Table 2 Results of mean displacement d (in mm) of specimens underwent cyclic loading test, and equality of variance

From: Loading capacity of dynamic knee spacers: a comparison between hand-moulded and COPAL spacers

 

Mean ± SD

 

variance

Cyclic load

COPAL (n = 4)

hand-moulded (n = 4)

p-value

p-value

[30 400]

0.36 ± 0.14

0.36 ± 0.20

0.99

0.56

[30 600]

0.63 ± 0.30

0.48 ± 0.19

0.42

0.33

[30 800]

0.83 ± 0.40

0.65 ± 0.30

0.48

0.67

[30 1000]

1.01 ± 0.44

0.84 ± 0.43

0.62

0.96

[30 1200]

1.19 ± 0.57

0.89 ± 0.30

0.36

0.26