Table 1 Material properties of FE models’ components

From: Stepwise reduction of bone mineral density increases the risk of cage subsidence in oblique lumbar interbody fusion patients biomechanically: an in-silico study

Components

Elastic modulus (MPa)

Poisson’s ratio

Cross-section (mm2)

Mesh sizes (mm)

References

Bony structures

 Cortical (Normal BMD)

Exx = 11,300

Eyy = 11,300

Ezz = 22,000

Gxy = 3800

Gyz = 5400

Gxz = 5400

Vxy = 0.484

Vyz = 0.203

Vxz = 0.203

/

1.7

[14, 17]

 Cancellous (Normal BMD)

Exx = 140

Eyy = 140

Ezz = 200

Gxy = 48.3

Gyz = 48.3

Gxz = 48.3

Vxy = 0.45

Vyz = 0.315

Vxz = 0.315

/

2.6

[14, 17]

 Bony endplates (Normal BMD)

12,000

0.3

/

1.2

[14, 17]

 Cortical (Slight reduction of BMD)

Exx = 9436

Eyy = 9436

Ezz = 18,370

Gxy = 3173

Gyz = 4509

Gxz = 4509

Vxy = 0.484

Vyz = 0.203

Vxz = 0.203

/

1.7

[14, 17]

 Cancellous (Slight reduction of BMD)

Exx = 93.8

Eyy = 93.8

Ezz = 150

Gxy = 32.36

Gyz = 36.23

Gxz = 36.23/

Vxy = 0.45

Vyz = 0.315

Vxz = 0.315

/

2.6

[14, 17]

 Bony endplates (Slight reduction of BMD)

10,035

0.3

/

1.2

[14, 17]

 Cortical (Significant reduction of BMD)

Exx = 7571

Eyy = 7571

Ezz = 14,740

Gxy = 2546

Gyz = 3618

Gxz = 3618

Vxy = 0.484

Vyz = 0.203

Vxz = 0.203

 

1.7

[14, 17]

 Cancellous (Significant reduction of BMD)

Exx = 47.6

Eyy = 47.6

Ezz = 100

Gxy = 16.42

Gyz = 24.15

Gxz = 24.15

Vxy = 0.45

Vyz = 0.315

Vxz = 0.315

/

2.6

[14, 17]

 Bony endplates (Significant reduction of BMD)

8070

0.3

/

1.2

[14, 17]

Non bony structures

 Annulus

Hypoelastic material

/

2.0

[18, 26]

 Nucleus

1

0.49

/

1.7

[18, 26]

 Cartilages

10

0.4

 

0.7

[18, 26]

Ligaments

 Anterior longitudinal

ligaments

Calibrated load-deformation curved under different loading conditions

0.3

60

/

[16, 17]

 Posterior longitudinal

ligaments

Calibrated load-deformation curved under different loading conditions

0.3

21

/

[16, 17]

 Ligamentum flavum

Calibrated load-deformation curved under different loading conditions

0.3

60

/

[16, 17]

 Interspinous

ligaments

Calibrated load-deformation curved under different loading conditions

0.3

40

/

[16, 17]

 Supraspinous

ligaments

Calibrated load-deformation curved under different loading conditions

0.3

30

/

[16, 17]

 Intertransverse

ligaments

Calibrated load-deformation curved under different loading conditions

0.3

10

/

[16, 17]

 Capsular

7.5 (≤25%)

32.9 (>25%)

0.3

67.5

/

[16, 17]

Internal fixation devices

 PEEK OLIF Cage

3500

0.3

/

2.0

[16, 17]

 Titanium alloy screw

110,000

0.3

/

0.5

[16, 17]

  1. Abbreviations: FE Finite element, BMD Bone mineral density, PEEK Polyetheretherketone, OLIF Oblique lumbar interbody fusion