Skip to main content

Table 3 Finite element analysis results of three internal fixation models

From: Finite element comparative analysis of three different internal fixation methods in the treatment of Pauwels type III femoral neck fractures

Parameter

CLS

DHS + DS

FNS

Stress of the proximal fracture fragment (MPa)

30.664

25.404

35.651

Displacement of the proximal fragment (mm)

1.817

1.255

1.828

Internal fixation stress (MPa)

95.596

111.380

281.2

Displacement of internal fixation (mm)

1.667

1.1563

1.642

Displacement of fracture surface at the broken end (mm)

1.281

0.887

1.286