Skip to main content

Table 3 Difference in postoperative radiographic measurements between stem types

From: Restoration of hip geometry after total hip arthroplasty: retrospective comparison of two short stems and one straight stem

 

Group I (Proxy®)

N = 99

Group II (Optimys®)

N = 62

Group III (Corail®)

N = 90

P-value

(one-way ANOVA)

P-value (t-test)

Centrum-Collum-Diaphysis-Angle

 Postoperative CCD Angle (mean + SD; in degrees)

138.6 ± 5.1

134.0 ± 5.0

136.7 ± 2.9

 < 0.001

I vs II: < 0.001

I vs III: 0.003

II vs III: < 0.001

 Difference in CCD Angle (mean + SD; in degrees)

9.3 ± 7.2

6.8 ± 6.9

8.6 ± 7.3

0.097

 

Femoral Offset

 Postoperative Femoral Offset (mean + SD; in mm)

40.9 ± 6.9

43.1 ± 7.4

41.5 ± 5.7

0.125

 

 Difference in Femoral Offset (mean + SD; in mm)

1.6 ± 6.9

4.1 ± 7.8

4.9 ± 7.2

0.006

I vs II: 0.033

I vs III: 0.002

II vs III: 0.534

Acetabular Offset

 Postoperative Acetabular Offset (mean + SD; in mm)

34.7 ± 4.5

35.2 ± 3.6

31.9 ± 3.6

 < 0.001

I vs II: 0.417

I vs III: < 0.001

II vs III: < 0.001

 Difference in Acetabular Offset (mean + SD; in mm)

-1.4 ± 3.6

-2.0 ± 4.4

-3.3 ± 5.3

0.011

I vs II: 0.394

I vs III: 0.003

II vs III: 0.071

Total Offset

 Postoperative Total Offset (mean + SD: in mm)

75.6 ± 7.7

78.3 ± 8.0

73.4 ± 6.4

 < 0.001

I vs II: 0.024

I vs III: 0.042

II vs III: < 0.001

 Difference in Total Offset (mean + SD; in mm)

0.2 ± 6.3

2.1 ± 6.6

1.6 ± 7.2

0.177

 
  1. Legend: bold significant result