Table 2 Max contact pressure and wear data for Posterior Tibial Slope degree variation wrt cycle

From: Computational wear of knee implant polyethylene insert surface under continuous dynamic loading and posterior tibial slope variation based on cadaver experiments with comparative verification

 

(Tibial Slope) Flexion

3 Degrees

5 Degrees

7 Degrees

Cycle

Press (MPa)

Wear (mm)

Press (MPa)

Wear (mm)

Press (MPa)

Wear (mm)

100 k

2.0316

0.0061495

2.0295

0.0061434

2.0264

0.0061341

1 m

1.8544

0.04654

1.8526

0.046483

1.85

0.046397

3 m

1.5098

0.099388

1.5068

0.099245

1.5043

0.099073

10 m

1.0399

0.20896

1.0374

0.20867

1.04

0.20822

20 m

0.85564

0.30619

0.85496

0.30573

0.85338

0.30494

30 m

0.66597

0.38078

0.66875

0.38014

0.64804

0.37913