed tomography (CT) Hounsfield units (HU) has been developed as a modern trustworthy approach to measuring bone quality. HU values have been found to be favorably correlated with both vertebral compressive strength and DEXA-measured BMD [58]. Although the HU values offer the advantage of self-selected area measurement, there remains a lack of established diagnostic standards for osteoporosis, particularly in localized regions.

The preservation of the vertebral endplate and avoidance of intervertebral over-distraction have been demonstrated in studies to possess superior biomechanical properties, effectively preventing bone graft collapse, reducing kyphotic deformity, and maintaining intervertebral space height [48, 55, 59,60,61,62]. The endplate is a laminar porous structure composed of fused bone trabeculae. It contains pore-like forms of varying sizes, more in the central region and less in the periphery, which are the anatomical basis for the nutritional supply of the intervertebral disc and the maintenance of stress stability in the vertebral body. Therefore, the location of the implant also has an impact on the subsidence. The graft should be placed in the anterior periphery position of the endplate [60]. In this study, both groups of cases were operated by the same physician, who consistently followed identical procedures for implant insertion and end plate disposition in both groups. Consequently, potential surgical errors resulting from different surgeons were effectively eliminated.

Additionally, it has been postulated by certain researchers that the extent of corpectomy may serve as a risk factor for severe subsidence [33], with a higher likelihood of subsidence occurring in the inferior vertebrae, potentially attributed to the biomechanical forces acting upon them.

The maintenance of cervical lordosis is associated with subsidence, as previous studies have suggested that implant subsidence may contribute to the development of kyphosis [63]. Li et al. reported that a larger T1 Slope may be associated with subsidence [48]. An et al. also believed that more subsidence was observed when the T1 slope was larger, and the CL was corrected markedly [64]. The increase of CL and the large obliquity of the T1 slope were regarded as an inducer of subsidence. In our study, although both groups of patients demonstrated improvement in CL compared to their pre-operative status, there was no statistically significant difference observed between the two groups at each follow-up time point. This lack of significance may be attributed to the non-statistically significant difference in subsidence rate between the two groups.

The association between sagittal parameters and surgical outcomes extends well beyond that. The NDI score serves as a pivotal indicator for evaluating clinical outcomes. Rao et al. discovered that the mismatch between T1 slope and C2-C7 lordosis (T1s-CL) was associated with a more unfavorable postoperative NDI score compared to the matched group in laminoplasty studies [65]. Similarly, Lan et al. emphasized the significance of C2-7 SVA as a pivotal parameter for predicting the surgical outcome in patients with cervical kyphosis in their study [66].

Although both groups achieved comparable clinical and radiological outcomes in our study, the 3DP group exhibited a higher cost profile, with greater total hospital expenses compared to the TMC group. To be specific, the cost of internal fixation and the other costs of the two groups were similar. However, the average prosthesis cost of 3DP was nearly 3 times of TMC, even including bone graft. China spent USD 298 per person on health care and the per capita disposable income was USD 5,189 in 2022. The difference between the graft cost of 3DP and TMC is roughly 40% of annual per capita income (USD 2170 vs. USD 5189). Therefore, cost-effectiveness should be carefully considered in the selection of implants.

Our study had several limitations, including its single-center retrospective design with a relatively small sample size. To enhance the persuasiveness and confirm these findings, it is crucial to conduct a randomized multi-center prospective study with larger sample size. Furthermore, our study did not elucidate the specific relationship between local BMD, smoking, obesity, other comorbidities, and subsidence. Additionally, we excluded only patients with osteoporosis and did not compare bone mass between the 3DP group and the TMC group. Lastly, further research is needed to investigate the long-term effect of subsidence on post-operative cervical sagittal balance as well as its association with symptoms.

Conclusions

In the treatment of single-segment DCSM with ACCF, both 3DP OTS prosthesis and TMC achieved satisfactory outcomes, yet no statistically significant differences were observed between the two implants in terms of clinical and radiological results. The purported claim of reducing implant subsidence by the 3DP OTS prosthesis was not substantiated. However, it is worth noting that the 3DP OTS prosthesis incurred higher costs compared to TMC.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author upon reasonable request.

References

  1. Tetreault L, Goldstein CL, Arnold P, Harrop J, Hilibrand A, Nouri A, et al. Degenerative Cervical Myelopathy: A Spectrum of Related Disorders Affecting the Aging Spine. Neurosurgery. 2015;77(Suppl 4):S51–67.

    Article  PubMed  Google Scholar 

  2. Baptiste DC, Fehlings MG. Pathophysiology of cervical myelopathy. Spine J. 2006;6(6 Suppl):190S–197S.

    Article  PubMed  Google Scholar 

  3. Cloward RB. The anterior surgical approach to the cervical spine: the Cloward Procedure: past, present, and future. The presidential guest lecture, Cervical Spine Research Society. Spine (Phila Pa 1976). 1988;13:823–7.

    Article  CAS  PubMed  Google Scholar 

  4. Zhang Y, Quan Z, Zhao Z, Luo X, Tang K, Li J, et al. Evaluation of anterior cervical reconstruction with titanium mesh cages versus nano-hydroxyapatite/polyamide66 cages after 1- or 2-level corpectomy for multilevel cervical spondylotic myelopathy: a retrospective study of 117 patients. PLoS ONE. 2014;9:e96265.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Ashkenazi E, Smorgick Y, Rand N, Millgram MA, Mirovsky Y, Floman Y. Anterior decompression combined with corpectomies and discectomies in the management of multilevel cervical myelopathy: a hybrid decompression and fixation technique. J Neurosurg Spine. 2005;3:205–9.

    Article  PubMed  Google Scholar 

  6. Wei-bing X, Wun-Jer S, Gang L, Yue Z, Ming-xi J, Lian-shun J. Reconstructive techniques study after anterior decompression of multilevel cervical spondylotic myelopathy. J Spinal Disord Tech. 2009;22:511–5.

    Article  PubMed  Google Scholar 

  7. Goffin J, Geusens E, Vantomme N, Quintens E, Waerzeggers Y, Depreitere B, et al. Long-Term Follow-Up After Interbody Fusion of the Cervical Spine. J Spinal Disord. 2004;17:79–85.

    Article  Google Scholar 

  8. Malloy KM, Hilibrand AS. Autograft versus allograft in degenerative cervical disease. Clin Orthop Relat Res. 2002;(394):27–38.

  9. Silber JS, Anderson DG, Daffner SD, Brislin BT, Leland JM, Hilibrand AS, et al. Donor Site Morbidity After Anterior Iliac Crest Bone Harvest for Single-Level Anterior Cervical Discectomy and Fusion. Spine. 2003;28:134–9.

    Article  PubMed  Google Scholar 

  10. Nakase H, Park Y-S, Kimura H, Sakaki T, Morimoto T. Complications and long-term follow-up results in titanium mesh cage reconstruction after cervical corpectomy. J Spinal Disord Tech. 2006;19:353–7.

    Article  PubMed  Google Scholar 

  11. Ventola CL. Medical Applications for 3D Printing: Current and Projected Uses. P T. 2014;39:704–11.

    PubMed  PubMed Central  Google Scholar 

  12. Eltorai AEM, Nguyen E, Daniels AH. Three-Dimensional Printing in Orthopedic Surgery. Orthopedics. 2015;38:684–7.

    Article  PubMed  Google Scholar 

  13. Wilcox B, Mobbs RJ, Wu A-M, Phan K. Systematic review of 3D printing in spinal surgery: the current state of play. J Spine Surg. 2017;3:433–43.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Xu N, Wei F, Liu X, Jiang L, Cai H, Li Z, et al. Reconstruction of the Upper Cervical Spine Using a Personalized 3D-Printed Vertebral Body in an Adolescent With Ewing Sarcoma. Spine (Phila Pa 1976). 2016;41:E50–54.

    Article  PubMed  Google Scholar 

  15. Choy WJ, Parr WCH, Phan K, Walsh WR, Mobbs RJ. 3-dimensional printing for anterior cervical surgery: a review. J Spine Surg. 2018;4:757–69.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Burnard JL, Parr WCH, Choy WJ, Walsh WR, Mobbs RJ. 3D-printed spine surgery implants: a systematic review of the efficacy and clinical safety profile of patient-specific and off-the-shelf devices. Eur Spine J. 2020;29:1248–60.

    Article  PubMed  Google Scholar 

  17. Martelli N, Serrano C, van den Brink H, Pineau J, Prognon P, Borget I, et al. Advantages and disadvantages of 3-dimensional printing in surgery: A systematic review. Surgery. 2016;159:1485–500.

    Article  PubMed  Google Scholar 

  18. Tack P, Victor J, Gemmel P, Annemans L. 3D-printing techniques in a medical setting: a systematic literature review. Biomed Eng Online. 2016;15:115.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Malik HH, Darwood ARJ, Shaunak S, Kulatilake P, El-Hilly AA, Mulki O, et al. Three-dimensional printing in surgery: a review of current surgical applications. J Surg Res. 2015;199:512–22.

    Article  PubMed  Google Scholar 

  20. Amelot A, Colman M, Loret J-E. Vertebral body replacement using patient-specific three–dimensional-printed polymer implants in cervical spondylotic myelopathy: an encouraging preliminary report. Spine J. 2018;18:892–9.

    Article  PubMed  Google Scholar 

  21. Li X, Wang Y, Zhao Y, Liu J, Xiao S, Mao K. Multilevel 3D Printing Implant for Reconstructing Cervical Spine With Metastatic Papillary Thyroid Carcinoma. Spine (Phila Pa 1976). 2017;42:E1326–30.

    Article  PubMed  Google Scholar 

  22. Liu J, Ebraheim NA, Haman SP, Sanford CG, Sairyo K, Faizan A, et al. How the Increase of the Cervical Disc Space Height Affects the Facet Joint: An Anatomy Study. Spine. 2006;31:E350–4.

    Article  PubMed  Google Scholar 

  23. Wang S, Zhu J, Sun K, Chen R, Cao J, You R, et al. Clinical impact and imaging results after a modified procedure of ACDF: a prospective case-controlled study based on ninety cases with two-year follow-up. BMC Musculoskelet Disord. 2021;22:605.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Riew KD, Yang JJ, Chang D-G, Park S-M, Yeom JS, Lee JS, et al. What is the most accurate radiographic criterion to determine anterior cervical fusion? Spine J. 2019;19:469–75.

    Article  PubMed  Google Scholar 

  25. Ji C, Yu S, Yan N, Wang J, Hou F, Hou T, et al. Risk factors for subsidence of titanium mesh cage following single-level anterior cervical corpectomy and fusion. BMC Musculoskelet Disord. 2020;21:32.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  26. Song K-S, Piyaskulkaew C, Chuntarapas T, Buchowski JM, Kim HJ, Park MS, et al. Dynamic Radiographic Criteria for Detecting Pseudarthrosis Following Anterior Cervical Arthrodesis. J Bone Joint Surg. 2014;96:557–63.

    Article  PubMed  Google Scholar 

  27. Phillips FM, Carlson G, Emery SE, Bohlman HH. Anterior cervical pseudarthrosis. Natural history and treatment. Spine (Phila Pa 1976). 1997;22:1585–9.

    Article  CAS  PubMed  Google Scholar 

  28. Song K-S, Chaiwat P, Kim HJ, Mesfin A, Park S-M, Riew KD. Anterior cervical fusion assessment using reconstructed computed tomographic scans: surgical confirmation of 254 segments. Spine (Phila Pa 1976). 2013;38:2171–7.

    Article  PubMed  Google Scholar 

  29. Lu T, Liu C, Yang B, Liu J, Zhang F, Wang D, et al. Single-Level Anterior Cervical Corpectomy and Fusion Using a New 3D-Printed Anatomy-Adaptive Titanium Mesh Cage for Treatment of Cervical Spondylotic Myelopathy and Ossification of the Posterior Longitudinal Ligament: A Retrospective Case Series Study. Med Sci Monit. 2017;23:3105–14.

    Article  PubMed  PubMed Central  Google Scholar 

  30. Mobbs RJ, Choy WJ, Wilson P, McEvoy A, Phan K, Parr WCH. L5 En-Bloc Vertebrectomy with Customized Reconstructive Implant: Comparison of Patient-Specific Versus Off-the-Shelf Implant. World Neurosurg. 2018;112:94–100.

    Article  PubMed  Google Scholar 

  31. Arts MP, Peul WC. Vertebral body replacement systems with expandable cages in the treatment of various spinal pathologies: a prospectively followed case series of 60 patients. Neurosurgery. 2008;63:537–44. discussion 544-545.

    Article  PubMed  Google Scholar 

  32. Lau D, Song Y, Guan Z, La Marca F, Park P. Radiological outcomes of static vs expandable titanium cages after corpectomy: a retrospective cohort analysis of subsidence. Neurosurgery. 2013;72:529–39. discussion 528-529.

    Article  PubMed  Google Scholar 

  33. Chen Y, Chen D, Guo Y, Wang X, Lu X, He Z, et al. Subsidence of titanium mesh cage: a study based on 300 cases. J Spinal Disord Tech. 2008;21:489–92.

    Article  PubMed  Google Scholar 

  34. Mobbs RJ, Parr WCH, Choy WJ, McEvoy A, Walsh WR, Phan K. Anterior Lumbar Interbody Fusion Using a Personalized Approach: Is Custom the Future of Implants for Anterior Lumbar Interbody Fusion Surgery? World Neurosurg. 2019;124:452–458.e1.

    Article  PubMed  Google Scholar 

  35. Pinder EM, Sharp DJ. Cage subsidence after anterior cervical discectomy and fusion using a cage alone or combined with anterior plate fixation. J Orthop Surg (Hong Kong). 2016;24:97–100.

    Article  CAS  PubMed  Google Scholar 

  36. Wu J, Luo D, Ye X, Luo X, Yan L, Qian H. Anatomy-related risk factors for the subsidence of titanium mesh cage in cervical reconstruction after one-level corpectomy. Int J Clin Exp Med. 2015;8:7405–11.

    PubMed  PubMed Central  Google Scholar 

  37. Rao PJ, Pelletier MH, Walsh WR, Mobbs RJ. Spine interbody implants: material selection and modification, functionalization and bioactivation of surfaces to improve osseointegration. Orthop Surg. 2014;6:81–9.

    Article  PubMed  PubMed Central  Google Scholar 

  38. Chong E, Pelletier MH, Mobbs RJ, Walsh WR. The design evolution of interbody cages in anterior cervical discectomy and fusion: a systematic review. BMC Musculoskelet Disord. 2015;16:99.

    Article  PubMed  PubMed Central  Google Scholar 

  39. Geetha M, Singh AK, Asokamani R, Gogia AK. Ti based biomaterials, the ultimate choice for orthopaedic implants – A review. Prog Mater Sci. 2009;54:397–425.

    Article  CAS  Google Scholar 

  40. Duan H, Cao C, Wang X, Tao J, Li C, Xin H, et al. Magnesium-alloy rods reinforced bioglass bone cement composite scaffolds with cortical bone-matching mechanical properties and excellent osteoconductivity for load-bearing bone in vivo regeneration. Sci Rep. 2020;10:18193.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  41. Vaezi M, Black C, Gibbs DMR, Oreffo ROC, Brady M, Moshrefi-Torbati M, et al. Characterization of New PEEK/HA Composites with 3D HA Network Fabricated by Extrusion Freeforming. Molecules. 2016;21:687.

    Article  PubMed  PubMed Central  Google Scholar 

  42. Wang C, Xu D, Li S, Yi C, Zhang X, He Y, et al. Effect of Pore Size on the Physicochemical Properties and Osteogenesis of Ti6Al4V Porous Scaffolds with Bionic Structure. ACS Omega. 2020;5:28684–92.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  43. Kwon SW, Kim CH, Chung CK, Park TH, Woo SH, Lee S-J, et al. The Formation of Extragraft Bone Bridging after Anterior Cervical Discectomy and Fusion: A Finite Element Analysis. J Korean Neurosurg Soc. 2017;60:611–9.

    Article  PubMed  PubMed Central  Google Scholar 

  44. Schlenk RP, Stewart T, Benzel EC. The biomechanics of iatrogenic spinal destabilization and implant failure. Neurosurg Focus. 2003;15:E2.

    Article  PubMed  Google Scholar 

  45. Fürderer S, Schöllhuber F, Rompe JD. Eysel P [Effect of design and implantation technique on risk of progressive sintering of various cervical vertebrae cages]. Orthopade. 2002;31:466–71.

    Article  PubMed  Google Scholar 

  46. Akamaru T, Kawahara N, Sakamoto J, Yoshida A, Murakami H, Hato T, et al. The transmission of stress to grafted bone inside a titanium mesh cage used in anterior column reconstruction after total spondylectomy: a finite-element analysis. Spine (Phila Pa 1976). 2005;30:2783–7.

    Article  PubMed  Google Scholar 

  47. Sun K, Sun J, Wang S, Xu X, Wang Y, Xu T, et al. Placement of Titanium Mesh in Hybrid Decompression Surgery to Avoid Graft Subsidence in Treatment of Three-Level Cervical Spondylotic Myelopathy: Cephalad or Caudal? Med Sci Monit. 2018;24:9479–87.

    Article  PubMed  PubMed Central  Google Scholar 

  48. Zhang B, Li S, Miao D, Zhao C, Wang L. Risk Factors of Cage Subsidence in Patients with Ossification of Posterior Longitudinal Ligament (OPLL) After Anterior Cervical Discectomy and Fusion. Med Sci Monit. 2018;24:4753–9.

    Article  PubMed  PubMed Central  Google Scholar 

  49. Fürderer S, Schöllhuber F, Rompe J-D, Eysel P. Einfluss von Design und Implantationstechnik auf das Risiko der progredienten Sinterung verschiedener HWS-Cages. Orthopäde. 2002;31:466–71.

    Article  PubMed  Google Scholar 

  50. Moussa A, Tanzer M, Pasini D. Cervical fusion cage computationally optimized with porous architected Titanium for minimized subsidence. J Mech Behav Biomed Mater. 2018;85:134–51.

    Article  CAS  PubMed  Google Scholar 

  51. Zhang Y-W, Deng L, Zhang X-X, Yu X-L, Ai Z-Z, Mei Y-X, et al. Three-Dimensional Printing-Assisted Cervical Anterior Bilateral Pedicle Screw Fixation of Artificial Vertebral Body for Cervical Tuberculosis. World Neurosurg. 2019;127:25–30.

    Article  PubMed  Google Scholar 

  52. Brenke C, Dostal M, Scharf J, Weiß C, Schmieder K, Barth M. Influence of cervical bone mineral density on cage subsidence in patients following stand-alone anterior cervical discectomy and fusion. Eur Spine J. 2015;24:2832–40.

    Article  PubMed  Google Scholar 

  53. Hasegawa K, Abe M, Washio T, Hara T. An experimental study on the interface strength between titanium mesh cage and vertebra in reference to vertebral bone mineral density. Spine (Phila Pa 1976). 2001;26:957–63.

    Article  CAS  PubMed  Google Scholar 

  54. Tempel ZJ, Gandhoke GS, Okonkwo DO, Kanter AS. Impaired bone mineral density as a predictor of graft subsidence following minimally invasive transpsoas lateral lumbar interbody fusion. Eur Spine J. 2015;24(Suppl 3):414–9.

    Article  PubMed  Google Scholar 

  55. Lim TH, Kwon H, Jeon CH, Kim JG, Sokolowski M, Natarajan R, et al. Effect of endplate conditions and bone mineral density on the compressive strength of the graft-endplate interface in anterior cervical spine fusion. Spine (Phila Pa 1976). 2001;26:951–6.

    Article  CAS  PubMed  Google Scholar 

  56. Natarajan RN, Chen BH, An HS, Andersson GB. Anterior cervical fusion: a finite element model study on motion segment stability including the effect of osteoporosis. Spine (Phila Pa 1976). 2000;25:955–61.

    Article  CAS  PubMed  Google Scholar 

  57. Lee JH, Lee J-H, Park JW, Shin YH. The insertional torque of a pedicle screw has a positive correlation with bone mineral density in posterior lumbar pedicle screw fixation. J Bone Joint Surg Br. 2012;94:93–7.

    Article  CAS  PubMed  Google Scholar 

  58. Schreiber JJ, Anderson PA, Rosas HG, Buchholz AL, Au AG. Hounsfield units for assessing bone mineral density and strength: a tool for osteoporosis management. J Bone Joint Surg Am. 2011;93:1057–63.

    Article  PubMed  Google Scholar 

  59. Daubs MD. Early failures following cervical corpectomy reconstruction with titanium mesh cages and anterior plating. Spine (Phila Pa 1976). 2005;30:1402–6.

    Article  PubMed  Google Scholar 

  60. Truumees E, Demetropoulos CK, Yang KH, Herkowitz HN. Effects of Disc Height and Distractive Forces on Graft Compression in an Anterior Cervical Discectomy Model. Spine. 2002;27:2441–5.

    Article  PubMed  Google Scholar 

  61. Colman MW, Guss A, Bachus KN, Spiker WR, Lawrence BD, Brodke DS. Fixed-Angle, Posteriorly Connected Anterior Cage Reconstruction Improves Stiffness and Decreases Cancellous Subsidence in a Spondylectomy Model. Spine (Phila Pa 1976). 2016;41:E519–523.

    Article  PubMed  Google Scholar 

  62. Jang H-J, Chin D-K, Kim K-H, Park J-Y. Does Graft Position Affect Subsidence After Anterior Cervical Discectomy and Fusion? Global Spine J. 2022;12:668–76.

    Article  PubMed  Google Scholar 

  63. Lee Y-S, Kim Y-B, Park S-W. Risk factors for postoperative subsidence of single-level anterior cervical discectomy and fusion: the significance of the preoperative cervical alignment. Spine (Phila Pa 1976). 2014;39:1280–7.

    Article  PubMed  Google Scholar 

  64. An TY, Kim J-Y, Lee Y-S. Risk Factors and Radiologic Changes in Subsidence after Single-Level Anterior Cervical Corpectomy: A Minimum Follow-Up of 2 Years. Korean J Neurotrauma. 2021;17:126.

    Article  PubMed  PubMed Central  Google Scholar 

  65. Rao H, Huang Y, Lan Z, Xu Z, Li G, Xu W. Does Preoperative T1 Slope and Cervical Lordosis Mismatching Affect Surgical Outcomes After Laminoplasty in Patients with Cervical Spondylotic Myelopathy? World Neurosurgery. 2019;130:e687–93.

    Article  PubMed  Google Scholar 

  66. Lan Z, Wu Z, Huang Y, Xu W. Increased Neck Tilt/T1 slope ratio may play an important role in patients with cervical kyphosis. BMC Musculoskelet Disord. 2021;22:785.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

Not applicable.

Informed consent

Informed consent was obtained upon the final follow-up by written consent.

Disclosures

The authors have no financial or competing interests in relation to this work. This study was approved by the Ethical review committee of The Second Xiangya Hospital of Central South University.

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

All surgeries were performed by KL. The manuscript was written by HHY and revised by KL. The clinical results were evaluated by WXX and CZJ, and disagreements were resolved by discussion and consensus with LYC. The Radiological results were evaluated by PCY and YH, and disagreements were discussed with LGH. The Form 36 (SF-36) were completed by LXY. The statistical calculation was completed by ZO. The pictures and tables are done by FL.

Corresponding author

Correspondence to Lei Kuang.

Ethics declarations

Ethics approval and consent to participate

All procedures performed in studies involving human participants were in accordance with the ethical standards of the Ethics Committee of The Second Xiangya Hospital of Central South University.

This study was approved by The Ethics Committee of The Second Xiangya Hospital of Central South University (No. S043). Written informed consent to participate in the study was obtained from each participant (2023) Ethical review【Scientific】NO.(043).

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

He, H., Fan, L., Lü, G. et al. Myth or fact: 3D-printed off-the-shelf prosthesis is superior to titanium mesh cage in anterior cervical corpectomy and fusion?. BMC Musculoskelet Disord 25, 96 (2024). https://doi.org/10.1186/s12891-024-07213-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-024-07213-7

Keywords