Skip to main content

Table 1 Intratester Reliability of the measurement of Thickness of the Transverse Abdominis muscle at three Lumbar levels at rest and during the hollowing maneuver

From: Extended field-of-view ultrasound imaging is reliable for measuring Transversus Abdominis muscle size at rest and during contraction

Resting

 

Right Mean (SD) cm

Left Mean (SD) cm

ICC (95% CI)

SEM (cm) (95% CI for muscle size)

% SEM

Minimal Difference (cm)

Level

  

Right

Left

Right

Left

Right

Left

Right

Left

L1

0.394 (0.096)

0.404 (0.099)

0.985

(0.967–0.994)

0.976

(0.947–0.991)

0.012

(0.37–0.42)

0.015

(0.374–0.434)

3.046

3.713

0.033

0.043

L3

0.370 (0.078)

0.393 (0.108)

0.971

(0.936–.989)

0.969

(0.933–0.988)

0.016

(0.339–0.403)

0.019

(0.355–0.430)

3.846

4.835

0.044

0.053

L5

0.355 (0.071)

0.340 (0.091)

0.926

(0.838–0.972)

0.976

(0.947–0.991)

0.019

(0.32–0.39)

0.014

(0.311–0.367)

5.352

4.118

0.053

0.039

Average value (SD)

0.376 (0.09)

0.97 (0.02)

0.02 (0.003)

4.15 (0.83)

0.04 (0.008)

Hollowing

 

Right Mean (SD) cm

Left Mean (SD) cm

ICC (95% CI)

SEM (95% CI for muscle size)

%SEM

Minimal Difference (cm)

Level

  

Right

Left

Right

Left

Right

Left

Right

Left

L1

0.541 (0.153)

0.567 (0.148)

0.972

(0.938–0.989)

0.973

(0.942–0.990)

0.025

(0.49–0.59)

0.024

(0.519–0.615)

4.621

4.233

0.071

0.067

L3

0.523 (0.162)

0.561 (0.157)

0.976

(0.947–0.991)

0.977

(0.951–0.992)

0.025

(0.48–0.58)

0.024

(0.515–0.610)

4.780

4.278

0.069

0.066

L5

0.494 (0.186)

0.545 (0.246)

0.977

(0.949–0.991)

0.992

(0.983–0.997)

0.028

(0.44–0.55)

0.022

(0.502–0.588)

5.668

4.037

0.078

0.061

Average value (SD)

0.539 (0.18)

0.98 (0.01)

0.02 (0.002)

4.60 (0.59)

0.07 (0.006)