Table 1 Summary of patients’ characteristics and related findings

From: Liner dissociation in a large-diameter ceramic-bearing acetabular component: a report of five cases

Patient number

Gender

Age at index procedure (years)

BMI (kg/m2)

Cup size/head diameter

Cup inclination/anteversion angles on CT evaluation

Time to dissociation diagnosis (months)

Sudden noise

Sudden pain

Pneumarthrosis on radiographic evaluation

Revision surgery

Joint fluid at revision surgery

Soft tissue changes/signs of ARMD

1

Male

59

32.8

58 mm/48 mm

42°/16°

78

Yes

Yes

Yes

Yes

Large amount

Yes

2

Male

49

34.2

58 mm/48 mm

44°/7°

42

Yes

Yes

Yes

Yes

Minimal

No

3

Male

55

46.2

56 mm/44 mm

38°/27°

60

Yes

Yes

Yes

No after 1 year

-

-

4

Male

39

41.5

58 mm/48 mm

45°/9°

69

Yes

Yes

Yes

Yes

Minimal

No

5

Male

53

27.4

56 mm/44 mm

39°/25°

80

Yes

Yes

Yes

Yes

Large amount

Yes