From: Comparison of radiological and clinical outcomes, complications, and implant removals in anatomically pre-contoured clavicle plates versus reconstruction plates – a propensity score matched retrospective cohort study of 106 patients

Fig. 3

A patient with a comminuted shaft fracture treated with an RLCP. At three months, the implant had broken and the fracture displayed non-union. The patient continued to have mechanical pain and crepitation upon movement. The fracture healed after revision with an ALCP and bone grafting