Skip to main content

Table 1 Results of specimens that underwent the axial load-to-failure test, and equality of variance

From: Loading capacity of dynamic knee spacers: a comparison between hand-moulded and COPAL spacers

 

Mean ± SD

 

variance

 

COPAL (n = 3)

hand-moulded (n = 4)

p-value

p-value

Maximum Load (N)

5202.0 ± 486.9

4509.7 ± 1092.6

0.32

0.36

Stiffness (N/mm)

1098.3 ± 201.5

1008.7 ± 275.4

0.64

0.56