Skip to main content

Table 1 Results of the motion parameters

From: Complex biomechanical properties of non-augmented and augmented pedicle screws in human vertebrae with reduced bone density

 

Group 1 (non-augmented)

Group 2 (augmented)

Parameter

Screw head

Screw tip

Screw head

Screw tip

Pdisp

0.5 (0.3/1.3)*

−0.7 (− 0.6/−1.0)

0.3 (0.3/0.4)*

− 0.6 (− 0.3/− 0.8)

Pdefl

1.4° (1.3°/3.0°)*

1.2° (1.1°/1.4°)*

Pstiff

216 (76/332)

140 (105/162)*

295 (285/385)

159 (129/297)*

Cdisp

1.7 (0.5/2.9)*

−1.2 (−0.5/− 1.9)*

0.3 (0.2/0.3)*

−0.1(− 0.1/− 0.2)*

DEFL

7.0° (2.2°/9.6°)**

1.9° (1.6°/2.0°)**

STIFF

80 (63/316)**

122 (83/277)*

475 (417/602)**

298 (263/340)*

Cmov start

0.6 (0.5/0.6)

0.9 (0.8/1.2)*

0.5 (0.3/0.5)

0.7 (0.6/0.9)*

Cmov end

0.4 (0.3/0.4)

0.7 (0.5/0.8)

0.6 (0.3/0.6)

0.6 (0.5/0.7)

PDmax

0.14 (0.07/0.42)*

0.17 (0.11/0.45)*

0.02 (0.01/0.02)*

0.05 (0.03/0.05)*

  1. Subsidence under preload (Pdisp, n = 12) and its deflection (Pdefl, n = 12) and stiffness (Pstiff, n = 12), subsidence under cyclic loading (Cdisp, n = 10), total deflection (DEFL, n = 10), total stiffness (STIFF, n = 10), screw loosening (Cmov, n = 10), maximum progress of displacement (PDmax, n = 10). Stiffness is in N/mm, and all other values are in mm (Q25/Q75). Significantly different * (p < 0.05), ** (p < 0.01) between group 1 (non-augmented) and group 2 (augmented). The outliers were eliminated from cyclic analysis