Skip to main content

Table 3 Peak and through-range measurement differences between DorsaVi Version 6 and Vicon systems

From: Concurrent validity of DorsaVi wireless motion sensor system Version 6 and the Vicon motion analysis system during lifting

Flexion

Root mean square error (SD)

95% CI

Mean difference

Lower limit of agreement

Upper limit of agreement

Symmetrical Peak Flexion

     

Global-T12-angle

0.55 (± 0.49)

0.46–0.64

-0.90

-6.80

5.00

Global-S2-angle

0.44 (± 0.34)

0.38–0.50

0.60

-2.16

3.36

Relative-lumbar-angle

0.68 (± 0.55)

0.58–0.78

-1.20

-8.06

5.67

Asymmetrical Peak Flexion

     

Global-T12-angle

2.06 (± 1.59)

1.91–2.21

-1.59

-8.66

5.48

Global-S2-angle

2.00 (± 1.63)

1.85–2.15

-0.60

-7.00

5.79

Relative-lumbar-angle

1.24 (± 1.08)

1.14–1.34

-0.84

-8.55

6.88

Symmetrical Through-Range

     

Global-T12-angle

0.86 (± 0.78)

0.84–0.88

   

Global-S2-angle

0.90 (± 0.84)

0.88–0.92

   

Relative-lumbar-angle

1.34 (± 1.25)

1.31–1.37

   

Asymmetrical Through-Range

     

Global-T12-angle

1.84 (± 1.58)

1.82–1.86

   

Global-S2-angle

1.90 (± 1.65)

1.88–1.92

   

Relative-lumbar-angle

1.70 (± 1.54)

1.68–1.72