Repetition time (TR) | 5500 ms |
Echo time (TE) | 104 ms |
Echo-planar imaging (EPI) factor | 192 |
Number of acquisitions (averages) | 3 |
Number of concatenations | 1 |
Number of slices | 17 |
Matrix (frequency x phase) | 192 × 192 |
Field of view (FoV) | 350 mm × 350 mm |
Slice thickness | 4.0 mm |
Interslice gap | 0.0 mm |
Voxel size | 1.8 mm × 1.8 mm × 4.0 mm |
Receiver bandwidth | 1628 Hz/pixel |
Phase encoding direction | Anterior to posterior |
Saturation pulses | Anterior, 30 mm |
Acquisition time | 3 min 48 s |
Coverage | From above Th12 to below S2 |
Phase oversampling | 0% |
Fat saturation technique | Chemical shift-selective pre-pulse |
Parallel acquisition techniques (PAT) mode | None |
Distortion correction filters | Yes |
b values | 50, 400, 800 s/mm2 |
Diffusion weightings (b values) | 3 |
Diffusion encoding scheme | Bipolar |