Skip to main content

Table 1 The material properties used in the finite element models

From: Applicable safety analysis and biomechanical study of iliosacral triangular osteosynthesis

Materials

Young’s modulus (MPa)

Poisson’s ratio(u)

Reference

ligament

Stiffness (N/mm)

Number of elements

Reference

Titanium screw/rod/plate

110,000

0.3

[24]

Anterior sacroiliac ligament

700

10 × 2

[25]

Cortical bone (Pelvis)

17,000

0.3

[23]

Posterior sacroiliac ligament (long)

1000

4 × 2

[25]

Cancellous bone (Pelvis)

100

0.2

[23]

Posterior sacroiliac ligament (short)

400

10 × 2

[25]

Cortical bone (Lumbar)

12,000

0.3

[23]

Interosseous sacroiliac ligament

2800

4 × 2

[25]

Cancellous bone (Lumbar)

100

0.2

[23]

Sacrospinous ligament

1400

5 × 2

[25]

Posterior bony elements

3500

0.25

[24]

Sacrotuberous ligament

1500

5 × 2

[25]

Articular cartilage

10

0.4

[24]

Superior pubic ligament

500

1 × 1

[25]

Bony endplate

1000

0.4

[24]

Accurate pubic ligament

500

1 × 1

[25]

Cartilage endplate

25

0.25

[24]

Iliolumbar ligament

1000

4 × 2

[25]

Nucleus pulposus

1

0.499

[24]

    

Matrix of annulus fibrosus

4.2

0.45

[24]

    

Fibers of annulus fibrosus

450

0.3

[24]

    

Interpubic disc

5

0.45

[24]

   Â