Table 2 Comparison of the minimal detectable changes with the lowest standard significant differences *

From: Reliability of the standard goniometry and diagrammatic recording of finger joint angles: a comparative study with healthy subjects and non-professional raters

Position, Goniometer

MDC

Lowest significant standard difference (in degrees)

I-A

II-A

Rater 1 (I-B)

Rater 2 (I-B)

Rater 3 (II-B)

Rater 4 (II-Bcpa)

EXT, PGn

11.5

9.3

14

9

9

9

SGn

11.6

9.0

9

9

9

9

FLEX, PGn

9.5

11.5

9

9

9

9

SGn

9.9

8.7

13

9

9

5

  1. MDC = minimal detectable changes obtained from the corresponding SEMs in Table 1; I-A, II-A, I-B, and II-B = appropriate study parts; EXT, FLEX = position of imitated incomplete extension and flexion, respectively; PGn = paper goniometer; SGn = standard finger goniometer; * the lowest difference between the angles of the try-angles application of which resulted in significant differences between the corresponding observed PIP joint angles according to Wilcoxon signed ranks test (see Additional file 6 for details).