Skip to main content

Table 2 Interobserver reliability ICC of both methods in terms of angular and translational errors

From: Navigation-assisted suture anchor insertion for arthroscopic rotator cuff repair

Conventional

angular error

Investigator 2

Investigator 3

Investigator 4

Investigator 5

Investigator 1

0.405

fair

0.715

substantial

0.045

slight agreement

0.010

slight agreement

Investigator 2

  

0.732

substantial

0.725

substantial

0.615

substantial

Investigator 3

    

0.577

moderate

0.167

slight agreement

Investigator 4

      

0.436

moderate

translational error

Investigator 2

Investigator 3

Investigator 4

Investigator 5

Investigator 1

0.756

substantial

0.761

substantial

0.898

almost perfect

−0.396

slight agreement

Investigator 2

  

0.722

substantial

0.800

substantial

−0.873

slight agreement

Investigator 3

    

0.878

almost perfect

−0.580

slight agreement

Investigator 4

      

−0.607

slight agreement

NAS

angular error

Investigator 2

Investigator 3

Investigator 4

Investigator 5

Investigator 1

0.883

almost perfect

1.000

almost perfect

0.742

substantial

0.879

almost perfect

Investigator 2

  

0.883

almost perfect

0.864

almost perfect

0.979

almost perfect

Investigator 3

    

0.742

substantial

0.890

almost perfect

Investigator 4

      

0.773

substantial

translational error

Investigator 2

Investigator 3

Investigator 4

Investigator 5

Investigator 1

0.839

almost perfect

0.762

substantial

0.806

substantial

0.902

almost perfect

Investigator 2

  

0.769

substantial

0.668

substantial

0.934

almost perfect

Investigator 3

    

0.477

moderate

0.757

substantial

Investigator 4

      

0.879

almost perfect

  1. 0–0.2 slight agreement
  2. 0.21–0.4 fair
  3. 0.41–0.6 moderate
  4. 0.61–0.8 substantial
  5. 0.81–100 almost perfect