Table 2 Effect of tested parameters on the apparent stiffness (Kz) of the proximal femoral samples

From: Which experimental procedures influence the apparent proximal femoral stiffness? A parametric study

Parameter

Avg. %Diff

SigMethod

P

EffectSize

Re-fixation

-3.74

NA

NA

NA

Pre-loading

0.70

Wilx.T

0.45

NA

Re-adjustment

-5.35

Std.T

0.38

0.15

Storage

25.46

Wilx.T

HTML 0061000.00

NA

μCT scanning

-3.92

Std.T

0.92

0.02

Stance-Adduction

10.92

Std.T

0.11

0.55

Stance-Abduction

-33.22

Std.T

HTML 0061000.00

2.54

Fall-Adduction

-4.90

Wilx.T

0.17

NA

Fall-Abduction

12.91

Wilx.T

HTML 0061000.02

NA

  1. The first half of the table pertains to the unwanted sources of uncertainties, while the second half contains the controlled alteration of the test conditions. Based on the normality test results for the samples, either Student’s or Wilcoxon T-test was used for the significance test. Significant results are highlighted by green background. For the re-fixation parameter, due to lack of sufficient sample size, only the average % Diff is reported. The average repetition error (CV%) for all load cases was 1.53% (95% confidence interval [0.32, 2.75])