Table 2 Comparison of radiographic parameters measured on full-length radiographs of the lower limb taken in the standing and supine status

From: Factor affecting the discrepancy in the coronal alignment of the lower limb between the standing and supine radiographs

Variablesa

Full-length standing radiographb

Full-length supine radiographb

P Valuec

Hip-knee-ankle angle, °

5.2 ± 4.0

4.3 ± 3.6

<  0.001

Limb alignment

 Varus/ Neutral/ Valgus

118/ 27/ 2

107/ 38/ 2

0.056

Medial proximal tibial angle, °

86.0 ± 2.1

85.7 ± 2.3

0.083

Lateral distal femoral angle, °

88.1 ± 2.0

88.3 ± 2.0

<  0.001

Joint line convergence angle, °

3.2 ± 2.4

1.1 ± 1.9

<  0.001

Kellgren-Lawrence grade

 0/ 1/ 2/ 3/ 4 (observer 1)

44/ 37/ 23/ 25/ 18

43/ 40/ 28/ 23/ 13

0.007

 0/ 1/ 2/ 3/ 4 (observer 2)

30/ 44/ 26/ 20/ 27

31/ 44/ 31/ 18/ 23

0.023

  1. a The average value of the observers’ measurements was used for continuous variables, and each measurement value of observers was used separately for categorical variables
  2. b Data are presented as the mean ± standard deviation or number of limbs
  3. c Paired t test was used for continuous variables, and McNemar-Bowker’s test was used for categorical variables